Published in: Journal of Interventional Cardiac Electrophysiology 8/2023

08-02-2023 | Atrioventricular Block

Long-term follow-up of selective and non-selective His bundle pacing leads in patients with atrioventricular block

Authors: Maria Hee Jung Park Frausing, Aleksander Laust Bæk, Jens Kristensen, Christian Gerdes, Jens Cosedis Nielsen, Mads Brix Kronborg

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 8/2023

Login to get access

Abstract

Background

His bundle pacing (HBP) is a novel treatment with limited knowledge on long-term outcome. We aimed to assess long-term safety and effectiveness of HBP in patients with atrioventricular block treated with HBP and a back-up right ventricular pacing (RVP) lead.

Methods

We included 38 patients from a completed single-center, randomized controlled cross-over trial designed to compare left ventricular (LV) function after 12 months of HBP vs. RVP conducted between September 2007 and August 2011. Lead performance beyond the 2-year study period was assessed based on a retrospective review of capture thresholds, sensing, impedance, energy consumption, and rate of HBP interruption.

Results

Patients were followed for a mean of 7 ± 4 years. Both at baseline and during follow-up, HBP leads displayed significantly higher capture thresholds than RVP leads (P < 0.001), multifold higher energy consumption (P < 0.001), and lower sensing amplitudes (P < 0.001). During follow-up, 17 (53%) HBP leads were deactivated or abandoned. The principal cause for HBP interruption was high pacing thresholds in patients with preserved LVEF during RVP. Device longevity was shorter than that of contemporary cohorts treated with dual-chamber pacing or CRT, and time to first device exchange was 6.8 ± 1.5 years. No lead dislodgements occurred, but four patients (10%) developed device-related infections requiring device extraction.

Conclusion

HBP was interrupted in > 50% of patients during long-term follow-up. The principal cause was unacceptably high capture thresholds and no significant difference in LV function with HBP compared with RVP. Device longevity was shorter, and infection rates were higher than anticipated.

