Accuracy of the intermittently scanned continuous glucose monitoring system in critically ill patients: a prospective, multicenter, observational study

Authors: Weifeng Huang, Siwan Li, Jingyi Lu, Yun Shen, Yaxin Wang, Yufei Wang, Kaixuan Feng, Xiaoli Huang, Yan Zou, Linjie Hu, Yihan Lu, Jian Zhou, Yingchuan Li

Published in: Endocrine | Issue 3/2022

Login to get access

Abstract

Objective

Continuous glucose monitoring (CGM) has the potential to improve glucose control in the intensive care unit (ICU) setting. We sought to evaluate the accuracy of the intermittently scanned CGM (isCGM) system in critically ill patients.

Research design and methods

Adult patients were consecutively enrolled from three ICUs from August 2020 to January 2021. The performance of FreeStyle Libre Pro was evaluated against the venous blood glucose samples as a reference. Numerical accuracy was examined by the mean absolute relative difference (MARD), the Bland–Altman analysis, and the International Organization for Standardization criteria. Clinical accuracy was assessed by performing the Clarke and consensus error grid analysis.

Results

A total of 122 patients were included and 3416 matched glucose pairs were analyzed. The overall MARD was 18.0%, and the highest MARD (33.1%) was observed in the hypoglycemic range (<70 mg/dL). The Bland–Altman analysis revealed a mean bias of −11.7 mg/dL, with the 95% limits of agreement of −73.0 to 49.5 mg/dL. The percentages of isCGM glucose values within ±15%/15, ±20%/20, and ±30%/30 mg/dL were 49.8%, 64.7%, and 84.5%, respectively. The Clarke and consensus error grid analysis showed acceptable clinical accuracy with 98.5% and 98.8% of glucose values falling into zones A and B.

Conclusions

Our study demonstrated suboptimal overall accuracy of isCGM for critically ill patients. Whether the adjunctive use of isCGM could improve glucose management and health outcomes in the critically ill needs further investigation.

