Assessment of a continuous blood gas monitoring system in animals during circulatory stress

Authors: Sandro Gelsomino, Roberto Lorusso, Ugolino Livi, Stefano Romagnoli, Salvatore Mario Romano, Rocco Carella, Fabiana Lucà, Giuseppe Billè, Francesco Matteucci, Attilio Renzulli, Gil Bolotin, Giuseppe De Cicco, Pierluigi Stefàno, Jos Maessen, Gian Franco Gensini

Published in: BMC Anesthesiology | Issue 1/2011

Login to get access

Abstract

Background

The study was aimed to determine the measurement accuracy of The CDI™ blood parameter monitoring system 500 (Terumo Cardiovascular Systems Corporation, Ann Arbor MI) in the real-time continuous measurement of arterial blood gases under different cardiocirculatory stress conditions

Methods

Inotropic stimulation (Dobutamine 2.5 and 5 μg/kg/min), vasoconstriction (Arginine-vasopressin 4, 8 and 16 IU/h), hemorrhage (-10%, -20%, -35%, and -50% of the theoretical volemia), and volume resuscitation were induced in ten swine (57.4 ± 10.7 Kg).Intermittent blood gas assessments were carried out using a routine gas analyzer at any experimental phase and compared with values obtained at the same time settings during continuous monitoring with CDI™ 500 system. The Bland-Altman analysis was employed.

Results

Bias and precision for pO2 were - 0.06 kPa and 0.22 kPa, respectively (r2 = 0.96); pCO2 - 0.02 kPa and 0.15 kPa, respectively; pH -0.001 and 0.01 units, respectively ( r2 = 0.96). The analysis showed very good agreement for SO2 (bias 0.04,precision 0.33, r2 = 0.95), Base excess (bias 0.04,precision 0.28, r2 = 0.98), HCO3 (bias 0.05,precision 0.62, r2 = 0.92),hemoglobin (bias 0.02,precision 0.23, r2 = 0.96) and K+ (bias 0.02, precision 0.27, r2 = 0.93). The sensor was reliable throughout the experiment during hemodynamic variations.

