01-11-2011 | Glaucoma

A new rebound tonometer for home monitoring of intraocular pressure

Authors: André Rosentreter, Kerstin S. Jablonski, Anne C. Mellein, Stergiani Gaki, Arno Hueber, Thomas S. Dietlein

Published in: Graefe's Archive for Clinical and Experimental Ophthalmology | Issue 11/2011

Login to get access

Abstract

Background

To compare intraocular pressure (IOP) measurements obtained by the Icare ONE rebound tonometer (RTONE) and the Goldmann applanation tonometer (GAT) in healthy persons and glaucoma patients in a prospective study, and to investigate the influence of central corneal thickness (CCT).

Methods

Measurements on 126 right eyes were obtained by three equally skilled ophthalmologists with each of the above-mentioned tonometers. In addition, patients measured their own IOP with the RTONE (RTONE(p)). The means and standard deviation for all tonometers were compared. Agreement between the tonometers was calculated using the Bland–Altman method.

Results

A total of 95 (75.3%) patients were able to perform correct self-tonometry. Mean IOPs obtained were 17.1 ± 5.9 mmHg (RTONE performed by ophthalmologist: RTONE (o)), 17.3 ± 5.6 mmHg (RTONE(p)) and 16.5 ± 5.1 mmHg (GAT). Correlation analysis indicated a good correlation between IOP readings obtained using RTONE(o) and RTONE(p) (ρ = 0.916; p < 0.001) and RTONE(o) and GAT (ρ = 0.901; p < 0.001). Bland–Altman analysis revealed a mean difference (bias) between RTONE(o) and RTONE(p), between RTONE(o) and GAT, and between RTONE(p) and GAT of −0.2, 0.6, and 0.8 mmHg, respectively, with 95% limits of agreement of −5.0 to 4.5, –4.4 to 5.6, and −4.6 to 6.1 mmHg, respectively. The difference between RTONE(o) and GAT significantly increased with increasing CCT (ρ = 0.004), with a 10% increase in CCT resulting in a 1.8% increase in the difference.

