Published in: BMC Ophthalmology 1/2017

Open Access 01-12-2017 | Research article

Comparison of intraocular pressure as measured by three different non-contact tonometers and goldmann applanation tonometer for non-glaucomatous subjects

Authors: Seung Pil Bang, Chong Eun Lee, Yu Cheol Kim

Published in: BMC Ophthalmology | Issue 1/2017

Login to get access

Abstract

Background

To compare the measurement of intraocular pressure (IOP) among the three different non-contact tonometers (NCT) and the Goldmann applanation tonometer (GAT) for non-glaucomatous subjects.

Methods

In 52 eyes of 52 non-glaucomatous subjects, IOP was measured sequentially with the Canon TX-20P, the Nidek NT-530P, the Topcon CT-1P, and the GAT at the same time. We evaluated the IOP-measurement agreement among the tonometers as well as the factors affecting the measurements.

Results

A significant positive correlation was shown between the IOP values obtained with GAT and each NCT. The Canon TX-20P showed statistically the most significant agreement with the GAT (ICC 0.906, 95% CI 0.837–0.946). In an analysis of the Bland-Altman plots, the Canon TX-20P also showed the largest mean bias (1.38 mmHg) but the narrowest limits of agreement (LoA) (95% LoA; ± 3.43 mmHg). The Topcon CT-1P showed the smallest mean bias (0.48 mmHg) but the widest LoA (95% LoA; ± 4.16 mmHg). The Topcon CT-1P and Nidek NT-530P both showed a significantly positive correlation between increasing central corneal thickness (CCT) and increasing IOP.

