Skip to main content
Top
Published in: Journal of Anesthesia 2/2010

01-04-2010 | Short Communication

Comparison of tracheal intubation by the Macintosh laryngoscope and Pentax-AWS (Airway Scope) during chest compression: a manikin study

Authors: Nobuyasu Komasawa, Ryusuke Ueki, Hajime Nomura, Motoi Itani, Yoshiroh Kaminoh

Published in: Journal of Anesthesia | Issue 2/2010

Login to get access

Abstract

To compare tracheal intubation with the Pentax Airway Scope (AWS) and the Macintosh laryngoscope (McL) during chest compression, 25 anesthesiologists (including 12 specialists having >5 years of experience and 13 trainees having <2 years of experience) performed tracheal intubation using either the McL or the AWS, with or without chest compression, on a manikin. Using the McL, both specialists and trainees took a significantly longer time (P < 0.01) to secure the airway with chest compression (17.3 ± 3.7 and 22.5 ± 8.0, respectively) and than without chest compression (11.3 ± 2.9 and 13.9 ± 4.4 s, respectively). No significant difference was observed in time needed to secure the airway using the AWS with or without chest compression in both groups. From the standpoint of experience, time to complete intubation for specialists using the McL during chest compression was significantly shorter than that for trainees. In contrast, the difference in time to complete intubation with the AWS during chest compression was not significantly different between the two groups. Based on these results, we conclude that the use of the AWS may reduce the time needed to secure the airway during chest compression.
Literature
1.
go back to reference Hazinski MF, Nadkarni VM, Hickey RW, O’Connor R, Becker LB, Zaritsky A. Major changes in the 2005 AHA Guidelines for CPR and ECC: reaching the tipping point for change. Circulation. 2005;112:206–11.CrossRef Hazinski MF, Nadkarni VM, Hickey RW, O’Connor R, Becker LB, Zaritsky A. Major changes in the 2005 AHA Guidelines for CPR and ECC: reaching the tipping point for change. Circulation. 2005;112:206–11.CrossRef
2.
go back to reference Gatward JJ, Thomas MJ, Nolan JP, Cook TM. Effect of chest compressions on the time taken to insert airway devices in a manikin. Br J Anaesth. 2008;100:351–6.CrossRefPubMed Gatward JJ, Thomas MJ, Nolan JP, Cook TM. Effect of chest compressions on the time taken to insert airway devices in a manikin. Br J Anaesth. 2008;100:351–6.CrossRefPubMed
3.
go back to reference Katz SH, Falk JL. Misplaced endotracheal tubes by paramedics in an urban emergency medical services system. Ann Emerg Med. 2001;37:32–7.CrossRefPubMed Katz SH, Falk JL. Misplaced endotracheal tubes by paramedics in an urban emergency medical services system. Ann Emerg Med. 2001;37:32–7.CrossRefPubMed
4.
go back to reference Jones JH, Murphy MP, Dickson RL, Somerville GG, Brizendine EJ. Emergency physician-verified out-of-hospital intubation: miss rates by paramedics. Acad Emerg Med. 2004;11:707–9.PubMed Jones JH, Murphy MP, Dickson RL, Somerville GG, Brizendine EJ. Emergency physician-verified out-of-hospital intubation: miss rates by paramedics. Acad Emerg Med. 2004;11:707–9.PubMed
5.
go back to reference Koyama J, Aoyama T, Kusano Y, Seguchi T, Kawagishi K, Iwashita T, et al. Description and first clinical application of Airway Scope for tracheal intubation. J Neurosurg Anesthesiol. 2006;18:247–50.CrossRefPubMed Koyama J, Aoyama T, Kusano Y, Seguchi T, Kawagishi K, Iwashita T, et al. Description and first clinical application of Airway Scope for tracheal intubation. J Neurosurg Anesthesiol. 2006;18:247–50.CrossRefPubMed
6.
go back to reference Asai T, Liu EH, Matsumoto S, Hirabayashi Y, Seo N, Suzuki A, et al. Use of the Pentax-AWS in 293 patients with difficult airways. Anesthesiology. 2009;110:898–904.CrossRefPubMed Asai T, Liu EH, Matsumoto S, Hirabayashi Y, Seo N, Suzuki A, et al. Use of the Pentax-AWS in 293 patients with difficult airways. Anesthesiology. 2009;110:898–904.CrossRefPubMed
7.
go back to reference Hirabayashi Y, Seo N. Airway Scope: early clinical experience in 405 patients. J Anesth. 2008;22:81–5.CrossRefPubMed Hirabayashi Y, Seo N. Airway Scope: early clinical experience in 405 patients. J Anesth. 2008;22:81–5.CrossRefPubMed
8.
go back to reference Hirabayashi Y. Airway Scope versus Macintosh laryngoscope: a manikin study. Emerg Med J. 2007;24:357–8.CrossRefPubMed Hirabayashi Y. Airway Scope versus Macintosh laryngoscope: a manikin study. Emerg Med J. 2007;24:357–8.CrossRefPubMed
9.
go back to reference Suzuki A, Toyama Y, Katsumi N, Kunisawa T, Sasaki R, Hirota K, et al. The Pentax-AWS® rigid indirect video laryngoscope: clinical assessment of performance in 320 cases. Anaesthesia. 2008;63:641–7.CrossRefPubMed Suzuki A, Toyama Y, Katsumi N, Kunisawa T, Sasaki R, Hirota K, et al. The Pentax-AWS® rigid indirect video laryngoscope: clinical assessment of performance in 320 cases. Anaesthesia. 2008;63:641–7.CrossRefPubMed
10.
go back to reference Komatsu R, Kamata K, Hamada K, Sessler DI, Ozaki M. Airway scope and Stylet scope for tracheal intubation in a simulated difficult airway. Anesth Analg. 2009;108:273–9.CrossRefPubMed Komatsu R, Kamata K, Hamada K, Sessler DI, Ozaki M. Airway scope and Stylet scope for tracheal intubation in a simulated difficult airway. Anesth Analg. 2009;108:273–9.CrossRefPubMed
11.
12.
go back to reference Hirabayashi Y, Seo N. Tracheal intubation by non-anesthesia residents using the Pentax-AWS airway scope and Macintosh laryngoscope. J Clin Anesth. 2009;21:268–71.CrossRefPubMed Hirabayashi Y, Seo N. Tracheal intubation by non-anesthesia residents using the Pentax-AWS airway scope and Macintosh laryngoscope. J Clin Anesth. 2009;21:268–71.CrossRefPubMed
13.
go back to reference Hirabayashi Y. Ease of use of the Airway Scope vs the Bullard laryngoscope: a manikin study. Can J Anaesth. 2007;54:397–8.CrossRefPubMed Hirabayashi Y. Ease of use of the Airway Scope vs the Bullard laryngoscope: a manikin study. Can J Anaesth. 2007;54:397–8.CrossRefPubMed
Metadata
Title
Comparison of tracheal intubation by the Macintosh laryngoscope and Pentax-AWS (Airway Scope) during chest compression: a manikin study
Authors
Nobuyasu Komasawa
Ryusuke Ueki
Hajime Nomura
Motoi Itani
Yoshiroh Kaminoh
Publication date
01-04-2010
Publisher
Springer Japan
Published in
Journal of Anesthesia / Issue 2/2010
Print ISSN: 0913-8668
Electronic ISSN: 1438-8359
DOI
https://doi.org/10.1007/s00540-010-0874-9

Other articles of this Issue 2/2010

Journal of Anesthesia 2/2010 Go to the issue