Capsule endoscopy with flexible spectral imaging color enhancement reduces the bile pigment effect and improves the detectability of small bowel lesions

Authors: Eiji Sakai, Hiroki Endo, Shingo Kato, Tetsuya Matsuura, Wataru Tomeno, Leo Taniguchi, Takashi Uchiyama, Yasuo Hata, Eiji Yamada, Hidenori Ohkubo, Takuma Higrashi, Kunihiro Hosono, Hirokazu Takahashi, Atsushi Nakajima

Published in: BMC Gastroenterology | Issue 1/2012

Login to get access

Abstract

Background

Capsule endoscopy with flexible spectral imaging color enhancement (CE-FICE) has been reported to improve the visualization and detection of small-bowel lesions, however, its clinical usefulness is still not established. Therefore, we conducted a study to evaluate whether CE-FICE contributes to improve the detectability of small-bowel lesions by CE trainees.

Methods

Four gastroenterology trainees without prior CE experience were asked to read and interpret 12 CE videos. Each of the videos was read by conventional visualization method and under three different FICE settings. To evaluate whether the lesion recognition ability of the CE trainees could be improved by the FICE technology, the lesion detection rate under each of the three FICE settings was compared with that by conventional CE. CE trainees tend to miss small-bowel lesions in bile-pigment-positive condition, therefore we evaluated whether CE-FICE contributes to reducing the bile-pigment effect. The bile-pigment condition was determined by the color values around the small-bowel lesions according to the results of the receiver-operating-characteristic analysis. Moreover, we also evaluated whether poor bowel preparion might affect the accuracy of lesion recognition by CE-FICE.

Results

Of a total of 60 angioectasias, CE trainees identified 26 by conventional CE, 40 under FICE setting 1, 38 under FICE setting 2, and 31 under FICE setting 3. Of a total of 82 erosions/ulcerations, CE trainees identified 38 by conventional CE, 62 under FICE setting 1, 60 under FICE setting 2, and 20 under FICE setting 3. Compared with conventional CE, FICE settings 1 and 2 significantly improved the detectability of angioectasia (P = 0.0017 and P = 0.014, respectively) and erosions/ulcerations (P = 0.0012 and P = 0.0094, respectively). Although the detectability of small-bowel lesions by conventional CE (P = 0.020) and under FICE setting 2 (P = 0.0023) was reduced by the presence of bile-pigments, that under FICE setting 1 was not affected (P = 0.59). Our results also revealed that in poor bowel visibility conditions, CE-FICE yielded a high rate of false-positive findings.

