Skip to main content
Top
Published in: BMC Anesthesiology 1/2020

Open Access 01-12-2020 | Blood Pressure Measurement | Research article

Non-invasive continuous blood pressure monitoring (ClearSight™ system) during shoulder surgery in the beach chair position: a prospective self-controlled study

Authors: Konrad Chachula, Florian Lieb, Florian Hess, Joellen Welter, Nicole Graf, Alexander Dullenkopf

Published in: BMC Anesthesiology | Issue 1/2020

Login to get access

Abstract

Background

The beach chair position that is commonly used in shoulder surgery is associated with relative hypovolemia, which leads to a reduction in arterial blood pressure. The effects of patient positioning on the accuracy of non-invasive continuous blood pressure monitoring with the ClearSight™ system (CS-BP; Edwards Lifesciences, Irvine CA, USA) have not been studied extensively. Our research aim was to assess agreement levels between CS-BP measurements with traditional blood pressure monitoring techniques.

Methods

For this prospective self-controlled study, we included 20 consecutively treated adult patients undergoing elective shoulder surgery in the beach chair position. We performed Bland-Altman analyses to determine agreement levels between blood pressure values from CS-BP and standard non-invasive (NIBP) methods. Perioperative measurements were done in both the supine (as reference) and beach chair surgical positions. Additionally, we compared invasive blood pressure (IBP) measurements with both the non-invasive methods (CS-BP and NIBP) in a sub-group of patients (n = 10) who required arterial blood pressure monitoring.

Results

We analyzed 229 data points (116 supine, 113 beach chair) from the entire cohort; per patient measurements were based on surgical length (range 3–9 supine, 2–10 beach chair). The mean difference (±SD; 95% limits of agreement) in the mean arterial pressure (MAP) between CS-BP and NIBP was − 0.9 (±11.0; − 24.0–22.2) in the beach chair position and − 4.9 mmHg (±11.8; − 28.0–18.2) when supine. In the sub-group, the difference between CS-BP and IBP in the beach chair position was − 1.6 mmHg (±16.0; − 32.9–29.7) and − 2.8 mmHg (±15.3; − 32.8–27.1) in the supine position. Between NIBP and IBP, we detected a difference of 3.0 mmHg (±9.1; − 20.8–14.7) in the beach chair position, and 4.6 mmHg (±13.3; − 21.4–30.6) in the supine position.

Conclusions

We found clinically acceptable mean differences in MAP measurements between the ClearSight™ and non-invasive oscillometric blood pressure systems when patients were in either the supine or beach chair position. For all comparisons of the monitoring systems and surgical positions, the standard deviations and limits of agreement were wide.

