Assessment of two-dimensional (2D) and three-dimensional (3D) lower limb measurements in adults: Comparison of micro-dose and low-dose biplanar radiographs

Authors: Andrea B. Rosskopf, Christian W. A. Pfirrmann, Florian M. Buck

Published in: European Radiology | Issue 9/2016

Login to get access

Abstract

Objective

To evaluate reliability of 2D and 3D lower limb measurements in adults using micro-dose compared to low-dose biplanar radiographs(BPR).

Materials and methods

One hundred patients (mean 54.9 years) were examined twice using micro-dose and low-dose BPR. Length and mechanical axis of lower limbs were measured on the antero-posterior(ap) micro-dose and low-dose images by two independent readers. Femoral and tibial torsions of 50 patients were measured by two independent readers using reconstructed 3D-models based on the micro-dose and low-dose BPR. Intermethod and interreader agreements were calculated using descriptive statistics, intraclass-correlation-coefficient(ICC), and Bland-Altman analysis.

Results

Mean interreader-differences on micro-dose were 0.3 cm(range 0-1.0)/ 0.7°(0-2.9) for limb length/axis and 0.4 cm (0-1.0)/0.8°(0-3.3) on low-dose BPR. Mean intermethod-difference was 0.04 cm ± 0.2/0.04° ± 0.6 for limb length/axis. Interreader-ICC for limb length/axis was 0.999/0.991 on micro-dose and 0.999/0.987 on low-dose BPR. Interreader-ICC for micro-dose was 0.879/0.826 for femoral/ tibial torsion, for low-dose BPR was 0.924/0.909. Mean interreader-differences on micro-dose/low-dose BPR were 3°(0-13°)/2°(0°–12°) for femoral and 4°(0-18°)/3°(0°–10°) for tibial torsion. Mean intermethod-difference was -0.1° ± 5.0/-0.4° ± 2.9 for femoral/tibial torsion. Mean dose-area-product was significantly lower (9.9 times;p < 0.001) for micro-dose BPR.

Conclusion

2D-and 3D-measurements of lower limbs based on micro-dose BPR are reliable and provide a 10-times lower radiation dose.

