Accuracy of fit for cobaltchromium bar over two implants fabricated with different manufacturing techniques: an in-vitro study

Authors: Hossam I. Nassar, Ayman Fateen

Published in: BMC Oral Health | Issue 1/2023

Login to get access

Abstract

Objective

The purpose of the invitro research was to compare the fit of Cobalt Chromium customized bar fabricated with different manufacturing processes cast metal bar, milled bar and 3D printed bar using scanning electron microscope.

Materials and methods

Clear epoxy resin molds were prepared. In each mold two parallel implants with a 14 mm distance from each other were embedded. Thirty Co-Cr custom bars were constructed and were divided equally into three groups: Group (I) (Co-Cr conv), group (II) milled bar (Co-Cr milled), and group (III) printed bar (Co-Cr print). The marginal fit at implant-abutment interface was scanned using scanning electron microscope (SEM).

Results

There was a significant difference between the three studied groups regarding marginal misfit the between implant and fabricated bars with p-value < 0.001. The highest value of micro-gap distance was found in Co-Cr conventional group (7.95 ± 2.21 μm) followed by Co-Cr 3D printed group (4.98 ± 1.73) and the lower value were found in Co-Cr milled (3.22 ± 0.75).

Conclusion

The marginal fit of milled, 3D printed and conventional cast for Co-Cr alloy were within the clinically acceptable range of misfit. CAD/CAM milled Co-Cr bar revealed a superior internal fit at the implant-abutment interface. This was followed by selective laser melting (SLM) 3D printed bar and the least fit was shown for customized bar with the conventional lost wax technique.
Literature
1.
go back to reference Jung RE, Pjetursson BE, Glauser R, Zembic A, Zwahlen M, Lang NP. A systematic review of the 5-year survival and complication rates of implant‐supported single crowns. Clin Oral Implants Res. 2008;19(2):119–30.CrossRefPubMed Jung RE, Pjetursson BE, Glauser R, Zembic A, Zwahlen M, Lang NP. A systematic review of the 5-year survival and complication rates of implant‐supported single crowns. Clin Oral Implants Res. 2008;19(2):119–30.CrossRefPubMed
2.
go back to reference Sailer I, Philipp A, Zembic A, Pjetursson BE, Hämmerle CH, Zwahlen M. A systematic review of the performance of ceramic and metal implant abutments supporting fixed implant reconstructions. Clin Oral Implants Res. 2009;20:4–31.CrossRefPubMed Sailer I, Philipp A, Zembic A, Pjetursson BE, Hämmerle CH, Zwahlen M. A systematic review of the performance of ceramic and metal implant abutments supporting fixed implant reconstructions. Clin Oral Implants Res. 2009;20:4–31.CrossRefPubMed
3.
go back to reference Alqutaibi AY, Aboalrejal AN. Microgap and micromotion at the implant abutment interface cause marginal bone loss around dental implant but more evidence is needed. J Evid Based Dent Pract. 2018;18(2):171–2.CrossRefPubMed Alqutaibi AY, Aboalrejal AN. Microgap and micromotion at the implant abutment interface cause marginal bone loss around dental implant but more evidence is needed. J Evid Based Dent Pract. 2018;18(2):171–2.CrossRefPubMed
4.
go back to reference Liu Y, Wang J. Influences of microgap and micromotion of implant–abutment interface on marginal bone loss around implant neck. Arch Oral Biol. 2017;83:153–60.CrossRefPubMed Liu Y, Wang J. Influences of microgap and micromotion of implant–abutment interface on marginal bone loss around implant neck. Arch Oral Biol. 2017;83:153–60.CrossRefPubMed