Conclusions

CE-FICE may reduce the bile-pigment effect and improve the detectability of small-bowel lesions by CE trainees; the reliability of CE-FICE may be improved by good bowel preparation.
Appendix
Available only for authorised users
Literature
1.
go back to reference Iddan G, Merson G, Glukhovsky A, Swain P: Wireless capsule endoscopy. Nature. 2000, 405: 17-CrossRef Iddan G, Merson G, Glukhovsky A, Swain P: Wireless capsule endoscopy. Nature. 2000, 405: 17-CrossRef
2.
go back to reference Meron GD: The development of the swallowable video capsule (M2A). Gastrointest Endosc. 2000, 52: 817-9. 10.1067/mge.2000.110204.CrossRefPubMed Meron GD: The development of the swallowable video capsule (M2A). Gastrointest Endosc. 2000, 52: 817-9. 10.1067/mge.2000.110204.CrossRefPubMed
3.
go back to reference Lewis BS, Swain P: Capsule endoscopy in the evaluation of patients with suspected small intestinal bleeding: results of a pilot study. Gastrointest Endosc. 2002, 56: 349-53. 10.1016/S0016-5107(02)70037-0.CrossRefPubMed Lewis BS, Swain P: Capsule endoscopy in the evaluation of patients with suspected small intestinal bleeding: results of a pilot study. Gastrointest Endosc. 2002, 56: 349-53. 10.1016/S0016-5107(02)70037-0.CrossRefPubMed
4.
go back to reference Pennazio M, Santucci R, Rondonotti E, Abbiati C, Beccari G, Rossini FP, De Franchis R: Outcome of patients with obscure gastrointestinal bleeding after capsule endoscopy; report of 100 consecutive cases. Gastroenterology. 2004, 126: 643-53. 10.1053/j.gastro.2003.11.057.CrossRefPubMed Pennazio M, Santucci R, Rondonotti E, Abbiati C, Beccari G, Rossini FP, De Franchis R: Outcome of patients with obscure gastrointestinal bleeding after capsule endoscopy; report of 100 consecutive cases. Gastroenterology. 2004, 126: 643-53. 10.1053/j.gastro.2003.11.057.CrossRefPubMed
5.
go back to reference Apostolopoulos P, Liatosos C, Gralnek IM, Giannakoulopoulou E, Alexandrakis G, Kalantzis C, Gabriel P, Kalantzis N: The role of wireless capsule endoscopy in investigating unexplained iron deficiency anemia after negative endoscopic evaluation of the upper and lower gastrointestinal tract. Endoscopy. 2006, 38: 1127-32. 10.1055/s-2006-944736.CrossRefPubMed Apostolopoulos P, Liatosos C, Gralnek IM, Giannakoulopoulou E, Alexandrakis G, Kalantzis C, Gabriel P, Kalantzis N: The role of wireless capsule endoscopy in investigating unexplained iron deficiency anemia after negative endoscopic evaluation of the upper and lower gastrointestinal tract. Endoscopy. 2006, 38: 1127-32. 10.1055/s-2006-944736.CrossRefPubMed
6.
go back to reference Endo H, Hosono K, Inamori M, Kato S, Nozaki Y, Yoneda K, et al: Incidence of small bowel injury induced by low-dose aspirin: a crossover study using capsule endoscopy in healthy volunteers. Digestion. 2009, 79: 44-51. 10.1159/000204465.CrossRefPubMed Endo H, Hosono K, Inamori M, Kato S, Nozaki Y, Yoneda K, et al: Incidence of small bowel injury induced by low-dose aspirin: a crossover study using capsule endoscopy in healthy volunteers. Digestion. 2009, 79: 44-51. 10.1159/000204465.CrossRefPubMed
7.
go back to reference Endo H, Hosono K, Inamori M, Nozaki Y, Yoneda K, Fujita K, Takahashi H, Yoneda M, Abe Y, Kirikoshi H, Kobayashi N, Kubota K, Saito S, Ohya T, Hisatomi K, Teratani T, Matsuhashi N, Nakajima A: Characteristics of small bowel injury in symptomatic chronic low-dose aspirin users: the experience of two medical centers in capsule endoscopy. J Gastroenterol. 2009, 44: 544-9. 10.1007/s00535-009-0040-z.CrossRefPubMed Endo H, Hosono K, Inamori M, Nozaki Y, Yoneda K, Fujita K, Takahashi H, Yoneda M, Abe Y, Kirikoshi H, Kobayashi N, Kubota K, Saito S, Ohya T, Hisatomi K, Teratani T, Matsuhashi N, Nakajima A: Characteristics of small bowel injury in symptomatic chronic low-dose aspirin users: the experience of two medical centers in capsule endoscopy. J Gastroenterol. 2009, 44: 544-9. 10.1007/s00535-009-0040-z.CrossRefPubMed