Conclusions

Measurements obtained with the RTONE, either by an ophthalmologist or by the patient, showed an excellent correlation with those provided by applanation tonometry. RTONE generally tends to overestimate IOP compared to GAT readings and displays a dependence on CCT.
This study was registered with the DRKS (German Clinical Trials Register; www.​germanctr.​de; DRKS00000478).
Literature
2.
go back to reference Kass MA (1996) Standardizing the measurement of intraocular pressure for clinical research. Guidelines from the Eye Care Technology Forum. Ophthalmology 103:183–185PubMed Kass MA (1996) Standardizing the measurement of intraocular pressure for clinical research. Guidelines from the Eye Care Technology Forum. Ophthalmology 103:183–185PubMed
3.
go back to reference Kontiola AI (1997) A new electromechanical method for measuring intraocular pressure. Doc Ophthalmol 93:265–276CrossRef Kontiola AI (1997) A new electromechanical method for measuring intraocular pressure. Doc Ophthalmol 93:265–276CrossRef
4.
go back to reference Danias J, Kontiola AI, Filippopoulos T, Mittag T (2003) Method for the noninvasive measurement of intraocular pressure in mice. Invest Ophthalmol Vis Sci 44:1138–1141PubMedCrossRef Danias J, Kontiola AI, Filippopoulos T, Mittag T (2003) Method for the noninvasive measurement of intraocular pressure in mice. Invest Ophthalmol Vis Sci 44:1138–1141PubMedCrossRef
5.
go back to reference Abraham LM, Epasinghe NCR, Selva D, Casson R (2008) Comparison of the Icare® rebound tonometer with the Goldmann applanation tonometer by experienced and inexperience tonometrists. Eye 22:503–506PubMedCrossRef Abraham LM, Epasinghe NCR, Selva D, Casson R (2008) Comparison of the Icare® rebound tonometer with the Goldmann applanation tonometer by experienced and inexperience tonometrists. Eye 22:503–506PubMedCrossRef
6.
go back to reference Sahin A, Basmak H, Niyaz L, Yildirim N (2007) Reproducibility and tolerability of the Icare rebound tonometer in school children. J Glaucoma 16:185–188PubMedCrossRef Sahin A, Basmak H, Niyaz L, Yildirim N (2007) Reproducibility and tolerability of the Icare rebound tonometer in school children. J Glaucoma 16:185–188PubMedCrossRef
7.
go back to reference Martinez-de-la-Casa JM, Garcia-Feijoo J, Castillo A, Garcia-Sanchez J (2005) Reproducibility and clinical evaluation of rebound tonometry. Invest Ophthalmol Vis Sci 46:4578–4580PubMedCrossRef Martinez-de-la-Casa JM, Garcia-Feijoo J, Castillo A, Garcia-Sanchez J (2005) Reproducibility and clinical evaluation of rebound tonometry. Invest Ophthalmol Vis Sci 46:4578–4580PubMedCrossRef
8.
go back to reference Dielemans I, Vingerling JR, Hofman A, Grobbee DE, de Jong PT (1994) Reliability of intraocular pressure measurement with the Goldmann applanation tonometer in epidemiological studies. Graefes Arch Clin Exp Ophthalmol 232:141–144PubMedCrossRef Dielemans I, Vingerling JR, Hofman A, Grobbee DE, de Jong PT (1994) Reliability of intraocular pressure measurement with the Goldmann applanation tonometer in epidemiological studies. Graefes Arch Clin Exp Ophthalmol 232:141–144PubMedCrossRef
9.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMedCrossRef Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMedCrossRef
10.
go back to reference van der Jagt LH, Jansonius LM (2005) Three portable tonometers, the TGDc-01, the Icare and the Tonopen XL, compared with each other and with Goldmann applanation tonometry. Ophthalmic Physiol Opt 25:429–435PubMedCrossRef van der Jagt LH, Jansonius LM (2005) Three portable tonometers, the TGDc-01, the Icare and the Tonopen XL, compared with each other and with Goldmann applanation tonometry. Ophthalmic Physiol Opt 25:429–435PubMedCrossRef
11.
go back to reference Iliev ME, Goldblum D, Katsoulis K, Amstutz C, Frueh B (2006) Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness. Br J Ophthalmol 90:833–835PubMedCrossRef Iliev ME, Goldblum D, Katsoulis K, Amstutz C, Frueh B (2006) Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness. Br J Ophthalmol 90:833–835PubMedCrossRef
12.
go back to reference Fernandes P, Diaz-Rey JA, Queiros A, Gonzalez-Meijome JM, Jorge J (2005) Comparison of the Icare rebound tonometer with the Goldmann tonometer in a normal population. Ophthalmic Physiol Opt 25:436–440PubMedCrossRef Fernandes P, Diaz-Rey JA, Queiros A, Gonzalez-Meijome JM, Jorge J (2005) Comparison of the Icare rebound tonometer with the Goldmann tonometer in a normal population. Ophthalmic Physiol Opt 25:436–440PubMedCrossRef
13.
