Assessment of hand kinematics using inertial and magnetic sensors

Authors: Henk G Kortier, Victor I Sluiter, Daniel Roetenberg, Peter H Veltink

Published in: Journal of NeuroEngineering and Rehabilitation | Issue 1/2014

Login to get access

Abstract

Background

Assessment of hand kinematics is important when evaluating hand functioning. Major drawbacks of current sensing glove systems are lack of rotational observability in particular directions, labour intensive calibration methods which are sensitive to wear and lack of an absolute hand orientation estimate.

Methods

We propose an ambulatory system using inertial sensors that can be placed on the hand, fingers and thumb. It allows a full 3D reconstruction of all finger and thumb joints as well as the absolute orientation of the hand. The system was experimentally evaluated for the static accuracy, dynamic range and repeatability.

Results

The RMS position norm difference of the fingertip compared to an optical system was 5±0.5 mm (mean ± standard deviation) for flexion-extension and 12.4±3.0 mm for combined flexion-extension abduction-adduction movements of the index finger. The difference between index and thumb tips during a pinching movement was 6.5±2.1 mm. The dynamic range of the sensing system and filter was adequate to reconstruct full 80 degrees movements of the index finger performed at 116 times per minute, which was limited by the range of the gyroscope. Finally, the reliability study showed a mean range difference over five subjects of 1.1±0.4 degrees for a flat hand test and 1.8±0.6 degrees for a plastic mold clenching test, which is smaller than other reported data gloves.

