Skip to main content
Top
Published in: Journal of Orthopaedic Surgery and Research 1/2015

Open Access 01-12-2015 | Research article

A comparison between a hinged plate and screw system and a conventional tension-band plate and screw system used for correction of an angular deformity of the lower limb: an animal study

Authors: Zhenkai Wu, Dahang Zhao, Li Zhao, Jianlin Liu, Hai Li, Jie Zhu, Fengcang Ma, Daniel Edward Porter

Published in: Journal of Orthopaedic Surgery and Research | Issue 1/2015

Login to get access

Abstract

Background

The purpose of the animal study is to introduce a newly designed hinged plate and screw system for correction of angular deformities of the lower limbs. The technique was compared with the use of a conventional tension-band plate and screw system.

Methods

This is a randomized controlled animal trial. Eight 3-month-old Bama miniature pigs were used to establish animal models. In each animal, one leg was randomly allocated into study group and another leg into control group. Legs of the study group were corrected with a hinged plate and screw system, and legs of the control group were corrected with a conventional tension-band plate and screw system. The corrective rates of medial slope angle, medial proximal tibial angle, and angle of the two arms of the hinged plate were measured. Residual stress on the implants was also evaluated. A P < 0.05 was statistical significant.

Results

At the final measurements of 18 weeks, the mean corrective rates of medial slope angle, medial proximal tibial angle, and angle of the two arms of the study group were 0.71°/week, 0.85°/week, and 2.18°/week, respectively; the data in the control group were 0.84°/week, 0.89°/week, and 2.13°/week, respectively. No significant difference was found between the groups regarding the mean corrective rates of the angles (P < 0.05). The mean residual stress in the study group was 643.35 MPa, and measurement in the control group was 1,273.63 MPa, with a significant difference (P < 0.05).

