01-02-2016 | Original Article

Comparison of image quality and radiation dose between an image-intensifier system and a newer-generation flat-panel detector system — technical phantom measurements and evaluation of clinical imaging in children

Authors: Meike Weis, Claudia Hagelstein, Theo Diehm, Stefan O. Schoenberg, K. Wolfgang Neff

Published in: Pediatric Radiology | Issue 2/2016

Login to get access

Abstract

Background

Many image-intensifier fluoroscopy systems have been replaced by flat-panel detectors in recent years.

Objective

To compare the level of contrast, image resolution and radiation dose between an image-intensifier and a newer-generation flat-panel detector system in a pediatric radiology unit.

Materials and methods

We compared two systems — a conventional image intensifier and a newer-generation flat-panel system. We measured image quality and radiation dose using a technical phantom. Additionally, we retrospectively compared age-matched fluoroscopic pediatric voiding cystourethrography (n = 15) and upper gastrointestinal investigations (n = 25).

Results

In phantom studies image contrast was equal while image resolution was higher and mean radiation dose lower using the flat-panel system (P < 0.0001). In pediatric investigations, mean dose area product was significantly reduced on the flat-panel system for upper gastrointestinal investigation (45 ± 38 μGy*m2 vs. 11 ± 9 μGy*m2; P < 0.0001) and for voiding cystourethrography (18 ± 20 μGy*m2 vs. 10 ± 12 μGy*m2; P = 0.04).