Conclusion

Compared to existing data gloves, this research showed that inertial and magnetic sensors are of interest for ambulatory analysis of the human hand and finger kinematics in terms of static accuracy, dynamic range and repeatability. It allows for estimation of multi-degree of freedom joint movements using low-cost sensors.
Literature
1.
go back to reference Erol A, Bebis G, Nicolescu M, Boyle RD, Twombly X: Vision-based hand pose estimation: a review. Comput Vis Image Underst 2007,108(1–2):52-73.CrossRef Erol A, Bebis G, Nicolescu M, Boyle RD, Twombly X: Vision-based hand pose estimation: a review. Comput Vis Image Underst 2007,108(1–2):52-73.CrossRef
2.
go back to reference Stenger B, Thayananthan A, Torr PHS, Cipolla R: Model-based hand tracking using a hierarchical bayesian filter. IEEE Trans Pattern Anal Mach Intell 2006,28(9):1372-1384.CrossRefPubMed Stenger B, Thayananthan A, Torr PHS, Cipolla R: Model-based hand tracking using a hierarchical bayesian filter. IEEE Trans Pattern Anal Mach Intell 2006,28(9):1372-1384.CrossRefPubMed
3.
go back to reference Dipietro L, Sabatini AM, Dario P: A survey of glove-based systems and their applications. Syst Man Cybernet Part C Appl Rev IEEE Trans 2008,38(4):461-482.CrossRef Dipietro L, Sabatini AM, Dario P: A survey of glove-based systems and their applications. Syst Man Cybernet Part C Appl Rev IEEE Trans 2008,38(4):461-482.CrossRef
4.
go back to reference Dipietro L, Sabatini AM, Dario P: Evaluation of an instrumented glove for hand-movement acquisition. J Rehabil Res Dev 2003,40(2):179-190.CrossRefPubMed Dipietro L, Sabatini AM, Dario P: Evaluation of an instrumented glove for hand-movement acquisition. J Rehabil Res Dev 2003,40(2):179-190.CrossRefPubMed
6.
go back to reference Ashbrook D, Baudisch P, White S: Nenya: subtle and eyes-free mobile input with a magnetically-tracked finger ring. In CHI ‘11: Proceedings of the SIGCHI Conference on Human Factors in Computing Systems. Vancouver: ACM Request Permissions; 2011. Ashbrook D, Baudisch P, White S: Nenya: subtle and eyes-free mobile input with a magnetically-tracked finger ring. In CHI ‘11: Proceedings of the SIGCHI Conference on Human Factors in Computing Systems. Vancouver: ACM Request Permissions; 2011.
7.
go back to reference Rowe JB, Friedman N, Bachman M, Reinkensmeyer DJ: The Manumeter: a non-obtrusive wearable device for monitoring spontaneous use of the wrist and fingers. In Rehabilitation Robotics (ICORR) 2013 IEEE International Conference on. Seattle: IEEE; 2013:1-6.CrossRef Rowe JB, Friedman N, Bachman M, Reinkensmeyer DJ: The Manumeter: a non-obtrusive wearable device for monitoring spontaneous use of the wrist and fingers. In Rehabilitation Robotics (ICORR) 2013 IEEE International Conference on. Seattle: IEEE; 2013:1-6.CrossRef
8.
go back to reference Roetenberg D, Luinge HJ, Baten CTM, Veltink PH: Compensation of magnetic disturbances improves inertial and magnetic sensing of human body segment orientation. Neural Syst Rehabil Eng IEEE Trans 2005,13(3):395-405.CrossRef Roetenberg D, Luinge HJ, Baten CTM, Veltink PH: Compensation of magnetic disturbances improves inertial and magnetic sensing of human body segment orientation. Neural Syst Rehabil Eng IEEE Trans 2005,13(3):395-405.CrossRef
9.
go back to reference Hernandez-Rebollar JL, Kyriakopoulos N, Lindeman RW: The AcceleGlove: a whole-hand input device for virtual reality. In SIGGRAPH ’02 SIGGRAPH 2002 Conference Abstracts and Applications. San Antonio: ACM; 2002. Hernandez-Rebollar JL, Kyriakopoulos N, Lindeman RW: The AcceleGlove: a whole-hand input device for virtual reality. In SIGGRAPH ’02 SIGGRAPH 2002 Conference Abstracts and Applications. San Antonio: ACM; 2002.
10.
go back to reference Kim YS, Soh BS, Lee SG: A new wearable input device: SCURRY. Ind Electron IEEE Trans 2005,52(6):1490-1499.CrossRef Kim YS, Soh BS, Lee SG: A new wearable input device: SCURRY. Ind Electron IEEE Trans 2005,52(6):1490-1499.CrossRef
11.
go back to reference Wu G, van der Helm FCT, Veeger HEJ, Makhsous M, van Roy P, Anglin C, Nagels J, Karduna AR, McQuade K, Wang X, Werner FW, Bucholz B: ISB recommendation on definitions of joint coordinate systems of various joints for the reporting of human joint motion–Part II: shoulder, elbow, wrist and hand. J Biomech 2005,38(5):981-992.CrossRefPubMed Wu G, van der Helm FCT, Veeger HEJ, Makhsous M, van Roy P, Anglin C, Nagels J, Karduna AR, McQuade K, Wang X, Werner FW, Bucholz B: ISB recommendation on definitions of joint coordinate systems of various joints for the reporting of human joint motion–Part II: shoulder, elbow, wrist and hand. J Biomech 2005,38(5):981-992.CrossRefPubMed
12.
go back to reference Kuipers JB: Quaternions and Rotation Sequences. Princeton: Princeton University press; 1999. Kuipers JB: Quaternions and Rotation Sequences. Princeton: Princeton University press; 1999.
13.
go back to reference Crassidis JL, Markley FL, Cheng Y: Survey of nonlinear attitude estimation methods. J Guid Control Dyn 2007, 30: 12.CrossRef Crassidis JL, Markley FL, Cheng Y: Survey of nonlinear attitude estimation methods. J Guid Control Dyn 2007, 30: 12.CrossRef
14.