Key points

Lower limb length and mechanical axis can be reliably measured with micro-dose.
Femoral and tibial torsion can be reliably assessed with micro-dose.
Micro-dose allows a huge reduction of radiation exposure.
Literature
1.
go back to reference Giles LG, Taylor JR (1981) Low-back pain associated with leg length inequality. Spine (Phila Pa 1976) 6:510–521CrossRef Giles LG, Taylor JR (1981) Low-back pain associated with leg length inequality. Spine (Phila Pa 1976) 6:510–521CrossRef
2.
go back to reference Brady RJ, Dean JB, Skinner TM, Gross MT (2003) Limb length inequality: clinical implications for assessment and intervention. J Orthop Sports Phys Ther 33:221–234CrossRefPubMed Brady RJ, Dean JB, Skinner TM, Gross MT (2003) Limb length inequality: clinical implications for assessment and intervention. J Orthop Sports Phys Ther 33:221–234CrossRefPubMed
4.
go back to reference Sharma L, Song J, Felson DT, Cahue S, Shamiyeh E, Dunlop DD (2001) The role of knee alignment in disease progression and functional decline in knee osteoarthritis. JAMA 286:188–195CrossRefPubMed Sharma L, Song J, Felson DT, Cahue S, Shamiyeh E, Dunlop DD (2001) The role of knee alignment in disease progression and functional decline in knee osteoarthritis. JAMA 286:188–195CrossRefPubMed
5.
go back to reference Parikh S, Noyes FR (2011) Patellofemoral disorders: role of computed tomography and magnetic resonance imaging in defining abnormal rotational lower limb alignment. Sports Health 3:158–169CrossRefPubMedPubMedCentral Parikh S, Noyes FR (2011) Patellofemoral disorders: role of computed tomography and magnetic resonance imaging in defining abnormal rotational lower limb alignment. Sports Health 3:158–169CrossRefPubMedPubMedCentral
6.
go back to reference Jeffery RS, Morris RW, Denham RA (1991) Coronal alignment after total knee replacement. J Bone Joint Surg Br 73:709–714PubMed Jeffery RS, Morris RW, Denham RA (1991) Coronal alignment after total knee replacement. J Bone Joint Surg Br 73:709–714PubMed
7.
go back to reference Pape D, Seil R, Adam F, Rupp S, Kohn D, Lobenhoffer P (2004) Imaging and preoperative planning of osteotomy of tibial head osteotomy. Orthopade 33:122–134CrossRefPubMed Pape D, Seil R, Adam F, Rupp S, Kohn D, Lobenhoffer P (2004) Imaging and preoperative planning of osteotomy of tibial head osteotomy. Orthopade 33:122–134CrossRefPubMed
8.
go back to reference Radtke K, Becher C, Noll Y, Ostermeier S (2010) Effect of limb rotation on radiographic alignment in total knee arthroplasties. Arch Orthop Trauma Surg 130:451–457CrossRefPubMed Radtke K, Becher C, Noll Y, Ostermeier S (2010) Effect of limb rotation on radiographic alignment in total knee arthroplasties. Arch Orthop Trauma Surg 130:451–457CrossRefPubMed
9.
go back to reference Fujimaki H, Inaba Y, Kobayashi N, Tezuka T, Hirata Y, Saito T (2013) Leg length discrepancy and lower limb alignment after total hip arthroplasty in unilateral hip osteoarthritis patients. J Orthop Sci 18:969–976CrossRefPubMed Fujimaki H, Inaba Y, Kobayashi N, Tezuka T, Hirata Y, Saito T (2013) Leg length discrepancy and lower limb alignment after total hip arthroplasty in unilateral hip osteoarthritis patients. J Orthop Sci 18:969–976CrossRefPubMed
10.
go back to reference Buck FM, Guggenberger R, Koch PP, Pfirrmann CW (2012) Femoral and tibial torsion measurements with 3D models based on low-dose biplanar radiographs in comparison with standard CT measurements. AJR Am J Roentgenol 199:W607–W612CrossRefPubMed Buck FM, Guggenberger R, Koch PP, Pfirrmann CW (2012) Femoral and tibial torsion measurements with 3D models based on low-dose biplanar radiographs in comparison with standard CT measurements. AJR Am J Roentgenol 199:W607–W612CrossRefPubMed
11.
go back to reference Rosskopf AB, Ramseier LE, Sutter R, Pfirrmann CW, Buck FM (2014) Femoral and tibial torsion measurement in children and adolescents: comparison of 3D models based on low-dose biplanar radiography and low-dose CT. AJR Am J Roentgenol 202:W285–W291CrossRefPubMed Rosskopf AB, Ramseier LE, Sutter R, Pfirrmann CW, Buck FM (2014) Femoral and tibial torsion measurement in children and adolescents: comparison of 3D models based on low-dose biplanar radiography and low-dose CT. AJR Am J Roentgenol 202:W285–W291CrossRefPubMed
12.
go back to reference Deschenes S, Charron G, Beaudoin G et al (2010) Diagnostic imaging of spinal deformities: reducing patients radiation dose with a new slot-scanning X-ray imager. Spine (Phila Pa 1976) 35:989–994CrossRef Deschenes S, Charron G, Beaudoin G et al (2010) Diagnostic imaging of spinal deformities: reducing patients radiation dose with a new slot-scanning X-ray imager. Spine (Phila Pa 1976) 35:989–994CrossRef
13.
go back to reference Dietrich TJ, Pfirrmann CW, Schwab A, Pankalla K, Buck FM (2013) Comparison of radiation dose, workflow, patient comfort and financial break-even of standard digital radiography and a novel biplanar low-dose X-ray system for upright full-length lower limb and whole spine radiography. Skeletal Radiol 42:959–967CrossRefPubMed Dietrich TJ, Pfirrmann CW, Schwab A, Pankalla K, Buck FM (2013) Comparison of radiation dose, workflow, patient comfort and financial break-even of standard digital radiography and a novel biplanar low-dose X-ray system for upright full-length lower limb and whole spine radiography. Skeletal Radiol 42:959–967CrossRefPubMed
14.
go back to reference Kalifa G, Charpak Y, Maccia C et al (1998) Evaluation of a new low-dose digital x-ray device: first dosimetric and clinical results in children. Pediatr Radiol 28:557–561CrossRefPubMed Kalifa G, Charpak Y, Maccia C et al (1998) Evaluation of a new low-dose digital x-ray device: first dosimetric and clinical results in children. Pediatr Radiol 28:557–561CrossRefPubMed
