01-02-2011 | Symposium: Papers Presented at the Hip Society Meetings 2010

Acetabular UHMWPE Survival and Wear Changes With Different Manufacturing Techniques

Authors: John B. Meding, MD, E. Michael Keating, MD, Kenneth E. Davis, MS

Published in: Clinical Orthopaedics and Related Research® | Issue 2/2011

Login to get access

Abstract

Background

Polyethylene wear may be affected by the type of polyethylene resin, manufacturing technique, degree of thermal stabilization, and sterilization technique.

Questions/purposes

We therefore compared femoral head penetration into the PE and cup survival using the same cup system with different PE resins, manufacturing, and sterilization techniques.

Methods

Our study group consisted of 1912 THAs performed using the same uncemented cup and identical 28-mm cobalt-chrome heads. The polyethylene varied as follows: Group 1 (94 cups), GUR 4150 resin, ram-extruded, sterilized in air, no barrier packaging; Group 2 (74 cups), same as Group 1 but sterilized in argon; Group 3 (75 cups), Himont 1900 resin, compression-molded bar stock, sterilized in argon, no barrier packaging; Group 4 (620 cups), same as Group 3 except with barrier packing; Group 5 (711 cups), GUR 1050 resin, compression-molded bar stock, sterilized in argon gas with barrier packaging; and Group 6 (338 cups), GUR 1050 resin, compression-molded bar stock, sterilized in argon with barrier packaging, irradiated with 50 kGy, heated below melting temperature, machined, and finally placed in nonbarrier packaging with gas plasma sterilization. Minimum followup was 2 years (average, 7 years; range, 2–17 years).

Results

Femoral head penetration averaged 0.05 mm per year for Groups 5 and 6 and was substantially lower than for Groups 1 to 4. Cup survival was higher at seven years in Groups 3, 4, and 5, and at 10 years in group 4 when compared to groups 1, 2, and 3.

Conclusions

We observed lower FHP rates and higher cup survival with polyethylene machined from direct compression-molded bar stock, sterilized in argon gas, with barrier packaging.