Conclusions

Continuous blood gas analysis with the CDI™ 500 system was reliable and it might represent a new useful tool to accurately and timely monitor gas exchange in critically ill patients. Nonetheless, our findings need to be confirmed by larger studies to prove its reliability in the clinical setting.
Appendix
Available only for authorised users
Literature
1.
go back to reference Venkatetesh B: Continuous Intra-arterial Blood Gas Monitoring. Critical Care and Resuscitation. 1999, 1: 140-50. Venkatetesh B: Continuous Intra-arterial Blood Gas Monitoring. Critical Care and Resuscitation. 1999, 1: 140-50.
2.
go back to reference Sasse SA, Chen PA, Mahutte CK: Variability of arterial blood gas values over time in stable medical ICU patients. Chest. 1994, 106: 187-193. 10.1378/chest.106.1.187.CrossRefPubMed Sasse SA, Chen PA, Mahutte CK: Variability of arterial blood gas values over time in stable medical ICU patients. Chest. 1994, 106: 187-193. 10.1378/chest.106.1.187.CrossRefPubMed
3.
go back to reference Clutton-Brock TH, Venkatesh B: Blood Gas Monitoring. Monitoring in Anaestesia and Intensive Care. Edited by: Hutton P, Prys-Roberts C. 1994, London: WB Saunders, 242-55. First Clutton-Brock TH, Venkatesh B: Blood Gas Monitoring. Monitoring in Anaestesia and Intensive Care. Edited by: Hutton P, Prys-Roberts C. 1994, London: WB Saunders, 242-55. First
4.
go back to reference Schreur A, Niles S, Ploessl J: Use of the CDI blood parameter monitoring system 500 for continuous blood gas measurement during extracorporeal membrane oxygenation simulation. J Extra Corpor Technol. 2005, 37 (4): 377-80.PubMedPubMedCentral Schreur A, Niles S, Ploessl J: Use of the CDI blood parameter monitoring system 500 for continuous blood gas measurement during extracorporeal membrane oxygenation simulation. J Extra Corpor Technol. 2005, 37 (4): 377-80.PubMedPubMedCentral
5.
go back to reference Trowbridge CC, Vasquez M, Stammers AH, Glogowski K, Tremain K, Niimi K, Muhle M, Yiang T: The effects of continuous blood gas monitoring during cardiopulmonary bypass: A prospective, randomized study- Part I. The Journal of extra-corporeal technology. 2000, 32 (3): 120-128.PubMed Trowbridge CC, Vasquez M, Stammers AH, Glogowski K, Tremain K, Niimi K, Muhle M, Yiang T: The effects of continuous blood gas monitoring during cardiopulmonary bypass: A prospective, randomized study- Part I. The Journal of extra-corporeal technology. 2000, 32 (3): 120-128.PubMed
6.
go back to reference Romagnoli S, Romano SM, Bevilacqua S, Ciappi F, Lazzeri C, Peris A, Dini D, Gelsomino S: Cardiac oputput by arterial pulse contur: reliability under hemodynamic derangements. Interact Cardiovasc Interact Cardiovasc Thorac Surg. 2009, 8 (6): 642-6. 10.1510/icvts.2008.200451.CrossRefPubMed Romagnoli S, Romano SM, Bevilacqua S, Ciappi F, Lazzeri C, Peris A, Dini D, Gelsomino S: Cardiac oputput by arterial pulse contur: reliability under hemodynamic derangements. Interact Cardiovasc Interact Cardiovasc Thorac Surg. 2009, 8 (6): 642-6. 10.1510/icvts.2008.200451.CrossRefPubMed
7.
go back to reference Zangrillo A, Maj G, Monaco F, Scandroglio AM, Nuzzi M, Plumari V, Virzo I, Bignami E, Casiraghi G, Landoni G: Cardiac index validation using the pressure recording analytic method in unstable patients. J Cardiothorac Vasc Anesth. 2010, 24 (2): 265-9. 10.1053/j.jvca.2009.09.019.CrossRefPubMed Zangrillo A, Maj G, Monaco F, Scandroglio AM, Nuzzi M, Plumari V, Virzo I, Bignami E, Casiraghi G, Landoni G: Cardiac index validation using the pressure recording analytic method in unstable patients. J Cardiothorac Vasc Anesth. 2010, 24 (2): 265-9. 10.1053/j.jvca.2009.09.019.CrossRefPubMed
8.
