Published in: Osteoporosis International 10/2016

01-10-2016 | Original Article

Clinical cone beam computed tomography compared to high-resolution peripheral computed tomography in the assessment of distal radius bone

Authors: C. de Charry, S. Boutroy, R. Ellouz, F. Duboeuf, R. Chapurlat, H. Follet, J. B. Pialat

Published in: Osteoporosis International | Issue 10/2016

Login to get access

Abstract

Summary

Clinical cone beam computed tomography (CBCT) was compared to high-resolution peripheral quantitative computed tomography (HR-pQCT) for the assessment of ex vivo radii. Strong correlations were found for geometry, volumetric density, and trabecular structure. Using CBCT, bone architecture assessment was feasible but compared to HR-pQCT, trabecular parameters were overestimated whereas cortical ones were underestimated.

Introduction

HR-pQCT is the most widely used technique to assess bone microarchitecture in vivo. Yet, this technology has been only applicable at peripheral sites, in only few research centers. Clinical CBCT is more widely available but quantitative assessment of the bone structure is usually not performed. We aimed to compare the assessment of bone structure with CBCT (NewTom 5G, QR, Verona, Italy) and HR-pQCT (XtremeCT, Scanco Medical AG, Brüttisellen, Switzerland).

Methods

Twenty-four distal radius specimens were scanned with these two devices with a reconstructed voxel size of 75 μm for Newtom 5G and 82 μm for XtremeCT, respectively. A rescaling-registration scheme was used to define the common volume of interest. Cortical and trabecular compartments were separated using a semiautomated double contouring method. Density and microstructure were assessed with the HR-pQCT software on both modality images.

Results

Strong correlations were found for geometry parameters (r = 0.98–0.99), volumetric density (r = 0.91–0.99), and trabecular structure (r = 0.94–0.99), all p < 0.001. Correlations were lower for cortical microstructure (r = 0.80–0.89), p < 0.001. However, absolute differences were observed between modalities for all parameters, with an overestimation of the trabecular structure (trabecular number, 1.62 ± 0.37 vs. 1.47 ± 0.36 mm−1) and an underestimation of the cortical microstructure (cortical porosity, 3.3 ± 1.3 vs. 4.4 ± 1.4 %) assessed on CBCT images compared to HR-pQCT images.