Conclusion

There was a statistically significant correlation between each of the three different NCT and the GAT measurements. IOP measured with the Canon TX-20P and Topcon CT-1P tended to be higher, and with the Nidek NT-530P lower, than with the GAT. Practitioners need to know the properties of their own NCTs and their respective measurement tendencies.
Literature
1.
go back to reference Hansen MK. Clinical comparison of the XPERT non-contact tonometer and the conventional Goldmann applanation tonometer. Acta Ophthalmol Scand. 1995;73(2):176–80.CrossRefPubMed Hansen MK. Clinical comparison of the XPERT non-contact tonometer and the conventional Goldmann applanation tonometer. Acta Ophthalmol Scand. 1995;73(2):176–80.CrossRefPubMed
2.
go back to reference Beatty S, Nischal KK, Jones H, Eagling EM. Effect of applanation tonometry on mean corneal curvature. J Cataract Refract Surg. 1996;22(7):970–1.CrossRefPubMed Beatty S, Nischal KK, Jones H, Eagling EM. Effect of applanation tonometry on mean corneal curvature. J Cataract Refract Surg. 1996;22(7):970–1.CrossRefPubMed
3.
go back to reference Cho P, Lui T. Comparison of the performance of the Nidek NT-2000 noncontact tonometer with the Keeler Pulsair 2000 and the Goldmann applanation tonometer. Optom Vis Sci. 1997;74(1):51–8.CrossRefPubMed Cho P, Lui T. Comparison of the performance of the Nidek NT-2000 noncontact tonometer with the Keeler Pulsair 2000 and the Goldmann applanation tonometer. Optom Vis Sci. 1997;74(1):51–8.CrossRefPubMed
4.
go back to reference Jorge J, Diaz-Rey JA, Gonzalez-Meijome JM, Almeida JB, Parafita MA. Clinical performance of the Reichert AT550: a new non-contact tonometer. Ophthalmic Physiol Opt. 2002;22(6):560–4.CrossRefPubMed Jorge J, Diaz-Rey JA, Gonzalez-Meijome JM, Almeida JB, Parafita MA. Clinical performance of the Reichert AT550: a new non-contact tonometer. Ophthalmic Physiol Opt. 2002;22(6):560–4.CrossRefPubMed
5.
go back to reference Garcia Resua C, Giraldez Fernandez MJ, Cervino Exposito A, Gonzalez Perez J, Yebra-Pimentel E. Clinical evaluation of the new TGDc-01 "PRA" palpebral tonometer: comparison with contact and non-contact tonometry. Optom Vis Sci. 2005;82(2):143–50.CrossRefPubMed Garcia Resua C, Giraldez Fernandez MJ, Cervino Exposito A, Gonzalez Perez J, Yebra-Pimentel E. Clinical evaluation of the new TGDc-01 "PRA" palpebral tonometer: comparison with contact and non-contact tonometry. Optom Vis Sci. 2005;82(2):143–50.CrossRefPubMed
6.
go back to reference van der Jagt LH, Jansonius NM. Three portable tonometers, the TGDc-01, the ICARE and the Tonopen XL, compared with each other and with Goldmann applanation tonometry. Ophthalmic Physiol Opt. 2005;25(5):429–35.CrossRefPubMed van der Jagt LH, Jansonius NM. Three portable tonometers, the TGDc-01, the ICARE and the Tonopen XL, compared with each other and with Goldmann applanation tonometry. Ophthalmic Physiol Opt. 2005;25(5):429–35.CrossRefPubMed
7.
go back to reference Shields MB. The non-contact tonometer. Its value and limitations. Surv Ophthalmol. 1980;24(4):211–9.CrossRefPubMed Shields MB. The non-contact tonometer. Its value and limitations. Surv Ophthalmol. 1980;24(4):211–9.CrossRefPubMed
8.
go back to reference Moreno-Montanes J, Gomez-Demmel E, Lajara-Blesa J, Aliseda-Perez de Madrid D. Comparative study of three non-contact tonometers and the Goldmann tonometer. Ophthalmologica. 1994;208(3):115–8.CrossRefPubMed Moreno-Montanes J, Gomez-Demmel E, Lajara-Blesa J, Aliseda-Perez de Madrid D. Comparative study of three non-contact tonometers and the Goldmann tonometer. Ophthalmologica. 1994;208(3):115–8.CrossRefPubMed
9.
