Volar locking plate fixation of distal radius fractures: use of an intra-operative ‘carpal shoot through’ view to identify dorsal compartment and distal radioulnar joint screw penetration

Authors: Daniel Marsland, Chris M. Hobbs, Philip S. Sauvé

Published in: HAND | Issue 4/2014

Login to get access

Abstract

Introduction

We report our experience using a ‘carpal shoot through’ view of the distal radius to identify dorsal compartment screw penetration intra-operatively when performing volar plating of the distal radius.

Methods

A prospective study of 42 patients (mean age 56 years) with acute distal radius fractures treated with open reduction internal fixation was undertaken. Surgical fixation was performed using a volar locking plate in all patients. After plate application, inclined posteroanterior and lateral radiographs were taken followed by the carpal shoot through view.

Results

In six cases (14 %), the carpal shoot through view revealed dorsal screw protrusion, which was not detectable on standard PA and lateral views. In one case, a screw had penetrated the distal radioulnar joint (DRUJ), which was only apparent on the shoot through view. The overall screw exchange rate was 17 %.

Conclusions

Using the hand and carpus to minimise the contrast in X-ray penetration, the dorsal cortex of the distal radius may be imaged intra-operatively and dorsal compartment screw penetration detected in cases with significant multifragmentation when screw measurement is difficult. This view potentially reduces the risk of post-operative pain and extensor tendon injury and also provides excellent visualisation of the DRUJ.
Literature
1.
go back to reference Al-Rashid M, Theivendran K, Craigen MA. Delayed ruptures of the extensor tendon secondary to the use of volar locking compression plates for distal radial fractures. J Bone Joint Surg. 2006;88B:1610–2. Al-Rashid M, Theivendran K, Craigen MA. Delayed ruptures of the extensor tendon secondary to the use of volar locking compression plates for distal radial fractures. J Bone Joint Surg. 2006;88B:1610–2.
2.
go back to reference Arora R, Lutz M, Hennerbichler A, et al. Complications following internal fixation of unstable distal radius fracture with a palmar locking-plate. J Orthop Trauma. 2007;21:316–22.PubMedCrossRef Arora R, Lutz M, Hennerbichler A, et al. Complications following internal fixation of unstable distal radius fracture with a palmar locking-plate. J Orthop Trauma. 2007;21:316–22.PubMedCrossRef
3.
go back to reference Benson EC, DeCarvalho A, Mikola FA, et al. Two potential causes of EPL rupture after distal radius volar plate flexion. Clin Orthop Relat Res. 2006;451:218–22.PubMedCrossRef Benson EC, DeCarvalho A, Mikola FA, et al. Two potential causes of EPL rupture after distal radius volar plate flexion. Clin Orthop Relat Res. 2006;451:218–22.PubMedCrossRef
4.
go back to reference Bianchi S, van Aaken J, Glauser T, et al. Screw impingement on the extensor tendons in distal radius fractures treated by volar plating: sonographic appearance. Am J Roentgenol. 2008;191:W199–203.CrossRef Bianchi S, van Aaken J, Glauser T, et al. Screw impingement on the extensor tendons in distal radius fractures treated by volar plating: sonographic appearance. Am J Roentgenol. 2008;191:W199–203.CrossRef
5.
go back to reference Campbell DA. Open reduction and internal fixation of intra articular and unstable fractures of the distal radius using the AO distal radius plate. J Hand Surg. 2000;25B:528–34. Campbell DA. Open reduction and internal fixation of intra articular and unstable fractures of the distal radius using the AO distal radius plate. J Hand Surg. 2000;25B:528–34.
6.
