Skip to main content
Top
Published in: BMC Musculoskeletal Disorders 1/2022

Open Access 01-12-2022 | Bone Tumor | Research

Blood titanium levels in patients with large and sliding titanium implants

Authors: Martina Tognini, Harry Hothi, Stewart Tucker, Edel Broomfield, Masood Shafafy, Panos Gikas, Anna Di Laura, Johann Henckel, Alister Hart

Published in: BMC Musculoskeletal Disorders | Issue 1/2022

Login to get access

Abstract

Background

Titanium, which is known to be a highly biologically inert element, is one of the most commonly used metals in orthopaedic implants. While cobalt and chromium blood metal ion testing is routinely used in the clinical monitoring of patients with metal-on-metal hip implants, much less is known about the levels of titanium in patients with other implant types. The aim of this study was to better understand the normal ranges of blood titanium levels in patients implanted with large and sliding titanium constructs by comparison with reference levels from conventional titanium hips.

Methods

This study examined data collected from 136 patients. Over a period of 24 months, whole blood samples were collected from 41 patients implanted with large titanium implants: long (range 15 to 30 cm) spine rods with a sliding mechanism (“spine rods”, n = 18), long bone tumour implants (“tumour implants”, n = 13) and 3D-printed customised massive acetabular defect implants (“massive acetabular implants”, n = 10). This data was compared with standard, uncemented primary titanium hip implants (“standard hips”, 15 cm long) (n = 95). Clinical, imaging and blood titanium levels data were collected for all patients and compared statistically between the different groups.

Results

The median (range) of blood titanium levels of the standard hip, spine rods, femoral tumour implants and massive acetabular implants were 1.2 ppb (0.6–4.9), 9.7 ppb (4.0–25.4), 2.6 ppb (0.4–104.4) and 5.7 ppb (1.6–31.5) respectively. Spine rods and massive acetabular implants had significantly greater blood titanium levels compared to the standard hips group (p < 0.001).