go back to reference Garcia-Resua C, Gonzalez-Meijome JM, Gilino J, Yebra-Pimentel E (2006) Accuracy of the new Icare rebound tonometer vs. other portable tonometers in healthy eyes. Optom Vis Sci 83:102–107PubMedCrossRef Garcia-Resua C, Gonzalez-Meijome JM, Gilino J, Yebra-Pimentel E (2006) Accuracy of the new Icare rebound tonometer vs. other portable tonometers in healthy eyes. Optom Vis Sci 83:102–107PubMedCrossRef
14.
go back to reference Martinez-de-la-Casa JM, Garcia-Feijoo J, Vico E, Fernandez-Vidal A, Benitez del Castillo JM, Wasfi M, Garcia-Sanchez J (2006) Effect of corneal thickness on dynamic contour, rebound, and Goldmann tonometry. Ophthalmology 113:2156–2162PubMedCrossRef Martinez-de-la-Casa JM, Garcia-Feijoo J, Vico E, Fernandez-Vidal A, Benitez del Castillo JM, Wasfi M, Garcia-Sanchez J (2006) Effect of corneal thickness on dynamic contour, rebound, and Goldmann tonometry. Ophthalmology 113:2156–2162PubMedCrossRef
15.
go back to reference Munkwitz S, Elkarmouty A, Hoffmann EM, Pfeiffer N, Thieme H (2008) Comparison of the Icare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range. Graefes Arch Clin Exp Ophthalmol 246:875–879PubMedCrossRef Munkwitz S, Elkarmouty A, Hoffmann EM, Pfeiffer N, Thieme H (2008) Comparison of the Icare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range. Graefes Arch Clin Exp Ophthalmol 246:875–879PubMedCrossRef
16.
go back to reference Poostchi A, Mitchell R, Nicholas S, Purdie G, Wells A (2009) The Icare rebound tonometer: comparisons with Goldmann tonometry, and influence of central corneal thickness. Clin Exp Ophthalmol 37:687–691CrossRef Poostchi A, Mitchell R, Nicholas S, Purdie G, Wells A (2009) The Icare rebound tonometer: comparisons with Goldmann tonometry, and influence of central corneal thickness. Clin Exp Ophthalmol 37:687–691CrossRef
17.
go back to reference Emara B, Probst LE, Tingey DP, Kennedy DW, Willms LJ, Machat J (1998) Correlation of intraocular pressure and central corneal thickness in normal myopic eyes and after laser in situ keratomileusis. J Cataract Refract Surg 24:1320–1325PubMed Emara B, Probst LE, Tingey DP, Kennedy DW, Willms LJ, Machat J (1998) Correlation of intraocular pressure and central corneal thickness in normal myopic eyes and after laser in situ keratomileusis. J Cataract Refract Surg 24:1320–1325PubMed
18.
go back to reference Lam AK, Wu R, Wang Z, Woo V, Chan E, Tam K, Chau R, Wong KK (2010) Effect of laser in situ keratomileusis on rebound tonometry and Goldmann applanation tonometry. J Cataract Refract Surg 36:631–636PubMedCrossRef Lam AK, Wu R, Wang Z, Woo V, Chan E, Tam K, Chau R, Wong KK (2010) Effect of laser in situ keratomileusis on rebound tonometry and Goldmann applanation tonometry. J Cataract Refract Surg 36:631–636PubMedCrossRef
19.
go back to reference Schipper I, Senn P, Oyo-Szerenyi K, Peter R (2000) Central and peripheral pressure measurements with the Goldmann tonometer and Tono-Pen after photorefractive keratectomy for myopia. J Cataract Refract Surg 26:929–933PubMedCrossRef Schipper I, Senn P, Oyo-Szerenyi K, Peter R (2000) Central and peripheral pressure measurements with the Goldmann tonometer and Tono-Pen after photorefractive keratectomy for myopia. J Cataract Refract Surg 26:929–933PubMedCrossRef
20.
go back to reference Ehlers N, Bramsen T, Sperling S (1975) Applanation tonometry and central corneal thickness. Acta Ophthalmol 53:34–43 Ehlers N, Bramsen T, Sperling S (1975) Applanation tonometry and central corneal thickness. Acta Ophthalmol 53:34–43
21.
go back to reference Kohlhaas M, Boehm AG, Spoerl E, Pürsten A, Grein HJ, Pillunat LE (2006) Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry. Arch Ophthalmol 124:471–476PubMedCrossRef Kohlhaas M, Boehm AG, Spoerl E, Pürsten A, Grein HJ, Pillunat LE (2006) Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry. Arch Ophthalmol 124:471–476PubMedCrossRef
22.
go back to reference Whitacre MM, Stein RA, Hassanein K (1993) The effect of corneal thickness on applanation tonometry. Am J Ophthalmol 115:592–596PubMed Whitacre MM, Stein RA, Hassanein K (1993) The effect of corneal thickness on applanation tonometry. Am J Ophthalmol 115:592–596PubMed
23.
go back to reference Martinez-de-la-Casa JM, Garcia-Feijoo J, Saenz-Frances VG, Fernandez-Vidal A, Mendez-Hernandez C, Garcia-Sanchez J (2009) Comparison of rebound tonometer and Goldmann handheld applanation tonometer in congenital glaucoma. J Glaucoma 18:49–52PubMedCrossRef Martinez-de-la-Casa JM, Garcia-Feijoo J, Saenz-Frances VG, Fernandez-Vidal A, Mendez-Hernandez C, Garcia-Sanchez J (2009) Comparison of rebound tonometer and Goldmann handheld applanation tonometer in congenital glaucoma. J Glaucoma 18:49–52PubMedCrossRef