go back to reference Clement H, Pichler W, Nelson D, et al. Morphometric analysis of Lister's tubercle and its consequences on volar plate fixation of distal radius fractures. J Hand Surg. 2008;33A:1716–9. Clement H, Pichler W, Nelson D, et al. Morphometric analysis of Lister's tubercle and its consequences on volar plate fixation of distal radius fractures. J Hand Surg. 2008;33A:1716–9.
7.
go back to reference Drobetz H, Kutscha-Lissberg E. Osteosynthesis of distal radial fractures with a volar locking screw plate system. Int Orthop. 2003;27:1–6.PubMedCentralPubMed Drobetz H, Kutscha-Lissberg E. Osteosynthesis of distal radial fractures with a volar locking screw plate system. Int Orthop. 2003;27:1–6.PubMedCentralPubMed
8.
go back to reference Failla JM, Koniuch MP, Moed BR. Extensor pollicis longus rupture at the tip of a prominent fixation screw: report of three cases. J Hand Surg. 1993;18A:648–51. Failla JM, Koniuch MP, Moed BR. Extensor pollicis longus rupture at the tip of a prominent fixation screw: report of three cases. J Hand Surg. 1993;18A:648–51.
9.
go back to reference Fernandez DL, Geissler WB. Treatment of displaced articular fractures of the radius. J Hand Surg. 1991;16A:375–84. Fernandez DL, Geissler WB. Treatment of displaced articular fractures of the radius. J Hand Surg. 1991;16A:375–84.
10.
go back to reference Haug LC, Glodny B, Deml C, et al. A new radiological method to detect dorsally penetrating screws when using volar locking plates in distal radial fractures. The dorsal horizon view. Bone Joint J. 2013;95-B(8):1101–5.PubMedCrossRef Haug LC, Glodny B, Deml C, et al. A new radiological method to detect dorsally penetrating screws when using volar locking plates in distal radial fractures. The dorsal horizon view. Bone Joint J. 2013;95-B(8):1101–5.PubMedCrossRef
11.
go back to reference Joseph SJ, Harvey JN. The dorsal horizon view: detecting screw protrusion at the distal radius. J Hand Surg. 2011;36A:1691–3. Joseph SJ, Harvey JN. The dorsal horizon view: detecting screw protrusion at the distal radius. J Hand Surg. 2011;36A:1691–3.
12.
go back to reference Jupiter JB, Fernandez DL, Toh CL, et al. Operative treatment of volar intra-articular fractures of the distal end of the radius. J Bone Joint Surg. 1996;78A:1817–28. Jupiter JB, Fernandez DL, Toh CL, et al. Operative treatment of volar intra-articular fractures of the distal end of the radius. J Bone Joint Surg. 1996;78A:1817–28.
13.
go back to reference Kamano M, Honda Y, Kazuki K, et al. Palmar plating for dorsally displaced fractures of the distal radius. Clin Orthop Relat Res. 2002;397:403–8.PubMedCrossRef Kamano M, Honda Y, Kazuki K, et al. Palmar plating for dorsally displaced fractures of the distal radius. Clin Orthop Relat Res. 2002;397:403–8.PubMedCrossRef
14.
go back to reference Kambouroglou GK, Axelrod TS. Complications of the AO/ASIF titanium distal radius plate system in internal fixation of the distal radius: a brief report. J Hand Surg. 1998;23A:737–41. Kambouroglou GK, Axelrod TS. Complications of the AO/ASIF titanium distal radius plate system in internal fixation of the distal radius: a brief report. J Hand Surg. 1998;23A:737–41.
15.
go back to reference Koh S, Morris RP, Patterson RM, et al. Volar fixation for dorsally angulated extra-articular fractures of the distal radius: a biomechanical study. J Hand Surg. 2006;31A:771–9. Koh S, Morris RP, Patterson RM, et al. Volar fixation for dorsally angulated extra-articular fractures of the distal radius: a biomechanical study. J Hand Surg. 2006;31A:771–9.
16.
go back to reference Lee SK, Bae KW, Choy WS. Use of the radial groove view intra-operatively to prevent damage to the extensor pollicis longus tendon by protruding screws during volar plating of a distal radial fracture. Bone Joint J. 2013;95-B(10):1372–6.PubMedCrossRef Lee SK, Bae KW, Choy WS. Use of the radial groove view intra-operatively to prevent damage to the extensor pollicis longus tendon by protruding screws during volar plating of a distal radial fracture. Bone Joint J. 2013;95-B(10):1372–6.PubMedCrossRef