Conclusion

The newer flat-panel system performs at lower dose levels with equal to better image quality and therefore seems to be the more suitable technique for pediatric fluoroscopy in comparison to image-intensifier systems.
Literature
1.
go back to reference Partridge J, McGahan G, Causton S et al (2006) Radiation dose reduction without compromise of image quality in cardiac angiography and intervention with the use of a flat panel detector without an antiscatter grid. Heart 92:507–510CrossRefPubMedCentralPubMed Partridge J, McGahan G, Causton S et al (2006) Radiation dose reduction without compromise of image quality in cardiac angiography and intervention with the use of a flat panel detector without an antiscatter grid. Heart 92:507–510CrossRefPubMedCentralPubMed
3.
go back to reference Hammer GP, Seidenbusch MC, Schneider K et al (2009) A cohort study of childhood cancer incidence after postnatal diagnostic X-ray exposure. Radiat Res 171:504–512CrossRefPubMed Hammer GP, Seidenbusch MC, Schneider K et al (2009) A cohort study of childhood cancer incidence after postnatal diagnostic X-ray exposure. Radiat Res 171:504–512CrossRefPubMed
4.
go back to reference Justino H (2006) The ALARA concept in pediatric cardiac catheterization: techniques and tactics for managing radiation dose. Pediatr Radiol 36:146–153CrossRefPubMedCentralPubMed Justino H (2006) The ALARA concept in pediatric cardiac catheterization: techniques and tactics for managing radiation dose. Pediatr Radiol 36:146–153CrossRefPubMedCentralPubMed
5.
go back to reference Darling S, Sammer M, Chapman T et al (2011) Physician documentation of fluoroscopy time in voiding cystourethrography reports correlates with lower fluoroscopy times: a surrogate marker of patient radiation exposure. AJR Am J Roentgenol 196:W777–W780CrossRefPubMed Darling S, Sammer M, Chapman T et al (2011) Physician documentation of fluoroscopy time in voiding cystourethrography reports correlates with lower fluoroscopy times: a surrogate marker of patient radiation exposure. AJR Am J Roentgenol 196:W777–W780CrossRefPubMed
6.
go back to reference Malekzadeh M, Bahreyni Toossi MT, Alamdaran SA et al (2012) Comparison between patient dose arising from photofluorographic and standard fluoroscopic voiding cystourethrography in children with urinary tract infection [corrected]. Nephrourol Mon 4:541–544CrossRefPubMed Malekzadeh M, Bahreyni Toossi MT, Alamdaran SA et al (2012) Comparison between patient dose arising from photofluorographic and standard fluoroscopic voiding cystourethrography in children with urinary tract infection [corrected]. Nephrourol Mon 4:541–544CrossRefPubMed
8.
go back to reference Rommel N, Selleslagh M, Hoffman I et al (2014) Objective assessment of swallow function in children with suspected aspiration using pharyngeal automated impedance manometry. J Pediatr Gastroenterol Nutr 58:789–794PubMed Rommel N, Selleslagh M, Hoffman I et al (2014) Objective assessment of swallow function in children with suspected aspiration using pharyngeal automated impedance manometry. J Pediatr Gastroenterol Nutr 58:789–794PubMed
9.
go back to reference Baijens L, Barikroo A, Pilz W (2013) Intrarater and interrater reliability for measurements in videofluoroscopy of swallowing. Eur J Radiol 82:1683–1695CrossRefPubMed Baijens L, Barikroo A, Pilz W (2013) Intrarater and interrater reliability for measurements in videofluoroscopy of swallowing. Eur J Radiol 82:1683–1695CrossRefPubMed
10.
go back to reference Hatakeyama Y, Kakeda S, Ohnari N et al (2007) Reduction of radiation dose for cerebral angiography using flat panel detector of direct conversion type: a vascular phantom study. AJNR Am J Neuroradiol 28:645–650PubMed Hatakeyama Y, Kakeda S, Ohnari N et al (2007) Reduction of radiation dose for cerebral angiography using flat panel detector of direct conversion type: a vascular phantom study. AJNR Am J Neuroradiol 28:645–650PubMed
11.
go back to reference Wiesinger B, Stutz A, Schmehl J et al (2012) Comparison of digital flat-panel detector and conventional angiography machines: evaluation of stent detection rates, visibility scores, and dose-area products. AJR Am J Roentgenol 198:946–954CrossRefPubMed Wiesinger B, Stutz A, Schmehl J et al (2012) Comparison of digital flat-panel detector and conventional angiography machines: evaluation of stent detection rates, visibility scores, and dose-area products. AJR Am J Roentgenol 198:946–954CrossRefPubMed
12.
go back to reference Miraglia R, Maruzzelli L, Tuzzolino F et al (2013) Radiation exposure in biliary procedures performed to manage anastomotic strictures in pediatric liver transplant recipients: comparison between radiation exposure levels using an image intensifier and a flat-panel detector-based system. Cardiovasc Intervent Radiol 36:1670–1676CrossRefPubMed Miraglia R, Maruzzelli L, Tuzzolino F et al (2013) Radiation exposure in biliary procedures performed to manage anastomotic strictures in pediatric liver transplant recipients: comparison between radiation exposure levels using an image intensifier and a flat-panel detector-based system. Cardiovasc Intervent Radiol 36:1670–1676CrossRefPubMed
13.
go back to reference Miraglia R, Maruzzelli L, Cortis K et al (2015) Comparison between radiation exposure levels using an image intensifier and a flat-panel detector-based system in image-guided central venous catheter placement in children weighing less than 10 kg. Pediatr Radiol 45:235–240CrossRefPubMed Miraglia R, Maruzzelli L, Cortis K et al (2015) Comparison between radiation exposure levels using an image intensifier and a flat-panel detector-based system in image-guided central venous catheter placement in children weighing less than 10 kg. Pediatr Radiol 45:235–240CrossRefPubMed
14.
go back to reference Tsapaki V, Kottou S, Kollaros N et al (2004) Comparison of a conventional and a flat-panel digital system in interventional cardiology procedures. Br J Radiol 77:562–567CrossRefPubMed Tsapaki V, Kottou S, Kollaros N et al (2004) Comparison of a conventional and a flat-panel digital system in interventional cardiology procedures. Br J Radiol 77:562–567CrossRefPubMed
15.
