Open Access 01-12-2023 | Cardiotomy | Research

Long-term ECMO, efficiency and performance of EUROSETS adult A.L.ONE ECMO oxygenator

Authors: Ignazio Condello, Roberto Lorusso, Giuseppe Nasso, Giuseppe Speziale

Published in: Journal of Cardiothoracic Surgery | Issue 1/2023

Login to get access

Abstract

Background

The management of the oxygenator can be prolonged in the long-term procedures especially during extracorporeal membrane oxygenation (ECMO) for bridge to transplant or bridge to recovery. Long-term use often involves an overrun of the time of use with respect to certification of the oxygenating module of 14 days, for the maintenance of performance and efficiency of the oxygenator. The evaluation of the long-term oxygenator efficiency is complex and depends on the: patient pathology, ECMO configuration, the management of coagulation and anticoagulation, materials selection and circuit components, the structure, design and performance of the oxygenator. In this context we investgated the long-term performance of the A.L.ONE Eurosets ECMO oxygenator in relation to the parameters prodromal to replacement.

Methods

We retrospectively collected eight years data from Anthea Hospital GVM Care & Research, Bari, Italy on the long-term use exceeding 14 days of Eurosets A.L.ONE ECMO Adult oxygenator in Polymetylpentene fiber, for ECMO procedures, including the procedures: Veno Arterial (VA) ECMO post-cardiotomy or not, veno-venous (VV) ECMO. The primary end points were the evaluation of Gas Transfer: oxygen partial pressure (PO2) post oxygenator, Carbon dioxide partial pressure (PCO2) post oxygenator, the oxygen transfer across the oxygenator membrane V′O2, differential CO2 content across oxygenator; Pressure monitoring: oxygenator pressure Drop in relation to Blood flow rate (BFR) (ΔP); Hematologic values: Hemoglobin, Fibrinogen, Platelets, aPTT, D-Dimer, LDH.

Results

Nine VA ECMO patients who used the oxygenator for 18.5 days and two VV ECMO patients who used the oxygenators for 17.2 days on the seventeenth days reported average values PaO2 (267 ± 29 mmHg); PaCO2 (34 ± 4 mmHg) with gas blender values set to 3.8 ± 0.6 L/min of air and a FiO2 of 78 ± 5%; the transfer across the oxygenator membrane V′O2 was 189 ± 43 (ml/min/m2). The mean peak of partial pressure of carbon dioxide from the gas exhaust of oxygenator (PECO2) was 38 ± 4 mmHg; differential CO2 across the oxygenator “pre-oxygenator PCO2–post-oxygenator PCO2” (18 ± 6 mmHg); the mean blood flow rate (BFR) 4.5 ± 0.6 (L/minute); the pump revolution per minutes mean maximum rate was 4254 ± 345 (RPM); the mean pressure drop (ΔP) was 76 ± 12 mmHg; the mean peak of d-dimers (DDs) was 23.6 ± 0.8 mg / dL; the mean peak of LDH was 230 ± 55 (mg/dl); fibrinogen mean peak 223 ± 40 (mg/dl).

