Skip to main content
Top
Published in: Journal of Clinical Monitoring and Computing 5/2017

Open Access 01-10-2017 | Original Research

Analysis and compensation for errors in electrical impedance tomography images and ventilation-related measures due to serial data collection

Authors: Rebecca J. Yerworth, Inéz Frerichs, Richard Bayford

Published in: Journal of Clinical Monitoring and Computing | Issue 5/2017

Login to get access

Abstract

Electrical impedance tomography (EIT) is increasingly being used as a bedside tool for monitoring regional lung ventilation. However, most clinical systems use serial data collection which, if uncorrected, results in image distortion, particularly at high breathing rates. The objective of this study was to determine the extent to which this affects derived parameters. Raw EIT data were acquired with the GOE-MF II EIT device (CareFusion, Höchberg, Germany) at a scan rate of 13 images/s during both spontaneous breathing and mechanical ventilation. Boundary data for periods of undisturbed tidal breathing were corrected for serial data collection errors using a Fourier based algorithm. Images were reconstructed for both the corrected and original data using the GREIT algorithm, and parameters describing the filling characteristics of the right and left lung derived on a breath by breath basis. Values from the original and corrected data were compared using paired t-tests. Of the 33 data sets, 23 showed significant differences in filling index for at least one region, 11 had significant differences in calculated tidal impedance change and 12 had significantly different filling fractions (p = 0.05). We conclude that serial collection errors should be corrected before image reconstruction to avoid clinically misleading results.
Appendix
Available only for authorised users
Footnotes
1
For a breathing rate of 60 breaths per minute, i.e. 1 breath per second, if inspiration and expiration phases are equal there are 6–7 data points during the inspiratory phase and 6–7 during the expiratory. This corresponds to a change of >1/6th (17 %) of the inspired volume occurring during a single frame of data collection. However the inspiratory phase is often shorter than the expiratory, so this is an underestimate of the physiological changes present.
 
2
Most recently developed commercial EIT devices do not offer specific electrode interfaces for neonates, therefore these older EIT devices are the only ones available for this patient population.
 
