Published in: Graefe's Archive for Clinical and Experimental Ophthalmology 6/2008

01-06-2008 | Glaucoma

Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range

Authors: S. Munkwitz, A. Elkarmouty, E. M. Hoffmann, N. Pfeiffer, H. Thieme

Published in: Graefe's Archive for Clinical and Experimental Ophthalmology | Issue 6/2008

Login to get access

Abstract

Background

The aim of this study was to compare the intraocular pressure (IOP) results measured by the iCare rebound tonometer with those obtained by the Goldmann applanation tonometer (GAT) over a wide range of IOP values. Furthermore, the comfort level of the iCare measurement was evaluated.

Method

The study included 75 eyes of 75 patients. The patients were divided into three groups (7–15 mmHg n = 25, 16–22 mmHg n = 25, 23–60 mmHg n = 25). The measurements were taken by two independent observers in a masked fashion. All patients were asked about discomfort during the iCare measurement. To establish the agreement between the two devices, a Bland-Altman analysis was performed.

Results

Overall, the 95% confidence interval of the differences between the two devices was −8.67 to 10.25 mmHg and in 62.7%, the iCare measurement was within ±3 mmHg of the GAT measurements. The distribution of the differences in IOP was similar, from 7–22 mmHg. In the higher IOP range (23–60 mmHg), however, the deviation was almost twice as large. The measurement with the iCare tonometer was well tolerated; 100% of the patients denied any discomfort.