17.
go back to reference Lee HC, Wong YS, Chan BK, et al. Fixation of distal radius fractures using AO titanium volar distal radius plate. Hand Surg. 2003;8:7–15.PubMedCrossRef Lee HC, Wong YS, Chan BK, et al. Fixation of distal radius fractures using AO titanium volar distal radius plate. Hand Surg. 2003;8:7–15.PubMedCrossRef
18.
go back to reference Leung F, Zhu L, Ho H, et al. Palmar plate fixation of AO type C2 fracture of distal radius using a locking compression plate—a biomechanical study in a cadaveric model. J Hand Surg. 2003;28B:263–6. Leung F, Zhu L, Ho H, et al. Palmar plate fixation of AO type C2 fracture of distal radius using a locking compression plate—a biomechanical study in a cadaveric model. J Hand Surg. 2003;28B:263–6.
19.
go back to reference Lowry KJ, Gainor BJ, Hoskins JS. Extensor tendon rupture secondary to the AO/ASIF titanium distal radius plate without associated plate failure: a case report. Am J Orthop. 2000;29:789–91.PubMed Lowry KJ, Gainor BJ, Hoskins JS. Extensor tendon rupture secondary to the AO/ASIF titanium distal radius plate without associated plate failure: a case report. Am J Orthop. 2000;29:789–91.PubMed
20.
go back to reference Maschke SD, Evans PJ, Schub D, et al. Radiographic evaluation of dorsal screw penetration after volar fixed-angle plating of the distal radius: a cadaveric study. Hand (NY). 2007;2:144–50.CrossRef Maschke SD, Evans PJ, Schub D, et al. Radiographic evaluation of dorsal screw penetration after volar fixed-angle plating of the distal radius: a cadaveric study. Hand (NY). 2007;2:144–50.CrossRef
21.
go back to reference McCall TA, Conrad B, Badman B, et al. Volar versus dorsal fixed-angle fixation of dorsally unstable extra-articular distal radius fractures: a biomechanic study. J Hand Surg. 2007;32A:806–12. McCall TA, Conrad B, Badman B, et al. Volar versus dorsal fixed-angle fixation of dorsally unstable extra-articular distal radius fractures: a biomechanic study. J Hand Surg. 2007;32A:806–12.
22.
go back to reference Musgrave DS, Idler RS. Volar fixation of dorsally displaced distal radius fractures using the 2.4-mm locking compression plates. J Hand Surg. 2005;30A:743–9. Musgrave DS, Idler RS. Volar fixation of dorsally displaced distal radius fractures using the 2.4-mm locking compression plates. J Hand Surg. 2005;30A:743–9.
23.
go back to reference Nana AD, Joshi A, Lichtman DM. Plating of the distal radius. J Am Acad Orthop Surg. 2005;13:159–71.PubMed Nana AD, Joshi A, Lichtman DM. Plating of the distal radius. J Am Acad Orthop Surg. 2005;13:159–71.PubMed
24.
go back to reference Orbay JL, Fernandez DL. Volar fixation for dorsally displaced fractures of the distal radius: a preliminary report. J Hand Surg. 2002;27A:205–15. Orbay JL, Fernandez DL. Volar fixation for dorsally displaced fractures of the distal radius: a preliminary report. J Hand Surg. 2002;27A:205–15.
25.
go back to reference Orbay JL, Fernandez DL. Volar fixed-angle plate fixation for unstable distal radius fractures in the elderly patient. J Hand Surg. 2004;29A:96–102. Orbay JL, Fernandez DL. Volar fixed-angle plate fixation for unstable distal radius fractures in the elderly patient. J Hand Surg. 2004;29A:96–102.
26.
go back to reference Osada D, Fujita S, Tamai K, et al. Biomechanics in uniaxial compression of three distal radius volar plates. J Hand Surg. 2004;29A:446–51. Osada D, Fujita S, Tamai K, et al. Biomechanics in uniaxial compression of three distal radius volar plates. J Hand Surg. 2004;29A:446–51.
27.
go back to reference Osada D, Viegas SF, Shah MA, et al. Comparison of different distal radius dorsal and volar fracture fixation plates: a biomechanical study. J Hand Surg. 2003;28A:94–104. Osada D, Viegas SF, Shah MA, et al. Comparison of different distal radius dorsal and volar fracture fixation plates: a biomechanical study. J Hand Surg. 2003;28A:94–104.
