Published in: BMC Anesthesiology 1/2018

Open Access 01-12-2018 | Research article

Cell salvage using the continuous autotransfusion device CATSmart – an observational bicenter technical evaluation

Authors: Simone Lindau, Madeline Kohlhaas, Michael Nosch, Suma Choorapoikayil, Kai Zacharowski, Patrick Meybohm

Published in: BMC Anesthesiology | Issue 1/2018

Login to get access

Abstract

Background

The use of cell salvage and autologous blood transfusion has become an important method of blood conservation. So far, there are no clinical data about the performance of the continuous autotransfusion device CATSmart.

Methods

In total, 74 patients undergoing either cardiac or orthopedic surgery were included in this prospective, bicenter and observational technical evaluation to validate red cell separation process and washout quality of CATSmart. The target of red cell separation process was defined as a hematocrit value in the packed red cell unit of 55–75% and of washout quality of 80–100% removal ratio.

Results

Hematocrit values measured by CATSmart and laboratory analysis were 78.5% [71.3%; 84.0%] and 73.7% [67.5%; 75.5%], respectively. Removal ratios for platelets 94.7% [88.2%; 96.7%], free hemoglobin 89.3% [85.2%; 94.9%], albumin 97.9% [96.6%; 98.5%], heparin 99.9% [99.9%; 100.0%], and potassium 92.5% [90.8%; 95.0%] were within the target range while removal of white blood cells was slightly worse 72.4% [57.9%; 87.3%].

Conclusion

The new autotransfusion device enables sufficient red cell separation and washout quality.
Literature
2.
go back to reference Alberts M, Groom RC, Walczak R, Kramer R, Karpiel A, Dieter J, et al. In vitro evaluation of the Fresenius Kabi CATSmart autotransfusion system. J Extra Corpor Technol. 2017;49:107–11.PubMedPubMedCentral Alberts M, Groom RC, Walczak R, Kramer R, Karpiel A, Dieter J, et al. In vitro evaluation of the Fresenius Kabi CATSmart autotransfusion system. J Extra Corpor Technol. 2017;49:107–11.PubMedPubMedCentral
5.
go back to reference Bundesärztekammer. Richtlinie zur Gewinnung von Blut und Blutbestandteilen und zur Anwendung von Blutprodukten (Richtlinie Hämotherapie);17.02.2017. Bundesärztekammer. Richtlinie zur Gewinnung von Blut und Blutbestandteilen und zur Anwendung von Blutprodukten (Richtlinie Hämotherapie);17.02.2017.
Metadata
Title
Cell salvage using the continuous autotransfusion device CATSmart – an observational bicenter technical evaluation
Authors
Simone Lindau
Madeline Kohlhaas
Michael Nosch
Suma Choorapoikayil
Kai Zacharowski
Patrick Meybohm
Publication date
01-12-2018
Publisher
BioMed Central
Published in
BMC Anesthesiology / Issue 1/2018
Electronic ISSN: 1471-2253
DOI
https://doi.org/10.1186/s12871-018-0651-0

Other articles of this Issue 1/2018

BMC Anesthesiology 1/2018 Go to the issue