Skip to main content
Top
Published in: Osteoporosis International 4/2007

01-04-2007 | Original Article

A study of the long-term precision of dual-energy X-ray absorptiometry bone densitometers and implications for the validity of the least-significant-change calculation

Author: T. N. Hangartner

Published in: Osteoporosis International | Issue 4/2007

Login to get access

Abstract

Introduction

Short-term precision is often quoted and used as the most important performance parameter of a dual-energy X-ray absorptiometry (DXA) scanner; however, long-term precision has a more profound impact on patient monitoring. Long-term precision refers to the combination of in-vivo precision errors and long-term equipment stability.

Methods

To monitor long-term equipment stability, a phantom was designed with four inserts ranging in bone-mineral density from 0.5 to 3.3 g/cm2. This phantom was used to monitor the equipment stability of four modern fan-beam densitometers, two each from Hologic and GE/Lunar, over a 4-year period. Manufacturer-recommended quality assurance (QA) procedures were performed, and the scanners stayed within manufacturer-specified tolerances throughout the study.

Results and conclusion

During the 4-year period, the Hologic scanners were observed to cause clinically insignificant BMD shifts (maximum of 0.34%), whereas the GE/Lunar scanners revealed BMD shifts that were clinically significant (1.5% and 2.1%). As a result, using least-significant-change (LSC) calculations based only on short-term in-vivo precision studies for monitoring patients is not valid for the two GE/Lunar densitometers due to the poorer long-term stability they exhibited.
Literature
1.
go back to reference Lenchik L, Kiebzak GM, Blunt BA (2002) What is the role of serial bone mineral density measurements in patient management? J Clin Densitom 5(Suppl):S29–S38PubMedCrossRef Lenchik L, Kiebzak GM, Blunt BA (2002) What is the role of serial bone mineral density measurements in patient management? J Clin Densitom 5(Suppl):S29–S38PubMedCrossRef
2.
go back to reference Bonnick SL, Lewis L (2005) Bone densitometry for technologists. Humana Press, Totowa Bonnick SL, Lewis L (2005) Bone densitometry for technologists. Humana Press, Totowa
3.
go back to reference Khan KM, Henzell SL, Broderick C, Prince RL, Saul A, Lomman J, Wark JD (1997) Instrument performance in bone density testing at five Australian centres. Aust N Z J Med 27:526–530PubMed Khan KM, Henzell SL, Broderick C, Prince RL, Saul A, Lomman J, Wark JD (1997) Instrument performance in bone density testing at five Australian centres. Aust N Z J Med 27:526–530PubMed
4.
go back to reference Orwoll ES, Oviatt SK, The Nafarelin/Bone Study Group (1991) Longitudinal precision of dual-energy X-ray absorptiometry in a multicenter study. J Bone Miner Res 6:191–197PubMedCrossRef Orwoll ES, Oviatt SK, The Nafarelin/Bone Study Group (1991) Longitudinal precision of dual-energy X-ray absorptiometry in a multicenter study. J Bone Miner Res 6:191–197PubMedCrossRef
5.
go back to reference Reginster JY, Deroisy R, Zegels B, Jupsin I, Albert A, Franchimont P (1995) Long-term performance in vitro and in vivo of dual-energy X-ray absorptiometry. Clin Rheumatol 14:180–186PubMedCrossRef Reginster JY, Deroisy R, Zegels B, Jupsin I, Albert A, Franchimont P (1995) Long-term performance in vitro and in vivo of dual-energy X-ray absorptiometry. Clin Rheumatol 14:180–186PubMedCrossRef
6.
go back to reference Patel R, Blake GM, Rymer J, Fogelman I (2000) Long-term precision of DXA scanning assessed over 7 years in 40 postmenopausal women. Osteoporos Int 11:68–75PubMedCrossRef Patel R, Blake GM, Rymer J, Fogelman I (2000) Long-term precision of DXA scanning assessed over 7 years in 40 postmenopausal women. Osteoporos Int 11:68–75PubMedCrossRef
7.