Trial registration

This study was prospectively registered at the German Clinical Trial Register (www.DRKS.de; DRKS00013773). Registered 26/01/2018.
Literature
1.
go back to reference Netea RT, Lenders JW, Smits P, Thien T. Influence of body and arm position on blood pressure readings: an overview. J Hypertens. 2003;21:237–41.CrossRef Netea RT, Lenders JW, Smits P, Thien T. Influence of body and arm position on blood pressure readings: an overview. J Hypertens. 2003;21:237–41.CrossRef
2.
go back to reference Zeuzem-Lampert C, Groene P, Brummer V, Hofmann-Kiefer K. Cardiorespiratory effects of perioperative positioning techniques. Anaesthesist. 2019;68:805–13.CrossRef Zeuzem-Lampert C, Groene P, Brummer V, Hofmann-Kiefer K. Cardiorespiratory effects of perioperative positioning techniques. Anaesthesist. 2019;68:805–13.CrossRef
3.
go back to reference Gillespie R, Shishani Y, Streit J, Wanner JP, McCrum C, Syed T, Haas A, Gobezie R. The safety of controlled hypotension for shoulder arthroscopy in the beach-chair position. J Bone Joint Surg Am. 2012;94:1284–90.CrossRef Gillespie R, Shishani Y, Streit J, Wanner JP, McCrum C, Syed T, Haas A, Gobezie R. The safety of controlled hypotension for shoulder arthroscopy in the beach-chair position. J Bone Joint Surg Am. 2012;94:1284–90.CrossRef
4.
go back to reference Jiang J, Zhou R, Li B, Xue F. Is deliberate hypotension a safe technique for orthopedic surgery?: a systematic review and meta-analysis of parallel randomized controlled trials. J Orthop Surg Res. 2019;14:409.CrossRef Jiang J, Zhou R, Li B, Xue F. Is deliberate hypotension a safe technique for orthopedic surgery?: a systematic review and meta-analysis of parallel randomized controlled trials. J Orthop Surg Res. 2019;14:409.CrossRef
5.
go back to reference Monk TG, Bronsert MR, Henderson WG, Mangione MP, Sum-Ping ST, Bentt DR, Nguyen JD, Richman JS, Meguid RA, Hammermeister KE. Association between intraoperative hypotension and hypertension and 30-day postoperative mortality in noncardiac surgery. Anesthesiology. 2015;123:307–19.CrossRef Monk TG, Bronsert MR, Henderson WG, Mangione MP, Sum-Ping ST, Bentt DR, Nguyen JD, Richman JS, Meguid RA, Hammermeister KE. Association between intraoperative hypotension and hypertension and 30-day postoperative mortality in noncardiac surgery. Anesthesiology. 2015;123:307–19.CrossRef
6.
go back to reference Sessler DI, Meyhoff CS, Zimmerman NM, Mao G, Leslie K, Vasquez SM, Balaji P, Alvarez-Garcia J, Cavalcanti AB, Parlow JL, Rahate PV, Seeberger MD, Gossetti B, Walker SA, Premchand RK, Dahl RM, Duceppe E, Rodseth R, Botto F, Devereaux PJ. Period-dependent associations between hypotension during and for four days after noncardiac surgery and a composite of myocardial infarction and death: a substudy of the POISE-2 trial. Anesthesiology. 2018;128:317–27.CrossRef Sessler DI, Meyhoff CS, Zimmerman NM, Mao G, Leslie K, Vasquez SM, Balaji P, Alvarez-Garcia J, Cavalcanti AB, Parlow JL, Rahate PV, Seeberger MD, Gossetti B, Walker SA, Premchand RK, Dahl RM, Duceppe E, Rodseth R, Botto F, Devereaux PJ. Period-dependent associations between hypotension during and for four days after noncardiac surgery and a composite of myocardial infarction and death: a substudy of the POISE-2 trial. Anesthesiology. 2018;128:317–27.CrossRef
7.
go back to reference Maheshwari K, Ahuja S, Khanna AK, Mao G, Perez-Protto S, Farag E, Turan A, Kurz A, Sessler DI. Association between perioperative hypotension and delirium in postoperative critically ill patients: a retrospective cohort analysis. Anesth Analg. 2020;130:636–43.CrossRef Maheshwari K, Ahuja S, Khanna AK, Mao G, Perez-Protto S, Farag E, Turan A, Kurz A, Sessler DI. Association between perioperative hypotension and delirium in postoperative critically ill patients: a retrospective cohort analysis. Anesth Analg. 2020;130:636–43.CrossRef
8.