Clinical trial registration

ChiCTR2100042036, Chinese Clinical Trial Registry.
Appendix
Available only for authorised users
Literature
1.
go back to reference G. van den Berghe, P. Wouters, F. Weekers et al. Intensive insulin therapy in critically ill patients. N. Engl. J. Med. 345(19), 1359–1367 (2001)CrossRefPubMed G. van den Berghe, P. Wouters, F. Weekers et al. Intensive insulin therapy in critically ill patients. N. Engl. J. Med. 345(19), 1359–1367 (2001)CrossRefPubMed
2.
go back to reference American Diabetes Association, 15. Diabetes care in the hospital: standards of medical care in diabetes—2021. Diabetes Care 44(Supplement 1), S211–S220 (2021)CrossRef American Diabetes Association, 15. Diabetes care in the hospital: standards of medical care in diabetes—2021. Diabetes Care 44(Supplement 1), S211–S220 (2021)CrossRef
3.
4.
go back to reference R.W. Beck, T. Riddlesworth, K. Ruedy et al. Effect of continuous glucose monitoring on glycemic control in adults with type 1 diabetes using insulin injections: the DIAMOND randomized clinical trial. JAMA 317(4), 371–378 (2017)CrossRefPubMed R.W. Beck, T. Riddlesworth, K. Ruedy et al. Effect of continuous glucose monitoring on glycemic control in adults with type 1 diabetes using insulin injections: the DIAMOND randomized clinical trial. JAMA 317(4), 371–378 (2017)CrossRefPubMed
5.
go back to reference R.A. Vigersky, S.J. Fonda, M. Chellappa et al. Short- and long-term effects of real-time continuous glucose monitoring in patients with type 2 diabetes. Diabetes Care 35(1), 32–38 (2012)CrossRefPubMed R.A. Vigersky, S.J. Fonda, M. Chellappa et al. Short- and long-term effects of real-time continuous glucose monitoring in patients with type 2 diabetes. Diabetes Care 35(1), 32–38 (2012)CrossRefPubMed
6.
go back to reference A.M. Gómez, G.E. Umpierrez, O.M. Muñoz et al. Continuous glucose monitoring versus capillary point-of-care testing for inpatient glycemic control in type 2 diabetes patients hospitalized in the general ward and treated with a basal bolus insulin regimen. J. Diabetes Sci. Technol. 10(2), 325–329 (2015)CrossRefPubMedPubMedCentral A.M. Gómez, G.E. Umpierrez, O.M. Muñoz et al. Continuous glucose monitoring versus capillary point-of-care testing for inpatient glycemic control in type 2 diabetes patients hospitalized in the general ward and treated with a basal bolus insulin regimen. J. Diabetes Sci. Technol. 10(2), 325–329 (2015)CrossRefPubMedPubMedCentral
7.
go back to reference S. Rijkenberg, S.C. van Steen, J.H. DeVries et al. Accuracy and reliability of a subcutaneous continuous glucose monitoring device in critically ill patients. J. Clin. Monit. Comput. 32(5), 953–964 (2018)CrossRefPubMed S. Rijkenberg, S.C. van Steen, J.H. DeVries et al. Accuracy and reliability of a subcutaneous continuous glucose monitoring device in critically ill patients. J. Clin. Monit. Comput. 32(5), 953–964 (2018)CrossRefPubMed
8.
go back to reference P. Ancona, G.M. Eastwood, L. Lucchetta et al. The performance of flash glucose monitoring in critically ill patients with diabetes. Crit. Care Resusc. 19(2), 167–174 (2017)PubMed P. Ancona, G.M. Eastwood, L. Lucchetta et al. The performance of flash glucose monitoring in critically ill patients with diabetes. Crit. Care Resusc. 19(2), 167–174 (2017)PubMed
9.
go back to reference F. Schierenbeck, A. Franco-Cereceda, J. Liska, Accuracy of 2 different continuous glucose monitoring systems in patients undergoing cardiac surgery. J. Diabetes Sci. Technol. 11(1), 108–116 (2017)CrossRefPubMed F. Schierenbeck, A. Franco-Cereceda, J. Liska, Accuracy of 2 different continuous glucose monitoring systems in patients undergoing cardiac surgery. J. Diabetes Sci. Technol. 11(1), 108–116 (2017)CrossRefPubMed
10.
go back to reference E. Kotzapanagiotou, E. Tsotridou, E. Volakli et al. Evaluation of continuous flash glucose monitoring in a pediatric ICU setting. J. Clin. Monit. Comput. 34(4), 843–852 (2020)CrossRefPubMed E. Kotzapanagiotou, E. Tsotridou, E. Volakli et al. Evaluation of continuous flash glucose monitoring in a pediatric ICU setting. J. Clin. Monit. Comput. 34(4), 843–852 (2020)CrossRefPubMed
11.
go back to reference M.C. Perez-Guzman, E. Duggan, S. Gibanica et al. Continuous glucose monitoring in the operating room and cardiac intensive care unit. Diabetes Care 44(3), e50–e52 (2021)CrossRefPubMedPubMedCentral M.C. Perez-Guzman, E. Duggan, S. Gibanica et al. Continuous glucose monitoring in the operating room and cardiac intensive care unit. Diabetes Care 44(3), e50–e52 (2021)CrossRefPubMedPubMedCentral
12.