go back to reference Tonnu PA, Ho T, Sharma K, White E, Bunce C, Garway-Heath D. A comparison of four methods of tonometry: method agreement and interobserver variability. Br J Ophthalmol. 2005;89(7):847–50.CrossRefPubMedPubMedCentral Tonnu PA, Ho T, Sharma K, White E, Bunce C, Garway-Heath D. A comparison of four methods of tonometry: method agreement and interobserver variability. Br J Ophthalmol. 2005;89(7):847–50.CrossRefPubMedPubMedCentral
10.
go back to reference Jansson U, Brautaset R, Cerviño A, Nilsson M. A comparison of the canon TX-20P™ non-contact tonometer and pachymeter in healthy eyes. Int J Ophthalmic Pract. 2012;3(3):96–102.CrossRef Jansson U, Brautaset R, Cerviño A, Nilsson M. A comparison of the canon TX-20P™ non-contact tonometer and pachymeter in healthy eyes. Int J Ophthalmic Pract. 2012;3(3):96–102.CrossRef
11.
go back to reference Tranchina L, Lombardo M, Oddone F, Serrao S, Schiano Lomoriello D, Ducoli P. Influence of corneal biomechanical properties on intraocular pressure differences between an air-puff tonometer and the Goldmann applanation tonometer. J Glaucoma. 2013;22(5):416–21.CrossRefPubMed Tranchina L, Lombardo M, Oddone F, Serrao S, Schiano Lomoriello D, Ducoli P. Influence of corneal biomechanical properties on intraocular pressure differences between an air-puff tonometer and the Goldmann applanation tonometer. J Glaucoma. 2013;22(5):416–21.CrossRefPubMed
12.
go back to reference Anton A, Neuburger M, Bohringer D, Jordan JF. Comparative measurement of intraocular pressure by Icare tonometry and Airpuff tonometry in healthy subjects and patients wearing therapeutic soft contact lenses. Graefes Arch Clin Exp Ophthalmol. 2013;251(7):1791–5.CrossRefPubMed Anton A, Neuburger M, Bohringer D, Jordan JF. Comparative measurement of intraocular pressure by Icare tonometry and Airpuff tonometry in healthy subjects and patients wearing therapeutic soft contact lenses. Graefes Arch Clin Exp Ophthalmol. 2013;251(7):1791–5.CrossRefPubMed
13.
go back to reference Wei W, Fan Z, Wang L, Li Z, Jiao W, Li Y. Correlation analysis between central corneal thickness and intraocular pressure in juveniles in northern China: the Jinan city eye study. PLoS One. 2014;9(8):e104842.CrossRefPubMedPubMedCentral Wei W, Fan Z, Wang L, Li Z, Jiao W, Li Y. Correlation analysis between central corneal thickness and intraocular pressure in juveniles in northern China: the Jinan city eye study. PLoS One. 2014;9(8):e104842.CrossRefPubMedPubMedCentral
14.
go back to reference Garcia-Resua C, Pena-Verdeal H, Minones M, Giraldez MJ, Yebra-Pimentel E. Reliability of the non-contact tono-pachymeter Tonopachy NT-530P in healthy eyes. Clin Exp Optom. 2013;96(3):286–94.CrossRefPubMed Garcia-Resua C, Pena-Verdeal H, Minones M, Giraldez MJ, Yebra-Pimentel E. Reliability of the non-contact tono-pachymeter Tonopachy NT-530P in healthy eyes. Clin Exp Optom. 2013;96(3):286–94.CrossRefPubMed
15.
go back to reference Fujimura F, Kamiya K, Fujiwara K, Shoji N, Shimizu K. Repeatability and reproducibility of measurements using a NT-530P noncontact tono/pachymeter and correlation of central corneal thickness with intraocular pressure. Biomed Res Int. 2013;2013:370592.CrossRefPubMedPubMedCentral Fujimura F, Kamiya K, Fujiwara K, Shoji N, Shimizu K. Repeatability and reproducibility of measurements using a NT-530P noncontact tono/pachymeter and correlation of central corneal thickness with intraocular pressure. Biomed Res Int. 2013;2013:370592.CrossRefPubMedPubMedCentral
16.
go back to reference Ogbuehi KC. Assessment of the accuracy and reliability of the Topcon CT80 non-contact tonometer. Clin Exp Optom. 2006;89(5):310–4.CrossRefPubMed Ogbuehi KC. Assessment of the accuracy and reliability of the Topcon CT80 non-contact tonometer. Clin Exp Optom. 2006;89(5):310–4.CrossRefPubMed
17.
