01-07-2017 | Original Article

Clinical evaluation of TOF versus non-TOF on PET artifacts in simultaneous PET/MR: a dual centre experience

Authors: Edwin E. G. W. ter Voert, Patrick Veit-Haibach, Sangtae Ahn, Florian Wiesinger, M. Mehdi Khalighi, Craig S. Levin, Andrei H. Iagaru, Greg Zaharchuk, Martin Huellner, Gaspar Delso

Published in: European Journal of Nuclear Medicine and Molecular Imaging | Issue 7/2017

Login to get access

Abstract

Purpose

Our objective was to determine clinically the value of time-of-flight (TOF) information in reducing PET artifacts and improving PET image quality and accuracy in simultaneous TOF PET/MR scanning.

Methods

A total 65 patients who underwent a comparative scan in a simultaneous TOF PET/MR scanner were included. TOF and non-TOF PET images were reconstructed, clinically examined, compared and scored. PET imaging artifacts were categorized as large or small implant-related artifacts, as dental implant-related artifacts, and as implant-unrelated artifacts. Differences in image quality, especially those related to (implant) artifacts, were assessed using a scale ranging from 0 (no artifact) to 4 (severe artifact).

Results

A total of 87 image artifacts were found and evaluated. Four patients had large and eight patients small implant-related artifacts, 27 patients had dental implants/fillings, and 48 patients had implant-unrelated artifacts. The average score was 1.14 ± 0.82 for non-TOF PET images and 0.53 ± 0.66 for TOF images (p < 0.01) indicating that artifacts were less noticeable when TOF information was included.

