Open Access 01-12-2014 | Research article

Which activity monitor to use? Validity, reproducibility and user friendliness of three activity monitors

Authors: Brenda AJ Berendsen, Marike RC Hendriks, Kenneth Meijer, Guy Plasqui, Nicolaas C Schaper, Hans HCM Savelberg

Published in: BMC Public Health | Issue 1/2014

Login to get access

Abstract

Background

Health is associated with amount of daily physical activity. Recently, the identification of sedentary time as an independent factor, has gained interest. A valid and easy to use activity monitor is needed to objectively investigate the relationship between physical activity, sedentary time and health. We compared validity and reproducibility of physical activity measurement and posture identification of three activity monitors, as well as user friendliness.

Methods

Healthy volunteers wore three activity monitors simultaneously: ActivPAL3, ActiGraphGT3X and CAM. Data were acquired under both controlled (n = 5) and free-living conditions (n = 9). The controlled laboratory measurement, that included standardized walking intensity and posture allocation, was performed twice. User friendliness was evaluated with a questionnaire. Posture classification was compared with direct observation (controlled measurement) and with diaries (free living). Accelerometer intensity accuracy was tested by correlations with walking speed. User friendliness was compared between activity monitors.

Results

Reproducibility was at least substantial in all monitors. The difference between the two CAM measurements increased with walking intensity. Amount of correct posture classification by ActivPAL3 was 100.0% (kappa 0.98), 33.9% by ActiGraphGT3X (kappa 0.29) and 100.0% by CAM (kappa 0.99). Correlations between accelerometer intensity and walking speed were 0.98 for ActivPAL3, 1.00 for ActiGraphGT3X and 0.98 for CAM. ICCs between activity monitors and diary were 0.98 in ActivPAL3, 0.59 and 0.96 in ActiGraphGT3X and 0.98 in CAM. ActivPAL3 and ActiGraphGT3X had higher user friendliness scores than the CAM.

