Published in: Journal of Cardiothoracic Surgery 1/2013

Open Access 01-12-2013 | Research article

Accuracy of the Masimo Pronto-7® system in patients with left ventricular assist device

Authors: Christian S Bruells, Ares K Menon, Rolf Rossaint, Andreas Goetzenich, Michael Czaplik, Norbert Zoremba, Rüdiger Autschbach, Gereon Schaelte

Published in: Journal of Cardiothoracic Surgery | Issue 1/2013

Login to get access

Abstract

Background

The Masimo Pronto-7® calculates hemoglobin (Hb) values using the pulsoximetry technique and a variety of mathematical algorithms analyzing the pulse waveform. Although this system has demonstrated a high level of accuracy in average patients, the performance might be altered in special patient populations. Regarding patients with left ventricular cardiac failure, a rotary blood pump generates a constant, continuous, non-pulsatile flow to improve effective cardiac output. Due to this alteration in both, blood flow and arterial blood pressure we hypothesized a reduced accuracy of the Masimo Pronto-7® to detect Hb in patients with left ventricular cardiac failure. To test our hypothesis, we evaluated the Pronto-7®SpHb system in outpatients after continuous-flow-left ventricular assist device (cf-LVAD) implantation (HeartMate II, Thoratec).

Methods

21 cf-LVAD outpatients from the Clinic for Cardiac, Thoracic and Vascular Surgery were investigated during routine follow up examinations. After venous blood samples were drawn, the Pronto-7® sensor was attached to one randomly selected finger of one hand. The collected SpHb data were compared with Hb values measured by our central laboratory. The difference between the methods was determined using Bland – Altman analysis. The study was registered in the DRKS (DRKS00004415).

Results

In all cf-LVAD patients evaluated, the Pronto-7® successfully detected SpHb values. Using Bland – Altman analysis, a bias of 0.14 g/dl (95% upper and lower limits of agreement ± 2.76 g/dl) was calculated.

