Open Access 01-12-2022 | Research

Cerclage performance analysis – a biomechanical comparison of different techniques and materials

Authors: L. M. Hägerich, F. G. E. Dyrna, J. C. Katthagen, P. A. Michel, L. F. Heilmann, A. Frank, M. J. Raschke, B. Schliemann, O. Riesenbeck

Published in: BMC Musculoskeletal Disorders | Issue 1/2022

Login to get access

Abstract

Background

Wire cerclages play a fundamental role in fracture fixation. With an increasing variety of designs being commercially available the question arises which cerclage should be used.
This study investigates the biomechanical properties of metallic and non-metallic cerclages and their different application-types. Furthermore, potential influence of muscular interposition between bone and cerclage constructs was tested.

Methods

Samples of the following four different cerclage types were tested on 3D printed models of human humeri as well as on human cadaveric humeri with and without muscular interposition:
Titanium Cable Cerclage (CC), Steel Wire Cerclage (SWC), Suture Tape (ST), Suture Tape Cerclage (STC) with both single- (sSTC) and double-loop application (dSTC). A preinstalled self-locking mechanism secured by the provided tensioner in the STCs being the main difference to the STs.
Cyclic loading was performed to 1 kN and then linearly to a maximum load of 3 kN.
Statistical analysis was performed using either one-way ANOVA and post-hoc Tukey or Kruskal–Wallis and post-hoc Dunn test depending on normalization of data (p < 0.05).

Results

Whilst all cerclage options could withstand high loads during failure testing, only within the CC and dSTC group, all samples reached the maximal testing load of 3000 N without any failure. The SWC reached 2977.5 ± 63.6 N, the ST 1970.8 ± 145.9 N, and the sSTC 1617.0 ± 341.6 N on average.
Neither muscular interposition nor bone quality showed to have a negative influence on the biomechanical properties of the cerclage constructs, presenting no significant differences.