go back to reference Pearson D, Cawte SA (1997) Long-term quality control of DXA: a comparison of Shewhart rules and Cusum charts. Osteoporos Int 7:338–343PubMedCrossRef Pearson D, Cawte SA (1997) Long-term quality control of DXA: a comparison of Shewhart rules and Cusum charts. Osteoporos Int 7:338–343PubMedCrossRef
8.
go back to reference Wapniarz M, Lehmann R, Randerath O, Baedeker S, John W, Klein K, Allolio B (1994) Precision of dual X-ray absorptiometry and peripheral computed tomography using mobile densitometry units. Calcif Tissue Int 54:219–223PubMedCrossRef Wapniarz M, Lehmann R, Randerath O, Baedeker S, John W, Klein K, Allolio B (1994) Precision of dual X-ray absorptiometry and peripheral computed tomography using mobile densitometry units. Calcif Tissue Int 54:219–223PubMedCrossRef
9.
go back to reference Lang T, Takada M, Gee R, Wu C, Li J, Hayashi-Clark C, Schoen S, March V, Genant HK (1997) A preliminary evaluation of the lunar expert-XL for bone densitometry and vertebral morphometry. J Bone Miner Res 12:136–143PubMedCrossRef Lang T, Takada M, Gee R, Wu C, Li J, Hayashi-Clark C, Schoen S, March V, Genant HK (1997) A preliminary evaluation of the lunar expert-XL for bone densitometry and vertebral morphometry. J Bone Miner Res 12:136–143PubMedCrossRef
10.
go back to reference Binkley N, Bilezikian JP, Kendler DL, Leib ES, Lewiecki EM, Petak SM (2006) Official positions of the International Society for Clinical Densitometry and executive summary of the 2005 Position Development Conference. J Clin Densitom 9:4–14PubMedCrossRef Binkley N, Bilezikian JP, Kendler DL, Leib ES, Lewiecki EM, Petak SM (2006) Official positions of the International Society for Clinical Densitometry and executive summary of the 2005 Position Development Conference. J Clin Densitom 9:4–14PubMedCrossRef
11.
go back to reference Bonnick SL, Johnston CC Jr, Kleerekoper M, Lindsay R, Miller P, Sherwood L and Siris E (2001) Importance of precision in bone density measurements. J Clin Densitom 4:105–110PubMedCrossRef Bonnick SL, Johnston CC Jr, Kleerekoper M, Lindsay R, Miller P, Sherwood L and Siris E (2001) Importance of precision in bone density measurements. J Clin Densitom 4:105–110PubMedCrossRef
12.
go back to reference Engelke K, Gluer CC, Genant HK (1995) Factors influencing short-term precision of dual X-ray bone absorptiometry (DXA) of spine and femur. Calcif Tissue Int 56:19–25PubMedCrossRef Engelke K, Gluer CC, Genant HK (1995) Factors influencing short-term precision of dual X-ray bone absorptiometry (DXA) of spine and femur. Calcif Tissue Int 56:19–25PubMedCrossRef
13.
go back to reference Morgan SL, Abercrombie W, Lee JY (2003) Need for precision studies at individual institutions and assessment of size of regions of interest on serial DXA scans. J Clin Densitom 6:97–101PubMedCrossRef Morgan SL, Abercrombie W, Lee JY (2003) Need for precision studies at individual institutions and assessment of size of regions of interest on serial DXA scans. J Clin Densitom 6:97–101PubMedCrossRef
14.
go back to reference El Maghraoui A, Do Santos Zounon AA, Jroundi I, Nouijai A, Ghazi M, Achemlal L, Bezza A, Tazi MA, Abouqual R (2005) Reproducibility of bone mineral density measurements using dual X-ray absorptiometry in daily clinical practice. Osteoporos Int 16:1742–1748PubMedCrossRef El Maghraoui A, Do Santos Zounon AA, Jroundi I, Nouijai A, Ghazi M, Achemlal L, Bezza A, Tazi MA, Abouqual R (2005) Reproducibility of bone mineral density measurements using dual X-ray absorptiometry in daily clinical practice. Osteoporos Int 16:1742–1748PubMedCrossRef
Metadata
Title
A study of the long-term precision of dual-energy X-ray absorptiometry bone densitometers and implications for the validity of the least-significant-change calculation
Author
T. N. Hangartner
Publication date
01-04-2007
Publisher
Springer-Verlag
Published in
Osteoporosis International / Issue 4/2007
Print ISSN: 0937-941X
Electronic ISSN: 1433-2965
DOI
https://doi.org/10.1007/s00198-006-0280-1

Other articles of this Issue 4/2007

Osteoporosis International 4/2007 Go to the issue