go back to reference Cicolini G, Pizzi C, Palma E, Bucci M, Schioppa F, Mezzetti A, Manzoli L. Differences in blood pressure by body position (supine, Fowler's, and sitting) in hypertensive subjects. Am J Hypertens. 2011;24:1073–9.CrossRef Cicolini G, Pizzi C, Palma E, Bucci M, Schioppa F, Mezzetti A, Manzoli L. Differences in blood pressure by body position (supine, Fowler's, and sitting) in hypertensive subjects. Am J Hypertens. 2011;24:1073–9.CrossRef
9.
go back to reference Netea RT, Lenders JW, Smits P, Thien T. Both body and arm position significantly influence blood pressure measurement. J Hum Hypertens. 2003;17:459–62.CrossRef Netea RT, Lenders JW, Smits P, Thien T. Both body and arm position significantly influence blood pressure measurement. J Hum Hypertens. 2003;17:459–62.CrossRef
10.
go back to reference Truijen J, van Lieshout JJ, Wesselink WA, Westerhof BE. Noninvasive continuous hemodynamic monitoring. J Clin Monit Comput. 2012;26:267–78.CrossRef Truijen J, van Lieshout JJ, Wesselink WA, Westerhof BE. Noninvasive continuous hemodynamic monitoring. J Clin Monit Comput. 2012;26:267–78.CrossRef
11.
go back to reference Chung E, Chen G, Alexander B, Cannesson M. Non-invasive continuous blood pressure monitoring: a review of current applications. Front Med. 2013;7:91–101.CrossRef Chung E, Chen G, Alexander B, Cannesson M. Non-invasive continuous blood pressure monitoring: a review of current applications. Front Med. 2013;7:91–101.CrossRef
12.
go back to reference Raggi EP, Sakai T. Update on finger-application-type noninvasive continuous hemodynamic monitors (CNAP and ccNexfin): physical principles, validation, and clinical use. Semin Cardiothorac Vasc Anesth. 2017;21:321–9.CrossRef Raggi EP, Sakai T. Update on finger-application-type noninvasive continuous hemodynamic monitors (CNAP and ccNexfin): physical principles, validation, and clinical use. Semin Cardiothorac Vasc Anesth. 2017;21:321–9.CrossRef
13.
go back to reference Meidert AS, Nold JS, Hornung R, Paulus AC, Zwissler B, Czerner S. The impact of continuous non-invasive arterial blood pressure monitoring on blood pressure stability during general anaesthesia in orthopaedic patients: a randomised trial. Europ J Anaesth. 2017;34:716–22.CrossRef Meidert AS, Nold JS, Hornung R, Paulus AC, Zwissler B, Czerner S. The impact of continuous non-invasive arterial blood pressure monitoring on blood pressure stability during general anaesthesia in orthopaedic patients: a randomised trial. Europ J Anaesth. 2017;34:716–22.CrossRef
14.
go back to reference Michard F, Giglio MT, Brienza N. Perioperative goal-directed therapy with uncalibrated pulse contour methods: impact on fluid management and postoperative outcome. Br J Anaesth. 2017;119:22–30.CrossRef Michard F, Giglio MT, Brienza N. Perioperative goal-directed therapy with uncalibrated pulse contour methods: impact on fluid management and postoperative outcome. Br J Anaesth. 2017;119:22–30.CrossRef
15.
go back to reference Rogge DE, Nicklas JY, Schon G, Grothe O, Haas SA, Reuter DA, Saugel B. Continuous noninvasive arterial pressure monitoring in obese patients during bariatric surgery: an evaluation of the vascular unloading technique (Clearsight system). Anesth Analg. 2019;128:477–83.CrossRef Rogge DE, Nicklas JY, Schon G, Grothe O, Haas SA, Reuter DA, Saugel B. Continuous noninvasive arterial pressure monitoring in obese patients during bariatric surgery: an evaluation of the vascular unloading technique (Clearsight system). Anesth Analg. 2019;128:477–83.CrossRef
16.
go back to reference Fischer MO, Avram R, Carjaliu I, Massetti M, Gerard JL, Hanouz JL, Fellahi JL. Non-invasive continuous arterial pressure and cardiac index monitoring with Nexfin after cardiac surgery. Br J Anaesth. 2012;109:514–21.CrossRef Fischer MO, Avram R, Carjaliu I, Massetti M, Gerard JL, Hanouz JL, Fellahi JL. Non-invasive continuous arterial pressure and cardiac index monitoring with Nexfin after cardiac surgery. Br J Anaesth. 2012;109:514–21.CrossRef