8.
go back to reference Mylonaki M, FritscheRavens A, Swain P: Wireless capsule endoscopy: a comparison with push enteroscopy with gastroscopy and colonoscopy negative gastrointestinal bleeding. Gut. 2003, 52: 1122-6. 10.1136/gut.52.8.1122.CrossRefPubMedPubMedCentral Mylonaki M, FritscheRavens A, Swain P: Wireless capsule endoscopy: a comparison with push enteroscopy with gastroscopy and colonoscopy negative gastrointestinal bleeding. Gut. 2003, 52: 1122-6. 10.1136/gut.52.8.1122.CrossRefPubMedPubMedCentral
9.
go back to reference Saurin JC, Delvaux M, Gaudin JL, Fassler I, Villareio J, Vahedi K, Bitoun A, Canard JM, Souquet JC, Ponchon T, Florent C, Gay G: Daiagnostic values of endoscopic capsule in patients obscure bleeding: blinded comparison with video push-enteroscopy. Endoscopy. 2003, 35: 576-84.CrossRefPubMed Saurin JC, Delvaux M, Gaudin JL, Fassler I, Villareio J, Vahedi K, Bitoun A, Canard JM, Souquet JC, Ponchon T, Florent C, Gay G: Daiagnostic values of endoscopic capsule in patients obscure bleeding: blinded comparison with video push-enteroscopy. Endoscopy. 2003, 35: 576-84.CrossRefPubMed
10.
go back to reference Constamagna G, Shah SK, Riccioni ME, Foschia F, Mutignani M, Perri V, Vecchioli A, Brizi MG, Picciocchi A, Marano P: A prospective trial comparing small bowel radiographs and video capsule endoscopy for suspected small bowel disease. Gastroenterology. 2002, 123: 999-1005. 10.1053/gast.2002.35988.CrossRef Constamagna G, Shah SK, Riccioni ME, Foschia F, Mutignani M, Perri V, Vecchioli A, Brizi MG, Picciocchi A, Marano P: A prospective trial comparing small bowel radiographs and video capsule endoscopy for suspected small bowel disease. Gastroenterology. 2002, 123: 999-1005. 10.1053/gast.2002.35988.CrossRef
11.
go back to reference Saperas E, Dot J, Videla S, Alvarez-Castells A, Perez-Lafuente M, Armengol JR, Malagelada JR: Capsule endoscopy versus computed tomography or standard angiography for the diagnosis of obscure gastrointestinal bleeding. Am J Gastroenterol. 2007, 102: 731-7. 10.1111/j.1572-0241.2007.01058.x.CrossRefPubMed Saperas E, Dot J, Videla S, Alvarez-Castells A, Perez-Lafuente M, Armengol JR, Malagelada JR: Capsule endoscopy versus computed tomography or standard angiography for the diagnosis of obscure gastrointestinal bleeding. Am J Gastroenterol. 2007, 102: 731-7. 10.1111/j.1572-0241.2007.01058.x.CrossRefPubMed
12.
go back to reference Viazis N, Sgouros S, Papaxoinis K, Vlachogiannakos J, Bergele C, Sklavus P, Panari A, Avgerinos A: Bowel preparation increases the diagnostic yield of capsule endoscopy: A prospective, randomized, controlled study. Gastrointest Endosc. 2004, 60: 534-8. 10.1016/S0016-5107(04)01879-6.CrossRefPubMed Viazis N, Sgouros S, Papaxoinis K, Vlachogiannakos J, Bergele C, Sklavus P, Panari A, Avgerinos A: Bowel preparation increases the diagnostic yield of capsule endoscopy: A prospective, randomized, controlled study. Gastrointest Endosc. 2004, 60: 534-8. 10.1016/S0016-5107(04)01879-6.CrossRefPubMed
13.
go back to reference Endo H, Kondo Y, Inamori M, Ohya TR, Yanagawa T, Asayama M, Hisatoshi K, Teratani T, Yoneda M, Nakajima A, Matsuhashi N: Ingesting 500 ml polyethylene glycol solution during capsule endoscopy improves the image quality and completion rate to the cecum. Dig Dis Sci. 2008, 53: 3201-5. 10.1007/s10620-008-0292-0.CrossRefPubMed Endo H, Kondo Y, Inamori M, Ohya TR, Yanagawa T, Asayama M, Hisatoshi K, Teratani T, Yoneda M, Nakajima A, Matsuhashi N: Ingesting 500 ml polyethylene glycol solution during capsule endoscopy improves the image quality and completion rate to the cecum. Dig Dis Sci. 2008, 53: 3201-5. 10.1007/s10620-008-0292-0.CrossRefPubMed
