Skip to main content
Top
Published in: BMC Gastroenterology 1/2011

Open Access 01-12-2011 | Research article

Comparative cost-efficiency of the EVOTECH endoscope cleaner and reprocessor versus manual cleaning plus automated endoscope reprocessing in a real-world Canadian hospital endoscopy setting

Authors: Lindy Forte, Cynthia Shum

Published in: BMC Gastroenterology | Issue 1/2011

Login to get access

Abstract

Background

Reprocessing of endoscopes generally requires labour-intensive manual cleaning followed by high-level disinfection in an automated endoscope reprocessor (AER). EVOTECH Endoscope Cleaner and Reprocessor (ECR) is approved for fully automated cleaning and disinfection whereas AERs require manual cleaning prior to the high-level disinfection procedure. The purpose of this economic evaluation was to determine the cost-efficiency of the ECR versus AER methods of endoscopy reprocessing in an actual practice setting.

Methods

A time and motion study was conducted at a Canadian hospital to collect data on the personnel resources and consumable supplies costs associated with the use of EVOTECH ECR versus manual cleaning followed by AER with Medivators DSD-201. Reprocessing of all endoscopes was observed and timed for both reprocessor types over three days. Laboratory staff members were interviewed regarding the consumption and cost of all disposable supplies and equipment. Exact Wilcoxon rank sum test was used for assessing differences in total cycle reprocessing time.

Results

Endoscope reprocessing was significantly shorter with the ECR than with manual cleaning followed by AER. The differences in median time were 12.46 minutes per colonoscope (p < 0.0001), 6.31 minutes per gastroscope (p < 0.0001), and 5.66 minutes per bronchoscope (p = 0.0040). Almost 2 hours of direct labour time was saved daily with the ECR. The total per cycle cost of consumables and labour for maintenance was slightly higher for EVOTECH ECR versus manual cleaning followed by AER ($8.91 versus $8.31, respectively). Including the cost of direct labour time consumed in reprocessing scopes, the per cycle and annual costs of using the EVOTECH ECR was less than the cost of manual cleaning followed by AER disinfection ($11.50 versus $11.88).