Conclusion

This study showed that titanium orthopaedic implants that are large and/or have a sliding mechanism have higher blood titanium levels compared to well-functioning, conventionally sized titanium hips. Reassuringly, the increased levels did not appear to induce adverse metal reactions. This study provides useful baseline data for future studies aimed at assessing blood titanium levels as a biomarker for implant function.
Literature
1.
go back to reference Katti KS, Verma D, Katti DR. Materials for joint replacement. Jt Replace Technol. 2008;81–104. Katti KS, Verma D, Katti DR. Materials for joint replacement. Jt Replace Technol. 2008;81–104.
2.
go back to reference Asri RIM, Harun WSW, Samykano M, Lah NAC, Ghani SAC, Tarlochan F, et al. Corrosion and surface modification on biocompatible metals: a review. Mater Sci Eng C. 2017;77:1261–74.CrossRef Asri RIM, Harun WSW, Samykano M, Lah NAC, Ghani SAC, Tarlochan F, et al. Corrosion and surface modification on biocompatible metals: a review. Mater Sci Eng C. 2017;77:1261–74.CrossRef
3.
go back to reference Swiatkowska I, Martin N, Hart AJ. Blood titanium level as a biomarker of orthopaedic implant wear. J Trace Elem Med Biol. 2018;2019(53):120–8. Swiatkowska I, Martin N, Hart AJ. Blood titanium level as a biomarker of orthopaedic implant wear. J Trace Elem Med Biol. 2018;2019(53):120–8.
4.
go back to reference Kaur M, Singh K. Review on titanium and titanium based alloys as biomaterials for orthopaedic applications. Mater Sci Eng C. 2019;102:844–62.CrossRef Kaur M, Singh K. Review on titanium and titanium based alloys as biomaterials for orthopaedic applications. Mater Sci Eng C. 2019;102:844–62.CrossRef
5.
go back to reference Lomer MCE, Hutchinson C, Volkert S, Greenfield SM, Catterall A, Thompson RPH, et al. Dietary sources of inorganic microparticles and their intake in healthy subjects and patients with Crohn’s disease. Br J Nutr. 2004;92:947–55.CrossRef Lomer MCE, Hutchinson C, Volkert S, Greenfield SM, Catterall A, Thompson RPH, et al. Dietary sources of inorganic microparticles and their intake in healthy subjects and patients with Crohn’s disease. Br J Nutr. 2004;92:947–55.CrossRef
6.
go back to reference Zhang T, Sze KY, Peng ZW, Cheung KMC, Lui YF, Wong YW, et al. Systematic investigation of metallosis associated with magnetically controlled growing rod implantation for early-onset scoliosis: Investigating metallosis in magnetically controlled growing rod surgery. Bone Jt J. 2020;102:1375–83.CrossRef Zhang T, Sze KY, Peng ZW, Cheung KMC, Lui YF, Wong YW, et al. Systematic investigation of metallosis associated with magnetically controlled growing rod implantation for early-onset scoliosis: Investigating metallosis in magnetically controlled growing rod surgery. Bone Jt J. 2020;102:1375–83.CrossRef
8.
go back to reference Ichinose S, Muneta T, Sekiya I, Itoh S, Aoki H, Tagami M. The study of metal ion release and cytotoxicity in Co-Cr-Mo and Ti-Al-V alloy in total knee prosthesis - scanning electron microscopic observation. J Mater Sci Mater Med. 2003;14:79–86.CrossRef Ichinose S, Muneta T, Sekiya I, Itoh S, Aoki H, Tagami M. The study of metal ion release and cytotoxicity in Co-Cr-Mo and Ti-Al-V alloy in total knee prosthesis - scanning electron microscopic observation. J Mater Sci Mater Med. 2003;14:79–86.CrossRef
9.
go back to reference Borde MD, Sapare S, Schutgens E, Ali C, Noordeen H. Analysis of serum levels of titanium and aluminium ions in patients with early onset scoliosis operated upon using the magnetic growing rod—a single centre study of 14 patients. Spine Deform. 2021;9:1473–8.CrossRef Borde MD, Sapare S, Schutgens E, Ali C, Noordeen H. Analysis of serum levels of titanium and aluminium ions in patients with early onset scoliosis operated upon using the magnetic growing rod—a single centre study of 14 patients. Spine Deform. 2021;9:1473–8.CrossRef
10.
go back to reference Yilgor C, Efendiyev A, Akbiyik F, Demirkiran G, Senkoylu A, Alanay A, et al. Metal ion release during growth-friendly instrumentation for early-onset scoliosis: a preliminary study. Spine Deform. 2018;6:48–53.CrossRef Yilgor C, Efendiyev A, Akbiyik F, Demirkiran G, Senkoylu A, Alanay A, et al. Metal ion release during growth-friendly instrumentation for early-onset scoliosis: a preliminary study. Spine Deform. 2018;6:48–53.CrossRef