5.
go back to reference Katsoulis J, Takeichi T, Sol Gaviria A, Peter L, Katsoulis K. Misfit of implant prostheses and its impact on clinical outcomes. Definition, assessment and a systematic review of the literature. Eur J Oral Implantol. 2017;10(Suppl 1):121–38.PubMed Katsoulis J, Takeichi T, Sol Gaviria A, Peter L, Katsoulis K. Misfit of implant prostheses and its impact on clinical outcomes. Definition, assessment and a systematic review of the literature. Eur J Oral Implantol. 2017;10(Suppl 1):121–38.PubMed
6.
go back to reference Priest G. Virtual-designed and computer-milled implant abutments. J Oral Maxillofac Surg. 2005;63(9):22–32.CrossRefPubMed Priest G. Virtual-designed and computer-milled implant abutments. J Oral Maxillofac Surg. 2005;63(9):22–32.CrossRefPubMed
7.
go back to reference Bondan JL, Oshima H, Shinkai RS, Mota EG, Meyer KR. Marginal fit analysis of premachined and castable UCLA abutments. Acta Odontológica Latinoamericana. 2009;22(2):139–42.PubMed Bondan JL, Oshima H, Shinkai RS, Mota EG, Meyer KR. Marginal fit analysis of premachined and castable UCLA abutments. Acta Odontológica Latinoamericana. 2009;22(2):139–42.PubMed
8.
go back to reference Silva CE, Soares S, Machado CM, Bergamo ET, Coelho PG, Witek L, Ramalho IS, Jalkh EB, Bonfante EA. Effect of CAD/CAM abutment height and cement type on the retention of zirconia crowns. Implant Dent. 2018;27(5):582–7.CrossRefPubMed Silva CE, Soares S, Machado CM, Bergamo ET, Coelho PG, Witek L, Ramalho IS, Jalkh EB, Bonfante EA. Effect of CAD/CAM abutment height and cement type on the retention of zirconia crowns. Implant Dent. 2018;27(5):582–7.CrossRefPubMed
9.
go back to reference Lopes AC, Machado CM, Bonjardim LR, Bergamo ET, Ramalho IS, Witek L, Coelho PG, Bonfante EA. The effect of CAD/CAM crown material and cement type on retention to implant abutments. J Prosthodont. 2019;28(2):e552–6.CrossRefPubMed Lopes AC, Machado CM, Bonjardim LR, Bergamo ET, Ramalho IS, Witek L, Coelho PG, Bonfante EA. The effect of CAD/CAM crown material and cement type on retention to implant abutments. J Prosthodont. 2019;28(2):e552–6.CrossRefPubMed
10.
go back to reference Gigandet M, Bigolin G, Faoro F, Bürgin W, Brägger U. Implants with original and non-original abutment connections. Clin Implant Dent Relat Res. 2014;16(2):303–11.CrossRefPubMed Gigandet M, Bigolin G, Faoro F, Bürgin W, Brägger U. Implants with original and non-original abutment connections. Clin Implant Dent Relat Res. 2014;16(2):303–11.CrossRefPubMed
11.
go back to reference Abdelhamed MI, Galley JD, Bailey MT, Johnston WM, Holloway J, McGlumphy E, et al. A comparison of Zirconia and Titanium abutments for Microleakage. Clin Implant Dent Relat Res. 2015;17(Suppl 2):e643–51.PubMed Abdelhamed MI, Galley JD, Bailey MT, Johnston WM, Holloway J, McGlumphy E, et al. A comparison of Zirconia and Titanium abutments for Microleakage. Clin Implant Dent Relat Res. 2015;17(Suppl 2):e643–51.PubMed
12.
go back to reference Molinero-Mourelle P, Cascos-Sanchez R, Yilmaz B, Lam WYH, Pow EHN, Del Río Highsmith J et al. Effect of fabrication technique on the Microgap of CAD/CAM Cobalt-Chrome and Zirconia abutments on a conical connection Implant: an in Vitro Study. Mater (Basel). 2021;14(9). Molinero-Mourelle P, Cascos-Sanchez R, Yilmaz B, Lam WYH, Pow EHN, Del Río Highsmith J et al. Effect of fabrication technique on the Microgap of CAD/CAM Cobalt-Chrome and Zirconia abutments on a conical connection Implant: an in Vitro Study. Mater (Basel). 2021;14(9).