Conclusions

Clinical CBCT devices are able to analyze large portions of distal bones with good spatial resolution and limited irradiation. However, compared to dedicated HR-pQCT, the assessment of microarchitecture by NewTom 5G dental CBCT showed some discrepancies, for density measurements mainly. Further technical developments are required to reach optimal assessment of bone characteristics.
Literature
2.
go back to reference Marshall D, Johnell O, Wedel H (1996) Meta-analysis of how well measures of bone mineral density predict occurrence of osteoporotic fractures. BMJ 312:1254–1259CrossRefPubMedPubMedCentral Marshall D, Johnell O, Wedel H (1996) Meta-analysis of how well measures of bone mineral density predict occurrence of osteoporotic fractures. BMJ 312:1254–1259CrossRefPubMedPubMedCentral
3.
go back to reference Nguyen TV, Center JR, Sambrook PN, Eisman JA (2001) Risk factors for proximal humerus, forearm, and wrist fractures in elderly men and women the Dubbo Osteoporosis Epidemiology Study. Am J Epidemiol 153:587–595. doi:10.1093/aje/153.6.587 CrossRefPubMed Nguyen TV, Center JR, Sambrook PN, Eisman JA (2001) Risk factors for proximal humerus, forearm, and wrist fractures in elderly men and women the Dubbo Osteoporosis Epidemiology Study. Am J Epidemiol 153:587–595. doi:10.​1093/​aje/​153.​6.​587 CrossRefPubMed
4.
go back to reference Sornay-Rendu E, Boutroy S, Munoz F, Delmas PD (2007) Alterations of cortical and trabecular architecture are associated with fractures in postmenopausal women, partially independent of decreased BMD measured by DXA: the OFELY study. J Bone Miner Res Off J Am Soc Bone Miner Res 22:425–433. doi:10.1359/jbmr.061206 CrossRef Sornay-Rendu E, Boutroy S, Munoz F, Delmas PD (2007) Alterations of cortical and trabecular architecture are associated with fractures in postmenopausal women, partially independent of decreased BMD measured by DXA: the OFELY study. J Bone Miner Res Off J Am Soc Bone Miner Res 22:425–433. doi:10.​1359/​jbmr.​061206 CrossRef
5.
go back to reference Hildebrand T, Laib A, Müller R et al (1999) Direct three-dimensional morphometric analysis of human cancellous bone: microstructural data from spine, femur, iliac crest, and calcaneus. J Bone Miner Res Off J Am Soc Bone Miner Res 14:1167–1174. doi:10.1359/jbmr.1999.14.7.1167 CrossRef Hildebrand T, Laib A, Müller R et al (1999) Direct three-dimensional morphometric analysis of human cancellous bone: microstructural data from spine, femur, iliac crest, and calcaneus. J Bone Miner Res Off J Am Soc Bone Miner Res 14:1167–1174. doi:10.​1359/​jbmr.​1999.​14.​7.​1167 CrossRef
6.
7.
go back to reference Ashe MC, Khan KM, Kontulainen SA et al (2006) Accuracy of pQCT for evaluating the aged human radius: an ashing, histomorphometry and failure load investigation. Osteoporos Int J Establ Result Coop Eur Found Osteoporos Natl Osteoporos Found USA 17:1241–1251. doi:10.1007/s00198-006-0110-5 CrossRef Ashe MC, Khan KM, Kontulainen SA et al (2006) Accuracy of pQCT for evaluating the aged human radius: an ashing, histomorphometry and failure load investigation. Osteoporos Int J Establ Result Coop Eur Found Osteoporos Natl Osteoporos Found USA 17:1241–1251. doi:10.​1007/​s00198-006-0110-5 CrossRef
8.
go back to reference Stauber M, Rapillard L, van Lenthe GH et al (2006) Importance of individual rods and plates in the assessment of bone quality and their contribution to bone stiffness. J Bone Miner Res Off J Am Soc Bone Miner Res 21:586–595. doi:10.1359/jbmr.060102 CrossRef Stauber M, Rapillard L, van Lenthe GH et al (2006) Importance of individual rods and plates in the assessment of bone quality and their contribution to bone stiffness. J Bone Miner Res Off J Am Soc Bone Miner Res 21:586–595. doi:10.​1359/​jbmr.​060102 CrossRef
9.
go back to reference Boutroy S, Bouxsein ML, Munoz F, Delmas PD (2005) In vivo assessment of trabecular bone microarchitecture by high-resolution peripheral quantitative computed tomography. J Clin Endocrinol Metab 90:6508–6515. doi:10.1210/jc.2005-1258 CrossRefPubMed Boutroy S, Bouxsein ML, Munoz F, Delmas PD (2005) In vivo assessment of trabecular bone microarchitecture by high-resolution peripheral quantitative computed tomography. J Clin Endocrinol Metab 90:6508–6515. doi:10.​1210/​jc.​2005-1258 CrossRefPubMed
10.
go back to reference Sornay-Rendu E, Cabrera-Bravo J-L, Boutroy S et al (2009) Severity of vertebral fractures is associated with alterations of cortical architecture in postmenopausal women. J Bone Miner Res Off J Am Soc Bone Miner Res 24:737–743. doi:10.1359/jbmr.081223 CrossRef Sornay-Rendu E, Cabrera-Bravo J-L, Boutroy S et al (2009) Severity of vertebral fractures is associated with alterations of cortical architecture in postmenopausal women. J Bone Miner Res Off J Am Soc Bone Miner Res 24:737–743. doi:10.​1359/​jbmr.​081223 CrossRef
16.
go back to reference Jaffray DA, Siewerdsen JH, Wong JW, Martinez AA (2002) Flat-panel cone-beam computed tomography for image-guided radiation therapy. Int J Radiat Oncol Biol Phys 53:1337–1349CrossRefPubMed Jaffray DA, Siewerdsen JH, Wong JW, Martinez AA (2002) Flat-panel cone-beam computed tomography for image-guided radiation therapy. Int J Radiat Oncol Biol Phys 53:1337–1349CrossRefPubMed