Level of Evidence

Level III Therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.
Literature
1.
go back to reference Bankston AB, Keating EM, Ranawat C, Faris PM, Ritter MA. Comparison of polyethylene war in machined versus moled polyethylene. Clin Orthop Relat Res. 1995;317:37–43.PubMed Bankston AB, Keating EM, Ranawat C, Faris PM, Ritter MA. Comparison of polyethylene war in machined versus moled polyethylene. Clin Orthop Relat Res. 1995;317:37–43.PubMed
2.
go back to reference Bargmann LS, Bargmann BC, Collier JP Currier BH, Mayor MB. Current sterilization and packaging methods for polyethylene. Clin Orthop Relat Res. 1999;369:49–58.CrossRefPubMed Bargmann LS, Bargmann BC, Collier JP Currier BH, Mayor MB. Current sterilization and packaging methods for polyethylene. Clin Orthop Relat Res. 1999;369:49–58.CrossRefPubMed
3.
go back to reference Cautilli GP, Beight J, Yao B, Hozack WJ. A prospective review of 303 cementless universal cups with emphasis on wear as the cause of failure. Semin Arthroplasty. 1994;5:25–29.PubMed Cautilli GP, Beight J, Yao B, Hozack WJ. A prospective review of 303 cementless universal cups with emphasis on wear as the cause of failure. Semin Arthroplasty. 1994;5:25–29.PubMed
4.
go back to reference Devane PA, Horne JG. Assessment of polyethylene wear in total hip replacement. Clin Orthop Relat Res. 1999;369:59–72.CrossRefPubMed Devane PA, Horne JG. Assessment of polyethylene wear in total hip replacement. Clin Orthop Relat Res. 1999;369:59–72.CrossRefPubMed
5.
go back to reference Dorr LD, Wan ZW, Shahrdar C, Sirianni L, Boutary M, Yun A. Clinical performance of a Durasul highly cross-linked polyethylene acetabular liner for total hip arthroplasty at five years. J Bone Joint Surg Am. 2005;87:1816–1821.CrossRefPubMed Dorr LD, Wan ZW, Shahrdar C, Sirianni L, Boutary M, Yun A. Clinical performance of a Durasul highly cross-linked polyethylene acetabular liner for total hip arthroplasty at five years. J Bone Joint Surg Am. 2005;87:1816–1821.CrossRefPubMed
6.
go back to reference Faris PM, Ritter MA, Pierce AL, Davis KE, Faris GW. Polyethylene sterilization and production affects wear in total hip arthroplasties. Clin Orthop Relat Res. 2009;453:305–308.CrossRef Faris PM, Ritter MA, Pierce AL, Davis KE, Faris GW. Polyethylene sterilization and production affects wear in total hip arthroplasties. Clin Orthop Relat Res. 2009;453:305–308.CrossRef
7.
go back to reference Garvin KL, Hartman CW, Mangla J, Murdoch N, Martell JM. Wear analysis in THA utilizing oxidized zirconium and cross-linked polyethylene. Clin Orthop Relat Res. 2009;467:141–145.CrossRefPubMed Garvin KL, Hartman CW, Mangla J, Murdoch N, Martell JM. Wear analysis in THA utilizing oxidized zirconium and cross-linked polyethylene. Clin Orthop Relat Res. 2009;467:141–145.CrossRefPubMed
8.
go back to reference Heisel C, Silva M, Dela Rosa MA, Schmalzried TP. Short-term in vivo wear of cross-linked polyethylene. J Bone Joint Surg Am. 2004;86:748–751.PubMed Heisel C, Silva M, Dela Rosa MA, Schmalzried TP. Short-term in vivo wear of cross-linked polyethylene. J Bone Joint Surg Am. 2004;86:748–751.PubMed
9.
go back to reference Hui AJ, McCalden RW, Martell JM, MacDonald SJ, Bourne RB, Rorabeck CH. Validation of two and three-dimensional radiographic techniques for measuring polyethylene wear after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:505–511.PubMed Hui AJ, McCalden RW, Martell JM, MacDonald SJ, Bourne RB, Rorabeck CH. Validation of two and three-dimensional radiographic techniques for measuring polyethylene wear after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:505–511.PubMed
10.
go back to reference Isaac DL, Forder J, Skrme AD, James SE. The Biomet bi-metric total hip arthroplasty and universal cup system. J Arthroplasty. 2007;22:697–700.CrossRefPubMed Isaac DL, Forder J, Skrme AD, James SE. The Biomet bi-metric total hip arthroplasty and universal cup system. J Arthroplasty. 2007;22:697–700.CrossRefPubMed
11.
go back to reference Jacobs CA, Christenson CP, Greenwald AS, McKellop H. Clinical performance of highly cross-linked polyethylene in total hip arthroplasty. J Bone Joint Surg Am. 2007;89:2779–2786.CrossRefPubMed Jacobs CA, Christenson CP, Greenwald AS, McKellop H. Clinical performance of highly cross-linked polyethylene in total hip arthroplasty. J Bone Joint Surg Am. 2007;89:2779–2786.CrossRefPubMed
12.
go back to reference Kurtz SM, Muratoglu OK, Evans M, Edidin AA. Advances in the processing, sterilization, and crosslinking of ultra-high molecular weight polyethylene for total joint arthroplasty. Biomaterials. 1999;20:1659–1688.CrossRefPubMed Kurtz SM, Muratoglu OK, Evans M, Edidin AA. Advances in the processing, sterilization, and crosslinking of ultra-high molecular weight polyethylene for total joint arthroplasty. Biomaterials. 1999;20:1659–1688.CrossRefPubMed
13.