Conclusions

The EVOTECH ECR was more efficient and less costly to use for the reprocessing of endoscopes than manual cleaning followed by AER disinfection. Although the cost of consumable supplies required to reprocess endoscopes with EVOTECH ECR was slightly higher, the value of the labour time saved with EVOTECH ECR more than offset the additional consumables cost. The increased efficiency with EVOTECH ECR could lead to even further cost-savings by shifting endoscopy laboratory personnel responsibilities but further study is required.
Literature
1.
go back to reference Seeff LC, Richards TB, Shapiro JA, Nadel MR, Manninen DL, Given LS, Dong FB, Winges LD, McKenna MT: How many endoscopies are performed for colorectal cancer screening? Results from CDC's survey of endoscopic capacity. Gastroenterol. 2004, 127: 1670-1677. 10.1053/j.gastro.2004.09.051.CrossRef Seeff LC, Richards TB, Shapiro JA, Nadel MR, Manninen DL, Given LS, Dong FB, Winges LD, McKenna MT: How many endoscopies are performed for colorectal cancer screening? Results from CDC's survey of endoscopic capacity. Gastroenterol. 2004, 127: 1670-1677. 10.1053/j.gastro.2004.09.051.CrossRef
2.
go back to reference Ernst A, Silvestri GA, Johnstone D: American College of Chest Physicians. Interventional pulmonary procedures: Guidelines from the American College of Chest Physicians. Chest. 2003, 123 (5): 1693-1717. 10.1378/chest.123.5.1693.CrossRefPubMed Ernst A, Silvestri GA, Johnstone D: American College of Chest Physicians. Interventional pulmonary procedures: Guidelines from the American College of Chest Physicians. Chest. 2003, 123 (5): 1693-1717. 10.1378/chest.123.5.1693.CrossRefPubMed
3.
go back to reference Society of Gastroenterology Nurses and Associates: Guideline for the use of high-level disinfectants and sterilants for reprocessing of flexible gastrointestinal endoscopes. Gastroenterol Nurs. 2003, 27: 198-206. Society of Gastroenterology Nurses and Associates: Guideline for the use of high-level disinfectants and sterilants for reprocessing of flexible gastrointestinal endoscopes. Gastroenterol Nurs. 2003, 27: 198-206.
4.
go back to reference Society of Gastroenterology Nurses and Associates: Standards of infection control and reprocessing of flexible gastrointestinal endoscopes. Gastroenterol Nurs. 2006, 29: 42-48. 10.1097/00001610-200601000-00007.CrossRef Society of Gastroenterology Nurses and Associates: Standards of infection control and reprocessing of flexible gastrointestinal endoscopes. Gastroenterol Nurs. 2006, 29: 42-48. 10.1097/00001610-200601000-00007.CrossRef
5.
go back to reference Mehta AC, Prakash UBS, Garland R, et al: Prevention of flexible bronchoscopy-associated infection. Chest. 2005, 128: 1742-1755. 10.1378/chest.128.3.1742.CrossRefPubMed Mehta AC, Prakash UBS, Garland R, et al: Prevention of flexible bronchoscopy-associated infection. Chest. 2005, 128: 1742-1755. 10.1378/chest.128.3.1742.CrossRefPubMed
6.
go back to reference Spach DH, Silverstein FE, Stamm WE: Transmission of infection by gastrointestinal endoscopy and bronchoscopy. Ann Intern Med. 1993, 118: 117-128.CrossRefPubMed Spach DH, Silverstein FE, Stamm WE: Transmission of infection by gastrointestinal endoscopy and bronchoscopy. Ann Intern Med. 1993, 118: 117-128.CrossRefPubMed
7.
go back to reference Alfa MJ, Olson N, Degagne P, Jackson M: A survey of reprocessing methods, residual viable bioburden, and soil levels in patient ready endoscopic retrograde choliangiopancreatography duodenoscopes used in Canadian centers. Infect Control Hosp Epidemiol. 2002, 23: 198-206. 10.1086/502035.CrossRefPubMed Alfa MJ, Olson N, Degagne P, Jackson M: A survey of reprocessing methods, residual viable bioburden, and soil levels in patient ready endoscopic retrograde choliangiopancreatography duodenoscopes used in Canadian centers. Infect Control Hosp Epidemiol. 2002, 23: 198-206. 10.1086/502035.CrossRefPubMed
8.
go back to reference Moses FM, Lee JS: Current GI endoscope disinfection and QA practices. Dig Dis Sci. 2004, 49: 1791-1797. 10.1007/s10620-004-9572-5.CrossRefPubMed Moses FM, Lee JS: Current GI endoscope disinfection and QA practices. Dig Dis Sci. 2004, 49: 1791-1797. 10.1007/s10620-004-9572-5.CrossRefPubMed
9.
go back to reference Bronowicki JP, Venard V, Botté C, Monhoven N, Gastin I, Choné L, Hudziak H, Rihn B, Delanoë C, LeFaou A, Bigard MA, Gaucher P: Patient-to-patient transmission of hepatitis C virus during colonoscopy. N Engl J Med. 1997, 337: 237-40. 10.1056/NEJM199707243370404.CrossRefPubMed Bronowicki JP, Venard V, Botté C, Monhoven N, Gastin I, Choné L, Hudziak H, Rihn B, Delanoë C, LeFaou A, Bigard MA, Gaucher P: Patient-to-patient transmission of hepatitis C virus during colonoscopy. N Engl J Med. 1997, 337: 237-40. 10.1056/NEJM199707243370404.CrossRefPubMed