Graphical Abstract

Appendix
Available only for authorised users
Literature
1.
go back to reference Yu CM, Fang F, Luo XX, Zhang Q, Azlan H, Razali O. Long-term follow-up results of the pacing to avoid cardiac enlargement (PACE) trial. Eur J Heart Fail. 2014;16(9):1016–25.CrossRefPubMed Yu CM, Fang F, Luo XX, Zhang Q, Azlan H, Razali O. Long-term follow-up results of the pacing to avoid cardiac enlargement (PACE) trial. Eur J Heart Fail. 2014;16(9):1016–25.CrossRefPubMed
2.
go back to reference Kiehl EL, Makki T, Kumar R, et al. Incidence and predictors of right ventricular pacing-induced cardiomyopathy in patients with complete atrioventricular block and preserved left ventricular systolic function. Heart Rhythm. 2016;13(12):2272–8.CrossRefPubMed Kiehl EL, Makki T, Kumar R, et al. Incidence and predictors of right ventricular pacing-induced cardiomyopathy in patients with complete atrioventricular block and preserved left ventricular systolic function. Heart Rhythm. 2016;13(12):2272–8.CrossRefPubMed
3.
go back to reference Tayal B, Fruelund P, Sogaard P, et al. Incidence of heart failure after pacemaker implantation: a nationwide Danish registry-based follow-up study. Eur Heart J. 2019;40(44):3641–8.CrossRefPubMed Tayal B, Fruelund P, Sogaard P, et al. Incidence of heart failure after pacemaker implantation: a nationwide Danish registry-based follow-up study. Eur Heart J. 2019;40(44):3641–8.CrossRefPubMed
4.
go back to reference Wilkoff BL, Cook JR, Epstein AE, et al. Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the dual chamber and VVI implantable defibrillator (DAVID) trial. JAMA. 2002;288(24):3115–23.CrossRefPubMed Wilkoff BL, Cook JR, Epstein AE, et al. Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the dual chamber and VVI implantable defibrillator (DAVID) trial. JAMA. 2002;288(24):3115–23.CrossRefPubMed
5.
go back to reference Kronborg MB, Mortensen PT, Poulsen SH, Gerdes JC, Jensen HK, Nielsen JC. His or para-His pacing preserves left ventricular function in atrioventricular block: a double-blind, randomized, crossover study. Europace. 2014;16(8):1189–96.CrossRefPubMed Kronborg MB, Mortensen PT, Poulsen SH, Gerdes JC, Jensen HK, Nielsen JC. His or para-His pacing preserves left ventricular function in atrioventricular block: a double-blind, randomized, crossover study. Europace. 2014;16(8):1189–96.CrossRefPubMed
6.
go back to reference Abdelrahman M, Subzposh FA, Beer D, et al. Clinical outcomes of His bundle pacing compared to right ventricular pacing. J Am Coll Cardiol. 2018;71(20):2319–30.CrossRefPubMed Abdelrahman M, Subzposh FA, Beer D, et al. Clinical outcomes of His bundle pacing compared to right ventricular pacing. J Am Coll Cardiol. 2018;71(20):2319–30.CrossRefPubMed
7.
go back to reference Vijayaraman P, Naperkowski A, Subzposh FA, et al. Permanent His-bundle pacing: long-term lead performance and clinical outcomes. Heart Rhythm. 2018;15(5):696–702.CrossRefPubMed Vijayaraman P, Naperkowski A, Subzposh FA, et al. Permanent His-bundle pacing: long-term lead performance and clinical outcomes. Heart Rhythm. 2018;15(5):696–702.CrossRefPubMed
8.
go back to reference Vijayaraman P, Dandamudi G, Lustgarten D, Ellenbogen KA. Permanent His bundle pacing: electrophysiological and echocardiographic observations from long-term follow-up. Pacing Clin Electrophysiol. 2017;40(7):883–91.CrossRefPubMed Vijayaraman P, Dandamudi G, Lustgarten D, Ellenbogen KA. Permanent His bundle pacing: electrophysiological and echocardiographic observations from long-term follow-up. Pacing Clin Electrophysiol. 2017;40(7):883–91.CrossRefPubMed
9.
go back to reference Teigeler T, Kolominsky J, Vo C, et al. Intermediate-term performance and safety of His-bundle pacing leads: a single-center experience. Heart Rhythm. 2021;18(5):743–9.CrossRefPubMed Teigeler T, Kolominsky J, Vo C, et al. Intermediate-term performance and safety of His-bundle pacing leads: a single-center experience. Heart Rhythm. 2021;18(5):743–9.CrossRefPubMed
10.
go back to reference Kronborg MB, Mortensen PT, Gerdes JC, Jensen HK, Nielsen JC. His and para-His pacing in AV block: feasibility and electrocardiographic findings. J Interv Card Electrophysiol. 2011;31(3):255–62.CrossRefPubMed Kronborg MB, Mortensen PT, Gerdes JC, Jensen HK, Nielsen JC. His and para-His pacing in AV block: feasibility and electrocardiographic findings. J Interv Card Electrophysiol. 2011;31(3):255–62.CrossRefPubMed
11.