14.
go back to reference Ben-Soussan E, Savoye G, Antonietti M, Ramirez S, Ducrotte P, Levebours E: Is a 2-liter PEG preparation useful before capsule endoscopy?. J Clin Gastroenterol. 2005, 39: 381-4. 10.1097/01.mcg.0000159271.43233.45.CrossRefPubMed Ben-Soussan E, Savoye G, Antonietti M, Ramirez S, Ducrotte P, Levebours E: Is a 2-liter PEG preparation useful before capsule endoscopy?. J Clin Gastroenterol. 2005, 39: 381-4. 10.1097/01.mcg.0000159271.43233.45.CrossRefPubMed
15.
go back to reference Franke A, Hummel F, Knebel P, Antoni C, Bocker U, Singer MV, Lohr M: Prospective evaluation of small bowel preparation with bisacodyl and sodium phosphate for capsule endoscopy. World J Gastroenterol. 2008, 14: 2061-4. 10.3748/wjg.14.2061.CrossRefPubMedPubMedCentral Franke A, Hummel F, Knebel P, Antoni C, Bocker U, Singer MV, Lohr M: Prospective evaluation of small bowel preparation with bisacodyl and sodium phosphate for capsule endoscopy. World J Gastroenterol. 2008, 14: 2061-4. 10.3748/wjg.14.2061.CrossRefPubMedPubMedCentral
16.
go back to reference Pohl J, May A, Rabenstein T, Pech O, Nguyen-Tat M, Fissler-Eckhoff A, Ell C: Comparison of computed virtual chromoendoscopy and conventional chromoendoscopy with acetic acid for detection of neoplasia in Barrett’s esophagus. Endoscopy. 2007, 39: 594-8. 10.1055/s-2007-966649.CrossRefPubMed Pohl J, May A, Rabenstein T, Pech O, Nguyen-Tat M, Fissler-Eckhoff A, Ell C: Comparison of computed virtual chromoendoscopy and conventional chromoendoscopy with acetic acid for detection of neoplasia in Barrett’s esophagus. Endoscopy. 2007, 39: 594-8. 10.1055/s-2007-966649.CrossRefPubMed
17.
go back to reference Pohl J, Nguyen-tat M, Pech O, May A, Rabenstein T, Ell C: Computed virtual chromoendoscopy for classification of small colorectal lesions; a prospective comparative study. Am J Gastroenterol. 2008, 103: 562-9. 10.1111/j.1572-0241.2007.01670.x.CrossRefPubMed Pohl J, Nguyen-tat M, Pech O, May A, Rabenstein T, Ell C: Computed virtual chromoendoscopy for classification of small colorectal lesions; a prospective comparative study. Am J Gastroenterol. 2008, 103: 562-9. 10.1111/j.1572-0241.2007.01670.x.CrossRefPubMed
18.
go back to reference Mouri R, Yoshida S, Tanaka S, Oka S, Yoshihara M, Chayama K: Evaluation and validation of computed virtual chromoendoscopy in early gastric cancer. Gastrointest Endosc. 2009, 69: 1052-8. 10.1016/j.gie.2008.08.032.CrossRefPubMed Mouri R, Yoshida S, Tanaka S, Oka S, Yoshihara M, Chayama K: Evaluation and validation of computed virtual chromoendoscopy in early gastric cancer. Gastrointest Endosc. 2009, 69: 1052-8. 10.1016/j.gie.2008.08.032.CrossRefPubMed
19.
go back to reference Pohl J, May A, Rabenstein T, Pech O, Ell C: Computed virtual chromoendoscopy: a new tool for enhancing tissue surface structures. Endoscopy. 2007, 39: 80-3. 10.1055/s-2006-945045.CrossRefPubMed Pohl J, May A, Rabenstein T, Pech O, Ell C: Computed virtual chromoendoscopy: a new tool for enhancing tissue surface structures. Endoscopy. 2007, 39: 80-3. 10.1055/s-2006-945045.CrossRefPubMed
20.
go back to reference Pohl J, Ashmoneit I, Schuhmann S, Ell C: Computed image modification for enhancement of small-bowel surface structures at video capsule endoscopy. Endoscopy. 2010, 42: 490-2. 10.1055/s-0029-1243994.CrossRefPubMed Pohl J, Ashmoneit I, Schuhmann S, Ell C: Computed image modification for enhancement of small-bowel surface structures at video capsule endoscopy. Endoscopy. 2010, 42: 490-2. 10.1055/s-0029-1243994.CrossRefPubMed
21.
