Published in: BMC Ophthalmology 1/2018

Open Access 01-12-2018 | Research article

Comparison of a new swept-source optical biometer with a partial coherence interferometry

Authors: Hyo Kyung Lee, Mee Kum Kim

Published in: BMC Ophthalmology | Issue 1/2018

Login to get access

Abstract

Background

The purpose of this study is to compare the biometric parameters and intraocular lens (IOL) power calculation by a new swept-source optical coherence tomography (SS-OCT) biometer with those by a partial coherence interferometry (PCI) biometer.

Methods

Medical records of 175 eyes from 175 patients were retrospectively reviewed. One of two monofocal IOLs (Tecnis ZCB00 or Acrysof SA60AT) were implanted in the eyes. Axial length (AL), mean keratometry (Km), J0, J45 and anterior chamber depth (ACD) were compared between PCI and SS-OCT biometers. The refractive mean error (ME) and refractive mean absolute error (MAE) were also compared. Examination failure rates were calculated in each device.

Results

Out of 175 eyes, 150 eyes were successfully examined by both devices. AL was measured slightly shorter when using SS-OCT than PCI biometer, while Km was measured higher (P < .0001, P = .03, respectively, paired t-test). J0, J45 and ACD were not significantly different between two devices. ME and MAE calculated using SRK-T, Hoffer Q, and Haigis formula were not significantly different except MAE calculated with Haigis formula for Tecnis ZCB00 IOLs (P = .03, paired t-test). The examination failure rates were 14.29 and 1.14% when using the PCI and SS-OCT biometers, respectively.