3
A more sophisticated breath detection approach would be needed for unsupervised use and non-tidal breathing.
 
4
Available from EIDORS http://​eidors3d.​sourceforge.​net/​ or by contacting the authors..
 
Literature
1.
go back to reference Grychtol B, Wolf GK, Adler A, et al. Towards lung EIT image segmentation: automatic classification of lung tissue state from analysis of EIT monitored recruitment manoeuvres. Physiol Meas. 2010;31:S31–43.CrossRefPubMed Grychtol B, Wolf GK, Adler A, et al. Towards lung EIT image segmentation: automatic classification of lung tissue state from analysis of EIT monitored recruitment manoeuvres. Physiol Meas. 2010;31:S31–43.CrossRefPubMed
2.
go back to reference Miedema M, de Jongh FH, Frerichs I, et al. Changes in Lung volume and ventilation during surfactant treatment in ventilated preterm infants. Am J Respir Crit Care Med. 2011;184:100–5.CrossRefPubMed Miedema M, de Jongh FH, Frerichs I, et al. Changes in Lung volume and ventilation during surfactant treatment in ventilated preterm infants. Am J Respir Crit Care Med. 2011;184:100–5.CrossRefPubMed
3.
go back to reference Adler A, Amato MB, Arnold JH, et al. Whither lung EIT: where are we, where do we want to go and what do we need to get there? Physiol Meas. 2012;33:679–94.CrossRefPubMed Adler A, Amato MB, Arnold JH, et al. Whither lung EIT: where are we, where do we want to go and what do we need to get there? Physiol Meas. 2012;33:679–94.CrossRefPubMed
4.
go back to reference Frerichs I, Becher T, Weiler N. Electrical impedanc tomography imaging of the cardiopulmonary system. Curr Opin Crit Care. 2014;20:323–32.CrossRefPubMed Frerichs I, Becher T, Weiler N. Electrical impedanc tomography imaging of the cardiopulmonary system. Curr Opin Crit Care. 2014;20:323–32.CrossRefPubMed
5.
go back to reference Brown BH, Seagar AD. The Sheffield data collection system. Clin Phys Physiol Meas: an official journal of the Hospital Physicists’ Association, Deutsche Gesellschaft fur Medizinische Physik and the European Federation of Organisations for Medical Physics 8 Suppl A 1987;91–97. Brown BH, Seagar AD. The Sheffield data collection system. Clin Phys Physiol Meas: an official journal of the Hospital Physicists’ Association, Deutsche Gesellschaft fur Medizinische Physik and the European Federation of Organisations for Medical Physics 8 Suppl A 1987;91–97.
6.
go back to reference Yerworth R, Bayford R. The effect of serial data collection on the accuracy of electrical impedance tomography images. Physiol Meas. 2013;34:659–69.CrossRefPubMed Yerworth R, Bayford R. The effect of serial data collection on the accuracy of electrical impedance tomography images. Physiol Meas. 2013;34:659–69.CrossRefPubMed
7.
go back to reference Adler A, Gagnon H, Dittmar J, et al Temporal reconstructions in EIT. In: Adler A, Grychtol B, editors. 15th international conference on biomedical applications of electrical impedance Tomography, Gananoque, Ontario, Canada, 2014 p. 14. Adler A, Gagnon H, Dittmar J, et al Temporal reconstructions in EIT. In: Adler A, Grychtol B, editors. 15th international conference on biomedical applications of electrical impedance Tomography, Gananoque, Ontario, Canada, 2014 p. 14.
8.
go back to reference Hartinger AE, Guardo R, Adler A, et al. Real-time management of faulty electrodes in electrical impedance tomography. IEEE Trans Biomed Eng. 2009;56:369–77.CrossRefPubMed Hartinger AE, Guardo R, Adler A, et al. Real-time management of faulty electrodes in electrical impedance tomography. IEEE Trans Biomed Eng. 2009;56:369–77.CrossRefPubMed
9.
go back to reference Rooney D, Friese M, Fraser JF, et al. Gravity-dependent ventilation distribution in rats measured with electrical impedance tomography. Physiol Meas. 2009;30:1075–85.CrossRefPubMed Rooney D, Friese M, Fraser JF, et al. Gravity-dependent ventilation distribution in rats measured with electrical impedance tomography. Physiol Meas. 2009;30:1075–85.CrossRefPubMed
10.
go back to reference Heinrich S, Schiffmann H, Frerichs A, et al. Body and head position effects on regional lung ventilation in infants: an electrical impedance tomography study. Intensive Care Med. 2006;32:1392–8.CrossRefPubMed Heinrich S, Schiffmann H, Frerichs A, et al. Body and head position effects on regional lung ventilation in infants: an electrical impedance tomography study. Intensive Care Med. 2006;32:1392–8.CrossRefPubMed
11.
go back to reference Frerichs I, Schmitz G, Pulletz S, et al. Reproducibility of regional lung ventilation distribution determined by electrical impedance tomography during mechanical ventilation. Physiol Meas. 2007;28:S261–7.CrossRefPubMed Frerichs I, Schmitz G, Pulletz S, et al. Reproducibility of regional lung ventilation distribution determined by electrical impedance tomography during mechanical ventilation. Physiol Meas. 2007;28:S261–7.CrossRefPubMed
12.
go back to reference Tingay DG, Wallace MJ, Bhatia R, et al. Surfactant before the first inflation at birth improves spatial distribution of ventilation and reduces lung injury in preterm lambs. J Appl Physiol. 2014;116:251–8.CrossRefPubMed Tingay DG, Wallace MJ, Bhatia R, et al. Surfactant before the first inflation at birth improves spatial distribution of ventilation and reduces lung injury in preterm lambs. J Appl Physiol. 2014;116:251–8.CrossRefPubMed
13.
go back to reference Bhatia R, Schmölzer G, Davis P, et al. Electrical impedance tomography can rapidly detect small pneumothoraces in surfactant-depleted piglets. Intensive Care Med. 2012;38:308–15.CrossRefPubMed Bhatia R, Schmölzer G, Davis P, et al. Electrical impedance tomography can rapidly detect small pneumothoraces in surfactant-depleted piglets. Intensive Care Med. 2012;38:308–15.CrossRefPubMed
14.
go back to reference Blankman P, Hasan D, van Mourik MS, et al. Ventilation distribution measured with EIT at varying levels of pressure support and neurally adjusted ventilatory assist in patients with ALI. Intensive Care Med. 2013;39:1057–62.CrossRefPubMed Blankman P, Hasan D, van Mourik MS, et al. Ventilation distribution measured with EIT at varying levels of pressure support and neurally adjusted ventilatory assist in patients with ALI. Intensive Care Med. 2013;39:1057–62.CrossRefPubMed
15.
go back to reference Frerichs I, Schiffmann H, Oehler R, et al. Distribution of lung ventilation in spontaneously breathing neonates lying in different body positions. Intensive Care Med. 2003;29:787–94.CrossRefPubMed Frerichs I, Schiffmann H, Oehler R, et al. Distribution of lung ventilation in spontaneously breathing neonates lying in different body positions. Intensive Care Med. 2003;29:787–94.CrossRefPubMed
16.
go back to reference Reifferscheid F, Elke G, Pulletz S, et al. Regional ventilation distribution determined by electrical impedance tomography: reproducibility and effects of posture and chest plane. Respirology. 2011;16:523–31.CrossRefPubMed Reifferscheid F, Elke G, Pulletz S, et al. Regional ventilation distribution determined by electrical impedance tomography: reproducibility and effects of posture and chest plane. Respirology. 2011;16:523–31.CrossRefPubMed
17.
go back to reference Adler A, Arnold J, Bayford R, et al. GREIT: a unified approach to 2D linear EIT reconstruction of lung images. Physiol Meas. 2009;30:S35–55.CrossRefPubMed Adler A, Arnold J, Bayford R, et al. GREIT: a unified approach to 2D linear EIT reconstruction of lung images. Physiol Meas. 2009;30:S35–55.CrossRefPubMed
18.
go back to reference Schnidrig S, Casaulta C, Schibler A, et al. Influence of end-expiratory level and tidal volume on gravitational ventilation distribution during tidal breathing in healthy adults. Eur J Appl Physiol. 2013;113:591–8.CrossRefPubMed Schnidrig S, Casaulta C, Schibler A, et al. Influence of end-expiratory level and tidal volume on gravitational ventilation distribution during tidal breathing in healthy adults. Eur J Appl Physiol. 2013;113:591–8.CrossRefPubMed
Metadata
Title
Analysis and compensation for errors in electrical impedance tomography images and ventilation-related measures due to serial data collection
Authors
Rebecca J. Yerworth
Inéz Frerichs
Richard Bayford
Publication date
01-10-2017
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 5/2017
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-016-9920-y

Other articles of this Issue 5/2017

Journal of Clinical Monitoring and Computing 5/2017 Go to the issue