go back to reference Imagawa H, Oka S, Tanaka S, Noda I, Higashiyama M, Sanomura Y, Shishido T, Yoshida S, Chayama K: Improved visibility of lesions of the small intestine via capsule endoscopy with computed virtual chromoendoscopy. Gastrointest Endsc. 2011, 73: 299-306. 10.1016/j.gie.2010.10.016.CrossRef Imagawa H, Oka S, Tanaka S, Noda I, Higashiyama M, Sanomura Y, Shishido T, Yoshida S, Chayama K: Improved visibility of lesions of the small intestine via capsule endoscopy with computed virtual chromoendoscopy. Gastrointest Endsc. 2011, 73: 299-306. 10.1016/j.gie.2010.10.016.CrossRef
22.
go back to reference Brotz C, Nandi N, Conn M, Daskalakis C, DiMarino M, Infantolino A, Katz LC, Schroeder T, Kastenberg D: A validation study of 3 grading systems to evaluate small-bowel cleaning for wireless capsule endoscopy: a quantitative index, a qualitative evaluation, and an overall adequacy assessment. Gastrointest Endosc. 2009, 69: 262-70. 10.1016/j.gie.2008.04.016.CrossRefPubMed Brotz C, Nandi N, Conn M, Daskalakis C, DiMarino M, Infantolino A, Katz LC, Schroeder T, Kastenberg D: A validation study of 3 grading systems to evaluate small-bowel cleaning for wireless capsule endoscopy: a quantitative index, a qualitative evaluation, and an overall adequacy assessment. Gastrointest Endosc. 2009, 69: 262-70. 10.1016/j.gie.2008.04.016.CrossRefPubMed
23.
go back to reference VanWeyenberg SJ, De Leest HT, Mulder CJ: Description of a novel grading system to assess the quality of bowel preparation in video capsule endoscopy. Endoscopy. 2011, 43: 406-11. 10.1055/s-0030-1256228.CrossRef VanWeyenberg SJ, De Leest HT, Mulder CJ: Description of a novel grading system to assess the quality of bowel preparation in video capsule endoscopy. Endoscopy. 2011, 43: 406-11. 10.1055/s-0030-1256228.CrossRef
24.
go back to reference Postgate A, Haycock A, Thomas-Gibson S, Fitzpatrick A, Bassett P, Preston S, Saunders BP, Fraser C: Computer-aided learning in capsule endoscopy leads to improvement in lesion recognition abilty. Gastrointest Endosc. 2009, 70: 310-6. 10.1016/j.gie.2008.11.043.CrossRefPubMed Postgate A, Haycock A, Thomas-Gibson S, Fitzpatrick A, Bassett P, Preston S, Saunders BP, Fraser C: Computer-aided learning in capsule endoscopy leads to improvement in lesion recognition abilty. Gastrointest Endosc. 2009, 70: 310-6. 10.1016/j.gie.2008.11.043.CrossRefPubMed
25.
go back to reference Sidhu R, Sakellariou P, McAlindon ME, Leeds JS, Shafiq K, Hoeroldt BS, Hopper AD, Karmo M, Salmon C, Elphick D, Ali A, Saunders DS: Is formal training necessary for capsule endoscopy? The largest gastroenterology trainee study with controls. Dig Liver Dis. 2008, 40: 298-302. 10.1016/j.dld.2007.11.022.CrossRefPubMed Sidhu R, Sakellariou P, McAlindon ME, Leeds JS, Shafiq K, Hoeroldt BS, Hopper AD, Karmo M, Salmon C, Elphick D, Ali A, Saunders DS: Is formal training necessary for capsule endoscopy? The largest gastroenterology trainee study with controls. Dig Liver Dis. 2008, 40: 298-302. 10.1016/j.dld.2007.11.022.CrossRefPubMed
26.
go back to reference Levinthal GN, Burke CA, Statisi JM: The accuracy of an endoscopy nurse in interpreting capsule endoscopy. Am J Gastroenterol. 2003, 98: 2669-71. 10.1111/j.1572-0241.2003.08726.x.CrossRefPubMed Levinthal GN, Burke CA, Statisi JM: The accuracy of an endoscopy nurse in interpreting capsule endoscopy. Am J Gastroenterol. 2003, 98: 2669-71. 10.1111/j.1572-0241.2003.08726.x.CrossRefPubMed
27.
go back to reference Bossa F, Cocomazzi G, Valvano MR, Andriulli A, Annesa V: Detection of abnormal lesions recorded by capsule endoscopy. A prospective study comparing endoscopist’s and nurse’s accuracy. Dig Liver Dis. 2006, 38: 599-602. 10.1016/j.dld.2006.03.019.CrossRefPubMed Bossa F, Cocomazzi G, Valvano MR, Andriulli A, Annesa V: Detection of abnormal lesions recorded by capsule endoscopy. A prospective study comparing endoscopist’s and nurse’s accuracy. Dig Liver Dis. 2006, 38: 599-602. 10.1016/j.dld.2006.03.019.CrossRefPubMed