24.
go back to reference Pakrou N, Gray T, Mills R, Landers J, Craig J (2008) Clinical comparison of the Icare tonometer and Goldmann applanation tonometry. J Glaucoma 17:43–47PubMedCrossRef Pakrou N, Gray T, Mills R, Landers J, Craig J (2008) Clinical comparison of the Icare tonometer and Goldmann applanation tonometry. J Glaucoma 17:43–47PubMedCrossRef
25.
go back to reference Brusini P, Salvetat ML, Zeppieri M, Tosoni C, Parisi L (2006) Comparison of Icare tonometer with Goldmann applanation tonometer in glaucoma patients. J Glaucoma 15:213–217PubMedCrossRef Brusini P, Salvetat ML, Zeppieri M, Tosoni C, Parisi L (2006) Comparison of Icare tonometer with Goldmann applanation tonometer in glaucoma patients. J Glaucoma 15:213–217PubMedCrossRef
26.
go back to reference Sacu S, Vass C, Schemper M, Rainer G (2004) Self-tonometry with the Ocuton S: evaluation of accuracy in glaucoma patients. Acta Ophthalmol Scand 82:405–409PubMedCrossRef Sacu S, Vass C, Schemper M, Rainer G (2004) Self-tonometry with the Ocuton S: evaluation of accuracy in glaucoma patients. Acta Ophthalmol Scand 82:405–409PubMedCrossRef
27.
go back to reference Leonardi M, Leuenberger P, Bertrand D, Bertsch A, Renaud P (2004) First steps toward noninvasive intraocular pressure monitoring with a sensing contact lens. Invest Ophthalmol Vis Sci 45:3113–3117PubMedCrossRef Leonardi M, Leuenberger P, Bertrand D, Bertsch A, Renaud P (2004) First steps toward noninvasive intraocular pressure monitoring with a sensing contact lens. Invest Ophthalmol Vis Sci 45:3113–3117PubMedCrossRef
28.
go back to reference Leonardi M, Pitchon EM, Bertsch A, Renaud P, Mermoud A (2009) Wireless contact lens sensor for intraocular pressure monitoring: assessment on enucleated pig eyes. Acta Ophthalmol 87:433–437PubMedCrossRef Leonardi M, Pitchon EM, Bertsch A, Renaud P, Mermoud A (2009) Wireless contact lens sensor for intraocular pressure monitoring: assessment on enucleated pig eyes. Acta Ophthalmol 87:433–437PubMedCrossRef
29.
go back to reference Hediger A, Kniestedt C, Zweifel S, Knecht P, Funk J, Kanngiesser H (2009) Continuous intraocular pressure measurement: first results with a pressure-sensitive contact lens. Ophthalmologe 106:1111–1115PubMedCrossRef Hediger A, Kniestedt C, Zweifel S, Knecht P, Funk J, Kanngiesser H (2009) Continuous intraocular pressure measurement: first results with a pressure-sensitive contact lens. Ophthalmologe 106:1111–1115PubMedCrossRef
30.
go back to reference Draeger J, Schwartz R, Deutsch C, Groenhoff S (1991) Clinical and experimental results with a new fully automatic self-tonometer. Fortschr Ophthalmol 88:304–307PubMed Draeger J, Schwartz R, Deutsch C, Groenhoff S (1991) Clinical and experimental results with a new fully automatic self-tonometer. Fortschr Ophthalmol 88:304–307PubMed
31.
go back to reference Draeger J, Groenhoff S, Hock B, Klemm M (1993) Optimizing the automatic self-tonometer by an acoustic control signal and changed fixation optics. Ophthalmologe 90:54–57PubMed Draeger J, Groenhoff S, Hock B, Klemm M (1993) Optimizing the automatic self-tonometer by an acoustic control signal and changed fixation optics. Ophthalmologe 90:54–57PubMed
32.
go back to reference Asrani S, Chatterjee A, Wallace DK, Santiago-Turla C, Stinnett S (2011) Evaluation of the Icare rebound tonometer as a home intraocular pressure monitoring device. J Glaucoma 20:74–79PubMedCrossRef Asrani S, Chatterjee A, Wallace DK, Santiago-Turla C, Stinnett S (2011) Evaluation of the Icare rebound tonometer as a home intraocular pressure monitoring device. J Glaucoma 20:74–79PubMedCrossRef
33.
go back to reference Chui WS, Lam A, Chen D, Chiu R (2008) The influence of corneal properties on rebound tonometry. Ophthalmology 115:80–84PubMedCrossRef Chui WS, Lam A, Chen D, Chiu R (2008) The influence of corneal properties on rebound tonometry. Ophthalmology 115:80–84PubMedCrossRef
Metadata
Title
A new rebound tonometer for home monitoring of intraocular pressure
Authors
André Rosentreter
Kerstin S. Jablonski
Anne C. Mellein
Stergiani Gaki
Arno Hueber
Thomas S. Dietlein
Publication date
01-11-2011
Publisher
Springer-Verlag
Published in
Graefe's Archive for Clinical and Experimental Ophthalmology / Issue 11/2011
Print ISSN: 0721-832X
Electronic ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-011-1785-7

Other articles of this Issue 11/2011

Graefe's Archive for Clinical and Experimental Ophthalmology 11/2011 Go to the issue