go back to reference Philbin N, Rice J, Gurney J, McGwin G, Arnaud F, Dong F, Johnson T, Flournoy WS, Ahlers S, Pearce LB, McCarron R, Freilich D: A hemoglobin-based oxygen carrier, bovine polymerized hemoglobin (HBOC-201) versus hetastarch (HEX) in a moderate severity hemorrhagic shock swine model with delayed evacuation. Resuscitation. 2005, 66: 367-8. 10.1016/j.resuscitation.2005.03.019.CrossRefPubMed Philbin N, Rice J, Gurney J, McGwin G, Arnaud F, Dong F, Johnson T, Flournoy WS, Ahlers S, Pearce LB, McCarron R, Freilich D: A hemoglobin-based oxygen carrier, bovine polymerized hemoglobin (HBOC-201) versus hetastarch (HEX) in a moderate severity hemorrhagic shock swine model with delayed evacuation. Resuscitation. 2005, 66: 367-8. 10.1016/j.resuscitation.2005.03.019.CrossRefPubMed
10.
go back to reference Altman DG, Bland JM: Measurement in medicine: the analysis of method comparison studies. Statistician. 1983, 32: 307-317. 10.2307/2987937.CrossRef Altman DG, Bland JM: Measurement in medicine: the analysis of method comparison studies. Statistician. 1983, 32: 307-317. 10.2307/2987937.CrossRef
11.
go back to reference Hatherill M, Tibby SM, Durward A, Rajah V, Murdoch IA: Continuous intra-arterial blood-gas monitoring in infants and children with cyanotic heart disease. Br J Anaesth. 1997, 79: 665-667.CrossRefPubMed Hatherill M, Tibby SM, Durward A, Rajah V, Murdoch IA: Continuous intra-arterial blood-gas monitoring in infants and children with cyanotic heart disease. Br J Anaesth. 1997, 79: 665-667.CrossRefPubMed
12.
go back to reference Shapiro BA: Arterial blood gas monitoring. Crit Care Clin. 1988, 4: 479-92.PubMed Shapiro BA: Arterial blood gas monitoring. Crit Care Clin. 1988, 4: 479-92.PubMed
13.
go back to reference Menzel M, Soukup J, Henze D, Engelbrecht K, Senderreck M, Scharf A, Rieger A, Grond S: Experiences with continuous intra-arterial blood gas monitoring:precision and drift of pure optode-system. Intensive Care Med. 2003, 29: 2180-86. 10.1007/s00134-003-1962-1.CrossRefPubMed Menzel M, Soukup J, Henze D, Engelbrecht K, Senderreck M, Scharf A, Rieger A, Grond S: Experiences with continuous intra-arterial blood gas monitoring:precision and drift of pure optode-system. Intensive Care Med. 2003, 29: 2180-86. 10.1007/s00134-003-1962-1.CrossRefPubMed
14.
go back to reference Bahrami KR, Rivera O, Miskell GT, Short BL: Continuous Blood gas monitoring using a in-dwelling optode method. Comparison to intermittent arterial blood gas sampling in ECMO patients. Journal of Perinatology. 2002, 22: 472-4. 10.1038/sj.jp.7210776.CrossRef Bahrami KR, Rivera O, Miskell GT, Short BL: Continuous Blood gas monitoring using a in-dwelling optode method. Comparison to intermittent arterial blood gas sampling in ECMO patients. Journal of Perinatology. 2002, 22: 472-4. 10.1038/sj.jp.7210776.CrossRef
15.
go back to reference Desiderio DP, Wong G, Shah NK, Liu J, Loughlin CJ, Bedford RF: A clinical evaluation of pulse oximetry during thoracic surgery. J Cardiothorac Anesth. 1990, 4: 30-4. 10.1016/0888-6296(90)90443-J.CrossRefPubMed Desiderio DP, Wong G, Shah NK, Liu J, Loughlin CJ, Bedford RF: A clinical evaluation of pulse oximetry during thoracic surgery. J Cardiothorac Anesth. 1990, 4: 30-4. 10.1016/0888-6296(90)90443-J.CrossRefPubMed
16.
go back to reference Reich DL, Timcenko A, Bodian CA, Kraidin J, Hofman J, DePerio M, Konstadt SN, Kurki T, Eisenkraft JB: Predictors of pulse oximetry data failure. Anesthesiology. 1996, 84: 859-64. 10.1097/00000542-199604000-00013.CrossRefPubMed Reich DL, Timcenko A, Bodian CA, Kraidin J, Hofman J, DePerio M, Konstadt SN, Kurki T, Eisenkraft JB: Predictors of pulse oximetry data failure. Anesthesiology. 1996, 84: 859-64. 10.1097/00000542-199604000-00013.CrossRefPubMed
17.
go back to reference Tremper KK, Barker SJ: Pulse oximetry. Anesthesiology. 1989, 70: 98-108. 10.1097/00000542-198901000-00019.CrossRefPubMed Tremper KK, Barker SJ: Pulse oximetry. Anesthesiology. 1989, 70: 98-108. 10.1097/00000542-198901000-00019.CrossRefPubMed