Conclusion

The Pronto-7® overestimated the actual Hb value in cf-LVAD outpatients with the HeartMate II. Due to this, we conclude that the system is suitable for screening in routine examinations and further analysis can be performed if needed. However, its use as an emergency tool is questionable because of the increased inaccuracy when Hb values are critically low.
Appendix
Available only for authorised users
Literature
1.
go back to reference Barker S, Badal J: The measurement of dyshemoglobins and total hemoglobin by pulse oximetry. Curr Opin Anaesthesiol. 2008, 21: 805-810. 10.1097/ACO.0b013e328316bb6f.CrossRefPubMed Barker S, Badal J: The measurement of dyshemoglobins and total hemoglobin by pulse oximetry. Curr Opin Anaesthesiol. 2008, 21: 805-810. 10.1097/ACO.0b013e328316bb6f.CrossRefPubMed
2.
go back to reference Berkow L, Rotolo S, Mirski E: Continuous noninvasive hemoglobin monitoring during complex spine surgery. Anesth Analg. 2011, 113: 1396-1402. 10.1213/ANE.0b013e318230b425.CrossRefPubMed Berkow L, Rotolo S, Mirski E: Continuous noninvasive hemoglobin monitoring during complex spine surgery. Anesth Analg. 2011, 113: 1396-1402. 10.1213/ANE.0b013e318230b425.CrossRefPubMed
3.
go back to reference Butwick A, Hilton G, Carvalho B: Non-invasive haemoglobin measurement in patients undergoing elective Caesarean section. Br J Anaesth. 2012, 108: 271-277. 10.1093/bja/aer373.CrossRefPubMed Butwick A, Hilton G, Carvalho B: Non-invasive haemoglobin measurement in patients undergoing elective Caesarean section. Br J Anaesth. 2012, 108: 271-277. 10.1093/bja/aer373.CrossRefPubMed
4.
go back to reference Macknet M, Allard M, Applegate R, Rook J: The accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-Oximetry in human subjects undergoing hemodilution. Anesth Analg. 2010, 111: 1424-1426. 10.1213/ANE.0b013e3181fc74b9.CrossRefPubMed Macknet M, Allard M, Applegate R, Rook J: The accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-Oximetry in human subjects undergoing hemodilution. Anesth Analg. 2010, 111: 1424-1426. 10.1213/ANE.0b013e3181fc74b9.CrossRefPubMed
5.
go back to reference Frasca D, Dahyot-Fizelier C, Catherine K, Levrat Q, Debaene B, Mimoz O: Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients. Crit Care Med. 2011, 39: 2277-2282. 10.1097/CCM.0b013e3182227e2d.CrossRefPubMed Frasca D, Dahyot-Fizelier C, Catherine K, Levrat Q, Debaene B, Mimoz O: Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients. Crit Care Med. 2011, 39: 2277-2282. 10.1097/CCM.0b013e3182227e2d.CrossRefPubMed
6.
go back to reference Slaughter M, Milano C, Russell S, Conte J, Feldman D, Sun B: Advanced heart failure treated with continuous-flow left ventricular assist device. N Engl J Med. 2009, 361: 2241-2251. 10.1056/NEJMoa0909938.CrossRefPubMed Slaughter M, Milano C, Russell S, Conte J, Feldman D, Sun B: Advanced heart failure treated with continuous-flow left ventricular assist device. N Engl J Med. 2009, 361: 2241-2251. 10.1056/NEJMoa0909938.CrossRefPubMed
7.
go back to reference Bland J, Altman D: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986, 8: 307-310.CrossRef Bland J, Altman D: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986, 8: 307-310.CrossRef
8.
go back to reference Zoremba N, Brulls C, Thiel V, Rohl A, Rossaint R: Pulse oximetry during intraaortic balloon pump application. Anesth Analg. 2011, 55: 322-327. Zoremba N, Brulls C, Thiel V, Rohl A, Rossaint R: Pulse oximetry during intraaortic balloon pump application. Anesth Analg. 2011, 55: 322-327.
9.
go back to reference Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B: Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue(R) with automated laboratory hemoglobin measurement. Anesthesiology. 2011, 115: 548-555. 10.1097/ALN.0b013e3182270c22.CrossRefPubMed Lamhaut L, Apriotesei R, Combes X, Lejay M, Carli P, Vivien B: Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue(R) with automated laboratory hemoglobin measurement. Anesthesiology. 2011, 115: 548-555. 10.1097/ALN.0b013e3182270c22.CrossRefPubMed
10.
go back to reference Miller R, Ward T, Shiboski S, Cohen N: A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg. 2011, 112: 858-863. 10.1213/ANE.0b013e31820eecd1.CrossRefPubMed Miller R, Ward T, Shiboski S, Cohen N: A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery. Anesth Analg. 2011, 112: 858-863. 10.1213/ANE.0b013e31820eecd1.CrossRefPubMed
11.
go back to reference Nguyen B, Vincent J, Nowak E, Coat M, Paleiron N, Gouny P: The accuracy of noninvasive hemoglobin measurement by multiwavelength pulse oximetry after cardiac surgery. Anesth Analg. 2011, 113: 1052-1057. 10.1213/ANE.0b013e31822c9679.CrossRefPubMed Nguyen B, Vincent J, Nowak E, Coat M, Paleiron N, Gouny P: The accuracy of noninvasive hemoglobin measurement by multiwavelength pulse oximetry after cardiac surgery. Anesth Analg. 2011, 113: 1052-1057. 10.1213/ANE.0b013e31822c9679.CrossRefPubMed
12.
go back to reference Gayat E, Aulagnier J, Matthieu E, Boisson M, Fischler M: Non-invasive measurement of hemoglobin: assessment of two different point-of-care technologies. PLoS One. 2012, 7: e30065-10.1371/journal.pone.0030065.CrossRefPubMedPubMedCentral Gayat E, Aulagnier J, Matthieu E, Boisson M, Fischler M: Non-invasive measurement of hemoglobin: assessment of two different point-of-care technologies. PLoS One. 2012, 7: e30065-10.1371/journal.pone.0030065.CrossRefPubMedPubMedCentral
13.
go back to reference Gayat E, Bodin A, Sportiello C, Boisson M, Dreyfus J, Mathieu E: Performance evaluation of a noninvasive hemoglobin monitoring device. Ann Emerg Med. 2011, 57: 330-333. 10.1016/j.annemergmed.2010.11.032.CrossRefPubMed Gayat E, Bodin A, Sportiello C, Boisson M, Dreyfus J, Mathieu E: Performance evaluation of a noninvasive hemoglobin monitoring device. Ann Emerg Med. 2011, 57: 330-333. 10.1016/j.annemergmed.2010.11.032.CrossRefPubMed
14.
go back to reference Coquin J, Dewitte A, Manach YL, Caujolle M, Joannes-Boyau O, Fleureau C: Precision of noninvasive hemoglobin-level measurement by pulse co-oximetry in patients admitted to intensive care units for severe gastrointestinal bleeds. Crit Care Med. 2012, 40: 2576-2582. 10.1097/CCM.0b013e318258fb4f.CrossRefPubMed Coquin J, Dewitte A, Manach YL, Caujolle M, Joannes-Boyau O, Fleureau C: Precision of noninvasive hemoglobin-level measurement by pulse co-oximetry in patients admitted to intensive care units for severe gastrointestinal bleeds. Crit Care Med. 2012, 40: 2576-2582. 10.1097/CCM.0b013e318258fb4f.CrossRefPubMed
15.
go back to reference Morey T, Gravenstein N, Rice M: Let’s think clinically instead of mathematically about device accuracy. Anesth Analg. 2011, 113: 89-91. 10.1213/ANE.0b013e318219a290.CrossRefPubMed Morey T, Gravenstein N, Rice M: Let’s think clinically instead of mathematically about device accuracy. Anesth Analg. 2011, 113: 89-91. 10.1213/ANE.0b013e318219a290.CrossRefPubMed
Metadata
Title
Accuracy of the Masimo Pronto-7® system in patients with left ventricular assist device
Authors
Christian S Bruells
Ares K Menon
Rolf Rossaint
Andreas Goetzenich
Michael Czaplik
Norbert Zoremba
Rüdiger Autschbach
Gereon Schaelte
Publication date
01-12-2013
Publisher
BioMed Central
Published in
Journal of Cardiothoracic Surgery / Issue 1/2013
Electronic ISSN: 1749-8090
DOI
https://doi.org/10.1186/1749-8090-8-159