18.
go back to reference Hashimoto K, Kawashima S, Kameoka S et al (2007) Comparison of image validity between cone beam computed tomography for dental use and multidetector row helical computed tomography. Dentomaxillofacial Radiol 36:465–471. doi:10.1259/dmfr/22818643 CrossRef Hashimoto K, Kawashima S, Kameoka S et al (2007) Comparison of image validity between cone beam computed tomography for dental use and multidetector row helical computed tomography. Dentomaxillofacial Radiol 36:465–471. doi:10.​1259/​dmfr/​22818643 CrossRef
20.
go back to reference Ludlow JB, Davies-Ludlow LE, Brooks SL (2003) Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus DS panoramic unit. Dentomaxillofac Radiol 32:229–234CrossRefPubMed Ludlow JB, Davies-Ludlow LE, Brooks SL (2003) Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus DS panoramic unit. Dentomaxillofac Radiol 32:229–234CrossRefPubMed
22.
go back to reference De Cock J, Mermuys K, Goubau J et al (2012) Cone-beam computed tomography: a new low dose, high resolution imaging technique of the wrist, presentation of three cases with technique. Skelet Radiol 41:93–96. doi:10.1007/s00256-011-1198-z CrossRef De Cock J, Mermuys K, Goubau J et al (2012) Cone-beam computed tomography: a new low dose, high resolution imaging technique of the wrist, presentation of three cases with technique. Skelet Radiol 41:93–96. doi:10.​1007/​s00256-011-1198-z CrossRef
23.
28.
go back to reference Ellouz R, Chapurlat R, van Rietbergen B et al (2014) Challenges in longitudinal measurements with HR-pQCT: evaluation of a 3D registration method to improve bone microarchitecture and strength measurement reproducibility. Bone 63:147–157. doi:10.1016/j.bone.2014.03.001 CrossRefPubMed Ellouz R, Chapurlat R, van Rietbergen B et al (2014) Challenges in longitudinal measurements with HR-pQCT: evaluation of a 3D registration method to improve bone microarchitecture and strength measurement reproducibility. Bone 63:147–157. doi:10.​1016/​j.​bone.​2014.​03.​001 CrossRefPubMed
29.
30.
go back to reference Laib A, Häuselmann HJ, Rüegsegger P (1998) In vivo high resolution 3D-QCT of the human forearm. Technol Heal Care Off J Eur Soc Eng Med 6:329–337 Laib A, Häuselmann HJ, Rüegsegger P (1998) In vivo high resolution 3D-QCT of the human forearm. Technol Heal Care Off J Eur Soc Eng Med 6:329–337
32.
go back to reference Krause M, Museyko O, Breer S et al (2014) Accuracy of trabecular structure by HR-pQCT compared to gold standard μCT in the radius and tibia of patients with osteoporosis and long-term bisphosphonate therapy. Osteoporos Int 25:1595–1606. doi:10.1007/s00198-014-2650-4 CrossRefPubMed Krause M, Museyko O, Breer S et al (2014) Accuracy of trabecular structure by HR-pQCT compared to gold standard μCT in the radius and tibia of patients with osteoporosis and long-term bisphosphonate therapy. Osteoporos Int 25:1595–1606. doi:10.​1007/​s00198-014-2650-4 CrossRefPubMed
37.
go back to reference Boutroy S, Van Rietbergen B, Sornay-Rendu E et al (2008) Finite element analysis based on in vivo HR-pQCT images of the distal radius is associated with wrist fracture in postmenopausal women. J Bone Miner Res Off J Am Soc Bone Miner Res 23:392–399. doi:10.1359/jbmr.071108 CrossRef Boutroy S, Van Rietbergen B, Sornay-Rendu E et al (2008) Finite element analysis based on in vivo HR-pQCT images of the distal radius is associated with wrist fracture in postmenopausal women. J Bone Miner Res Off J Am Soc Bone Miner Res 23:392–399. doi:10.​1359/​jbmr.​071108 CrossRef
38.
go back to reference Muller ME, Webber CE, Bouxsein ML (2003) Predicting the failure load of the distal radius. Osteoporos Int J Establ Result Coop Eur Found Osteoporos Natl Osteoporos Found USA 14:345–352. doi:10.1007/s00198-003-1380-9 CrossRef Muller ME, Webber CE, Bouxsein ML (2003) Predicting the failure load of the distal radius. Osteoporos Int J Establ Result Coop Eur Found Osteoporos Natl Osteoporos Found USA 14:345–352. doi:10.​1007/​s00198-003-1380-9 CrossRef
40.
go back to reference Pialat J, Burghardt AJ, Sode M et al (2012) Visual grading of motion induced image degradation in high resolution peripheral computed tomography: impact of image quality on measures of bone density and microarchitecture. Bone 50:111–118. doi:10.1016/j.bone.2011.10.003 CrossRefPubMed Pialat J, Burghardt AJ, Sode M et al (2012) Visual grading of motion induced image degradation in high resolution peripheral computed tomography: impact of image quality on measures of bone density and microarchitecture. Bone 50:111–118. doi:10.​1016/​j.​bone.​2011.​10.​003 CrossRefPubMed
Metadata
Title
Clinical cone beam computed tomography compared to high-resolution peripheral computed tomography in the assessment of distal radius bone
Authors
C. de Charry
S. Boutroy
R. Ellouz
F. Duboeuf
R. Chapurlat
H. Follet
J. B. Pialat
Publication date
01-10-2016
Publisher
Springer London
Published in
Osteoporosis International / Issue 10/2016
Print ISSN: 0937-941X
Electronic ISSN: 1433-2965
DOI
https://doi.org/10.1007/s00198-016-3609-4

Other articles of this Issue 10/2016

Osteoporosis International 10/2016 Go to the issue