18.
go back to reference Van de Louw A, Cracco C, Cerf C, Harf A, Duvaldestin P, Lemaire F, Brochard L: Accuracy of pulse oximetry in the intensive care unit. Intensive Care Med. 2001, 27: 1606-13. 10.1007/s001340101064.CrossRefPubMed Van de Louw A, Cracco C, Cerf C, Harf A, Duvaldestin P, Lemaire F, Brochard L: Accuracy of pulse oximetry in the intensive care unit. Intensive Care Med. 2001, 27: 1606-13. 10.1007/s001340101064.CrossRefPubMed
19.
go back to reference Benson JP, Venkatesh B, Patla V: Misleading information from pulse oximetry and the usefulness of continuous blood gas monitoring in a post cardiac surgery patient. Intensive Care Med. 1995, 21: 437-9. 10.1007/BF01707413.CrossRefPubMed Benson JP, Venkatesh B, Patla V: Misleading information from pulse oximetry and the usefulness of continuous blood gas monitoring in a post cardiac surgery patient. Intensive Care Med. 1995, 21: 437-9. 10.1007/BF01707413.CrossRefPubMed
20.
go back to reference Rithalia SV: Developments in transcutaneous blood gas monitoring: a review. J Med Eng Technol. 1991, 15: 143-53. 10.3109/03091909109023701.CrossRefPubMed Rithalia SV: Developments in transcutaneous blood gas monitoring: a review. J Med Eng Technol. 1991, 15: 143-53. 10.3109/03091909109023701.CrossRefPubMed
21.
go back to reference Gehrich J, Lubbers D, Opitz N, Hansman D, Miller W, Tusa J: Optical fluorescence and its applications to an intravascular blood gas monitoring system. IEE Trans Biomed Eng. 1986, 33: 117-132. 10.1109/TBME.1986.325886.CrossRef Gehrich J, Lubbers D, Opitz N, Hansman D, Miller W, Tusa J: Optical fluorescence and its applications to an intravascular blood gas monitoring system. IEE Trans Biomed Eng. 1986, 33: 117-132. 10.1109/TBME.1986.325886.CrossRef
22.
go back to reference Shapiro BA, Mahutte CK, Cane RD, Gilmour IJ: Clinical performance of a blood gas monitor: a prospective, multicenter trial. Crit Care Med. 1993, 21: 487-94. 10.1097/00003246-199304000-00005.CrossRefPubMed Shapiro BA, Mahutte CK, Cane RD, Gilmour IJ: Clinical performance of a blood gas monitor: a prospective, multicenter trial. Crit Care Med. 1993, 21: 487-94. 10.1097/00003246-199304000-00005.CrossRefPubMed
23.
go back to reference Mahutte CK, Holody M, Maxwell T, Chen P, Sasse A: Development of a patient dedicated on demand blood gas monitor in medical ICI patients. Am J Resp Crit Care Med. 1994, 149: 852-9.CrossRefPubMed Mahutte CK, Holody M, Maxwell T, Chen P, Sasse A: Development of a patient dedicated on demand blood gas monitor in medical ICI patients. Am J Resp Crit Care Med. 1994, 149: 852-9.CrossRefPubMed
24.
go back to reference Mahutte CK, Sasse A, Chen P, Holody M: Performance of a patient dedicated on demand blood gas monitor in medical ICI patients. Am J Resp Crit Care Med. 1994, 150: 865-9.CrossRefPubMed Mahutte CK, Sasse A, Chen P, Holody M: Performance of a patient dedicated on demand blood gas monitor in medical ICI patients. Am J Resp Crit Care Med. 1994, 150: 865-9.CrossRefPubMed
25.
go back to reference Mahutte CK, Sassoon CS, Muro JR, Hansmann DR, Maxwell TP, Miller WW, Yafuso M: Progress in the development of a fluorescent intravascular blood gas system in man. J Clin Monit. 1990, 6: 147-157. 10.1007/BF02828293.CrossRefPubMed Mahutte CK, Sassoon CS, Muro JR, Hansmann DR, Maxwell TP, Miller WW, Yafuso M: Progress in the development of a fluorescent intravascular blood gas system in man. J Clin Monit. 1990, 6: 147-157. 10.1007/BF02828293.CrossRefPubMed
26.
go back to reference Greenblott G, Tremper K, Barker J, Gerschultz S, Gebrich J: Continuous blood gas monitoring with an intra-arterial optode during one lung anesthesia. J Cardiothorac Vasc Anesth. 1991, 5: 365-367. 10.1016/1053-0770(91)90161-L.CrossRefPubMed Greenblott G, Tremper K, Barker J, Gerschultz S, Gebrich J: Continuous blood gas monitoring with an intra-arterial optode during one lung anesthesia. J Cardiothorac Vasc Anesth. 1991, 5: 365-367. 10.1016/1053-0770(91)90161-L.CrossRefPubMed