13.
go back to reference Pjetursson BE, Thoma D, Jung R, Zwahlen M, Zembic A. A systematic review of the survival and complication rates of implant-supported fixed dental prostheses (FDPs) after a mean observation period of at least 5 years. Clin Oral Implants Res. 2012;23(Suppl 6):22–38.CrossRefPubMed Pjetursson BE, Thoma D, Jung R, Zwahlen M, Zembic A. A systematic review of the survival and complication rates of implant-supported fixed dental prostheses (FDPs) after a mean observation period of at least 5 years. Clin Oral Implants Res. 2012;23(Suppl 6):22–38.CrossRefPubMed
14.
go back to reference Kim MJ, Choi YJ, Kim SK, Heo SJ, Koak JY. Marginal accuracy and internal fit of 3-D Printing Laser-Sintered Co-cr Alloy Copings. Mater (Basel). 2017;10(1). Kim MJ, Choi YJ, Kim SK, Heo SJ, Koak JY. Marginal accuracy and internal fit of 3-D Printing Laser-Sintered Co-cr Alloy Copings. Mater (Basel). 2017;10(1).
15.
go back to reference Abduo J. Fit of CAD/CAM implant frameworks: a comprehensive review. J Oral Implantology. 2014;40(6):758–66.CrossRefPubMed Abduo J. Fit of CAD/CAM implant frameworks: a comprehensive review. J Oral Implantology. 2014;40(6):758–66.CrossRefPubMed
16.
go back to reference Abduo J, Bennani V, Waddell N, Lyons K, Swain M. Assessing the fit of implant fixed prostheses: a critical review. Int J Oral Maxillofacial Implants. 2010;25(3). Abduo J, Bennani V, Waddell N, Lyons K, Swain M. Assessing the fit of implant fixed prostheses: a critical review. Int J Oral Maxillofacial Implants. 2010;25(3).
17.
go back to reference Blatz MB, Conejo J. The current state of chairside digital dentistry and materials. Dent Clin. 2019;63(2):175–97. Blatz MB, Conejo J. The current state of chairside digital dentistry and materials. Dent Clin. 2019;63(2):175–97.
18.
go back to reference Chen J, Xiao Z, Yangpeng S, Deng F, Zhiguang Z. Production of inter-connective porous dental implants by computer-aided design and metal three-dimensional printing. J Biomater Appl. 2020;34(9):1227–38.CrossRefPubMed Chen J, Xiao Z, Yangpeng S, Deng F, Zhiguang Z. Production of inter-connective porous dental implants by computer-aided design and metal three-dimensional printing. J Biomater Appl. 2020;34(9):1227–38.CrossRefPubMed
19.
go back to reference Pereira T, Kennedy JV, Potgieter J. A comparison of traditional manufacturing vs additive manufacturing, the best method for the job. Procedia Manuf. 2019;30:11–8.CrossRef Pereira T, Kennedy JV, Potgieter J. A comparison of traditional manufacturing vs additive manufacturing, the best method for the job. Procedia Manuf. 2019;30:11–8.CrossRef
20.
go back to reference Oliveira TT, Reis AC. Fabrication of dental implants by the additive manufacturing method: a systematic review. J Prosthet Dent. 2019;122(3):270–4.CrossRefPubMed Oliveira TT, Reis AC. Fabrication of dental implants by the additive manufacturing method: a systematic review. J Prosthet Dent. 2019;122(3):270–4.CrossRefPubMed
21.
go back to reference Wally ZJ, Haque AM, Feteira A, Claeyssens F, Goodall R, Reilly GC. Selective laser melting processed Ti6Al4V lattices with graded porosities for dental applications. J Mech Behav Biomed Mater. 2019;90:20–9.CrossRefPubMed Wally ZJ, Haque AM, Feteira A, Claeyssens F, Goodall R, Reilly GC. Selective laser melting processed Ti6Al4V lattices with graded porosities for dental applications. J Mech Behav Biomed Mater. 2019;90:20–9.CrossRefPubMed
22.