28.
go back to reference Imagawa H, Oka S, Tanaka S, Noda I, Higashiyama M, Sanomura Y, Shishido T, Yoshida S, Chayama K: Improved detectability of small-bowel lesions via capsule endoscopy with computed virtual chromoendoscopy: A pilot study. Scand J Gastroenterol. 2011, 46: 1133-7. 10.3109/00365521.2011.584899.CrossRefPubMedPubMedCentral Imagawa H, Oka S, Tanaka S, Noda I, Higashiyama M, Sanomura Y, Shishido T, Yoshida S, Chayama K: Improved detectability of small-bowel lesions via capsule endoscopy with computed virtual chromoendoscopy: A pilot study. Scand J Gastroenterol. 2011, 46: 1133-7. 10.3109/00365521.2011.584899.CrossRefPubMedPubMedCentral
29.
go back to reference Heinz GD, Hadithi M, Groenen MJ, Kuipers EJ, Jacobs MA, Mulder CJ: Double-balloon enteroscopy: indications, diagnostic yield, and complications in a series of 275 patients with suspected small-bowel disease. Endoscopy. 2006, 38: 42-8. 10.1055/s-2005-921188.CrossRef Heinz GD, Hadithi M, Groenen MJ, Kuipers EJ, Jacobs MA, Mulder CJ: Double-balloon enteroscopy: indications, diagnostic yield, and complications in a series of 275 patients with suspected small-bowel disease. Endoscopy. 2006, 38: 42-8. 10.1055/s-2005-921188.CrossRef
30.
go back to reference Ohmiya N, Yano Y, Yamamoto H, Arakawa D, Nakamura M, Honda W, Itoh A, Hirooka Y, Niwa Y, Maeda O, Ando T, Yao T, Matsui T, Iida M, Tanaka S, Chiba T, Sakamoto C, Sugano K, Goto H: Diagnosis and treatment of obscure GI bleeding at double balloon endoscopy. Gastrointest Endosc. 2007, 66: 72-7. 10.1016/j.gie.2007.05.041.CrossRef Ohmiya N, Yano Y, Yamamoto H, Arakawa D, Nakamura M, Honda W, Itoh A, Hirooka Y, Niwa Y, Maeda O, Ando T, Yao T, Matsui T, Iida M, Tanaka S, Chiba T, Sakamoto C, Sugano K, Goto H: Diagnosis and treatment of obscure GI bleeding at double balloon endoscopy. Gastrointest Endosc. 2007, 66: 72-7. 10.1016/j.gie.2007.05.041.CrossRef
Metadata
Title
Capsule endoscopy with flexible spectral imaging color enhancement reduces the bile pigment effect and improves the detectability of small bowel lesions
Authors
Eiji Sakai
Hiroki Endo
Shingo Kato
Tetsuya Matsuura
Wataru Tomeno
Leo Taniguchi
Takashi Uchiyama
Yasuo Hata
Eiji Yamada
Hidenori Ohkubo
Takuma Higrashi
Kunihiro Hosono
Hirokazu Takahashi
Atsushi Nakajima
Publication date
01-12-2012
Publisher
BioMed Central
Published in
BMC Gastroenterology / Issue 1/2012
Electronic ISSN: 1471-230X
DOI
https://doi.org/10.1186/1471-230X-12-83

Other articles of this Issue 1/2012

BMC Gastroenterology 1/2012 Go to the issue
Obesity Clinical Trial Summary

At a glance: The STEP trials

A round-up of the STEP phase 3 clinical trials evaluating semaglutide for weight loss in people with overweight or obesity.

Developed by: Springer Medicine

Highlights from the ACC 2024 Congress

Year in Review: Pediatric cardiology

Watch Dr. Anne Marie Valente present the last year's highlights in pediatric and congenital heart disease in the official ACC.24 Year in Review session.

Year in Review: Pulmonary vascular disease

The last year's highlights in pulmonary vascular disease are presented by Dr. Jane Leopold in this official video from ACC.24.

Year in Review: Valvular heart disease

Watch Prof. William Zoghbi present the last year's highlights in valvular heart disease from the official ACC.24 Year in Review session.

Year in Review: Heart failure and cardiomyopathies

Watch this official video from ACC.24. Dr. Biykem Bozkurt discusses last year's major advances in heart failure and cardiomyopathies.