Conclusions

The ActivPAL3 is valid, reproducible and user friendly. The posture classification by the ActiGraphGT3X is not valid, but reflection of walking intensity and user friendliness are good. The CAM is valid; however, reproducibility at higher walking intensity and user friendliness might cause problems. Further validity studies in free living are recommended.
Appendix
Available only for authorised users
Literature
1.
go back to reference World Health Organization: Global Health Risks: Mortality and Burden of Disease Attributable to Selected Major Risks. 2009, Geneva, Switzerland: WHO Press, 10-11. World Health Organization: Global Health Risks: Mortality and Burden of Disease Attributable to Selected Major Risks. 2009, Geneva, Switzerland: WHO Press, 10-11.
2.
go back to reference van der Ploeg HP, Chey T, Korda RJ, Banks E, Bauman A: Sitting time and all-cause mortality risk in 222 497 Australian adults. Arch Intern Med. 2012, 172: 494-500.CrossRefPubMed van der Ploeg HP, Chey T, Korda RJ, Banks E, Bauman A: Sitting time and all-cause mortality risk in 222 497 Australian adults. Arch Intern Med. 2012, 172: 494-500.CrossRefPubMed
3.
go back to reference Duvivier BM, Schaper NC, Bremers MA, van Crombrugge G, Menheere PP, Kars M, Savelberg HH: Minimal intensity physical activity (standing and walking) of longer duration improves insulin action and plasma lipids more than shorter periods of moderate to vigorous exercise (cycling) in sedentary subjects when energy expenditure is comparable. PLoS One. 2013, 8: e55542-CrossRefPubMedPubMedCentral Duvivier BM, Schaper NC, Bremers MA, van Crombrugge G, Menheere PP, Kars M, Savelberg HH: Minimal intensity physical activity (standing and walking) of longer duration improves insulin action and plasma lipids more than shorter periods of moderate to vigorous exercise (cycling) in sedentary subjects when energy expenditure is comparable. PLoS One. 2013, 8: e55542-CrossRefPubMedPubMedCentral
4.
go back to reference Ekelund U, Franks PW, Sharp S, Brage S, Wareham NJ: Increase in physical activity energy expenditure is associated with reduced metabolic risk independent of change in fatness and fitness. Diabetes Care. 2007, 30: 2101-2106.CrossRefPubMed Ekelund U, Franks PW, Sharp S, Brage S, Wareham NJ: Increase in physical activity energy expenditure is associated with reduced metabolic risk independent of change in fatness and fitness. Diabetes Care. 2007, 30: 2101-2106.CrossRefPubMed
5.
go back to reference Bankoski A, Harris TB, McClain JJ, Brychta RJ, Caserotti P, Chen KY, Berrigan D, Troiano RP, Koster A: Sedentary activity associated with metabolic syndrome independent of physical activity. Diabetes Care. 2011, 34: 497-503.CrossRefPubMedPubMedCentral Bankoski A, Harris TB, McClain JJ, Brychta RJ, Caserotti P, Chen KY, Berrigan D, Troiano RP, Koster A: Sedentary activity associated with metabolic syndrome independent of physical activity. Diabetes Care. 2011, 34: 497-503.CrossRefPubMedPubMedCentral
6.
go back to reference Hamilton MT, Hamilton DG, Zderic TW: Role of low energy expenditure and sitting in obesity, metabolic syndrome, type 2 diabetes, and cardiovascular disease. Diabetes. 2007, 56: 2655-2667.CrossRefPubMed Hamilton MT, Hamilton DG, Zderic TW: Role of low energy expenditure and sitting in obesity, metabolic syndrome, type 2 diabetes, and cardiovascular disease. Diabetes. 2007, 56: 2655-2667.CrossRefPubMed
7.
go back to reference Healy GN, Clark BK, Winkler EA, Gardiner PA, Brown WJ, Matthews CE: Measurement of adults’ sedentary time in population-based studies. Am J Prev Med. 2011, 41: 216-227.CrossRefPubMedPubMedCentral Healy GN, Clark BK, Winkler EA, Gardiner PA, Brown WJ, Matthews CE: Measurement of adults’ sedentary time in population-based studies. Am J Prev Med. 2011, 41: 216-227.CrossRefPubMedPubMedCentral
8.
go back to reference Thorp AA, Owen N, Neuhaus M, Dunstan DW: Sedentary behaviors and subsequent health outcomes in adults a systematic review of longitudinal studies, 1996–2011. Am J Prev Med. 2011, 41: 207-215.CrossRefPubMed Thorp AA, Owen N, Neuhaus M, Dunstan DW: Sedentary behaviors and subsequent health outcomes in adults a systematic review of longitudinal studies, 1996–2011. Am J Prev Med. 2011, 41: 207-215.CrossRefPubMed
9.
go back to reference Clark BK, Sugiyama T, Healy GN, Salmon J, Dunstan DW, Owen N: Validity and reliability of measures of television viewing time and other non-occupational sedentary behaviour of adults: a review. Obes Rev. 2009, 10: 7-16.CrossRefPubMed Clark BK, Sugiyama T, Healy GN, Salmon J, Dunstan DW, Owen N: Validity and reliability of measures of television viewing time and other non-occupational sedentary behaviour of adults: a review. Obes Rev. 2009, 10: 7-16.CrossRefPubMed
10.