Conclusion

Our study indicates that PET image artifacts are significantly mitigated with integration of TOF information in simultaneous PET/MR. The impact is predominantly seen in patients with significant artifacts due to metal implants.
Appendix
Available only for authorised users
Literature
1.
7.
go back to reference Levin C, Glover G, Deller T, McDaniel D, Peterson W, Maramraju SH. Prototype time-of-flight PET ring integrated with a 3T MRI system for simultaneous whole-body PET/MR imaging. JNM Meeting Abstracts. 2013;54:148. Levin C, Glover G, Deller T, McDaniel D, Peterson W, Maramraju SH. Prototype time-of-flight PET ring integrated with a 3T MRI system for simultaneous whole-body PET/MR imaging. JNM Meeting Abstracts. 2013;54:148.
19.
go back to reference Delso G, Khalighi M, Ter Voert E, Barbosa F, Sekine T, Hullner M, et al. Effect of time-of-flight information on PET/MR reconstruction artifacts: comparison of free-breathing versus breath-hold MR-based attenuation correction. Radiology. 2017:282;229–235. doi:10.1148/radiol.2016152509.CrossRefPubMed Delso G, Khalighi M, Ter Voert E, Barbosa F, Sekine T, Hullner M, et al. Effect of time-of-flight information on PET/MR reconstruction artifacts: comparison of free-breathing versus breath-hold MR-based attenuation correction. Radiology. 2017:282;229–235. doi:10.​1148/​radiol.​2016152509.CrossRefPubMed
23.
go back to reference Wollenweber SD, Ambwani S, Delso G, Lonn AHR, Mullick R, Wiesinger F, et al. Evaluation of an atlas-based PET head attenuation correction using PET/CT & MR patient data. IEEE Trans Nucl Sci. 2013;60:3383–3390. doi:10.1109/TNS.2013.2273417.CrossRef Wollenweber SD, Ambwani S, Delso G, Lonn AHR, Mullick R, Wiesinger F, et al. Evaluation of an atlas-based PET head attenuation correction using PET/CT & MR patient data. IEEE Trans Nucl Sci. 2013;60:3383–3390. doi:10.​1109/​TNS.​2013.​2273417.CrossRef
24.
go back to reference Martinez-Moller A, Souvatzoglou M, Delso G, Bundschuh RA, Chefd’hotel C, Ziegler SI, et al. Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data. J Nucl Med. 2009;50:520–526. doi:10.2967/jnumed.108.054726.CrossRefPubMed Martinez-Moller A, Souvatzoglou M, Delso G, Bundschuh RA, Chefd’hotel C, Ziegler SI, et al. Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data. J Nucl Med. 2009;50:520–526. doi:10.​2967/​jnumed.​108.​054726.CrossRefPubMed
25.
go back to reference Wollenweber SD, Ambwani S, Lonn AHR, Shanbhag DD, Thiruvenkadam S, Kaushik S, et al. Comparison of 4-class and continuous fat/water methods for whole-body, MR-based PET attenuation correction. IEEE Trans Nucl Sci. 2013;60:3391–3398. doi:10.1109/TNS.2013.2278759.CrossRef Wollenweber SD, Ambwani S, Lonn AHR, Shanbhag DD, Thiruvenkadam S, Kaushik S, et al. Comparison of 4-class and continuous fat/water methods for whole-body, MR-based PET attenuation correction. IEEE Trans Nucl Sci. 2013;60:3391–3398. doi:10.​1109/​TNS.​2013.​2278759.CrossRef
29.
go back to reference Sekine T, Ter Voert EE, Warnock G, Buck A, Huellner MW, Veit-Haibach P, et al. Clinical evaluation of zero-echo-time attenuation correction for brain 18F-FDG PET/MRI: comparison with atlas attenuation correction. J Nucl Med. 2016;57:1927–1932. doi:10.2967/jnumed.116.175398.CrossRefPubMed Sekine T, Ter Voert EE, Warnock G, Buck A, Huellner MW, Veit-Haibach P, et al. Clinical evaluation of zero-echo-time attenuation correction for brain 18F-FDG PET/MRI: comparison with atlas attenuation correction. J Nucl Med. 2016;57:1927–1932. doi:10.​2967/​jnumed.​116.​175398.CrossRefPubMed
30.
go back to reference Nuyts J, Dupont P, Stroobants S, Benninck R, Mortelmans L, Suetens P. Simultaneous maximum a posteriori reconstruction of attenuation and activity distributions from emission sinograms. IEEE Trans Med Imaging. 1999;18:393–403. doi:10.1109/42.774167.CrossRefPubMed Nuyts J, Dupont P, Stroobants S, Benninck R, Mortelmans L, Suetens P. Simultaneous maximum a posteriori reconstruction of attenuation and activity distributions from emission sinograms. IEEE Trans Med Imaging. 1999;18:393–403. doi:10.​1109/​42.​774167.CrossRefPubMed
33.
go back to reference Brendle C, Schmidt H, Oergel A, Bezrukov I, Mueller M, Schraml C, et al. Segmentation-based attenuation correction in positron emission tomography/magnetic resonance: erroneous tissue identification and its impact on positron emission tomography interpretation. Invest Radiol. 2015;50:339–346. doi:10.1097/rli.0000000000000131.CrossRefPubMed Brendle C, Schmidt H, Oergel A, Bezrukov I, Mueller M, Schraml C, et al. Segmentation-based attenuation correction in positron emission tomography/magnetic resonance: erroneous tissue identification and its impact on positron emission tomography interpretation. Invest Radiol. 2015;50:339–346. doi:10.​1097/​rli.​0000000000000131​.CrossRefPubMed