10.
go back to reference Agerton T, Valway S, Gore B, Pozsik C, Plikaytis B, Woodley C, Onorato I: Transmission of a highly drug-resistant strain (strain W1) of Mycobacterium tuberculosis. Community outbreak and nosocomial transmission via a contaminated bronchoscope. JAMA. 1997, 278: 1073-7. Agerton T, Valway S, Gore B, Pozsik C, Plikaytis B, Woodley C, Onorato I: Transmission of a highly drug-resistant strain (strain W1) of Mycobacterium tuberculosis. Community outbreak and nosocomial transmission via a contaminated bronchoscope. JAMA. 1997, 278: 1073-7.
12.
go back to reference Alfa MJ, Degagne P, Olson N, Fatima I: Correction: EVOTECH® endoscope cleaner and reprocessor (ECR) simulated-use and clinical-use evaluation of cleaning efficacy. BMC Infectious Diseases. 2010, 10: 291-10.1186/1471-2334-10-291.CrossRefPubMedCentral Alfa MJ, Degagne P, Olson N, Fatima I: Correction: EVOTECH® endoscope cleaner and reprocessor (ECR) simulated-use and clinical-use evaluation of cleaning efficacy. BMC Infectious Diseases. 2010, 10: 291-10.1186/1471-2334-10-291.CrossRefPubMedCentral
13.
go back to reference Advanced Sterilization Products: The EVOTECH™ Endoscope Cleaner and Reprocessor Product Monograph. December 2008 Advanced Sterilization Products: The EVOTECH™ Endoscope Cleaner and Reprocessor Product Monograph. December 2008
14.
go back to reference Minntech Corporation Business Group: Medivators™ Reprocessing Systems. User Manual. Single-shot Generator Kit for DSD-201™ DSD-91E™ Endoscope Reprocessors. 50090-946 Revision A. 2003 Minntech Corporation Business Group: Medivators™ Reprocessing Systems. User Manual. Single-shot Generator Kit for DSD-201™ DSD-91E™ Endoscope Reprocessors. 50090-946 Revision A. 2003
16.
go back to reference Kozarek RA, Bronowicki J-P, Bigard MA: Transmission of Hepatitis C Virus during Colonoscopy. New Engl J Med. 1997, 337: 1848-1849.CrossRefPubMed Kozarek RA, Bronowicki J-P, Bigard MA: Transmission of Hepatitis C Virus during Colonoscopy. New Engl J Med. 1997, 337: 1848-1849.CrossRefPubMed
18.
go back to reference Ofstead CL, Wetzler HP, Snyder AK, Horton RA: Endoscope Reprocessing Methods. A Prospective Study on the Impact of Human Factors and Automation. Gastroenterology Nursing. 2010, 33 (54): Ofstead CL, Wetzler HP, Snyder AK, Horton RA: Endoscope Reprocessing Methods. A Prospective Study on the Impact of Human Factors and Automation. Gastroenterology Nursing. 2010, 33 (54):
19.
go back to reference Alfa MJ, Olson N, Degagne P: Automated washing with the Reliance Endoscope Processing System and its equivalence to optimal manual cleaning. Am J Infect Control. 2006, 34: 561-570. 10.1016/j.ajic.2006.01.010.CrossRefPubMed Alfa MJ, Olson N, Degagne P: Automated washing with the Reliance Endoscope Processing System and its equivalence to optimal manual cleaning. Am J Infect Control. 2006, 34: 561-570. 10.1016/j.ajic.2006.01.010.CrossRefPubMed
Metadata
Title
Comparative cost-efficiency of the EVOTECH endoscope cleaner and reprocessor versus manual cleaning plus automated endoscope reprocessing in a real-world Canadian hospital endoscopy setting
Authors
Lindy Forte
Cynthia Shum
Publication date
01-12-2011
Publisher
BioMed Central
Published in
BMC Gastroenterology / Issue 1/2011
Electronic ISSN: 1471-230X
DOI
https://doi.org/10.1186/1471-230X-11-105

Other articles of this Issue 1/2011

BMC Gastroenterology 1/2011 Go to the issue
Obesity Clinical Trial Summary

At a glance: The STEP trials

A round-up of the STEP phase 3 clinical trials evaluating semaglutide for weight loss in people with overweight or obesity.

Developed by: Springer Medicine

Highlights from the ACC 2024 Congress

Year in Review: Pediatric cardiology

Watch Dr. Anne Marie Valente present the last year's highlights in pediatric and congenital heart disease in the official ACC.24 Year in Review session.

Year in Review: Pulmonary vascular disease

The last year's highlights in pulmonary vascular disease are presented by Dr. Jane Leopold in this official video from ACC.24.

Year in Review: Valvular heart disease

Watch Prof. William Zoghbi present the last year's highlights in valvular heart disease from the official ACC.24 Year in Review session.

Year in Review: Heart failure and cardiomyopathies

Watch this official video from ACC.24. Dr. Biykem Bozkurt discuss last year's major advances in heart failure and cardiomyopathies.