go back to reference Prince SA, Adamo KB, Hamel ME, Hardt J, Connor Gorber S, Tremblay M: A comparison of direct versus self-report measures for assessing physical activity in adults: a systematic review. Int J Behav Nutr Phys Act. 2008, 5: 56-CrossRefPubMedPubMedCentral Prince SA, Adamo KB, Hamel ME, Hardt J, Connor Gorber S, Tremblay M: A comparison of direct versus self-report measures for assessing physical activity in adults: a systematic review. Int J Behav Nutr Phys Act. 2008, 5: 56-CrossRefPubMedPubMedCentral
11.
go back to reference Foerster F, Fahrenberg J: Motion pattern and posture: correctly assessed by calibrated accelerometers. Behav Res Methods Instrum Comput. 2000, 32: 450-457.CrossRefPubMed Foerster F, Fahrenberg J: Motion pattern and posture: correctly assessed by calibrated accelerometers. Behav Res Methods Instrum Comput. 2000, 32: 450-457.CrossRefPubMed
12.
go back to reference Oliver M, Schofield GM, Badland HM, Shepherd J: Utility of accelerometer thresholds for classifying sitting in office workers. Prev Med. 2010, 51: 357-360.CrossRefPubMed Oliver M, Schofield GM, Badland HM, Shepherd J: Utility of accelerometer thresholds for classifying sitting in office workers. Prev Med. 2010, 51: 357-360.CrossRefPubMed
13.
go back to reference Carr LJ, Mahar MT: Accuracy of intensity and inclinometer output of three activity monitors for identification of sedentary behavior and light-intensity activity. J Obes 2012. 2012, 2012: 460271- Carr LJ, Mahar MT: Accuracy of intensity and inclinometer output of three activity monitors for identification of sedentary behavior and light-intensity activity. J Obes 2012. 2012, 2012: 460271-
14.
go back to reference Kozey-Keadle S, Libertine A, Lyden K, Staudenmayer J, Freedson PS: Validation of wearable monitors for assessing sedentary behavior. Med Sci Sports Exerc. 2011, 43: 1561-1567.CrossRefPubMed Kozey-Keadle S, Libertine A, Lyden K, Staudenmayer J, Freedson PS: Validation of wearable monitors for assessing sedentary behavior. Med Sci Sports Exerc. 2011, 43: 1561-1567.CrossRefPubMed
15.
go back to reference Kozey-Keadle S, Libertine A, Staudenmayer J, Freedson P: The Feasibility of Reducing and Measuring Sedentary Time among Overweight, Non-Exercising Office Workers. J Obes. 2012, 2012: 282303-CrossRefPubMed Kozey-Keadle S, Libertine A, Staudenmayer J, Freedson P: The Feasibility of Reducing and Measuring Sedentary Time among Overweight, Non-Exercising Office Workers. J Obes. 2012, 2012: 282303-CrossRefPubMed
16.
go back to reference Freedson P, Bowles HR, Troiano R, Haskell W: Assessment of physical activity using wearable monitors: recommendations for monitor calibration and use in the field. Med Sci Sports Exerc. 2012, 44: S1-4.CrossRefPubMedPubMedCentral Freedson P, Bowles HR, Troiano R, Haskell W: Assessment of physical activity using wearable monitors: recommendations for monitor calibration and use in the field. Med Sci Sports Exerc. 2012, 44: S1-4.CrossRefPubMedPubMedCentral
17.
go back to reference Annegarn J, Spruit MA, Uszko-Lencer NH, Vanbelle S, Savelberg HH, Schols AM, Wouters EF, Meijer K: Objective physical activity assessment in patients with chronic organ failure: a validation study of a new single-unit activity monitor. Arch Phys Med Rehabil. 2011, 92: 1852-1857. e1851CrossRefPubMed Annegarn J, Spruit MA, Uszko-Lencer NH, Vanbelle S, Savelberg HH, Schols AM, Wouters EF, Meijer K: Objective physical activity assessment in patients with chronic organ failure: a validation study of a new single-unit activity monitor. Arch Phys Med Rehabil. 2011, 92: 1852-1857. e1851CrossRefPubMed
18.
go back to reference McMahon GC, Brychta RJ, Chen KY: Validation of the Actigraph (GT3X) inclinometer function. Med Sci Sports Exerc. 2010, 42: 489-CrossRef McMahon GC, Brychta RJ, Chen KY: Validation of the Actigraph (GT3X) inclinometer function. Med Sci Sports Exerc. 2010, 42: 489-CrossRef
19.
go back to reference Grant PM, Ryan CG, Tigbe WW, Granat MH: The validation of a novel activity monitor in the measurement of posture and motion during everyday activities. Br J Sports Med. 2006, 40: 992-997.CrossRefPubMedPubMedCentral Grant PM, Ryan CG, Tigbe WW, Granat MH: The validation of a novel activity monitor in the measurement of posture and motion during everyday activities. Br J Sports Med. 2006, 40: 992-997.CrossRefPubMedPubMedCentral
20.
go back to reference Hart TL, Ainsworth BE, Tudor-Locke C: Objective and subjective measures of sedentary behavior and physical activity. Med Sci Sports Exerc. 2011, 43: 449-456.CrossRefPubMed Hart TL, Ainsworth BE, Tudor-Locke C: Objective and subjective measures of sedentary behavior and physical activity. Med Sci Sports Exerc. 2011, 43: 449-456.CrossRefPubMed
21.
go back to reference Godfrey A, Culhane KM, Lyons GM: Comparison of the performance of the activPAL Professional physical activity logger to a discrete accelerometer-based activity monitor. Med Eng Phys. 2007, 29: 930-934.CrossRefPubMed Godfrey A, Culhane KM, Lyons GM: Comparison of the performance of the activPAL Professional physical activity logger to a discrete accelerometer-based activity monitor. Med Eng Phys. 2007, 29: 930-934.CrossRefPubMed