go back to reference Lu S, Orr JF, Buchanan FJ. The influence of inert packaging on the shelf aging of gamma-irradiation sterilized ultra-high molecular weight polyethylene. Biomaterials. 2003;24:139–145.CrossRefPubMed Lu S, Orr JF, Buchanan FJ. The influence of inert packaging on the shelf aging of gamma-irradiation sterilized ultra-high molecular weight polyethylene. Biomaterials. 2003;24:139–145.CrossRefPubMed
14.
go back to reference Manning DW, Chiang PP, Martell JM, Galante JO, Harris WH. In vivo wear study of traditional and highly cross-linked polyethylene in total hip arthroplasty. J Arthroplasty. 2005;20:880–886.CrossRefPubMed Manning DW, Chiang PP, Martell JM, Galante JO, Harris WH. In vivo wear study of traditional and highly cross-linked polyethylene in total hip arthroplasty. J Arthroplasty. 2005;20:880–886.CrossRefPubMed
15.
go back to reference Martell JM, Berkson E, Berger R, Jacobs J. Comparison of two and three-dimensional computerized polyethylene wear analysis after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:1111–1117.PubMed Martell JM, Berkson E, Berger R, Jacobs J. Comparison of two and three-dimensional computerized polyethylene wear analysis after total hip arthroplasty. J Bone Joint Surg Am. 2003;85:1111–1117.PubMed
16.
go back to reference Martell JM, Sunjay B. Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg Am.1997;79:1635–1641.PubMed Martell JM, Sunjay B. Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg Am.1997;79:1635–1641.PubMed
17.
go back to reference McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty. J Bone Joint Surg Am. 2009;91:773–782.CrossRefPubMed McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB, Chess DG, Charron KD. Wear rate of highly cross-linked polyethylene in total hip arthroplasty. J Bone Joint Surg Am. 2009;91:773–782.CrossRefPubMed
18.
go back to reference McKellop H, Shen FW, Lu B, Campbell P, Salovey R. Effect of sterilization method and other modifications on the wear resistance of acetabular cups made of ultra-high molecular weight polyethylene: a hip-simulator study. J Bone Joint Surg Am. 2000;82:1708–1725.PubMed McKellop H, Shen FW, Lu B, Campbell P, Salovey R. Effect of sterilization method and other modifications on the wear resistance of acetabular cups made of ultra-high molecular weight polyethylene: a hip-simulator study. J Bone Joint Surg Am. 2000;82:1708–1725.PubMed
19.
go back to reference Oonishi H, Kim SC, Takao Y, Kyomoto M, Iwamoto M, Ueno M. Wear of highly cross-linked polyethylene acetabular cup in Japan. J Arthroplasty. 2006;21:944–949.CrossRefPubMed Oonishi H, Kim SC, Takao Y, Kyomoto M, Iwamoto M, Ueno M. Wear of highly cross-linked polyethylene acetabular cup in Japan. J Arthroplasty. 2006;21:944–949.CrossRefPubMed
20.
go back to reference Puolakka TJ, Laine HJ, Moilanen TP, Koivisto AM, Pajamaki KJ. Alarming wear of the first-generation polyethylene liner of the cementless porous-coated Biomet Universal cup. Acta Orthop Scand. 2001;72:72–77.CrossRef Puolakka TJ, Laine HJ, Moilanen TP, Koivisto AM, Pajamaki KJ. Alarming wear of the first-generation polyethylene liner of the cementless porous-coated Biomet Universal cup. Acta Orthop Scand. 2001;72:72–77.CrossRef
21.
go back to reference Rohrl A, Nivbrant B, Mingguo L, Hewitt B. In vivo wear and migration of highly cross-linked polyethylene cups. J Arthroplasty. 2005;20:409–413.CrossRefPubMed Rohrl A, Nivbrant B, Mingguo L, Hewitt B. In vivo wear and migration of highly cross-linked polyethylene cups. J Arthroplasty. 2005;20:409–413.CrossRefPubMed
22.
go back to reference Sychterz CJ, Engh CA, Shah N, Engh CA. Radiographic evaluation of penetration by the femoral head into the polyethylene liner over time. J Bone Joint Surg Am. 1997;79:1040–1046.PubMed Sychterz CJ, Engh CA, Shah N, Engh CA. Radiographic evaluation of penetration by the femoral head into the polyethylene liner over time. J Bone Joint Surg Am. 1997;79:1040–1046.PubMed
23.
go back to reference Wroblewski BM, Siney PD, Dowson D, Collin SN. Prospective clinical and joint simulator studies of a new total hip arthroplasty using alumina ceramic heads and cross-linked polyethylene cups. J Bone Joint Surg Br. 1996;78:280–285.PubMed Wroblewski BM, Siney PD, Dowson D, Collin SN. Prospective clinical and joint simulator studies of a new total hip arthroplasty using alumina ceramic heads and cross-linked polyethylene cups. J Bone Joint Surg Br. 1996;78:280–285.PubMed
Metadata
Title
Acetabular UHMWPE Survival and Wear Changes With Different Manufacturing Techniques
Authors
John B. Meding, MD
E. Michael Keating, MD
Kenneth E. Davis, MS
Publication date
01-02-2011
Publisher
Springer-Verlag
Published in
Clinical Orthopaedics and Related Research® / Issue 2/2011
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-010-1571-6

Other articles of this Issue 2/2011

Clinical Orthopaedics and Related Research® 2/2011 Go to the issue

Symposium: Papers Presented at the Hip Society Meetings

High Complication Rate With Anterior Total Hip Arthroplasties on a Fracture Table

Symposium: Papers Presented at the Hip Society Meetings 2010

Improving Cup Positioning Using a Mechanical Navigation Instrument