Conclusions

AL and Km don’t seem to be comparable between two devices, while J0, J45, and ACD do. IOL power calculation using SRK-T and Hoffer Q was correlated between the devices. The penetration ability of a SS-OCT biometer is superior.
Literature
1.
go back to reference Fontes BM, Fontes BM, Castro E. Intraocular lens power calculation by measuring axial length with partial optical coherence and ultrasonic biometry. Arq Bras Oftalmol. 2011;74(3):166–70.CrossRef Fontes BM, Fontes BM, Castro E. Intraocular lens power calculation by measuring axial length with partial optical coherence and ultrasonic biometry. Arq Bras Oftalmol. 2011;74(3):166–70.CrossRef
2.
go back to reference Whang WJ, Jung BJ, Oh TH, Byun YS, Joo CK. Comparison of postoperative refractive outcomes: IOLMaster(R) versus immersion ultrasound. Ophthalmic Surg Lasers Imaging. 2012;43(6):496–9.CrossRef Whang WJ, Jung BJ, Oh TH, Byun YS, Joo CK. Comparison of postoperative refractive outcomes: IOLMaster(R) versus immersion ultrasound. Ophthalmic Surg Lasers Imaging. 2012;43(6):496–9.CrossRef
3.
go back to reference Bhatt AB, Schefler AC, Feuer WJ, Yoo SH, Murray TG. Comparison of predictions made by the intraocular lens master and ultrasound biometry. Arch Ophthalmol. 2008;126(7):929–33.CrossRef Bhatt AB, Schefler AC, Feuer WJ, Yoo SH, Murray TG. Comparison of predictions made by the intraocular lens master and ultrasound biometry. Arch Ophthalmol. 2008;126(7):929–33.CrossRef
4.
go back to reference McAlinden C, Wang Q, Pesudovs K, Yang X, Bao F, Yu A, Lin S, Feng Y, Huang J. Axial length measurement failure rates with the IOLMaster and Lenstar LS 900 in eyes with cataract. PLoS One. 2015;10(6):e0128929.CrossRef McAlinden C, Wang Q, Pesudovs K, Yang X, Bao F, Yu A, Lin S, Feng Y, Huang J. Axial length measurement failure rates with the IOLMaster and Lenstar LS 900 in eyes with cataract. PLoS One. 2015;10(6):e0128929.CrossRef
5.
go back to reference Srivannaboon S, Chirapapaisan C, Chonpimai P, Loket S. Clinical comparison of a new swept-source optical coherence tomography-based optical biometer and a time-domain optical coherence tomography-based optical biometer. J Cataract Refract Surg. 2015;41(10):2224–32.CrossRef Srivannaboon S, Chirapapaisan C, Chonpimai P, Loket S. Clinical comparison of a new swept-source optical coherence tomography-based optical biometer and a time-domain optical coherence tomography-based optical biometer. J Cataract Refract Surg. 2015;41(10):2224–32.CrossRef
6.
go back to reference Kunert KS, Peter M, Blum M, Haigis W, Sekundo W, Schutze J, Buehren T. Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography-based biometer versus partial coherence interferometry and optical low-coherence reflectometry. J Cataract Refract Surg. 2016;42(1):76–83.CrossRef Kunert KS, Peter M, Blum M, Haigis W, Sekundo W, Schutze J, Buehren T. Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography-based biometer versus partial coherence interferometry and optical low-coherence reflectometry. J Cataract Refract Surg. 2016;42(1):76–83.CrossRef
7.
go back to reference Akman A, Asena L, Gungor SG. Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500. Br J Ophthalmol. 2016;100(9):1201–5.CrossRef Akman A, Asena L, Gungor SG. Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500. Br J Ophthalmol. 2016;100(9):1201–5.CrossRef
8.
go back to reference Yoo TK, Choi MJ, Lee HK, Seo KY, Kim EK, Kim T-i. Comparison of ocular biometry and refractive outcomes using IOL master 700, IOL master 500, and ultrasound. J Korean Ophthalmol Soc. 2017;58(5):523–9.CrossRef Yoo TK, Choi MJ, Lee HK, Seo KY, Kim EK, Kim T-i. Comparison of ocular biometry and refractive outcomes using IOL master 700, IOL master 500, and ultrasound. J Korean Ophthalmol Soc. 2017;58(5):523–9.CrossRef
9.
go back to reference Olsen T. Theoretical approach to intraocular lens calculation using Gaussian optics. J Cataract Refract Surg. 1987;13(2):141–5.CrossRef Olsen T. Theoretical approach to intraocular lens calculation using Gaussian optics. J Cataract Refract Surg. 1987;13(2):141–5.CrossRef
10.
go back to reference Norrby S. Sources of error in intraocular lens power calculation. J Cataract Refract Surg. 2008;34(3):368–76.CrossRef Norrby S. Sources of error in intraocular lens power calculation. J Cataract Refract Surg. 2008;34(3):368–76.CrossRef
11.
go back to reference Hoffer KJ, Hoffmann PC, Savini G. Comparison of a new optical biometer using swept-source optical coherence tomography and a biometer using optical low-coherence reflectometry. J Cataract Refract Surg. 2016;42(8):1165–72.CrossRef Hoffer KJ, Hoffmann PC, Savini G. Comparison of a new optical biometer using swept-source optical coherence tomography and a biometer using optical low-coherence reflectometry. J Cataract Refract Surg. 2016;42(8):1165–72.CrossRef
12.
go back to reference Karunaratne N. Comparison of the Pentacam equivalent keratometry reading and IOL master keratometry measurement in intraocular lens power calculations. Clin Exp Ophthalmol. 2013;41(9):825–34.CrossRef Karunaratne N. Comparison of the Pentacam equivalent keratometry reading and IOL master keratometry measurement in intraocular lens power calculations. Clin Exp Ophthalmol. 2013;41(9):825–34.CrossRef
13.
go back to reference Arriola-Villalobos P, Almendral-Gomez J, Garzon N, Ruiz-Medrano J, Fernandez-Perez C, Martinez-de-la-Casa JM, Diaz-Valle D. Agreement and clinical comparison between a new swept-source optical coherence tomography-based optical biometer and an optical low-coherence reflectometry biometer. Eye (Lon). 2016. Arriola-Villalobos P, Almendral-Gomez J, Garzon N, Ruiz-Medrano J, Fernandez-Perez C, Martinez-de-la-Casa JM, Diaz-Valle D. Agreement and clinical comparison between a new swept-source optical coherence tomography-based optical biometer and an optical low-coherence reflectometry biometer. Eye (Lon). 2016.
14.
go back to reference Vogel A, Dick HB, Krummenauer F. Reproducibility of optical biometry using partial coherence interferometry : intraobserver and interobserver reliability. J Cataract Refract Surg. 2001;27(12):1961–8.CrossRef Vogel A, Dick HB, Krummenauer F. Reproducibility of optical biometry using partial coherence interferometry : intraobserver and interobserver reliability. J Cataract Refract Surg. 2001;27(12):1961–8.CrossRef
15.
go back to reference Matsuo Y, Sakamoto T, Yamashita T, Tomita M, Shirasawa M, Terasaki H. Comparisons of choroidal thickness of normal eyes obtained by two different spectral-domain OCT instruments and one swept-source OCT instrument. Invest Ophthalmol Vis Sci. 2013;54(12):7630–6.CrossRef Matsuo Y, Sakamoto T, Yamashita T, Tomita M, Shirasawa M, Terasaki H. Comparisons of choroidal thickness of normal eyes obtained by two different spectral-domain OCT instruments and one swept-source OCT instrument. Invest Ophthalmol Vis Sci. 2013;54(12):7630–6.CrossRef
16.
go back to reference Sheard R. Optimising biometry for best outcomes in cataract surgery. Eye (Lond). 2014;28(2):118–25.CrossRef Sheard R. Optimising biometry for best outcomes in cataract surgery. Eye (Lond). 2014;28(2):118–25.CrossRef
17.
go back to reference Holladay JT, Prager TC, Chandler TY, Musgrove KH, Lewis JW, Ruiz RS. A three-part system for refining intraocular lens power calculations. J Cataract Refract Surg. 1988;14(1):17–24.CrossRef Holladay JT, Prager TC, Chandler TY, Musgrove KH, Lewis JW, Ruiz RS. A three-part system for refining intraocular lens power calculations. J Cataract Refract Surg. 1988;14(1):17–24.CrossRef
Metadata
Title
Comparison of a new swept-source optical biometer with a partial coherence interferometry
Authors
Hyo Kyung Lee
Mee Kum Kim
Publication date
01-12-2018
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2018
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-018-0936-6

Other articles of this Issue 1/2018

BMC Ophthalmology 1/2018 Go to the issue