go back to reference Farhood QK. Comparative evaluation of intraocular pressure with an air-puff tonometer versus a Goldmann applanation tonometer. Clin Ophthalmol. 2013;7:23–7.PubMed Farhood QK. Comparative evaluation of intraocular pressure with an air-puff tonometer versus a Goldmann applanation tonometer. Clin Ophthalmol. 2013;7:23–7.PubMed
18.
go back to reference Tamcelik N, Atalay E, Cicik E, Ozkok A. Comparability of Icare pro rebound Tonometer with Goldmann Applanation and noncontact Tonometer in a wide range of intraocular pressure and central corneal thickness. Ophthalmic Res. 2015;54(1):18–25.CrossRefPubMed Tamcelik N, Atalay E, Cicik E, Ozkok A. Comparability of Icare pro rebound Tonometer with Goldmann Applanation and noncontact Tonometer in a wide range of intraocular pressure and central corneal thickness. Ophthalmic Res. 2015;54(1):18–25.CrossRefPubMed
19.
go back to reference Garcia-Resua C, Giraldez Fernandez MJ, Yebra-Pimentel E, Garcia-Montero S. Clinical evaluation of the canon TX-10 noncontact tonometer in healthy eyes. Eur J Ophthalmol. 2010;20(3):523–30.PubMed Garcia-Resua C, Giraldez Fernandez MJ, Yebra-Pimentel E, Garcia-Montero S. Clinical evaluation of the canon TX-10 noncontact tonometer in healthy eyes. Eur J Ophthalmol. 2010;20(3):523–30.PubMed
20.
go back to reference Pelit A, Altan-Yaycioglu R, Pelit A, Akova YA. Effect of corneal thickness on intraocular pressure measurements with the Pascal dynamic contour, canon TX-10 non-contact and Goldmann applanation tonometers in healthy subjects. Clin Exp Optom. 2009;92(1):14–8.CrossRefPubMed Pelit A, Altan-Yaycioglu R, Pelit A, Akova YA. Effect of corneal thickness on intraocular pressure measurements with the Pascal dynamic contour, canon TX-10 non-contact and Goldmann applanation tonometers in healthy subjects. Clin Exp Optom. 2009;92(1):14–8.CrossRefPubMed
21.
go back to reference Yilmaz I, Altan C, Aygit ED, Alagoz C, Baz O, Ahmet S, Urvasizoglu S, Yasa D, Demirok A. Comparison of three methods of tonometry in normal subjects: Goldmann applanation tonometer, non-contact airpuff tonometer, and Tono-pen XL. Clin Ophthalmol. 2014;8:1069–74.CrossRefPubMedPubMedCentral Yilmaz I, Altan C, Aygit ED, Alagoz C, Baz O, Ahmet S, Urvasizoglu S, Yasa D, Demirok A. Comparison of three methods of tonometry in normal subjects: Goldmann applanation tonometer, non-contact airpuff tonometer, and Tono-pen XL. Clin Ophthalmol. 2014;8:1069–74.CrossRefPubMedPubMedCentral
22.
go back to reference Esen F, Kostek M, Emekli AS, Eraslan M. Double-organ bias in published randomized controlled trials of glaucoma. J Glaucoma. 2016;25(6):520–2.CrossRefPubMed Esen F, Kostek M, Emekli AS, Eraslan M. Double-organ bias in published randomized controlled trials of glaucoma. J Glaucoma. 2016;25(6):520–2.CrossRefPubMed
23.
go back to reference Almubrad TM, Ogbuehi KC. Clinical investigation of the effect of topical anesthesia on intraocular pressure. Clin Ophthalmol. 2007;1(3):305–9.PubMedPubMedCentral Almubrad TM, Ogbuehi KC. Clinical investigation of the effect of topical anesthesia on intraocular pressure. Clin Ophthalmol. 2007;1(3):305–9.PubMedPubMedCentral
24.
go back to reference AlMubrad TM, Ogbuehi KC. The effect of repeated applanation on subsequent IOP measurements. Clin Exp Optom. 2008;91(6):524–9.CrossRefPubMed AlMubrad TM, Ogbuehi KC. The effect of repeated applanation on subsequent IOP measurements. Clin Exp Optom. 2008;91(6):524–9.CrossRefPubMed
25.
go back to reference Almubrad TM, Ogbuehi KC. On repeated corneal applanation with the Goldmann and two non-contact tonometers. Clin Exp Optom. 2010;93(2):77–82.CrossRefPubMed Almubrad TM, Ogbuehi KC. On repeated corneal applanation with the Goldmann and two non-contact tonometers. Clin Exp Optom. 2010;93(2):77–82.CrossRefPubMed
26.