Conclusions

Compared to the conventional tension-band plate and screw system, the hinged plate and screw system may be more reliable for correction of angular deformities of the lower limb.
Literature
1.
go back to reference Paley D. Growth plate consideration. In: Paley D, editor. Principles of deformity correction. New York: Springer- Verlag Berlin Heidelberg; 2005. p. 695–716. Paley D. Growth plate consideration. In: Paley D, editor. Principles of deformity correction. New York: Springer- Verlag Berlin Heidelberg; 2005. p. 695–716.
2.
go back to reference Blount WP, Clarke GR. Control of bone growth by epiphyseal stapling: a preliminary report. J Bone Joint Surg Am. 1949;31:464–78. Blount WP, Clarke GR. Control of bone growth by epiphyseal stapling: a preliminary report. J Bone Joint Surg Am. 1949;31:464–78.
3.
go back to reference Stevens PM. Guided growth: 1933 to the present. Strat Traum Limb Recon. 2006;1:29–35.CrossRef Stevens PM. Guided growth: 1933 to the present. Strat Traum Limb Recon. 2006;1:29–35.CrossRef
4.
go back to reference Stevens PM. Guided growth for angular correction: a preliminary series using a tension band plate. J Pediatr Orthop. 2007;27:253–9.CrossRefPubMed Stevens PM. Guided growth for angular correction: a preliminary series using a tension band plate. J Pediatr Orthop. 2007;27:253–9.CrossRefPubMed
5.
go back to reference Boero S, Michelis MB, Riganti S. Use of the eight-Plate for angular correction of knee deformities due to idiopathic and pathologic physis: initiating treatment according to etiology. J Child Orthop. 2011;5:209–16.CrossRefPubMedCentralPubMed Boero S, Michelis MB, Riganti S. Use of the eight-Plate for angular correction of knee deformities due to idiopathic and pathologic physis: initiating treatment according to etiology. J Child Orthop. 2011;5:209–16.CrossRefPubMedCentralPubMed
6.
go back to reference Jelinek EM, Bittersohl B, Martiny F, Scharfstädt A, Krauspe R, Westhoff B. The 8-plate versus physeal stapling for temporary hemiepiphyseodesis correcting genu valgum and genu varum: a retrospective analysis of thirty five patients. Int Orthop. 2012;36:599–605.CrossRefPubMedCentralPubMed Jelinek EM, Bittersohl B, Martiny F, Scharfstädt A, Krauspe R, Westhoff B. The 8-plate versus physeal stapling for temporary hemiepiphyseodesis correcting genu valgum and genu varum: a retrospective analysis of thirty five patients. Int Orthop. 2012;36:599–605.CrossRefPubMedCentralPubMed
7.
go back to reference Burghardt RD, Herzenberg JE. Temporary hemiepiphysiodesis with the eight-Plate for angular deformities: mid-term results. J Orthop Sci. 2010;15:699–704.CrossRefPubMed Burghardt RD, Herzenberg JE. Temporary hemiepiphysiodesis with the eight-Plate for angular deformities: mid-term results. J Orthop Sci. 2010;15:699–704.CrossRefPubMed
8.
go back to reference Wiemann 4th JM, Tryon C, Szalay EA. Physeal stapling versus 8-plate hemiepiphysiodesis for guided correction of angular deformity about the knee. J Pediatr Orthop. 2009;29:481–5.CrossRefPubMed Wiemann 4th JM, Tryon C, Szalay EA. Physeal stapling versus 8-plate hemiepiphysiodesis for guided correction of angular deformity about the knee. J Pediatr Orthop. 2009;29:481–5.CrossRefPubMed
9.
go back to reference Gottliebsen M, Rahbek O, Hvid I, Davidsen M, Hellfritzsch MB, Møller-Madsen B. Hemiepiphysiodesis: similar treatment time for tension-band plating and for stapling: a randomized clinical trial on guided growth for idiopathic genu valgum. Acta Orthop. 2013;84:202–6.CrossRefPubMedCentralPubMed Gottliebsen M, Rahbek O, Hvid I, Davidsen M, Hellfritzsch MB, Møller-Madsen B. Hemiepiphysiodesis: similar treatment time for tension-band plating and for stapling: a randomized clinical trial on guided growth for idiopathic genu valgum. Acta Orthop. 2013;84:202–6.CrossRefPubMedCentralPubMed
10.
go back to reference Burghardt RD, Herzenberg JE, Standard SC, Paley D. Temporary hemiepiphyseal arrest using a screw and plate device to treat knee and ankle deformities in children: a preliminary report. J Child Orthop. 2008;2:187–97.CrossRefPubMedCentralPubMed Burghardt RD, Herzenberg JE, Standard SC, Paley D. Temporary hemiepiphyseal arrest using a screw and plate device to treat knee and ankle deformities in children: a preliminary report. J Child Orthop. 2008;2:187–97.CrossRefPubMedCentralPubMed
11.
go back to reference Burghardt R, Specht S, Herzenberg J. Mechanical failures of eight plate guided growth system for temporary hemiepiphysiodesis. J Pediatr Orthop. 2010;30:594–7.CrossRefPubMed Burghardt R, Specht S, Herzenberg J. Mechanical failures of eight plate guided growth system for temporary hemiepiphysiodesis. J Pediatr Orthop. 2010;30:594–7.CrossRefPubMed
12.
go back to reference Oto M, Yılmaz G, Bowen JR, Thacker M, Kruse R. Adolescent Blount disease in obese children treated by eight-plate hemiepiphysiodesis. Eklem Hastalik Cerrahisi. 2012;23:20–4.PubMed Oto M, Yılmaz G, Bowen JR, Thacker M, Kruse R. Adolescent Blount disease in obese children treated by eight-plate hemiepiphysiodesis. Eklem Hastalik Cerrahisi. 2012;23:20–4.PubMed
13.
go back to reference Schroerlucke S, Bertrand S, Clapp J, Bundy J, Gregg FO. Failure of Orthofix eight-Plate for the treatment of Blount disease. J Pediatr Orthop. 2009;29:57–60.CrossRefPubMed Schroerlucke S, Bertrand S, Clapp J, Bundy J, Gregg FO. Failure of Orthofix eight-Plate for the treatment of Blount disease. J Pediatr Orthop. 2009;29:57–60.CrossRefPubMed
14.
go back to reference Aykut US, Yazici M, Kandemir U, Gedikoglu G, Aksoy MC, Cil A, et al. The effect of temporary hemiepiphyseal stapling on the growth plate: a radiologic and immunohistochemical study in rabbits. J Pediatr Orthop. 2005;25:336–41.CrossRefPubMed Aykut US, Yazici M, Kandemir U, Gedikoglu G, Aksoy MC, Cil A, et al. The effect of temporary hemiepiphyseal stapling on the growth plate: a radiologic and immunohistochemical study in rabbits. J Pediatr Orthop. 2005;25:336–41.CrossRefPubMed
15.
go back to reference Kanellopoulos AD, Mavrogenis AF, Dovris D, Vlasis K, Burghart R, Soucacos PN, et al. Temporary hemiepiphysiodesis with blount staples and eight-plates in pigs. Orthopedics. 2011;34:263. Kanellopoulos AD, Mavrogenis AF, Dovris D, Vlasis K, Burghart R, Soucacos PN, et al. Temporary hemiepiphysiodesis with blount staples and eight-plates in pigs. Orthopedics. 2011;34:263.
16.
go back to reference Ahmed R, Faisal NH, Paradowska AM, Fitzpatrick ME, Khor KA. Neutron diffraction residual strain measurements in nanostructured hydroxyapatite coatings for orthopaedic implants. J Mech Behav Biomed Mater. 2011;4:2043–54.CrossRefPubMed Ahmed R, Faisal NH, Paradowska AM, Fitzpatrick ME, Khor KA. Neutron diffraction residual strain measurements in nanostructured hydroxyapatite coatings for orthopaedic implants. J Mech Behav Biomed Mater. 2011;4:2043–54.CrossRefPubMed
17.
go back to reference Schuh A, Zeller C, Holzwarth U, Kachler W, Wilcke G, Zeiler G, et al. Deep rolling of titanium rods for application in modular total hip arthroplasty. J Biomed Mater Res B Appl Biomater. 2007;81:330–5.CrossRefPubMed Schuh A, Zeller C, Holzwarth U, Kachler W, Wilcke G, Zeiler G, et al. Deep rolling of titanium rods for application in modular total hip arthroplasty. J Biomed Mater Res B Appl Biomater. 2007;81:330–5.CrossRefPubMed
Metadata
Title
A comparison between a hinged plate and screw system and a conventional tension-band plate and screw system used for correction of an angular deformity of the lower limb: an animal study
Authors
Zhenkai Wu
Dahang Zhao
Li Zhao
Jianlin Liu
Hai Li
Jie Zhu
Fengcang Ma
Daniel Edward Porter
Publication date
01-12-2015
Publisher
BioMed Central
Published in
Journal of Orthopaedic Surgery and Research / Issue 1/2015
Electronic ISSN: 1749-799X
DOI
https://doi.org/10.1186/s13018-015-0198-4

Other articles of this Issue 1/2015

Journal of Orthopaedic Surgery and Research 1/2015 Go to the issue