Abstract

Background

Robot-assisted spine surgery aims to improve the accuracy of screw placement. We compared the accuracy and safety between a novel robot and free hand in thoracolumbar pedicle screw placement.

Methods

Eighty patients scheduled to undergo robot-assisted (40 patients) and free-hand (40 patients) pedicle screw placement were included. The patients’ demographic characteristics, radiographic accuracy, and perioperative outcomes were compared. The accuracy of screw placement was based on cortical violation and screw deviation. Safety outcomes mainly included operative time, blood loss, revision, and complications.

Results

A total of 178 and 172 screws were placed in the robot-assisted and free-hand groups, respectively. The rate of perfect screw position (grade A) was higher in the robot-assisted group than in the free-hand group (91.0% vs. 75.6%; P < 0.001). The rate of clinically acceptable screw position (grades A and B) was also higher in the robot-assisted group than in the free-hand group (99.4% vs. 90.1%; P < 0.001). The robot-assisted group had significantly lower screw deviation than the free-hand group [1.46 (0.94, 1.95) mm vs. 2.48 (1.09, 3.74) mm, P < 0.001]. There was no robot abandonment in the robot-assisted group. No revision was required in any of the groups.

Conclusions

Robot-assisted pedicle screw placement is more accurate than free-hand placement. The second-generation TiRobot–assisted thoracolumbar pedicle screw placement is an accurate and safe procedure.
Trial registration retrospectively registered
Literature
1.
go back to reference Roy-Camille R, Saillant G, Mazel C. Internal fixation of the lumbar spine with pedicle screw plating. Clin Orthop Relat Res. 1986;203:7–17.CrossRef Roy-Camille R, Saillant G, Mazel C. Internal fixation of the lumbar spine with pedicle screw plating. Clin Orthop Relat Res. 1986;203:7–17.CrossRef
23.
go back to reference Lee NJ, Buchanan IA, Zuckermann SL, et al. What is the comparison in robot time per screw, radiation exposure, robot abandonment, screw accuracy, and clinical outcomes between percutaneous and open robot-assisted short lumbar fusion? A multicenter, propensity-matched analysis of 310 patients. Spine Phila Pa 1976. 2022;47(1):42–8. https://doi.org/10.1097/BRS.0000000000004132.CrossRefPubMed Lee NJ, Buchanan IA, Zuckermann SL, et al. What is the comparison in robot time per screw, radiation exposure, robot abandonment, screw accuracy, and clinical outcomes between percutaneous and open robot-assisted short lumbar fusion? A multicenter, propensity-matched analysis of 310 patients. Spine Phila Pa 1976. 2022;47(1):42–8. https://​doi.​org/​10.​1097/​BRS.​0000000000004132​.CrossRefPubMed
Metadata
Title
Comparison of accuracy and safety between second-generation TiRobot-assisted and free-hand thoracolumbar pedicle screw placement
Authors
Kai Yan
Qi Zhang
Wei Tian
Publication date
01-12-2022
Publisher
BioMed Central
Published in
BMC Surgery / Issue 1/2022
Electronic ISSN: 1471-2482
DOI
https://doi.org/10.1186/s12893-022-01723-8

Other articles of this Issue 1/2022

BMC Surgery 1/2022 Go to the issue