35.
go back to reference Davison H, ter Voert EE, de Galiza BF, Veit-Haibach P, Delso G. Incorporation of time-of-flight information reduces metal artifacts in simultaneous positron emission tomography/magnetic resonance imaging: a simulation study. Invest Radiol. 2015;50:423–429. doi:10.1097/RLI.0000000000000146.CrossRefPubMed Davison H, ter Voert EE, de Galiza BF, Veit-Haibach P, Delso G. Incorporation of time-of-flight information reduces metal artifacts in simultaneous positron emission tomography/magnetic resonance imaging: a simulation study. Invest Radiol. 2015;50:423–429. doi:10.​1097/​RLI.​0000000000000146​.CrossRefPubMed
37.
go back to reference Bai C, Kinahan PE, Brasse D, Comtat C, Townsend DW, Meltzer CC, et al. An analytic study of the effects of attenuation on tumor detection in whole-body PET oncology imaging. J Nucl Med. 2003;44:1855–1861.PubMed Bai C, Kinahan PE, Brasse D, Comtat C, Townsend DW, Meltzer CC, et al. An analytic study of the effects of attenuation on tumor detection in whole-body PET oncology imaging. J Nucl Med. 2003;44:1855–1861.PubMed
38.
go back to reference Turkington TG, Wilson JM. Attenuation artifacts and time-of-flight PET. IEEE Nuclear Science Symposium Conference Record (NSS/MIC), New York, NY: IEEE; 2009. p. 2997–2999. Turkington TG, Wilson JM. Attenuation artifacts and time-of-flight PET. IEEE Nuclear Science Symposium Conference Record (NSS/MIC), New York, NY: IEEE; 2009. p. 2997–2999.
40.
go back to reference Antoch G, Freudenberg LS, Beyer T, Bockisch A, Debatin JF. To enhance or not to enhance? 18F-FDG and CT contrast agents in dual-modality 18F-FDG PET/CT. J Nucl Med. 2004;45:56S–65S.PubMed Antoch G, Freudenberg LS, Beyer T, Bockisch A, Debatin JF. To enhance or not to enhance? 18F-FDG and CT contrast agents in dual-modality 18F-FDG PET/CT. J Nucl Med. 2004;45:56S–65S.PubMed
41.
go back to reference Zeimpekis KG, Barbosa F, Hullner M, ter Voert E, Davison H, Veit-Haibach P, et al. Clinical evaluation of PET image quality as a function of acquisition time in a new TOF-PET/MRI compared to TOF-PET/CT – initial results. Mol Imaging Biol. 2015;17:735–44. doi:10.1007/s11307-015-0845-5.CrossRefPubMed Zeimpekis KG, Barbosa F, Hullner M, ter Voert E, Davison H, Veit-Haibach P, et al. Clinical evaluation of PET image quality as a function of acquisition time in a new TOF-PET/MRI compared to TOF-PET/CT – initial results. Mol Imaging Biol. 2015;17:735–44. doi:10.​1007/​s11307-015-0845-5.CrossRefPubMed
49.
go back to reference Ladefoged C, Andersen F, Keller S, Löfgren J, Hansen A, Holm S, et al. PET/MR imaging of the pelvis in the presence of endoprostheses: reducing image artifacts and increasing accuracy through inpainting. Eur J Nucl Med Mol Imaging. 2013;40:594–601. doi:10.1007/s00259-012-2316-4.CrossRefPubMed Ladefoged C, Andersen F, Keller S, Löfgren J, Hansen A, Holm S, et al. PET/MR imaging of the pelvis in the presence of endoprostheses: reducing image artifacts and increasing accuracy through inpainting. Eur J Nucl Med Mol Imaging. 2013;40:594–601. doi:10.​1007/​s00259-012-2316-4.CrossRefPubMed
50.
go back to reference Schramm G, Maus J, Hofheinz F, Petr J, Lougovski A, Beuthien-Baumann B, et al. Evaluation and automatic correction of metal-implant-induced artifacts in MR-based attenuation correction in whole-body PET/MR imaging. Phys Med Biol. 2014;59:2713–2726. doi:10.1088/0031-9155/59/11/2713.CrossRefPubMed Schramm G, Maus J, Hofheinz F, Petr J, Lougovski A, Beuthien-Baumann B, et al. Evaluation and automatic correction of metal-implant-induced artifacts in MR-based attenuation correction in whole-body PET/MR imaging. Phys Med Biol. 2014;59:2713–2726. doi:10.​1088/​0031-9155/​59/​11/​2713.CrossRefPubMed
52.
go back to reference den Harder JC, van Yperen GH, Blume UA, Bos C. Off-resonance suppression for multispectral MR imaging near metallic implants. Magn Reson Med. 2015;73:233–243. doi:10.1002/mrm.25126.CrossRef den Harder JC, van Yperen GH, Blume UA, Bos C. Off-resonance suppression for multispectral MR imaging near metallic implants. Magn Reson Med. 2015;73:233–243. doi:10.​1002/​mrm.​25126.CrossRef
Metadata
Title
Clinical evaluation of TOF versus non-TOF on PET artifacts in simultaneous PET/MR: a dual centre experience
Authors
Edwin E. G. W. ter Voert
Patrick Veit-Haibach
Sangtae Ahn
Florian Wiesinger
M. Mehdi Khalighi
Craig S. Levin
Andrei H. Iagaru
Greg Zaharchuk
Martin Huellner
Gaspar Delso
Publication date
01-07-2017
Publisher
Springer Berlin Heidelberg
Published in
European Journal of Nuclear Medicine and Molecular Imaging / Issue 7/2017
Print ISSN: 1619-7070
Electronic ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-017-3619-2

Other articles of this Issue 7/2017

European Journal of Nuclear Medicine and Molecular Imaging 7/2017 Go to the issue