11.
go back to reference Siddiqi O, Urquhart JC, Rasoulinejad P. A systematic review of metal ion concentrations following instrumented spinal fusion. Spine Deform. 2021;9:13–40.CrossRef Siddiqi O, Urquhart JC, Rasoulinejad P. A systematic review of metal ion concentrations following instrumented spinal fusion. Spine Deform. 2021;9:13–40.CrossRef
12.
go back to reference Hart AJ, Sabah SA, Bandi AS, Maggiore P, Tarassoli P, Sampson B, et al. Sensitivity and specificity of blood cobalt and chromium metal ions for predicting failure of metal-on-metal hip replacement. J Bone Jt Surg Ser B. 2011;93 B:1308–13.CrossRef Hart AJ, Sabah SA, Bandi AS, Maggiore P, Tarassoli P, Sampson B, et al. Sensitivity and specificity of blood cobalt and chromium metal ions for predicting failure of metal-on-metal hip replacement. J Bone Jt Surg Ser B. 2011;93 B:1308–13.CrossRef
13.
go back to reference Swiatkowska I, Martin NG, Henckel J, Apthorp H, Hamshere J, Hart AJ. Blood and plasma titanium levels associated with well-functioning hip implants. J Trace Elem Med Biol. 2019;2020(57):9–17. Swiatkowska I, Martin NG, Henckel J, Apthorp H, Hamshere J, Hart AJ. Blood and plasma titanium levels associated with well-functioning hip implants. J Trace Elem Med Biol. 2019;2020(57):9–17.
14.
go back to reference Calderón SAL, Kuechle J, Raskin KA, Hornicek FJ. Lower extremity megaprostheses in orthopaedic oncology. J Am Acad Orthop Surg. 2018;26:e249–57.CrossRef Calderón SAL, Kuechle J, Raskin KA, Hornicek FJ. Lower extremity megaprostheses in orthopaedic oncology. J Am Acad Orthop Surg. 2018;26:e249–57.CrossRef
15.
go back to reference Di Laura A, Henckel J, Wescott R, Hothi H, Hart AJ. The effect of metal artefact on the design of custom 3D printed acetabular implants. 3D Print Med. 2020;6:1–11.CrossRef Di Laura A, Henckel J, Wescott R, Hothi H, Hart AJ. The effect of metal artefact on the design of custom 3D printed acetabular implants. 3D Print Med. 2020;6:1–11.CrossRef
16.
go back to reference Nuevo-Ordóñez Y, Montes-Bayón M, Blanco-González E, Paz-Aparicio J, Raimundez JD, Tejerina JM, et al. Titanium release in serum of patients with different bone fixation implants and its interaction with serum biomolecules at physiological levels. Anal Bioanal Chem. 2011;401:2747–54.CrossRef Nuevo-Ordóñez Y, Montes-Bayón M, Blanco-González E, Paz-Aparicio J, Raimundez JD, Tejerina JM, et al. Titanium release in serum of patients with different bone fixation implants and its interaction with serum biomolecules at physiological levels. Anal Bioanal Chem. 2011;401:2747–54.CrossRef
17.
go back to reference Engh CA, MacDonald SJ, Sritulanondha S, Thompson A, Naudie D, Engh CA. 2008 John Charnley award: metal ion levels after metal-on-metal total hip arthroplasty: a randomized trial. Clin Orthop Relat Res. 2009;467:101–11.CrossRef Engh CA, MacDonald SJ, Sritulanondha S, Thompson A, Naudie D, Engh CA. 2008 John Charnley award: metal ion levels after metal-on-metal total hip arthroplasty: a randomized trial. Clin Orthop Relat Res. 2009;467:101–11.CrossRef
18.
go back to reference Dunstan E, Sanghrajka AP, Tilley S, Unwin P, Blunn G, Cannon SR, et al. Metal ion levels after metal-on-metal proximal femoral replacements. A 30-year follow-up. J Bone Jt Surg Ser B. 2005;87:628–31.CrossRef Dunstan E, Sanghrajka AP, Tilley S, Unwin P, Blunn G, Cannon SR, et al. Metal ion levels after metal-on-metal proximal femoral replacements. A 30-year follow-up. J Bone Jt Surg Ser B. 2005;87:628–31.CrossRef
19.
go back to reference Sarmiento-González A, Marchante-Gayón JM, Tejerina-Lobo JM, Paz-Jiménez J, Sanz-Medel A. High-resolution ICP-MS determination of Ti, V, Cr Co, Ni, and Mo in human blood and urine of patients implanted with a hip or knee prosthesis. Anal Bioanal Chem. 2008;391:2583–9.CrossRef Sarmiento-González A, Marchante-Gayón JM, Tejerina-Lobo JM, Paz-Jiménez J, Sanz-Medel A. High-resolution ICP-MS determination of Ti, V, Cr Co, Ni, and Mo in human blood and urine of patients implanted with a hip or knee prosthesis. Anal Bioanal Chem. 2008;391:2583–9.CrossRef
20.
