Electrogram voltage and pacing threshold before ablation, measured by mini-electrodes, predict parameters indicative of transmural lesions in the human atrium
Authors:
Carla Lázaro, Teresa Barrio-López, Eduardo Castellanos, Mercedes Ortiz, Martín Arceluz, Jesús Almendral
Published in:
Journal of Interventional Cardiac Electrophysiology
|
Issue 3/2020
Login to get access
Abstract
Purpose
An important attenuation of the atrial signal recorded with mini-electrodes (ME) embedded in an 8-mm tip was associated with a transmural radiofrequency lesion. Our aim was to assess if parameters obtained from ME or conventional bipoles before applications predict successful atrial lesions.
Methods
We prospectively included 33 consecutive patients undergoing cavotricuspid isthmus (CTI) ablation. Electrogram voltages and pacing thresholds were measured with ME and conventional bipoles before and after radiofrequency (RF) applications. The time before the loss of capture during applications was recorded. Lesions were considered successful, in accordance with preclinical data, if ME voltage decreased > 54%.
Results
Of 207 applications, 107 could be analyzed. During applications, voltages decreased more in the ME than in the conventional bipoles (66.8 ± 26.1% vs 37.5 ± 42.5%, P = 0.001). Likewise, pacing threshold increased significantly more using the ME (86.3 ± 22.9% ME, 52.6 ± 35.6% conventional, P = 0.001). ME pre-ablation voltages were significantly higher and pacing thresholds significantly lower in successful lesions (voltage 0.88 ± 0.71 vs 0.26 ± 0.18 mV, P = 0.0001; threshold 1.6 ± 1.7 vs 2.8 ± 3.0, P = 0.04). Neither of these parameters with conventional bipoles nor time to loss of capture showed differences. A ME voltage > 0.33 mV and a pacing threshold < 1.5 mA predicted a successful lesion with 0.78 and 0.6 sensitivity and 0.78 and 0.59 specificity.
Conclusions
Certain pre-ablation parameters derived from ME such as electrogram voltage and pacing threshold differ from those obtained by a conventional configuration and can predict a successful atrial lesion.