22.
go back to reference Corder K, Brage S, Ekelund U: Accelerometers and pedometers: methodology and clinical application. Curr Opin Clin Nutr Metab Care. 2007, 10: 597-603.CrossRefPubMed Corder K, Brage S, Ekelund U: Accelerometers and pedometers: methodology and clinical application. Curr Opin Clin Nutr Metab Care. 2007, 10: 597-603.CrossRefPubMed
23.
go back to reference Mathie MJ, Coster AC, Lovell NH, Celler BG: Accelerometry: providing an integrated, practical method for long-term, ambulatory monitoring of human movement. Physiol Meas. 2004, 25: R1-20.CrossRefPubMed Mathie MJ, Coster AC, Lovell NH, Celler BG: Accelerometry: providing an integrated, practical method for long-term, ambulatory monitoring of human movement. Physiol Meas. 2004, 25: R1-20.CrossRefPubMed
24.
go back to reference Trost SG, McIver KL, Pate RR: Conducting accelerometer-based activity assessments in field-based research. Med Sci Sports Exerc. 2005, 37: S531-543.CrossRefPubMed Trost SG, McIver KL, Pate RR: Conducting accelerometer-based activity assessments in field-based research. Med Sci Sports Exerc. 2005, 37: S531-543.CrossRefPubMed
25.
go back to reference Landis JR, Koch GG: The measurement of observer agreement for categorical data. Biometrics. 1977, 33: 159-174.CrossRefPubMed Landis JR, Koch GG: The measurement of observer agreement for categorical data. Biometrics. 1977, 33: 159-174.CrossRefPubMed
26.
go back to reference Maddocks M, Petrou A, Skipper L, Wilcock A: Validity of three accelerometers during treadmill walking and motor vehicle travel. Br J Sports Med. 2010, 44: 606-608.CrossRefPubMed Maddocks M, Petrou A, Skipper L, Wilcock A: Validity of three accelerometers during treadmill walking and motor vehicle travel. Br J Sports Med. 2010, 44: 606-608.CrossRefPubMed
27.
go back to reference Dahlgren G, Carlsson D, Moorhead A, Hager-Ross C, McDonough SM: Test-retest reliability of step counts with the ActivPAL device in common daily activities. Gait Posture. 2010, 32: 386-390.CrossRefPubMed Dahlgren G, Carlsson D, Moorhead A, Hager-Ross C, McDonough SM: Test-retest reliability of step counts with the ActivPAL device in common daily activities. Gait Posture. 2010, 32: 386-390.CrossRefPubMed
28.
go back to reference Bouten CV, Sauren AA, Verduin M, Janssen JD: Effects of placement and orientation of body-fixed accelerometers on the assessment of energy expenditure during walking. Med Biol Eng Comput. 1997, 35: 50-56.CrossRefPubMed Bouten CV, Sauren AA, Verduin M, Janssen JD: Effects of placement and orientation of body-fixed accelerometers on the assessment of energy expenditure during walking. Med Biol Eng Comput. 1997, 35: 50-56.CrossRefPubMed
29.
go back to reference Troiano RP, Berrigan D, Dodd KW, Masse LC, Tilert T, McDowell M: Physical activity in the United States measured by accelerometer. Med Sci Sports Exerc. 2008, 40: 181-188.CrossRefPubMed Troiano RP, Berrigan D, Dodd KW, Masse LC, Tilert T, McDowell M: Physical activity in the United States measured by accelerometer. Med Sci Sports Exerc. 2008, 40: 181-188.CrossRefPubMed
30.
go back to reference Winkler EA, Gardiner PA, Clark BK, Matthews CE, Owen N, Healy GN: Identifying sedentary time using automated estimates of accelerometer wear time. Br J Sports Med. 2012, 46: 436-442.CrossRefPubMed Winkler EA, Gardiner PA, Clark BK, Matthews CE, Owen N, Healy GN: Identifying sedentary time using automated estimates of accelerometer wear time. Br J Sports Med. 2012, 46: 436-442.CrossRefPubMed
31.
go back to reference Choi L, Ward SC, Schnelle JF, Buchowski MS: Assessment of wear/nonwear time classification algorithms for triaxial accelerometer. Med Sci Sports Exerc. 2012, 44: 2009-2016.CrossRefPubMedPubMedCentral Choi L, Ward SC, Schnelle JF, Buchowski MS: Assessment of wear/nonwear time classification algorithms for triaxial accelerometer. Med Sci Sports Exerc. 2012, 44: 2009-2016.CrossRefPubMedPubMedCentral
32.
go back to reference Machado-Rodrigues AM, Figueiredo AJ, Mota J, Cumming SP, Eisenmann JC, Malina RM, Coelho ESMJ: Concurrent validation of estimated activity energy expenditure using a 3-day diary and accelerometry in adolescents. Scand J Med Sci Sports. 2012, 22: 259-264.CrossRefPubMed Machado-Rodrigues AM, Figueiredo AJ, Mota J, Cumming SP, Eisenmann JC, Malina RM, Coelho ESMJ: Concurrent validation of estimated activity energy expenditure using a 3-day diary and accelerometry in adolescents. Scand J Med Sci Sports. 2012, 22: 259-264.CrossRefPubMed
Metadata
Title
Which activity monitor to use? Validity, reproducibility and user friendliness of three activity monitors
Authors
Brenda AJ Berendsen
Marike RC Hendriks
Kenneth Meijer
Guy Plasqui
Nicolaas C Schaper
Hans HCM Savelberg
Publication date
01-12-2014
Publisher
BioMed Central
Published in
BMC Public Health / Issue 1/2014
Electronic ISSN: 1471-2458
DOI
https://doi.org/10.1186/1471-2458-14-749

Other articles of this Issue 1/2014

BMC Public Health 1/2014 Go to the issue