go back to reference Vendittoli PA, Roy A, Mottard S, Girard J, Lusignan D, Lavigne M. Metal ion release from bearing wear and corrosion with 28 mm and large-diameter metal-on-metal bearing articulations: a follow-up study. J Bone Jt Surg Ser B. 2010;92:12–9.CrossRef Vendittoli PA, Roy A, Mottard S, Girard J, Lusignan D, Lavigne M. Metal ion release from bearing wear and corrosion with 28 mm and large-diameter metal-on-metal bearing articulations: a follow-up study. J Bone Jt Surg Ser B. 2010;92:12–9.CrossRef
21.
go back to reference Omlor GW, Kretzer JP, Reinders J, Streit MR, Bruckner T, Gotterbarm T, et al. In vivo serum titanium ion levels following modular neck total hip arthroplasty-10 year results in 67 patients. Acta Biomater. 2013;9:6278–82.CrossRef Omlor GW, Kretzer JP, Reinders J, Streit MR, Bruckner T, Gotterbarm T, et al. In vivo serum titanium ion levels following modular neck total hip arthroplasty-10 year results in 67 patients. Acta Biomater. 2013;9:6278–82.CrossRef
22.
go back to reference Levine BR, Hsu AR, Skipor AK, Hallab NJ, Paprosky WG, Jacobs JJ. Effect of a second joint arthroplasty on metal ion levels after primary total hip arthroplasty. Am J Orthop (Belle Mead NJ). 2013;42:512–8. Levine BR, Hsu AR, Skipor AK, Hallab NJ, Paprosky WG, Jacobs JJ. Effect of a second joint arthroplasty on metal ion levels after primary total hip arthroplasty. Am J Orthop (Belle Mead NJ). 2013;42:512–8.
23.
go back to reference Gofton W, Beaule PE. Serum metal ions with a titanium modular neck total hip replacement system. J Arthroplasty. 2015;30:1781–6.CrossRef Gofton W, Beaule PE. Serum metal ions with a titanium modular neck total hip replacement system. J Arthroplasty. 2015;30:1781–6.CrossRef
24.
go back to reference Nam D, Keeney JA, Nunley RM, Johnson SR, Clohisy JC, Barrack RL. Metal ion concentrations in young, active patients following total hip arthroplasty with the use of modern bearing couples. J Arthroplasty. 2015;30:2227–32.CrossRef Nam D, Keeney JA, Nunley RM, Johnson SR, Clohisy JC, Barrack RL. Metal ion concentrations in young, active patients following total hip arthroplasty with the use of modern bearing couples. J Arthroplasty. 2015;30:2227–32.CrossRef
25.
go back to reference Yi Z, Bo Z, Bin S, Jing Y, Zongke Z, Fuxing P. Clinical results and metal ion levels after ceramic-on-metal total hip arthroplasty: a mean 50-month prospective single-center study. J Arthroplasty. 2016;31:438–41.CrossRef Yi Z, Bo Z, Bin S, Jing Y, Zongke Z, Fuxing P. Clinical results and metal ion levels after ceramic-on-metal total hip arthroplasty: a mean 50-month prospective single-center study. J Arthroplasty. 2016;31:438–41.CrossRef
26.
go back to reference Sampson B, Hart A. Clinical usefulness of blood metal measurements to assess the failure of metal-on-metal hip implants. Ann Clin Biochem. 2012;49:118–31.CrossRef Sampson B, Hart A. Clinical usefulness of blood metal measurements to assess the failure of metal-on-metal hip implants. Ann Clin Biochem. 2012;49:118–31.CrossRef
27.
go back to reference Borde MD, Sapare S, Schutgens E, Ali C, Noordeen H. Analysis of serum levels of titanium and aluminium ions in patients with early onset scoliosis operated upon using the magnetic growing rod—a single centre study of 14 patients. Spine Deform. 2021;9:1473–8.CrossRef Borde MD, Sapare S, Schutgens E, Ali C, Noordeen H. Analysis of serum levels of titanium and aluminium ions in patients with early onset scoliosis operated upon using the magnetic growing rod—a single centre study of 14 patients. Spine Deform. 2021;9:1473–8.CrossRef
28.
go back to reference Li Y, Graham CK, Robbins C, Caird MS, Farley FA. Elevated serum titanium levels in children with early onset scoliosis treated with growth-friendly instrumentation. J Pediatr Orthop. 2020;40:E420–3.CrossRef Li Y, Graham CK, Robbins C, Caird MS, Farley FA. Elevated serum titanium levels in children with early onset scoliosis treated with growth-friendly instrumentation. J Pediatr Orthop. 2020;40:E420–3.CrossRef
Metadata
Title
Blood titanium levels in patients with large and sliding titanium implants
Authors
Martina Tognini
Harry Hothi
Stewart Tucker
Edel Broomfield
Masood Shafafy
Panos Gikas
Anna Di Laura
Johann Henckel
Alister Hart
Publication date
01-12-2022
Publisher
BioMed Central
Keyword
Bone Tumor
Published in
BMC Musculoskeletal Disorders / Issue 1/2022
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-022-05717-8

Other articles of this Issue 1/2022

BMC Musculoskeletal Disorders 1/2022 Go to the issue