go back to reference Mangano C, Bianchi A, Mangano FG, Dana J, Colombo M, Solop I, Admakin O. Custom-made 3D printed subperiosteal titanium implants for the prosthetic restoration of the atrophic posterior mandible of elderly patients: a case series. 3D Print Med. 2020;6(1):1–4.PubMedCentralCrossRefPubMed Mangano C, Bianchi A, Mangano FG, Dana J, Colombo M, Solop I, Admakin O. Custom-made 3D printed subperiosteal titanium implants for the prosthetic restoration of the atrophic posterior mandible of elderly patients: a case series. 3D Print Med. 2020;6(1):1–4.PubMedCentralCrossRefPubMed
23.
go back to reference Mounir M, Atef M, Abou-Elfetouh A, Hakam MM. Titanium and polyether ether ketone (PEEK) patient-specific sub-periosteal implants: two novel approaches for rehabilitation of the severely atrophic anterior maxillary ridge. Int J Oral Maxillofac Surg. 2018;47(5):658–64.CrossRefPubMed Mounir M, Atef M, Abou-Elfetouh A, Hakam MM. Titanium and polyether ether ketone (PEEK) patient-specific sub-periosteal implants: two novel approaches for rehabilitation of the severely atrophic anterior maxillary ridge. Int J Oral Maxillofac Surg. 2018;47(5):658–64.CrossRefPubMed
24.
go back to reference Wang Y, Chen X, Zhang C, Feng W, Zhang P, Chen Y, Huang J, Luo Y, Chen J. Studies on the performance of selective laser melting porous dental implant by finite element model simulation, fatigue testing and in vivo experiments. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 2019;233(2):170 – 80. Wang Y, Chen X, Zhang C, Feng W, Zhang P, Chen Y, Huang J, Luo Y, Chen J. Studies on the performance of selective laser melting porous dental implant by finite element model simulation, fatigue testing and in vivo experiments. Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine. 2019;233(2):170 – 80.
25.
go back to reference Hartmann A, Hildebrandt H, Schmohl JU, Kämmerer PW. Evaluation of risk parameters in bone regeneration using a customized titanium mesh: results of a clinical study. Implant Dent. 2019;28(6):543–50.PubMed Hartmann A, Hildebrandt H, Schmohl JU, Kämmerer PW. Evaluation of risk parameters in bone regeneration using a customized titanium mesh: results of a clinical study. Implant Dent. 2019;28(6):543–50.PubMed
26.
go back to reference Hartmann A, Seiler M. Minimizing risk of customized titanium mesh exposures–a retrospective analysis. BMC Oral Health. 2020;20(1):1–9.CrossRef Hartmann A, Seiler M. Minimizing risk of customized titanium mesh exposures–a retrospective analysis. BMC Oral Health. 2020;20(1):1–9.CrossRef
27.
go back to reference Revilla-León M, Ceballos L, Martínez-Klemm I, Özcan M. Discrepancy of complete-arch titanium frameworks manufactured using selective laser melting and electron beam melting additive manufacturing technologies. J Prosthet Dent. 2018;120(6):942–7.CrossRefPubMed Revilla-León M, Ceballos L, Martínez-Klemm I, Özcan M. Discrepancy of complete-arch titanium frameworks manufactured using selective laser melting and electron beam melting additive manufacturing technologies. J Prosthet Dent. 2018;120(6):942–7.CrossRefPubMed
28.
go back to reference Revilla-León M, Ceballos L, Özcan M. Implant-Prosthodontic discrepancy of Complete-Arch Cobalt-Chromium Implant frameworks Manufactured through Selective Laser Melting Additive Manufacturing Technology using a coordinate measuring machine. Int J Oral Maxillofacial Implants. 2019;34(3). Revilla-León M, Ceballos L, Özcan M. Implant-Prosthodontic discrepancy of Complete-Arch Cobalt-Chromium Implant frameworks Manufactured through Selective Laser Melting Additive Manufacturing Technology using a coordinate measuring machine. Int J Oral Maxillofacial Implants. 2019;34(3).