Conclusion

All tested cerclage constructs produce reliable stability but differ in their resulting compression forces, in a simplified fracture model. Therefore, non-metallic cerclage alternatives can provide similar stability with less compression and stiffness to metallic cable constructs, but they may offer several advantages and could possibly provide future benefits. Especially, by offering more elasticity without losing overall stability, may offer a biologic benefit. Installing any cerclage constructs should be performed carefully, especially if poor bone quality is present, as the tightening process leads to high forces on the construct.
Appendix
Available only for authorised users
Literature
1.
go back to reference Karakoyun O, Sahin E, Erol MF, Karıksız M, Küçükkaya M. Effect of cable cerclage on regional blood circulation in rabbits: a scintigraphic study. J Orthop Surg (Hong Kong). 2016;24:367–9 SAGE PublicationsSage UK: London, England.CrossRefPubMed Karakoyun O, Sahin E, Erol MF, Karıksız M, Küçükkaya M. Effect of cable cerclage on regional blood circulation in rabbits: a scintigraphic study. J Orthop Surg (Hong Kong). 2016;24:367–9 SAGE PublicationsSage UK: London, England.CrossRefPubMed
2.
go back to reference Evans PE. Cerclage fixation of a fractured humerus in 1775. Fact or fiction? Clin Orthop Relat Res. 1983;174:138–42. Evans PE. Cerclage fixation of a fractured humerus in 1775. Fact or fiction? Clin Orthop Relat Res. 1983;174:138–42.
3.
go back to reference Hop JD, Callaghan JJ, Olejniczak JP, Pedersen DR, Brown TD, Johnston RC. The Frank Stinchfield Award. Contribution of cable debris generation to accelerated polyethylene wear. Clin Orthop Relat Res. 1997;344:20–32. Hop JD, Callaghan JJ, Olejniczak JP, Pedersen DR, Brown TD, Johnston RC. The Frank Stinchfield Award. Contribution of cable debris generation to accelerated polyethylene wear. Clin Orthop Relat Res. 1997;344:20–32.
4.
go back to reference Bauer TW, Ming J, D’Antonio JA, Morawa LG. Abrasive three-body wear of polyethylene caused by broken multifilament cables of a total hip prosthesis. A report of these cases. J of Bone Joint Surg. 1996;78:1244–7.CrossRef Bauer TW, Ming J, D’Antonio JA, Morawa LG. Abrasive three-body wear of polyethylene caused by broken multifilament cables of a total hip prosthesis. A report of these cases. J of Bone Joint Surg. 1996;78:1244–7.CrossRef
5.
go back to reference Peeters I, Depover A, Van Tongel A, De Wilde L. A review of metallic and non-metallic cerclage in orthopaedic surgery: is there still a place for metallic cerclage? Injury. 2019;50:1627–33.CrossRefPubMed Peeters I, Depover A, Van Tongel A, De Wilde L. A review of metallic and non-metallic cerclage in orthopaedic surgery: is there still a place for metallic cerclage? Injury. 2019;50:1627–33.CrossRefPubMed
6.
go back to reference Lenz M, Perren SM, Richards RG, Mückley T, Hofmann GO, Gueorguiev B, et al. Biomechanical performance of different cable and wire cerclage configurations. Int Orthop. 2013;37:125–30.CrossRefPubMed Lenz M, Perren SM, Richards RG, Mückley T, Hofmann GO, Gueorguiev B, et al. Biomechanical performance of different cable and wire cerclage configurations. Int Orthop. 2013;37:125–30.CrossRefPubMed
7.
go back to reference Wähnert D, Lenz M, Schlegel U, Perren S, Windolf M. Cerclage handling for improved fracture treatment. A biomechanical study on the twisting procedure. Acta Chir Orthop Traumatol Cech. 2011;78:208–14.PubMed Wähnert D, Lenz M, Schlegel U, Perren S, Windolf M. Cerclage handling for improved fracture treatment. A biomechanical study on the twisting procedure. Acta Chir Orthop Traumatol Cech. 2011;78:208–14.PubMed
8.
go back to reference Renner N, Wieser K, Lajtai G, Morrey ME, Meyer DC. Stainless steel wire versus FiberWire suture cerclage fixation to stabilize the humerus in total shoulder arthroplasty. J Shoulder Elbow Surg. 2014;23:1568–74.CrossRefPubMed Renner N, Wieser K, Lajtai G, Morrey ME, Meyer DC. Stainless steel wire versus FiberWire suture cerclage fixation to stabilize the humerus in total shoulder arthroplasty. J Shoulder Elbow Surg. 2014;23:1568–74.CrossRefPubMed
9.
go back to reference Angelini A, Battiato C. Past and present of the use of cerclage wires in orthopedics. Eur J Orthop Surg Traumatol. 2015;25:623–35 Springer Paris.CrossRefPubMed Angelini A, Battiato C. Past and present of the use of cerclage wires in orthopedics. Eur J Orthop Surg Traumatol. 2015;25:623–35 Springer Paris.CrossRefPubMed
10.
go back to reference Denard PJ, Nolte PC, Millett PJ, Adams CR, Liebler SAH, Rego G, et al. A tensionable suture-based cerclage is an alternative to stainless steel cerclage fixation for stabilization of a humeral osteotomy during shoulder arthroplasty. J Am Acad Orthop Surg. 2020;29:609.CrossRef Denard PJ, Nolte PC, Millett PJ, Adams CR, Liebler SAH, Rego G, et al. A tensionable suture-based cerclage is an alternative to stainless steel cerclage fixation for stabilization of a humeral osteotomy during shoulder arthroplasty. J Am Acad Orthop Surg. 2020;29:609.CrossRef
11.
go back to reference Michel PA, Katthagen JC, Schliemann B, Wilkens S, Frank A, Heilmann LF, et al. Biomechanical value of a protective proximal humeral cerclage in reverse total shoulder arthroplasty. J Clin Med. 2021;10(19):4600.CrossRefPubMedPubMedCentral Michel PA, Katthagen JC, Schliemann B, Wilkens S, Frank A, Heilmann LF, et al. Biomechanical value of a protective proximal humeral cerclage in reverse total shoulder arthroplasty. J Clin Med. 2021;10(19):4600.CrossRefPubMedPubMedCentral
12.
go back to reference Unnanuntana A, Saiyudthong N. Outcomes of cerclage wiring to manage intra-operative femoral fracture occurring during cementless hemiarthroplasty in older patients with femoral neck fractures. Int Orthop. 2019;43:2637–47 Springer Berlin Heidelberg.CrossRefPubMed Unnanuntana A, Saiyudthong N. Outcomes of cerclage wiring to manage intra-operative femoral fracture occurring during cementless hemiarthroplasty in older patients with femoral neck fractures. Int Orthop. 2019;43:2637–47 Springer Berlin Heidelberg.CrossRefPubMed
13.
go back to reference Fleischer J, Schleyer A, Nassutt R, Grittner U, Ojodu I, Hopp SJ. Biomechanical strength and failure mechanism of different tubercula refixation methods within the framework of an arthroplasty for shoulder fracture. Orthop Traumatol Surg Res. 2017;103:165–9.CrossRefPubMed Fleischer J, Schleyer A, Nassutt R, Grittner U, Ojodu I, Hopp SJ. Biomechanical strength and failure mechanism of different tubercula refixation methods within the framework of an arthroplasty for shoulder fracture. Orthop Traumatol Surg Res. 2017;103:165–9.CrossRefPubMed
14.
go back to reference Lenz M, Perren SM, Gueorguiev B, Richards RG, Krause F, Fernandez dell’Oca A, et al. Underneath the cerclage: an ex vivo study on the cerclage-bone interface mechanics. Arch Orthop Trauma Surg. 2012;132:1467–72.CrossRefPubMed Lenz M, Perren SM, Gueorguiev B, Richards RG, Krause F, Fernandez dell’Oca A, et al. Underneath the cerclage: an ex vivo study on the cerclage-bone interface mechanics. Arch Orthop Trauma Surg. 2012;132:1467–72.CrossRefPubMed
15.
go back to reference Thomsen M, Thomas P. Compatibility and allergies of osteosynthesis materials. Unfallchirurg Springer Medizin. 2017;120:116–21. Thomsen M, Thomas P. Compatibility and allergies of osteosynthesis materials. Unfallchirurg Springer Medizin. 2017;120:116–21.
Metadata
Title
Cerclage performance analysis – a biomechanical comparison of different techniques and materials
Authors
L. M. Hägerich
F. G. E. Dyrna
J. C. Katthagen
P. A. Michel
L. F. Heilmann
A. Frank
M. J. Raschke
B. Schliemann
O. Riesenbeck
Publication date
01-12-2022
Publisher
BioMed Central
Published in
BMC Musculoskeletal Disorders / Issue 1/2022
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-022-05983-6

Other articles of this Issue 1/2022

BMC Musculoskeletal Disorders 1/2022 Go to the issue