A new concept for verifying the isocentric alignment of the proton-rotational gantry for radiation control

Authors: Takahiro Kato, Yuhei Yamazaki, Tatsuhiko Sagara

Published in: Radiological Physics and Technology | Issue 1/2020

Login to get access

Abstract

The purpose of this study was to introduce the modified Winston–Lutz (mWL) test and to evaluate its feasibility. This is a new method to completely absorb the proton beam around the isocenter inside a phantom for radiation control. The mWL test was performed using a 14-cm-diameter acrylic Lucy 3D QA Phantom for a passive-scattering proton beam gantry. The energy of the unmodulated proton beam was adjusted such that the residual range was < 130 mm, and the energy of the proton beam was completely lost around the isocenter. The radiation field was formed with a multi-leaf collimator at 8.6 × 8.6 mm2 in the isocenter plane. The phantom was loaded with a 4-mm-diameter tungsten ball, and the EBT3 was set up at the isocenter. The proton beam was irradiated at gantry angles with 45° steps, and the isocenter deviation of the proton beam was measured and subsequently analyzed. Although the radiation field penumbra was blurred under the influence of scattered radiation due to placement in the phantom compared to the traditional WL test placed in the air, evaluation of the beam axis accuracy was possible. The results confirmed that the maximum total displacement was less than 0.9 mm, and the specifications of the device were met. The mWL test is feasible and effective to reduce the building activation in proton beam treatment facilities. Thus, it can be considered a useful method that sufficiently satisfies the shielding calculation conditions.
Literature
1.
go back to reference Klein EE, Hanley J, Bayouth J, Yin FF, Simon W, Dresser S, Serago C, Aguirre F, Ma L, Arjomandy B, Liu C, Sandin C, Holmes T; Task Group 142, American Association of Physicists in Medicine. Task Group 142 report: quality assurance of medical accelerators. Med Phys. 2009;36(9):4197–212. Klein EE, Hanley J, Bayouth J, Yin FF, Simon W, Dresser S, Serago C, Aguirre F, Ma L, Arjomandy B, Liu C, Sandin C, Holmes T; Task Group 142, American Association of Physicists in Medicine. Task Group 142 report: quality assurance of medical accelerators. Med Phys. 2009;36(9):4197–212.
2.
go back to reference Moyers MF, Lesyna W. Isocenter characteristics of an external ring proton gantry. Int J Radiat Oncol Biol Phys. 2004;60(5):1622–30.CrossRef Moyers MF, Lesyna W. Isocenter characteristics of an external ring proton gantry. Int J Radiat Oncol Biol Phys. 2004;60(5):1622–30.CrossRef
3.
go back to reference Barkhof J, Schut G, Flanz JB, Goitein M, Schippers JM. Verification of the alignment of a therapeutic radiation beam relative to its patient positioner. Med Phys. 1999;26(11):2429–37.CrossRef Barkhof J, Schut G, Flanz JB, Goitein M, Schippers JM. Verification of the alignment of a therapeutic radiation beam relative to its patient positioner. Med Phys. 1999;26(11):2429–37.CrossRef
4.
go back to reference Ciangaru G, Yang JN, Oliver PJ, Bues M, Zhu M, Nakagawa F, Chiba H, Nakamura S, Yoshino H, Umezawa M, Smith AR. Verification procedure for isocentric alignment of proton beams. J Appl Clin Med Phys. 2007;8(4):2671.CrossRef Ciangaru G, Yang JN, Oliver PJ, Bues M, Zhu M, Nakagawa F, Chiba H, Nakamura S, Yoshino H, Umezawa M, Smith AR. Verification procedure for isocentric alignment of proton beams. J Appl Clin Med Phys. 2007;8(4):2671.CrossRef
5.
go back to reference Fuse H, Sakae T, Terunuma T, Sato M, Aoki S. An isocenter detection and verification device for use in proton therapy. Rev. Sci. Instrum. 2012;83(12):125109.CrossRef Fuse H, Sakae T, Terunuma T, Sato M, Aoki S. An isocenter detection and verification device for use in proton therapy. Rev. Sci. Instrum. 2012;83(12):125109.CrossRef