29.
go back to reference Ciocca L, Meneghello R, Savio G, Scheda L, Monaco C, Gatto MR, Micarelli C, Baldissara P. Manufacturing of metal frameworks for full-arch dental restoration on implants: a comparison between milling and a novel hybrid technology. J Prosthodont. 2019;28(5):556–63.CrossRefPubMed Ciocca L, Meneghello R, Savio G, Scheda L, Monaco C, Gatto MR, Micarelli C, Baldissara P. Manufacturing of metal frameworks for full-arch dental restoration on implants: a comparison between milling and a novel hybrid technology. J Prosthodont. 2019;28(5):556–63.CrossRefPubMed
30.
go back to reference Kurbada A, Kurbadb S. CAD/CAM-based Implant Abutments Made of Lithium Disilicate CAD/CAM-basierte Implantatabutments aus Lithiumdisilikat. Int J Comput Dent. 2013;16:125–41. Kurbada A, Kurbadb S. CAD/CAM-based Implant Abutments Made of Lithium Disilicate CAD/CAM-basierte Implantatabutments aus Lithiumdisilikat. Int J Comput Dent. 2013;16:125–41.
31.
go back to reference Kapos T, Evans C. CAD/CAM technology for implant abutments, crowns, and superstructures. Int J Oral Maxillofacial Implants. 2014;29. Kapos T, Evans C. CAD/CAM technology for implant abutments, crowns, and superstructures. Int J Oral Maxillofacial Implants. 2014;29.
32.
go back to reference Mishra SK, Chowdhary R, Kumari S. Microleakage at the different implant abutment interface: a systematic review. J Clin Diagn Research: JCDR. 2017;11(6):ZE10. Mishra SK, Chowdhary R, Kumari S. Microleakage at the different implant abutment interface: a systematic review. J Clin Diagn Research: JCDR. 2017;11(6):ZE10.
33.
go back to reference Alves DC, de Carvalho PS, Elias CN, Vedovatto E, Martinez EF. In vitro analysis of the microbiological sealing of tapered implants after mechanical cycling. Clin Oral Invest. 2016;20(9):2437–45.CrossRef Alves DC, de Carvalho PS, Elias CN, Vedovatto E, Martinez EF. In vitro analysis of the microbiological sealing of tapered implants after mechanical cycling. Clin Oral Invest. 2016;20(9):2437–45.CrossRef
34.
go back to reference Mobilio N, Fasiol A, Franceschetti G, Catapano S. Marginal vertical fit along the implant-abutment interface: a microscope qualitative analysis. Dentistry J. 2016;4(3):31.CrossRef Mobilio N, Fasiol A, Franceschetti G, Catapano S. Marginal vertical fit along the implant-abutment interface: a microscope qualitative analysis. Dentistry J. 2016;4(3):31.CrossRef
35.
go back to reference Jörn D, Kohorst P, Besdo S, Borchers L, Stiesch M. Three-dimensional nonlinear finite element analysis and Microcomputed Tomography evaluation of Microgap formation in a Dental Implant under Oblique Loading. Int J Oral Maxillofacial Implants. 2016;31(3). Jörn D, Kohorst P, Besdo S, Borchers L, Stiesch M. Three-dimensional nonlinear finite element analysis and Microcomputed Tomography evaluation of Microgap formation in a Dental Implant under Oblique Loading. Int J Oral Maxillofacial Implants. 2016;31(3).
36.
go back to reference Rismanchian M, Hatami M, Badrian H, Khalighinejad N, Goroohi H. Evaluation of microgap size and microbial leakage in the connection area of 4 abutments with Straumann (ITI) implant. J Oral Implantology. 2012;38(6):677–85.CrossRefPubMed Rismanchian M, Hatami M, Badrian H, Khalighinejad N, Goroohi H. Evaluation of microgap size and microbial leakage in the connection area of 4 abutments with Straumann (ITI) implant. J Oral Implantology. 2012;38(6):677–85.CrossRefPubMed