go back to reference Chida K, Inaba Y, Saito H et al (2009) Radiation dose of interventional radiology system using a flat-panel detector. AJR Am J Roentgenol 193:1680–1685CrossRefPubMed Chida K, Inaba Y, Saito H et al (2009) Radiation dose of interventional radiology system using a flat-panel detector. AJR Am J Roentgenol 193:1680–1685CrossRefPubMed
16.
go back to reference Bogaert E, Bacher K, Lapere R et al (2009) Does digital flat detector technology tip the scale towards better image quality or reduced patient dose in interventional cardiology? Eur J Radiol 72:348–353CrossRefPubMed Bogaert E, Bacher K, Lapere R et al (2009) Does digital flat detector technology tip the scale towards better image quality or reduced patient dose in interventional cardiology? Eur J Radiol 72:348–353CrossRefPubMed
17.
go back to reference Wiesinger B, Kirchner S, Blumenstock G et al (2013) Difference in dose area product between analog image intensifier and digital flat panel detector in peripheral angiography and the effect of BMI. Röfo 185:153–159PubMed Wiesinger B, Kirchner S, Blumenstock G et al (2013) Difference in dose area product between analog image intensifier and digital flat panel detector in peripheral angiography and the effect of BMI. Röfo 185:153–159PubMed
18.
go back to reference Cohen M (2007) Are we doing enough to minimize fluoroscopic radiation exposure in children? Pediatr Radiol 37:1020–1024CrossRefPubMed Cohen M (2007) Are we doing enough to minimize fluoroscopic radiation exposure in children? Pediatr Radiol 37:1020–1024CrossRefPubMed
19.
go back to reference Sidhu M, Coley BD, Goske MJ et al (2009) Image Gently, Step Lightly: increasing radiation dose awareness in pediatric interventional radiology. Pediatr Radiol 39:1135–1138CrossRefPubMed Sidhu M, Coley BD, Goske MJ et al (2009) Image Gently, Step Lightly: increasing radiation dose awareness in pediatric interventional radiology. Pediatr Radiol 39:1135–1138CrossRefPubMed
20.
go back to reference Sidhu M, Goske MJ, Connolly B et al (2010) Image Gently, Step Lightly: promoting radiation safety in pediatric interventional radiology. AJR Am J Roentgenol 195:W299–W301CrossRefPubMed Sidhu M, Goske MJ, Connolly B et al (2010) Image Gently, Step Lightly: promoting radiation safety in pediatric interventional radiology. AJR Am J Roentgenol 195:W299–W301CrossRefPubMed
21.
go back to reference Schneider K, Krüger-Stollfuss I, Ernst G et al (2001) Paediatric fluoroscopy — a survey of children’s hospitals in Europe. I. Staffing, frequency of fluoroscopic procedures and investigation technique. Pediatr Radiol 31:238–246CrossRefPubMed Schneider K, Krüger-Stollfuss I, Ernst G et al (2001) Paediatric fluoroscopy — a survey of children’s hospitals in Europe. I. Staffing, frequency of fluoroscopic procedures and investigation technique. Pediatr Radiol 31:238–246CrossRefPubMed
22.
go back to reference Hsi RS, Dearn J, Dean M et al (2013) Effective and organ specific radiation doses from videourodynamics in children. J Urol 190:1364–1369CrossRefPubMed Hsi RS, Dearn J, Dean M et al (2013) Effective and organ specific radiation doses from videourodynamics in children. J Urol 190:1364–1369CrossRefPubMed
23.
go back to reference Riccabona M, Lobo ML, Willi U et al (2014) ESPR uroradiology task force and ESUR Paediatric Work Group — Imaging recommendations in paediatric uroradiology, part VI: childhood renal biopsy and imaging of neonatal and infant genital tract: minutes from the task force session at the annual ESPR Meeting 2012 in Athens on childhood renal biopsy and imaging neonatal genitalia. Pediatr Radiol 44:496–502CrossRefPubMed Riccabona M, Lobo ML, Willi U et al (2014) ESPR uroradiology task force and ESUR Paediatric Work Group — Imaging recommendations in paediatric uroradiology, part VI: childhood renal biopsy and imaging of neonatal and infant genital tract: minutes from the task force session at the annual ESPR Meeting 2012 in Athens on childhood renal biopsy and imaging neonatal genitalia. Pediatr Radiol 44:496–502CrossRefPubMed
24.
go back to reference Tompane T, Leong CW, Bush R et al (2013) Appropriateness of radiology procedures performed in children with gastrointestinal symptoms and conditions. Clin Gastroenterol Hepatol 12:970–977CrossRefPubMed Tompane T, Leong CW, Bush R et al (2013) Appropriateness of radiology procedures performed in children with gastrointestinal symptoms and conditions. Clin Gastroenterol Hepatol 12:970–977CrossRefPubMed
25.
go back to reference Hiorns MP, Ryan MM (2006) Current practice in paediatric videofluoroscopy. Pediatr Radiol 36:911–919CrossRefPubMed Hiorns MP, Ryan MM (2006) Current practice in paediatric videofluoroscopy. Pediatr Radiol 36:911–919CrossRefPubMed
26.
go back to reference Schneider K, Perlmutter N, Arthur R et al (2000) Micturition cystourethrography in paediatric patients in selected children’s hospitals in Europe: evaluation of fluoroscopy technique, image quality criteria and dose. Radiat Prot Dosim 90:197–201CrossRef Schneider K, Perlmutter N, Arthur R et al (2000) Micturition cystourethrography in paediatric patients in selected children’s hospitals in Europe: evaluation of fluoroscopy technique, image quality criteria and dose. Radiat Prot Dosim 90:197–201CrossRef
27.
go back to reference Hiorns MP, Saini A, Marsden PJ (2006) A review of current local dose-area product levels for paediatric fluoroscopy in a tertiary referral centre compared with national standards. Why are they so different? Br J Radiol 79:326–330CrossRefPubMed Hiorns MP, Saini A, Marsden PJ (2006) A review of current local dose-area product levels for paediatric fluoroscopy in a tertiary referral centre compared with national standards. Why are they so different? Br J Radiol 79:326–330CrossRefPubMed
Metadata
Title
Comparison of image quality and radiation dose between an image-intensifier system and a newer-generation flat-panel detector system — technical phantom measurements and evaluation of clinical imaging in children
Authors
Meike Weis
Claudia Hagelstein
Theo Diehm
Stefan O. Schoenberg
K. Wolfgang Neff
Publication date
01-02-2016
Publisher
Springer Berlin Heidelberg
Published in
Pediatric Radiology / Issue 2/2016
Print ISSN: 0301-0449
Electronic ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-015-3456-z

Other articles of this Issue 2/2016

Pediatric Radiology 2/2016 Go to the issue

Hermes

Hermes

Minisymposium: Advances in Fetal and Neonatal Imaging

Fetal and neonatal imaging minisymposium – 2015 executive summary