6.
go back to reference Hansen P, Hu D. An isocenter estimation tool for proton gantry alignment. Rev. Sc.i Instrum. 2017;88(12):125102.CrossRef Hansen P, Hu D. An isocenter estimation tool for proton gantry alignment. Rev. Sc.i Instrum. 2017;88(12):125102.CrossRef
7.
go back to reference Lutz W, Winston KR, Maleki N. A system for stereotactic radiosurgery with a linear accelerator. Int. J. Radiat. Oncol. Biol. Phys. 1988;14(2):373–81.CrossRef Lutz W, Winston KR, Maleki N. A system for stereotactic radiosurgery with a linear accelerator. Int. J. Radiat. Oncol. Biol. Phys. 1988;14(2):373–81.CrossRef
8.
go back to reference Du W, Johnson JL, Jiang W, Kudchadker RJ. On the selection of gantry and collimator angles for isocenter localization using Winston–Lutz tests. J. Appl. Clin. Med. Phys. 2016;17(1):167–78.CrossRef Du W, Johnson JL, Jiang W, Kudchadker RJ. On the selection of gantry and collimator angles for isocenter localization using Winston–Lutz tests. J. Appl. Clin. Med. Phys. 2016;17(1):167–78.CrossRef
9.
go back to reference Cadini F, Bolst D, Guatelli S, Beltran C, Jackson M, Rosenfeld AB. Neutron shielding for a new projected proton therapy facility: a Geant4 simulation study. Phys Med. 2016;32(12):1862–71.CrossRef Cadini F, Bolst D, Guatelli S, Beltran C, Jackson M, Rosenfeld AB. Neutron shielding for a new projected proton therapy facility: a Geant4 simulation study. Phys Med. 2016;32(12):1862–71.CrossRef
10.
go back to reference Sakurai H, Ishikawa H, Okumura T. Proton beam therapy in Japan: current and future status. Jpn. J. Clin. Oncol. 2016;46(10):885–92.CrossRef Sakurai H, Ishikawa H, Okumura T. Proton beam therapy in Japan: current and future status. Jpn. J. Clin. Oncol. 2016;46(10):885–92.CrossRef
11.
go back to reference Smith A, Gillin M, Bues M, Zhu XR, Suzuki K, Mohan R, Woo S, Lee A, Komaki R, Cox J, Hiramoto K, Akiyama H, Ishida T, Sasaki T, Matsuda K. The M. D. Anderson proton therapy system. Med Phys. 2009;36(9): 4068–83.CrossRef Smith A, Gillin M, Bues M, Zhu XR, Suzuki K, Mohan R, Woo S, Lee A, Komaki R, Cox J, Hiramoto K, Akiyama H, Ishida T, Sasaki T, Matsuda K. The M. D. Anderson proton therapy system. Med Phys. 2009;36(9): 4068–83.CrossRef
12.
go back to reference Wang N, Ghebremedhin A, Patyal B. Commissioning of a proton gantry equipped with dual X-ray imagers and a robotic patient positioner, and evaluation of the accuracy of single-beam image registration for this system. Med. Phys. 2015;42(6):2979–91.CrossRef Wang N, Ghebremedhin A, Patyal B. Commissioning of a proton gantry equipped with dual X-ray imagers and a robotic patient positioner, and evaluation of the accuracy of single-beam image registration for this system. Med. Phys. 2015;42(6):2979–91.CrossRef
13.
go back to reference Pidikiti R, Patel BC, Maynard MR, Dugas JP, Syh J, Sahoo N, Wu HT, Rosen LR. Commissioning of the world's first compact pencil-beam scanning proton therapy system. J. Appl. Clin. Med. Phys. 2018;19(1):94–105.CrossRef Pidikiti R, Patel BC, Maynard MR, Dugas JP, Syh J, Sahoo N, Wu HT, Rosen LR. Commissioning of the world's first compact pencil-beam scanning proton therapy system. J. Appl. Clin. Med. Phys. 2018;19(1):94–105.CrossRef
Metadata
Title
A new concept for verifying the isocentric alignment of the proton-rotational gantry for radiation control
Authors
Takahiro Kato
Yuhei Yamazaki
Tatsuhiko Sagara
Publication date
01-03-2020
Publisher
Springer Singapore
Published in
Radiological Physics and Technology / Issue 1/2020
Print ISSN: 1865-0333
Electronic ISSN: 1865-0341
DOI
https://doi.org/10.1007/s12194-019-00544-4

Other articles of this Issue 1/2020

Radiological Physics and Technology 1/2020 Go to the issue