37.
go back to reference Lops D, Meneghello R, Sbricoli L, Savio G, Bressan E, Stellini E. Precision of the connection between Implant and Standard or computer-aided Design/Computer-Aided Manufacturing abutments: a novel evaluation method. Int J Oral Maxillofacial Implants. 2018;33(1). Lops D, Meneghello R, Sbricoli L, Savio G, Bressan E, Stellini E. Precision of the connection between Implant and Standard or computer-aided Design/Computer-Aided Manufacturing abutments: a novel evaluation method. Int J Oral Maxillofacial Implants. 2018;33(1).
38.
go back to reference Kim MJ, Choi YJ, Kim SK, Heo SJ, Koak JY. Marginal accuracy and internal fit of 3-D printing laser-sintered co-cr alloy copings. Materials. 2017;10(1):93.PubMedCentralCrossRefPubMed Kim MJ, Choi YJ, Kim SK, Heo SJ, Koak JY. Marginal accuracy and internal fit of 3-D printing laser-sintered co-cr alloy copings. Materials. 2017;10(1):93.PubMedCentralCrossRefPubMed
39.
go back to reference Tsuge T, Hagiwara Y, Matsumura H. Marginal fit and microgaps of implant-abutment interface with internal anti-rotation configuration. Dent Mater J. 2008;27(1):29–34.CrossRefPubMed Tsuge T, Hagiwara Y, Matsumura H. Marginal fit and microgaps of implant-abutment interface with internal anti-rotation configuration. Dent Mater J. 2008;27(1):29–34.CrossRefPubMed
40.
go back to reference Baggi L, Di Girolamo M, Mirisola C, Calcaterra R. Microbiological evaluation of bacterial and mycotic seal in implant systems with different implant-abutment interfaces and closing torque values. Implant Dent. 2013;22(4):344–50.CrossRefPubMed Baggi L, Di Girolamo M, Mirisola C, Calcaterra R. Microbiological evaluation of bacterial and mycotic seal in implant systems with different implant-abutment interfaces and closing torque values. Implant Dent. 2013;22(4):344–50.CrossRefPubMed
41.
go back to reference Jeong KW, Kim SH, Şen N, Şermet IB, Gürler N. Sealing capability and marginal fit of titanium versus zirconia abutments with different connection designs. J Adv Prosthodont. 2019;11(2):105–11.CrossRef Jeong KW, Kim SH, Şen N, Şermet IB, Gürler N. Sealing capability and marginal fit of titanium versus zirconia abutments with different connection designs. J Adv Prosthodont. 2019;11(2):105–11.CrossRef
42.
go back to reference Vélez J, Peláez J, López-Suárez C, Agustín-Panadero R, Tobar C, Suárez MJ. Influence of implant connection, abutment design and screw insertion torque on implant-abutment misfit. J Clin Med. 2020;9(8):2365.PubMedCentralCrossRefPubMed Vélez J, Peláez J, López-Suárez C, Agustín-Panadero R, Tobar C, Suárez MJ. Influence of implant connection, abutment design and screw insertion torque on implant-abutment misfit. J Clin Med. 2020;9(8):2365.PubMedCentralCrossRefPubMed
43.
go back to reference Fernández M, Delgado L, Molmeneu M, García D, Rodríguez D. Analysis of the misfit of dental implant-supported prostheses made with three manufacturing processes. J Prosthet Dent. 2014;111(2):116–23.CrossRefPubMed Fernández M, Delgado L, Molmeneu M, García D, Rodríguez D. Analysis of the misfit of dental implant-supported prostheses made with three manufacturing processes. J Prosthet Dent. 2014;111(2):116–23.CrossRefPubMed
44.
go back to reference Gonzalo E, Vizoso B, Lopez-Suarez C, Diaz P, Pelaez J, Suarez MJ. Evaluation of milled titanium versus laser sintered co-cr abutments on the marginal misfit in internal implant-abutment connection. Materials. 2020;13(21):4873.PubMedCentralCrossRefPubMed Gonzalo E, Vizoso B, Lopez-Suarez C, Diaz P, Pelaez J, Suarez MJ. Evaluation of milled titanium versus laser sintered co-cr abutments on the marginal misfit in internal implant-abutment connection. Materials. 2020;13(21):4873.PubMedCentralCrossRefPubMed