27.
go back to reference Barker S, Hyatt J: Continuous measurement of intra-arterial pH, PaCO2 and PaO2 in the operating room. Anesth Analg. 1991, 73: 43-48.PubMed Barker S, Hyatt J: Continuous measurement of intra-arterial pH, PaCO2 and PaO2 in the operating room. Anesth Analg. 1991, 73: 43-48.PubMed
28.
go back to reference Shapiro BA, Cane RD, Chomka CM: Preliminary evaluation of an intra-arterial blood gas system in dogs and humans. Crit Care Med. 1989, 17: 455-460. 10.1097/00003246-198905000-00016.CrossRefPubMed Shapiro BA, Cane RD, Chomka CM: Preliminary evaluation of an intra-arterial blood gas system in dogs and humans. Crit Care Med. 1989, 17: 455-460. 10.1097/00003246-198905000-00016.CrossRefPubMed
29.
go back to reference Clark C, O'Brien J, McCulloch J: Early clinical experience with Gas-STAT. J Extra-Corporeal Technol. 1986, 18: 185-189. Clark C, O'Brien J, McCulloch J: Early clinical experience with Gas-STAT. J Extra-Corporeal Technol. 1986, 18: 185-189.
30.
go back to reference Pfeifer P, Pearson D, Clayton R: Clinical trial of the Continucath intra-arterial oxygen monitor. Anaesthesia. 1988, 43: 677-682. 10.1111/j.1365-2044.1988.tb04157.x.CrossRefPubMed Pfeifer P, Pearson D, Clayton R: Clinical trial of the Continucath intra-arterial oxygen monitor. Anaesthesia. 1988, 43: 677-682. 10.1111/j.1365-2044.1988.tb04157.x.CrossRefPubMed
31.
go back to reference Haller M, Kilger E, Briegel J, Frost H, Peter K: Continuous intra - arterial blood gas and pH monitoring in critically ill patients with severe respiratory failure: a prospective, criterion standard study. Crit Care Med. 1994, 22: 580-6. 10.1097/00003246-199404000-00012.CrossRefPubMed Haller M, Kilger E, Briegel J, Frost H, Peter K: Continuous intra - arterial blood gas and pH monitoring in critically ill patients with severe respiratory failure: a prospective, criterion standard study. Crit Care Med. 1994, 22: 580-6. 10.1097/00003246-199404000-00012.CrossRefPubMed
32.
go back to reference Venkatesh B, Clutton-Brock TH, Hendry SP: A multi-parameter sensor for continuous intra-arterial blood gas monitoring: a prospective evaluation. Crit Care Med. 1994, 22: 588-94. 10.1097/00003246-199404000-00013.CrossRefPubMed Venkatesh B, Clutton-Brock TH, Hendry SP: A multi-parameter sensor for continuous intra-arterial blood gas monitoring: a prospective evaluation. Crit Care Med. 1994, 22: 588-94. 10.1097/00003246-199404000-00013.CrossRefPubMed
33.
go back to reference Roupie EE, Brochard L, Lemaire FJ: Critical evaluation of a continuous intra-arterial blood gas system in critically ill patients. Intensive Care Med. 1996, 22: 1162-1168. 10.1007/BF01709330.CrossRefPubMed Roupie EE, Brochard L, Lemaire FJ: Critical evaluation of a continuous intra-arterial blood gas system in critically ill patients. Intensive Care Med. 1996, 22: 1162-1168. 10.1007/BF01709330.CrossRefPubMed
34.
go back to reference Ganter M, Zollinger A: Continuous intravascular blood gas monitoring : development, current techniques and clinical use of a commercial device. British Journal of Anaesthesia. 2003, 91 (3): 397-407. 10.1093/bja/aeg176.CrossRefPubMed Ganter M, Zollinger A: Continuous intravascular blood gas monitoring : development, current techniques and clinical use of a commercial device. British Journal of Anaesthesia. 2003, 91 (3): 397-407. 10.1093/bja/aeg176.CrossRefPubMed
35.
go back to reference Menzel M, Henze D, Soukup J, Engelbrecht K, Senderreck M, Clausen T, Radke J: Experiences with continuous intra-arterial blood gas monitoring. Minerva Anestesiol. 2001, 67: 325-31.PubMed Menzel M, Henze D, Soukup J, Engelbrecht K, Senderreck M, Clausen T, Radke J: Experiences with continuous intra-arterial blood gas monitoring. Minerva Anestesiol. 2001, 67: 325-31.PubMed