15.
go back to reference Ohl X, Stanchina C, Billuart F, Skalli W (2010) Shoulder bony landmarks location using the EOS low-dose stereoradiography system: a reproducibility study. Surg Radiol Anat 32:153–158CrossRefPubMed Ohl X, Stanchina C, Billuart F, Skalli W (2010) Shoulder bony landmarks location using the EOS low-dose stereoradiography system: a reproducibility study. Surg Radiol Anat 32:153–158CrossRefPubMed
16.
go back to reference Ilharreborde B, Ferrero E, Alison M, Mazda K (2015) EOS microdose protocol for the radiological follow-up of adolescent idiopathic scoliosis. Eur Spine J Ilharreborde B, Ferrero E, Alison M, Mazda K (2015) EOS microdose protocol for the radiological follow-up of adolescent idiopathic scoliosis. Eur Spine J
17.
go back to reference Chaibi Y, Cresson T, Aubert B et al (2012) Fast 3D reconstruction of the lower limb using a parametric model and statistical inferences and clinical measurements calculation from biplanar X-rays. Comput Methods Biomech Biomed Engin 15:457–466CrossRefPubMed Chaibi Y, Cresson T, Aubert B et al (2012) Fast 3D reconstruction of the lower limb using a parametric model and statistical inferences and clinical measurements calculation from biplanar X-rays. Comput Methods Biomech Biomed Engin 15:457–466CrossRefPubMed
18.
go back to reference Guggenberger R, Pfirrmann CW, Koch PP, Buck FM (2014) Assessment of lower limb length and alignment by biplanar linear radiography: comparison with supine CT and upright full-length radiography. AJR Am J Roentgenol 202:W161–W167CrossRefPubMed Guggenberger R, Pfirrmann CW, Koch PP, Buck FM (2014) Assessment of lower limb length and alignment by biplanar linear radiography: comparison with supine CT and upright full-length radiography. AJR Am J Roentgenol 202:W161–W167CrossRefPubMed
19.
go back to reference Guenoun B, Zadegan F, Aim F, Hannouche D, Nizard R (2012) Reliability of a new method for lower-extremity measurements based on stereoradiographic three-dimensional reconstruction. Orthop Traumatol Surg Res 98:506–513CrossRefPubMed Guenoun B, Zadegan F, Aim F, Hannouche D, Nizard R (2012) Reliability of a new method for lower-extremity measurements based on stereoradiographic three-dimensional reconstruction. Orthop Traumatol Surg Res 98:506–513CrossRefPubMed
20.
go back to reference Thelen P, Delin C, Folinais D, Radier C (2012) Evaluation of a new low-dose biplanar system to assess lower-limb alignment in 3D: a phantom study. Skeletal Radiol 41:1287–1293CrossRefPubMed Thelen P, Delin C, Folinais D, Radier C (2012) Evaluation of a new low-dose biplanar system to assess lower-limb alignment in 3D: a phantom study. Skeletal Radiol 41:1287–1293CrossRefPubMed
21.
go back to reference Meyrignac O, Moreno R, Baunin C et al (2015) Low-dose biplanar radiography can be used in children and adolescents to accurately assess femoral and tibial torsion and greatly reduce irradiation. Eur Radiol 25:1752–1760CrossRefPubMed Meyrignac O, Moreno R, Baunin C et al (2015) Low-dose biplanar radiography can be used in children and adolescents to accurately assess femoral and tibial torsion and greatly reduce irradiation. Eur Radiol 25:1752–1760CrossRefPubMed
22.
go back to reference Sugano N, Noble PC, Kamaric E (1998) A comparison of alternative methods of measuring femoral anteversion. J Comput Assist Tomogr 22:610–614CrossRefPubMed Sugano N, Noble PC, Kamaric E (1998) A comparison of alternative methods of measuring femoral anteversion. J Comput Assist Tomogr 22:610–614CrossRefPubMed
23.
go back to reference Waidelich HA, Strecker W, Schneider E (1992) Computed tomographic torsion-angle and length measurement of the lower extremity. The methods, normal values and radiation load. Rofo 157:245–251CrossRefPubMed Waidelich HA, Strecker W, Schneider E (1992) Computed tomographic torsion-angle and length measurement of the lower extremity. The methods, normal values and radiation load. Rofo 157:245–251CrossRefPubMed
24.
go back to reference Ohl X, Stanchina C, Billuart F, Skalli W Shoulder bony landmarks location using the EOS low-dose stereoradiography system: a reproducibility study. Surg Radiol Anat 32:153-158 Ohl X, Stanchina C, Billuart F, Skalli W Shoulder bony landmarks location using the EOS low-dose stereoradiography system: a reproducibility study. Surg Radiol Anat 32:153-158
25.
go back to reference Delin C, Silvera S, Bassinet C et al (2014) Ionizing radiation doses during lower limb torsion and anteversion measurements by EOS stereoradiography and computed tomography. Eur J Radiol 83:371–377CrossRefPubMed Delin C, Silvera S, Bassinet C et al (2014) Ionizing radiation doses during lower limb torsion and anteversion measurements by EOS stereoradiography and computed tomography. Eur J Radiol 83:371–377CrossRefPubMed
27.
go back to reference Doody MM, Lonstein JE, Stovall M, Hacker DG, Luckyanov N, Land CE (2000) Breast cancer mortality after diagnostic radiography: findings from the U.S. Scoliosis Cohort Study. Spine (Phila Pa 1976) 25:2052–2063CrossRef Doody MM, Lonstein JE, Stovall M, Hacker DG, Luckyanov N, Land CE (2000) Breast cancer mortality after diagnostic radiography: findings from the U.S. Scoliosis Cohort Study. Spine (Phila Pa 1976) 25:2052–2063CrossRef
Metadata
Title
Assessment of two-dimensional (2D) and three-dimensional (3D) lower limb measurements in adults: Comparison of micro-dose and low-dose biplanar radiographs
Authors
Andrea B. Rosskopf
Christian W. A. Pfirrmann
Florian M. Buck
Publication date
01-09-2016
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 9/2016
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-015-4166-5

Other articles of this Issue 9/2016

European Radiology 9/2016 Go to the issue