17.
go back to reference Martina JR, Westerhof BE, van Goudoever J, de Beaumont EM, Truijen J, Kim YS, Immink RV, Jobsis DA, Hollmann MW, Lahpor JR, de Mol BA, van Lieshout JJ. Noninvasive continuous arterial blood pressure monitoring with Nexfin (R). Anesthesiology. 2012;116:1092–103.CrossRef Martina JR, Westerhof BE, van Goudoever J, de Beaumont EM, Truijen J, Kim YS, Immink RV, Jobsis DA, Hollmann MW, Lahpor JR, de Mol BA, van Lieshout JJ. Noninvasive continuous arterial blood pressure monitoring with Nexfin (R). Anesthesiology. 2012;116:1092–103.CrossRef
18.
go back to reference Balzer F, Habicher M, Sander M, Sterr J, Scholz S, Feldheiser A, Muller M, Perka C, Treskatsch S. Comparison of the non-invasive Nexfin (R) monitor with conventional methods for the measurement of arterial blood pressure in moderate risk orthopaedic surgery patients. J Int Med Res. 2016;44:832–43.CrossRef Balzer F, Habicher M, Sander M, Sterr J, Scholz S, Feldheiser A, Muller M, Perka C, Treskatsch S. Comparison of the non-invasive Nexfin (R) monitor with conventional methods for the measurement of arterial blood pressure in moderate risk orthopaedic surgery patients. J Int Med Res. 2016;44:832–43.CrossRef
19.
go back to reference Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17:571–82.CrossRef Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17:571–82.CrossRef
20.
go back to reference Hohn A, Defosse JM, Becker S, Steffen C, Wappler F, Sakka SG. Non-invasive continuous arterial pressure monitoring with Nexfin does not sufficiently replace invasive measurements in critically ill patients. Br J Anaesth. 2013;111:178–84.CrossRef Hohn A, Defosse JM, Becker S, Steffen C, Wappler F, Sakka SG. Non-invasive continuous arterial pressure monitoring with Nexfin does not sufficiently replace invasive measurements in critically ill patients. Br J Anaesth. 2013;111:178–84.CrossRef
21.
go back to reference Silke B, McAuley D. Accuracy and precision of blood pressure determination with the Finapres: an overview using re-sampling statistics. J Hum Hypertens. 1998;12:403–9.CrossRef Silke B, McAuley D. Accuracy and precision of blood pressure determination with the Finapres: an overview using re-sampling statistics. J Hum Hypertens. 1998;12:403–9.CrossRef
22.
go back to reference Kuck K, Baker PD. Perioperative noninvasive blood pressure monitoring. Anesth Analg. 2018;127:408–11.CrossRef Kuck K, Baker PD. Perioperative noninvasive blood pressure monitoring. Anesth Analg. 2018;127:408–11.CrossRef
23.
go back to reference Kim SH, Lilot M, Sidhu KS, Rinehart J, Yu Z, Canales C, Cannesson M. Accuracy and precision of continuous noninvasive arterial pressure monitoring compared with invasive arterial pressure: a systematic review and meta-analysis. Anesthesiology. 2014;120:1080–97.CrossRef Kim SH, Lilot M, Sidhu KS, Rinehart J, Yu Z, Canales C, Cannesson M. Accuracy and precision of continuous noninvasive arterial pressure monitoring compared with invasive arterial pressure: a systematic review and meta-analysis. Anesthesiology. 2014;120:1080–97.CrossRef
24.
go back to reference Saugel B, Grothe O, Nicklas JY. Error grid analysis for arterial pressure method comparison studies. Anesth Analg. 2018;126:1177–85.CrossRef Saugel B, Grothe O, Nicklas JY. Error grid analysis for arterial pressure method comparison studies. Anesth Analg. 2018;126:1177–85.CrossRef
Metadata
Title
Non-invasive continuous blood pressure monitoring (ClearSight™ system) during shoulder surgery in the beach chair position: a prospective self-controlled study
Authors
Konrad Chachula
Florian Lieb
Florian Hess
Joellen Welter
Nicole Graf
Alexander Dullenkopf
Publication date
01-12-2020
Publisher
BioMed Central
Published in
BMC Anesthesiology / Issue 1/2020
Electronic ISSN: 1471-2253
DOI
https://doi.org/10.1186/s12871-020-01185-6

Other articles of this Issue 1/2020