go back to reference Y. Zhang, X. Liu, J. Zhang et al. Evaluation for the feasibility and accuracy of Freestyle Libre Flash Glucose Monitoring System used by COVID-19 patients in intensive care unit. J. Diabetes 13(7), 603–605 (2021)CrossRefPubMed Y. Zhang, X. Liu, J. Zhang et al. Evaluation for the feasibility and accuracy of Freestyle Libre Flash Glucose Monitoring System used by COVID-19 patients in intensive care unit. J. Diabetes 13(7), 603–605 (2021)CrossRefPubMed
13.
go back to reference D.C. Klonoff, J.L. Parkes, B.P. Kovatchev et al. Investigation of the accuracy of 18 marketed blood glucose monitors. Diabetes Care 41(8), 1681–1688 (2018)CrossRefPubMed D.C. Klonoff, J.L. Parkes, B.P. Kovatchev et al. Investigation of the accuracy of 18 marketed blood glucose monitors. Diabetes Care 41(8), 1681–1688 (2018)CrossRefPubMed
14.
go back to reference M.J. Rice, J.L. Smith, D.B. Coursin, Glucose measurement in the ICU: regulatory intersects reality. Crit. Care Med. 45(4), 741–743 (2017)CrossRefPubMed M.J. Rice, J.L. Smith, D.B. Coursin, Glucose measurement in the ICU: regulatory intersects reality. Crit. Care Med. 45(4), 741–743 (2017)CrossRefPubMed
15.
go back to reference V. Tyndall, R.H. Stimson, N.N. Zammitt et al. Marked improvement in HbA following commencement of flash glucose monitoring in people with type 1 diabetes. Diabetologia 62(8), 1349–1356 (2019)CrossRefPubMedPubMedCentral V. Tyndall, R.H. Stimson, N.N. Zammitt et al. Marked improvement in HbA following commencement of flash glucose monitoring in people with type 1 diabetes. Diabetologia 62(8), 1349–1356 (2019)CrossRefPubMedPubMedCentral
16.
go back to reference M. Fokkert, P. van Dijk, M. Edens et al. Improved well-being and decreased disease burden after 1-year use of flash glucose monitoring (FLARE-NL4). BMJ Open Diabetes Res. Care 7(1), e000809 (2019)CrossRefPubMedPubMedCentral M. Fokkert, P. van Dijk, M. Edens et al. Improved well-being and decreased disease burden after 1-year use of flash glucose monitoring (FLARE-NL4). BMJ Open Diabetes Res. Care 7(1), e000809 (2019)CrossRefPubMedPubMedCentral
17.
go back to reference S. Charleer, C. De Block, Van, L. Huffel et al. Quality of life and glucose control after 1 year of nationwide reimbursement of intermittently scanned continuous glucose monitoring in adults living with type 1 diabetes (FUTURE): a prospective observational real-world cohort study. Diabetes Care 43(2), 389–397 (2020)CrossRefPubMed S. Charleer, C. De Block, Van, L. Huffel et al. Quality of life and glucose control after 1 year of nationwide reimbursement of intermittently scanned continuous glucose monitoring in adults living with type 1 diabetes (FUTURE): a prospective observational real-world cohort study. Diabetes Care 43(2), 389–397 (2020)CrossRefPubMed
18.
go back to reference J. Bolinder, R. Antuna, P. Geelhoed-Duijvestijn et al. Novel glucose-sensing technology and hypoglycaemia in type 1 diabetes: a multicentre, non-masked, randomised controlled trial. Lancet 388(10057), 2254–2263 (2016)CrossRefPubMed J. Bolinder, R. Antuna, P. Geelhoed-Duijvestijn et al. Novel glucose-sensing technology and hypoglycaemia in type 1 diabetes: a multicentre, non-masked, randomised controlled trial. Lancet 388(10057), 2254–2263 (2016)CrossRefPubMed
19.
Metadata
Title
Accuracy of the intermittently scanned continuous glucose monitoring system in critically ill patients: a prospective, multicenter, observational study
Authors
Weifeng Huang
Siwan Li
Jingyi Lu
Yun Shen
Yaxin Wang
Yufei Wang
Kaixuan Feng
Xiaoli Huang
Yan Zou
Linjie Hu
Yihan Lu
Jian Zhou
Yingchuan Li
Publication date
13-10-2022
Publisher
Springer US
Published in
Endocrine / Issue 3/2022
Print ISSN: 1355-008X
Electronic ISSN: 1559-0100
DOI
https://doi.org/10.1007/s12020-022-03216-3

Other articles of this Issue 3/2022

Endocrine 3/2022 Go to the issue
Obesity Clinical Trial Summary

At a glance: The STEP trials

A round-up of the STEP phase 3 clinical trials evaluating semaglutide for weight loss in people with overweight or obesity.

Developed by: Springer Medicine

Highlights from the ACC 2024 Congress

Year in Review: Pediatric cardiology

Watch Dr. Anne Marie Valente present the last year's highlights in pediatric and congenital heart disease in the official ACC.24 Year in Review session.

Year in Review: Pulmonary vascular disease

The last year's highlights in pulmonary vascular disease are presented by Dr. Jane Leopold in this official video from ACC.24.

Year in Review: Valvular heart disease

Watch Prof. William Zoghbi present the last year's highlights in valvular heart disease from the official ACC.24 Year in Review session.

Year in Review: Heart failure and cardiomyopathies

Watch this official video from ACC.24. Dr. Biykem Bozkurt discuss last year's major advances in heart failure and cardiomyopathies.