Conclusions

The iCare tonometer is a mobile alternative to GAT in a low to moderate IOP range, but our findings show a greater deviation than previously reported. In high IOP values, measurements with the iCare tonometer do not correlate well with GAT.
Literature
1.
go back to reference The Advanced Glaucoma Intervention Study (AGIS) : 7 (2000) The relationship between control of intraocular pressure and visual field deterioration. The AGIS Investigators. Am J Ophthalmol 130(4):429–440CrossRef The Advanced Glaucoma Intervention Study (AGIS) : 7 (2000) The relationship between control of intraocular pressure and visual field deterioration. The AGIS Investigators. Am J Ophthalmol 130(4):429–440CrossRef
2.
go back to reference Internationaler Standard für Augentonometer ISO 8612 (2001) Beuth-Verlag GmbH, Berlin Internationaler Standard für Augentonometer ISO 8612 (2001) Beuth-Verlag GmbH, Berlin
3.
go back to reference Bechrakis E (1966) On spontaneous decrease of pressure in applanation tonometry. Ophthalmologica 151(5):604–614PubMedCrossRef Bechrakis E (1966) On spontaneous decrease of pressure in applanation tonometry. Ophthalmologica 151(5):604–614PubMedCrossRef
4.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310PubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310PubMed
5.
go back to reference Brusini P, Salvetat ML, Zeppieri M, Tosoni C, Parisi L (2006) Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients. J Glaucoma 15(3):213–217PubMedCrossRef Brusini P, Salvetat ML, Zeppieri M, Tosoni C, Parisi L (2006) Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients. J Glaucoma 15(3):213–217PubMedCrossRef
6.
go back to reference Chapman CR, Casey KL, Dubner R, Foley KM, Gracely RH, Reading AE (1985) Pain measurement: an overview. Pain 22(1):1–31PubMedCrossRef Chapman CR, Casey KL, Dubner R, Foley KM, Gracely RH, Reading AE (1985) Pain measurement: an overview. Pain 22(1):1–31PubMedCrossRef
7.
go back to reference Davies LN, Bartlett H, Mallen EA, Wolffsohn JS (2006) Clinical evaluation of rebound tonometer. Acta Ophthalmol Scand 84(2):206–209PubMedCrossRef Davies LN, Bartlett H, Mallen EA, Wolffsohn JS (2006) Clinical evaluation of rebound tonometer. Acta Ophthalmol Scand 84(2):206–209PubMedCrossRef
8.
go back to reference Dekking HM, Coster HD (1967) Dynamic tonometry. Ophthalmologica 154(1):59–74PubMed Dekking HM, Coster HD (1967) Dynamic tonometry. Ophthalmologica 154(1):59–74PubMed
9.
go back to reference Fernandes P, Diaz-Rey JA, Queiros A, Gonzalez-Meijome JM, Jorge J (2005) Comparison of the ICare rebound tonometer with the Goldmann tonometer in a normal population. Ophthalmic Physiol Opt 25(5):436–440PubMedCrossRef Fernandes P, Diaz-Rey JA, Queiros A, Gonzalez-Meijome JM, Jorge J (2005) Comparison of the ICare rebound tonometer with the Goldmann tonometer in a normal population. Ophthalmic Physiol Opt 25(5):436–440PubMedCrossRef
10.
go back to reference Goldmann H, Schmidt T (1957) About applanation tonometry. Ophthalmologica 134(4):221–242PubMed Goldmann H, Schmidt T (1957) About applanation tonometry. Ophthalmologica 134(4):221–242PubMed
11.
go back to reference Heijl A, Leske MC, Bengtsson B, Hyman L, Bengtsson B, Hussein M (2002) Reduction of intraocular pressure and glaucoma progression: results from the Early Manifest Glaucoma Trial. Arch Ophthalmol 120(10):1268–1279PubMed Heijl A, Leske MC, Bengtsson B, Hyman L, Bengtsson B, Hussein M (2002) Reduction of intraocular pressure and glaucoma progression: results from the Early Manifest Glaucoma Trial. Arch Ophthalmol 120(10):1268–1279PubMed
12.
go back to reference Iliev ME, Goldblum D, Katsoulis K, Amstutz C, Frueh B (2006) Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness. Br J Ophthalmol 90(7):833–835PubMedCrossRef Iliev ME, Goldblum D, Katsoulis K, Amstutz C, Frueh B (2006) Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness. Br J Ophthalmol 90(7):833–835PubMedCrossRef
13.
go back to reference Kohlhaas M, Boehm AG, Spoerl E, Pursten A, Grein HJ, Pillunat LE (2006) Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry. Arch Ophthalmol 124(4):471–476PubMedCrossRef Kohlhaas M, Boehm AG, Spoerl E, Pursten A, Grein HJ, Pillunat LE (2006) Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry. Arch Ophthalmol 124(4):471–476PubMedCrossRef
14.
go back to reference Kontiola A, Puska P (2004) Measuring intraocular pressure with the Pulsair 3000 and Rebound tonometers in elderly patients without an anesthetic. Graefes Arch Clin Exp Ophthalmol 242(1):3–7PubMedCrossRef Kontiola A, Puska P (2004) Measuring intraocular pressure with the Pulsair 3000 and Rebound tonometers in elderly patients without an anesthetic. Graefes Arch Clin Exp Ophthalmol 242(1):3–7PubMedCrossRef