go back to reference Liu JH, Bouligny RP, Kripke DF, Weinreb RN. Nocturnal elevation of intraocular pressure is detectable in the sitting position. Invest Ophthalmol Vis Sci. 2003;44(10):4439–42.CrossRefPubMed Liu JH, Bouligny RP, Kripke DF, Weinreb RN. Nocturnal elevation of intraocular pressure is detectable in the sitting position. Invest Ophthalmol Vis Sci. 2003;44(10):4439–42.CrossRefPubMed
27.
go back to reference Doughty MJ, Zaman ML. Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol. 2000;44(5):367–408.CrossRefPubMed Doughty MJ, Zaman ML. Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol. 2000;44(5):367–408.CrossRefPubMed
28.
go back to reference Hwang YH, Kim HK, Sohn YH, Namil Study Group KGS. Central corneal thickness in a Korean population: the Namil study. Invest Ophthalmol Vis Sci. 2012;53(11):6851–5.CrossRefPubMed Hwang YH, Kim HK, Sohn YH, Namil Study Group KGS. Central corneal thickness in a Korean population: the Namil study. Invest Ophthalmol Vis Sci. 2012;53(11):6851–5.CrossRefPubMed
29.
go back to reference Regine F, Scuderi GL, Cesareo M, Ricci F, Cedrone C, Nucci C. Validity and limitations of the Nidek NT-4000 non-contact tonometer: a clinical study. Ophthalmic Physiol Opt. 2006;26(1):33–9.CrossRefPubMed Regine F, Scuderi GL, Cesareo M, Ricci F, Cedrone C, Nucci C. Validity and limitations of the Nidek NT-4000 non-contact tonometer: a clinical study. Ophthalmic Physiol Opt. 2006;26(1):33–9.CrossRefPubMed
30.
go back to reference Recep OF, Hasiripi H, Cagil N, Sarikatipoglu H. Relation between corneal thickness and intraocular pressure measurement by noncontact and applanation tonometry. J Cataract Refract Surg. 2001;27(11):1787–91.CrossRefPubMed Recep OF, Hasiripi H, Cagil N, Sarikatipoglu H. Relation between corneal thickness and intraocular pressure measurement by noncontact and applanation tonometry. J Cataract Refract Surg. 2001;27(11):1787–91.CrossRefPubMed
31.
go back to reference Stodtmeister R. Applanation tonometry and correction according to corneal thickness. Acta Ophthalmol Scand. 1998;76(3):319–24.CrossRefPubMed Stodtmeister R. Applanation tonometry and correction according to corneal thickness. Acta Ophthalmol Scand. 1998;76(3):319–24.CrossRefPubMed
32.
go back to reference Babalola OE, Kehinde AV, Iloegbunam AC, Akinbinu T, Moghalu C, Onuoha I. A comparison of the Goldmann applanation and non-contact (Keeler Pulsair EasyEye) tonometers and the effect of central corneal thickness in indigenous African eyes. Ophthalmic Physiol Opt. 2009;29(2):182–8.CrossRefPubMed Babalola OE, Kehinde AV, Iloegbunam AC, Akinbinu T, Moghalu C, Onuoha I. A comparison of the Goldmann applanation and non-contact (Keeler Pulsair EasyEye) tonometers and the effect of central corneal thickness in indigenous African eyes. Ophthalmic Physiol Opt. 2009;29(2):182–8.CrossRefPubMed
33.
go back to reference Lee YG, Kim JH, Kim NR, Kim CY, Lee ES. Comparison between Tonopachy and other tonometric and pachymetric devices. Optom Vis Sci. 2011;88(7):843–9.CrossRefPubMed Lee YG, Kim JH, Kim NR, Kim CY, Lee ES. Comparison between Tonopachy and other tonometric and pachymetric devices. Optom Vis Sci. 2011;88(7):843–9.CrossRefPubMed
34.
go back to reference Whitacre MM, Stein R. Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol. 1993;38(1):1–30.CrossRefPubMed Whitacre MM, Stein R. Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol. 1993;38(1):1–30.CrossRefPubMed
Metadata
Title
Comparison of intraocular pressure as measured by three different non-contact tonometers and goldmann applanation tonometer for non-glaucomatous subjects
Authors
Seung Pil Bang
Chong Eun Lee
Yu Cheol Kim
Publication date
01-12-2017
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2017
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-017-0593-1

Other articles of this Issue 1/2017

BMC Ophthalmology 1/2017 Go to the issue