go back to reference Zanon F, Ellenbogen KA, Dandamudi G, et al. Permanent His-bundle pacing: a systematic literature review and meta-analysis. Europace. 2018;20(11):1819–26.CrossRefPubMed Zanon F, Ellenbogen KA, Dandamudi G, et al. Permanent His-bundle pacing: a systematic literature review and meta-analysis. Europace. 2018;20(11):1819–26.CrossRefPubMed
12.
go back to reference Birnie DH, Wang J, Alings M, et al. Risk factors for infections involving cardiac implanted electronic devices. J Am Coll Cardiol. 2019;74(23):2845–54.CrossRefPubMed Birnie DH, Wang J, Alings M, et al. Risk factors for infections involving cardiac implanted electronic devices. J Am Coll Cardiol. 2019;74(23):2845–54.CrossRefPubMed
13.
go back to reference Olsen T, Jorgensen OD, Nielsen JC, Thogersen AM, Philbert BT, Johansen JB. Incidence of device-related infection in 97 750 patients: clinical data from the complete Danish device-cohort (1982–2018). Eur Heart J. 2019;40(23):1862–9.CrossRefPubMedPubMedCentral Olsen T, Jorgensen OD, Nielsen JC, Thogersen AM, Philbert BT, Johansen JB. Incidence of device-related infection in 97 750 patients: clinical data from the complete Danish device-cohort (1982–2018). Eur Heart J. 2019;40(23):1862–9.CrossRefPubMedPubMedCentral
14.
15.
go back to reference Ebert M, Jander N, Minners J et al. Long-term impact of right ventricular pacing on left ventricular systolic function in pacemaker recipients with preserved ejection fraction: results from a large single-center registry. J Am Heart Assoc. 5(7) (2016). Ebert M, Jander N, Minners J et al. Long-term impact of right ventricular pacing on left ventricular systolic function in pacemaker recipients with preserved ejection fraction: results from a large single-center registry. J Am Heart Assoc. 5(7) (2016).
16.
go back to reference Curtis AB, Worley SJ, Adamson PB, et al. Biventricular pacing for atrioventricular block and systolic dysfunction. N Engl J Med. 2013;368(17):1585–93.CrossRefPubMed Curtis AB, Worley SJ, Adamson PB, et al. Biventricular pacing for atrioventricular block and systolic dysfunction. N Engl J Med. 2013;368(17):1585–93.CrossRefPubMed
17.
go back to reference Padala SK, Master VM, Terricabras M, et al. Initial experience, safety, and feasibility of left bundle branch area pacing: a multicenter prospective study. JACC Clin Electrophysiol. 2020;6(14):1773–82.CrossRefPubMed Padala SK, Master VM, Terricabras M, et al. Initial experience, safety, and feasibility of left bundle branch area pacing: a multicenter prospective study. JACC Clin Electrophysiol. 2020;6(14):1773–82.CrossRefPubMed
18.
go back to reference Kirkfeldt RE, Johansen JB, Nohr EA, Jorgensen OD, Nielsen JC. Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark. Eur Heart J. 2014;35(18):1186–94.CrossRefPubMed Kirkfeldt RE, Johansen JB, Nohr EA, Jorgensen OD, Nielsen JC. Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark. Eur Heart J. 2014;35(18):1186–94.CrossRefPubMed
19.
go back to reference Barakat AF, Inashvili A, Alkukhun L, et al. Use trends and adverse reports of SelectSecure 3830 lead implantations in the United States: implications for His bundle pacing. Circ Arrhythm Electrophysiol. 2020;13(7):e008577.CrossRefPubMed Barakat AF, Inashvili A, Alkukhun L, et al. Use trends and adverse reports of SelectSecure 3830 lead implantations in the United States: implications for His bundle pacing. Circ Arrhythm Electrophysiol. 2020;13(7):e008577.CrossRefPubMed
20.
go back to reference Glikson M, Nielsen JC, Kronborg MB, et al. 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: developed by the Task Force on cardiac pacing and cardiac resynchronization therapy of the European Society of Cardiology (ESC) With the special contribution of the European Heart Rhythm Association (EHRA). Europace. 2021;42(35):3427–520. Glikson M, Nielsen JC, Kronborg MB, et al. 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: developed by the Task Force on cardiac pacing and cardiac resynchronization therapy of the European Society of Cardiology (ESC) With the special contribution of the European Heart Rhythm Association (EHRA). Europace. 2021;42(35):3427–520.
Metadata
Title
Long-term follow-up of selective and non-selective His bundle pacing leads in patients with atrioventricular block
Authors
Maria Hee Jung Park Frausing
Aleksander Laust Bæk
Jens Kristensen
Christian Gerdes
Jens Cosedis Nielsen
Mads Brix Kronborg
Publication date
08-02-2023

Other articles of this Issue 8/2023

Journal of Interventional Cardiac Electrophysiology 8/2023 Go to the issue