15.
go back to reference Kontiola AI (2000) A new induction-based impact method for measuring intraocular pressure. Acta Ophthalmol Scand 78(2):142–145PubMedCrossRef Kontiola AI (2000) A new induction-based impact method for measuring intraocular pressure. Acta Ophthalmol Scand 78(2):142–145PubMedCrossRef
16.
go back to reference Kontiola AI (2003) Developing impact tonometers for clinical use and glaucoma research. Finland: Department of Ophthalmology, University of Helsinki Kontiola AI (2003) Developing impact tonometers for clinical use and glaucoma research. Finland: Department of Ophthalmology, University of Helsinki
17.
go back to reference Kontiola AI, Goldblum D, Mittag T, Danias J (2001) The induction/impact tonometer: a new instrument to measure intraocular pressure in the rat. Exp Eye Res 73(6):781–785PubMedCrossRef Kontiola AI, Goldblum D, Mittag T, Danias J (2001) The induction/impact tonometer: a new instrument to measure intraocular pressure in the rat. Exp Eye Res 73(6):781–785PubMedCrossRef
18.
go back to reference Krakau CE, Wilke K (1971) On repeated tonometry. Acta Ophthalmol (Copenh) 49(4):611–614CrossRef Krakau CE, Wilke K (1971) On repeated tonometry. Acta Ophthalmol (Copenh) 49(4):611–614CrossRef
19.
go back to reference Obbink J (1931) Onderzoek naar het verband tusschen inwendigen oogdruk en ballistische reacties. Thesis, Utrecht, The Netherlands Obbink J (1931) Onderzoek naar het verband tusschen inwendigen oogdruk en ballistische reacties. Thesis, Utrecht, The Netherlands
20.
go back to reference Parker VA, Herrtage J, Sarkies NJ (2001) Clinical comparison of the Keeler Pulsair 3000 with Goldmann applanation tonometry. Br J Ophthalmol 85(11):1303–1304PubMedCrossRef Parker VA, Herrtage J, Sarkies NJ (2001) Clinical comparison of the Keeler Pulsair 3000 with Goldmann applanation tonometry. Br J Ophthalmol 85(11):1303–1304PubMedCrossRef
21.
go back to reference Phelps CD, Phelps GK (1976) Measurement of intraocular pressure: a study of its reproducibility. Albrecht Von Graefes Arch Klin Exp Ophthalmol 198(1):39–43PubMedCrossRef Phelps CD, Phelps GK (1976) Measurement of intraocular pressure: a study of its reproducibility. Albrecht Von Graefes Arch Klin Exp Ophthalmol 198(1):39–43PubMedCrossRef
22.
go back to reference Sandhu SS, Chattopadhyay S, Birch MK, Ray-Chaudhuri N (2005) Frequency of goldmann applanation tonometer calibration error checks. J Glaucoma 14(3):215–218PubMedCrossRef Sandhu SS, Chattopadhyay S, Birch MK, Ray-Chaudhuri N (2005) Frequency of goldmann applanation tonometer calibration error checks. J Glaucoma 14(3):215–218PubMedCrossRef
23.
go back to reference Schreiber W, Vorwerk CK, Langenbucher A, Behrens-Baumann W, Viestenz A (2007) A comparison of rebound tonometry (ICare) with TonoPenXL and Goldmann applanation tonometry. Ophthalmologe 104:299–304PubMedCrossRef Schreiber W, Vorwerk CK, Langenbucher A, Behrens-Baumann W, Viestenz A (2007) A comparison of rebound tonometry (ICare) with TonoPenXL and Goldmann applanation tonometry. Ophthalmologe 104:299–304PubMedCrossRef
24.
go back to reference van der Jagt LH, Jansonius NM (2005) Three portable tonometers, the TGDc-01, the ICARE and the Tonopen XL, compared with each other and with Goldmann applanation tonometry*. Ophthalmic Physiol Opt 25(5):429–435PubMedCrossRef van der Jagt LH, Jansonius NM (2005) Three portable tonometers, the TGDc-01, the ICARE and the Tonopen XL, compared with each other and with Goldmann applanation tonometry*. Ophthalmic Physiol Opt 25(5):429–435PubMedCrossRef
25.
go back to reference Wessels IF, Oh Y (1990) Tonometer utilization, accuracy, and calibration under field conditions. Arch Ophthalmol 108(12):1709–1712PubMed Wessels IF, Oh Y (1990) Tonometer utilization, accuracy, and calibration under field conditions. Arch Ophthalmol 108(12):1709–1712PubMed
Metadata
Title
Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range
Authors
S. Munkwitz
A. Elkarmouty
E. M. Hoffmann
N. Pfeiffer
H. Thieme
Publication date
01-06-2008
Publisher
Springer-Verlag
Published in
Graefe's Archive for Clinical and Experimental Ophthalmology / Issue 6/2008
Print ISSN: 0721-832X
Electronic ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-007-0758-3

Other articles of this Issue 6/2008

Graefe's Archive for Clinical and Experimental Ophthalmology 6/2008 Go to the issue