Conclusions

The performance of the Eurosets A.L.ONE ECMO Adult polymethylpentene fiber oxygenator in our experience has proven efficiency in terms of O2 uptake and CO2 removal, blood fluid dynamics, metabolic compensation and heat exchange in the long-term treatment. The device was safe without iatrogenic problems over a period of 14 days in the patients undergoing ECMO VA and in all patients undergoing VV ECMO with continuous administration of anticoagulation therapy.
Literature
3.
go back to reference Dornia C, Philipp A, Bauer S, Lubnow M, Muller T, Lehle K, et al. Analysis of thrombotic deposits in extracorporeal membrane oxygenators by multidetector computed tomography. ASAIO J. 2014;60(6):652–6.CrossRefPubMed Dornia C, Philipp A, Bauer S, Lubnow M, Muller T, Lehle K, et al. Analysis of thrombotic deposits in extracorporeal membrane oxygenators by multidetector computed tomography. ASAIO J. 2014;60(6):652–6.CrossRefPubMed
4.
go back to reference Doyle AJ, Hunt BJ. Current understanding of how extracorporeal membrane oxygenators activate haemostasis and other blood components. Front Med (Lausanne). 2018;5:352.CrossRefPubMed Doyle AJ, Hunt BJ. Current understanding of how extracorporeal membrane oxygenators activate haemostasis and other blood components. Front Med (Lausanne). 2018;5:352.CrossRefPubMed
5.
go back to reference Lubnow M, Philipp A, Foltan M, Bull Enger T, Lunz D, Bein T, et al. Technical complications during veno-venous extracorporeal membrane oxygenation and their relevance predicting a system-exchange–retrospective analysis of 265 cases. PLoS ONE. 2014;9(12):e112316.CrossRefPubMedPubMedCentral Lubnow M, Philipp A, Foltan M, Bull Enger T, Lunz D, Bein T, et al. Technical complications during veno-venous extracorporeal membrane oxygenation and their relevance predicting a system-exchange–retrospective analysis of 265 cases. PLoS ONE. 2014;9(12):e112316.CrossRefPubMedPubMedCentral
6.
go back to reference Maul TM, Massicotte MP, Wearden PD. ECMO biocompatibility: surface coatings, anticoagulation, and coagulation monitoring. In: Firstenberg MS, editor. Extracorporeal membrane oxygenation: advances in therapy. Croatia: IntechOpen; 2016. p. 30–5. Maul TM, Massicotte MP, Wearden PD. ECMO biocompatibility: surface coatings, anticoagulation, and coagulation monitoring. In: Firstenberg MS, editor. Extracorporeal membrane oxygenation: advances in therapy. Croatia: IntechOpen; 2016. p. 30–5.
7.
go back to reference Hastings SM, Deshpande SR, Wagoner S, Maher K, Ku DN. Thrombosis in centrifugal pumps: location and composition in clinical and in vitrocircuits. Int J Artif Organs. 2016;39(4):200–4.CrossRefPubMed Hastings SM, Deshpande SR, Wagoner S, Maher K, Ku DN. Thrombosis in centrifugal pumps: location and composition in clinical and in vitrocircuits. Int J Artif Organs. 2016;39(4):200–4.CrossRefPubMed
8.
go back to reference Lehle K, Philipp A, Gleich O, Holzamer A, Muller T, Bein T, et al. Efficiency inextracorporeal membrane oxygenation-cellular deposits on polymethylpentene membranes increase resistance to blood flow and reduce gas exchange capacity. ASAIO J. 2008;54(6):612–7.CrossRefPubMed Lehle K, Philipp A, Gleich O, Holzamer A, Muller T, Bein T, et al. Efficiency inextracorporeal membrane oxygenation-cellular deposits on polymethylpentene membranes increase resistance to blood flow and reduce gas exchange capacity. ASAIO J. 2008;54(6):612–7.CrossRefPubMed
12.
go back to reference Zakhary B, Sheldrake J, Pellegrino V. Extracorporeal membrane oxygenation and V/Q ratios: an ex vivo analysis of CO2 clearance within the Maquet Quadrox-iD oxygenator. Perfusion. 2020;35(1):29–33.CrossRefPubMed Zakhary B, Sheldrake J, Pellegrino V. Extracorporeal membrane oxygenation and V/Q ratios: an ex vivo analysis of CO2 clearance within the Maquet Quadrox-iD oxygenator. Perfusion. 2020;35(1):29–33.CrossRefPubMed
Metadata
Title
Long-term ECMO, efficiency and performance of EUROSETS adult A.L.ONE ECMO oxygenator
Authors
Ignazio Condello
Roberto Lorusso
Giuseppe Nasso
Giuseppe Speziale
Publication date
01-12-2023
Publisher
BioMed Central
Published in
Journal of Cardiothoracic Surgery / Issue 1/2023
Electronic ISSN: 1749-8090
DOI
https://doi.org/10.1186/s13019-023-02190-9

Other articles of this Issue 1/2023

Journal of Cardiothoracic Surgery 1/2023 Go to the issue