go back to reference Yun X, Bachmann ER: Design, implementation, and experimental results of a quaternion-based Kalman filter for human body motion tracking. Robotics, IEEE Trans 2006,22(6):1216-1227.CrossRef Yun X, Bachmann ER: Design, implementation, and experimental results of a quaternion-based Kalman filter for human body motion tracking. Robotics, IEEE Trans 2006,22(6):1216-1227.CrossRef
15.
go back to reference Sabatini AM: Quaternion-based extended Kalman filter for determining orientation by inertial and magnetic sensing. Biomed Eng IEEE Trans 2006,53(7):1346-1356.CrossRef Sabatini AM: Quaternion-based extended Kalman filter for determining orientation by inertial and magnetic sensing. Biomed Eng IEEE Trans 2006,53(7):1346-1356.CrossRef
16.
go back to reference Roetenberg D, Slycke PJ, Veltink PH: Ambulatory position and orientation tracking fusing magnetic and inertial sensing. IEEE Trans Biomed Eng 2007,54(5):883-890.CrossRefPubMed Roetenberg D, Slycke PJ, Veltink PH: Ambulatory position and orientation tracking fusing magnetic and inertial sensing. IEEE Trans Biomed Eng 2007,54(5):883-890.CrossRefPubMed
17.
go back to reference Luinge HJ, Veltink PH, Baten CTM: Ambulatory measurement of arm orientation. J Biomech 2007, 40: 78-85.CrossRefPubMed Luinge HJ, Veltink PH, Baten CTM: Ambulatory measurement of arm orientation. J Biomech 2007, 40: 78-85.CrossRefPubMed
18.
go back to reference Gustafsson F: Statistical Sensor Fusion. Linköping: Studentlitteratur; 2010. Gustafsson F: Statistical Sensor Fusion. Linköping: Studentlitteratur; 2010.
19.
go back to reference Bortz JE: A new mathematical formulation for strapdown inertial navigation. IEEE Trans Aero Electron Syst 1971, 1: 61-66.CrossRef Bortz JE: A new mathematical formulation for strapdown inertial navigation. IEEE Trans Aero Electron Syst 1971, 1: 61-66.CrossRef
20.
go back to reference Skog I, Handel P, Nilsson JO, Rantakokko JGJ: Zero-velocity detection—an algorithm evaluation. Biomed Eng IEEE Trans 2010,57(11):2657-2666.CrossRef Skog I, Handel P, Nilsson JO, Rantakokko JGJ: Zero-velocity detection—an algorithm evaluation. Biomed Eng IEEE Trans 2010,57(11):2657-2666.CrossRef
21.
go back to reference Buchholz B, Thomas JA, Steven AG: Anthropometric data for describing the kinematics of the human hand. Ergonomics 1992,35(3):261-273.CrossRefPubMed Buchholz B, Thomas JA, Steven AG: Anthropometric data for describing the kinematics of the human hand. Ergonomics 1992,35(3):261-273.CrossRefPubMed
22.
go back to reference Brand PW, Hollister A: Clinical Mechanics of the Hand. St. Louis: Mosby Year Book; 1999. Brand PW, Hollister A: Clinical Mechanics of the Hand. St. Louis: Mosby Year Book; 1999.
23.
go back to reference Wise S, Gardner W, Sabelman E, Valainis E, Wong Y, Glass K, Drace J, Rosen JM: Evaluation of a fiber optic glove for se-automated goniometric measurements. J Rehabil Res Dev 1990.,27(4): Wise S, Gardner W, Sabelman E, Valainis E, Wong Y, Glass K, Drace J, Rosen JM: Evaluation of a fiber optic glove for se-automated goniometric measurements. J Rehabil Res Dev 1990.,27(4):
24.
go back to reference Williams NW, Penrose JMT, Caddy CM, Barnes E, Hose DR, Harley P: A goniometric glove for clinical hand assessment. J Hand Surg (British Eur Vol) 2000,25(2):200.CrossRef Williams NW, Penrose JMT, Caddy CM, Barnes E, Hose DR, Harley P: A goniometric glove for clinical hand assessment. J Hand Surg (British Eur Vol) 2000,25(2):200.CrossRef
25.
go back to reference Simone LK, Sundarrajan N, Luo X, Jia Y, Kamper DG: A low cost instrumented glove for extended monitoring and functional hand assessment. J Neurosci Methods 2007,160(2):335-348.CrossRefPubMed Simone LK, Sundarrajan N, Luo X, Jia Y, Kamper DG: A low cost instrumented glove for extended monitoring and functional hand assessment. J Neurosci Methods 2007,160(2):335-348.CrossRefPubMed
26.
go back to reference Gentner R, Classen J: Development and evaluation of a low-cost sensor glove for assessment of human finger movements in neurophysiological settings. J Neurosci Methods 2009, 178: 138-147.CrossRefPubMed Gentner R, Classen J: Development and evaluation of a low-cost sensor glove for assessment of human finger movements in neurophysiological settings. J Neurosci Methods 2009, 178: 138-147.CrossRefPubMed
27.
go back to reference Bullock IM, Borras J, Dollar AM: Assessing assumptions in kinematic hand models: a review. In Biomedical Robotics and Biomechatronics (BioRob) 2012 4th IEEE RAS & EMBS International Conference on. Rome: IEEE; 2012:139-146.CrossRef Bullock IM, Borras J, Dollar AM: Assessing assumptions in kinematic hand models: a review. In Biomedical Robotics and Biomechatronics (BioRob) 2012 4th IEEE RAS & EMBS International Conference on. Rome: IEEE; 2012:139-146.CrossRef
Metadata
Title
Assessment of hand kinematics using inertial and magnetic sensors
Authors
Henk G Kortier
Victor I Sluiter
Daniel Roetenberg
Peter H Veltink
Publication date
01-12-2014
Publisher
BioMed Central
Published in
Journal of NeuroEngineering and Rehabilitation / Issue 1/2014
Electronic ISSN: 1743-0003
DOI
https://doi.org/10.1186/1743-0003-11-70

Other articles of this Issue 1/2014

Journal of NeuroEngineering and Rehabilitation 1/2014 Go to the issue