36.
go back to reference Trowbridge CC, Vasquez M, Stammers AH, Glogowski K, Tremain K, Niimi K, Muhle M, Yiang T: The effects of continuous blood gas monitoring during cardiopulmonary bypass: A prospective, randomized study- Part II. The Journal of extra-corporeal technology. 2000, 32 (3): 129-137.PubMed Trowbridge CC, Vasquez M, Stammers AH, Glogowski K, Tremain K, Niimi K, Muhle M, Yiang T: The effects of continuous blood gas monitoring during cardiopulmonary bypass: A prospective, randomized study- Part II. The Journal of extra-corporeal technology. 2000, 32 (3): 129-137.PubMed
37.
go back to reference Southworth R, Sutton R, Mize S, Stammers AH, Fristoe LW, Cook D, Hostetler D, Richenbacher WE: Clinical evaluation of a new in-line continuous blood gas monitor. J Extra Corpor Technol. 1998, 30 (4): 166-70.PubMed Southworth R, Sutton R, Mize S, Stammers AH, Fristoe LW, Cook D, Hostetler D, Richenbacher WE: Clinical evaluation of a new in-line continuous blood gas monitor. J Extra Corpor Technol. 1998, 30 (4): 166-70.PubMed
38.
go back to reference Medicare MaCP: Regulations Implementing the Clinical Laboratory Improvement Amendments of 1988 (CLIA '88). Federal Register. 1992, 7002-7186. Medicare MaCP: Regulations Implementing the Clinical Laboratory Improvement Amendments of 1988 (CLIA '88). Federal Register. 1992, 7002-7186.
39.
go back to reference Venkatesh B, Clutton-Brock TH, Hendry SP: Evaluation of the Paratrend 7 intravascular blood gas monitor during cardiac surgery: comparison with the C4000 in-line blood gas monitor during cardiopulmonary bypass. J Cardiothorac Vasc Anesth. 1995, 9: 412-419. 10.1016/S1053-0770(05)80096-5.CrossRefPubMed Venkatesh B, Clutton-Brock TH, Hendry SP: Evaluation of the Paratrend 7 intravascular blood gas monitor during cardiac surgery: comparison with the C4000 in-line blood gas monitor during cardiopulmonary bypass. J Cardiothorac Vasc Anesth. 1995, 9: 412-419. 10.1016/S1053-0770(05)80096-5.CrossRefPubMed
40.
go back to reference Zollinger A, Spahn DR, Singer T, Zalunardo MP, Stoehr S, Weder W, Pasch T: Accuracy and clinical performance of a continuous intraarterial blood-gas monitoring system during thoracoscopic surgery. Br J Anaesth. 1997, 79: 47-52.CrossRefPubMed Zollinger A, Spahn DR, Singer T, Zalunardo MP, Stoehr S, Weder W, Pasch T: Accuracy and clinical performance of a continuous intraarterial blood-gas monitoring system during thoracoscopic surgery. Br J Anaesth. 1997, 79: 47-52.CrossRefPubMed
41.
go back to reference Abraham E, Gallagher TJ, Fink S: Clinical evaluation of a multiparameter intra-arterial blood-gas sensor. Intensive Care Med. 1996, 22: 507-51. 10.1007/BF01712177.CrossRefPubMed Abraham E, Gallagher TJ, Fink S: Clinical evaluation of a multiparameter intra-arterial blood-gas sensor. Intensive Care Med. 1996, 22: 507-51. 10.1007/BF01712177.CrossRefPubMed
42.
go back to reference Venkatesh B, Hendry SP: Continuous intra-arterial blood gas monitoring. Intensive Care Med. 1996, 22: 818-28. 10.1007/BF01709527.CrossRefPubMed Venkatesh B, Hendry SP: Continuous intra-arterial blood gas monitoring. Intensive Care Med. 1996, 22: 818-28. 10.1007/BF01709527.CrossRefPubMed
Metadata
Title
Assessment of a continuous blood gas monitoring system in animals during circulatory stress
Authors
Sandro Gelsomino
Roberto Lorusso
Ugolino Livi
Stefano Romagnoli
Salvatore Mario Romano
Rocco Carella
Fabiana Lucà
Giuseppe Billè
Francesco Matteucci
Attilio Renzulli
Gil Bolotin
Giuseppe De Cicco
Pierluigi Stefàno
Jos Maessen
Gian Franco Gensini
Publication date
01-12-2011
Publisher
BioMed Central
Published in
BMC Anesthesiology / Issue 1/2011
Electronic ISSN: 1471-2253
DOI
https://doi.org/10.1186/1471-2253-11-1

Other articles of this Issue 1/2011

BMC Anesthesiology 1/2011 Go to the issue