45.
go back to reference Do Nascimento C, Pita MS, de Souza Santos E, Monesi N, Pedrazzi V, de Albuquerque Junior RF, Ribeiro RF. Microbiome of titanium and zirconia dental implants abutments. Dent Mater. 2016;32(1):93–101.CrossRef Do Nascimento C, Pita MS, de Souza Santos E, Monesi N, Pedrazzi V, de Albuquerque Junior RF, Ribeiro RF. Microbiome of titanium and zirconia dental implants abutments. Dent Mater. 2016;32(1):93–101.CrossRef
46.
go back to reference El Haddad E, Giannì AB, Mancini GE, Cura F, Carinci F. Implant-Abutment leaking of replace conical connection nobel biocare® implant system. An in vitro study of the microbiological penetration from external environment to implant-abutment space. Oral & Implantology. 2016;9(2):76. El Haddad E, Giannì AB, Mancini GE, Cura F, Carinci F. Implant-Abutment leaking of replace conical connection nobel biocare® implant system. An in vitro study of the microbiological penetration from external environment to implant-abutment space. Oral & Implantology. 2016;9(2):76.
47.
go back to reference Papadiochou S, Pissiotis AL. Marginal adaptation and CAD-CAM technology: a systematic review of restorative material and fabrication techniques. J Prosthet Dent. 2018;119(4):545–51.CrossRefPubMed Papadiochou S, Pissiotis AL. Marginal adaptation and CAD-CAM technology: a systematic review of restorative material and fabrication techniques. J Prosthet Dent. 2018;119(4):545–51.CrossRefPubMed
48.
go back to reference Song B, Zhao X, Li S, Han C, Wei Q, Wen S, Liu J, Shi Y. Differences in microstructure and properties between selective laser melting and traditional manufacturing for fabrication of metal parts: a review. Front Mech Eng. 2015;10(2):111–25.CrossRef Song B, Zhao X, Li S, Han C, Wei Q, Wen S, Liu J, Shi Y. Differences in microstructure and properties between selective laser melting and traditional manufacturing for fabrication of metal parts: a review. Front Mech Eng. 2015;10(2):111–25.CrossRef
49.
go back to reference Yap CY, Chua CK, Dong ZL, Liu ZH, Zhang DQ, Loh LE, Sing SL. Review of selective laser melting: materials and applications. Appl Phys Reviews. 2015;2(4):041101.CrossRef Yap CY, Chua CK, Dong ZL, Liu ZH, Zhang DQ, Loh LE, Sing SL. Review of selective laser melting: materials and applications. Appl Phys Reviews. 2015;2(4):041101.CrossRef
50.
go back to reference Al Jabbari YS, Koutsoukis T, Barmpagadaki X, Zinelis S. Metallurgical and interfacial characterization of PFM Co–Cr dental alloys fabricated via casting, milling or selective laser melting. Dent Mater. 2014;30(4):e79–88.CrossRefPubMed Al Jabbari YS, Koutsoukis T, Barmpagadaki X, Zinelis S. Metallurgical and interfacial characterization of PFM Co–Cr dental alloys fabricated via casting, milling or selective laser melting. Dent Mater. 2014;30(4):e79–88.CrossRefPubMed
Metadata
Title
Accuracy of fit for cobaltchromium bar over two implants fabricated with different manufacturing techniques: an in-vitro study
Authors
Hossam I. Nassar
Ayman Fateen
Publication date
01-12-2023
Publisher
BioMed Central
Published in
BMC Oral Health / Issue 1/2023
Electronic ISSN: 1472-6831
DOI
https://doi.org/10.1186/s12903-023-03700-w

Other articles of this Issue 1/2023

BMC Oral Health 1/2023 Go to the issue