28.
go back to reference Ozer K, Toker S. Dorsal tangential view of the wrist to detect screw penetration to the dorsal cortex of the distal radius after volar fixed-angle plating. Hand (NY). 2011;6:190–3.CrossRef Ozer K, Toker S. Dorsal tangential view of the wrist to detect screw penetration to the dorsal cortex of the distal radius after volar fixed-angle plating. Hand (NY). 2011;6:190–3.CrossRef
29.
go back to reference Pace A, Cresswell T. Use of articular wrist views to assess intra-articular screw penetration in surgical fixation of distal radius fractures. J Hand Surg. 2010;35A:1015–8. Pace A, Cresswell T. Use of articular wrist views to assess intra-articular screw penetration in surgical fixation of distal radius fractures. J Hand Surg. 2010;35A:1015–8.
30.
go back to reference Rampoldi M, Marsico S. Complications of volar plating of distal radius fractures. Acta Orthop Belg. 2007;73:714–9.PubMed Rampoldi M, Marsico S. Complications of volar plating of distal radius fractures. Acta Orthop Belg. 2007;73:714–9.PubMed
31.
go back to reference Riddick AP, Hickey B, White SP. Accuracy of the skyline view for detecting dorsal cortical penetration during volar distal radius fixation. J Hand Surg. 2012;37B:407–11. Riddick AP, Hickey B, White SP. Accuracy of the skyline view for detecting dorsal cortical penetration during volar distal radius fixation. J Hand Surg. 2012;37B:407–11.
32.
go back to reference Rozental TD, Blazar PE. Functional outcome and complications after volar plating for dorsally displaced, unstable fractures of the distal radius. J Hand Surg. 2006;31A:359–65. Rozental TD, Blazar PE. Functional outcome and complications after volar plating for dorsally displaced, unstable fractures of the distal radius. J Hand Surg. 2006;31A:359–65.
33.
go back to reference Ruch DS, Papadonikolakis A. Volar versus dorsal plating in the management of intra-articular distal radius fractures. J Hand Surg. 2006;31A:9–16. Ruch DS, Papadonikolakis A. Volar versus dorsal plating in the management of intra-articular distal radius fractures. J Hand Surg. 2006;31A:9–16.
34.
go back to reference Simic PM, Weiland AJ. Fractures of the distal aspect of the radius: changes in treatment over the past two decades. Instr Course Lect. 2003;52:185–95.PubMed Simic PM, Weiland AJ. Fractures of the distal aspect of the radius: changes in treatment over the past two decades. Instr Course Lect. 2003;52:185–95.PubMed
35.
go back to reference Soong M, Got C, Katarincic J, et al. Fluoroscopic evaluation of intra-articular screw placement during locked volar plating of the distal radius: a cadaveric study. J Hand Surg. 2008;33A:1720–3. Soong M, Got C, Katarincic J, et al. Fluoroscopic evaluation of intra-articular screw placement during locked volar plating of the distal radius: a cadaveric study. J Hand Surg. 2008;33A:1720–3.
36.
go back to reference Thomas AD, Greenberg JA. Use of fluoroscopy in determining screw overshoot in the dorsal distal radius: a cadaveric study. J Hand Surg. 2009;34A:258–61. Thomas AD, Greenberg JA. Use of fluoroscopy in determining screw overshoot in the dorsal distal radius: a cadaveric study. J Hand Surg. 2009;34A:258–61.
37.
go back to reference Willis AA, Kutsumi K, Zobitz ME, et al. Internal fixation of dorsally displaced fractures of the distal part of the radius. A biomechanical analysis of volar plate fracture stability. J Bone Joint Surg. 2006;88A:2411–7. Willis AA, Kutsumi K, Zobitz ME, et al. Internal fixation of dorsally displaced fractures of the distal part of the radius. A biomechanical analysis of volar plate fracture stability. J Bone Joint Surg. 2006;88A:2411–7.
38.
go back to reference Wong-Chung J, Quinlan W. Rupture of extensor pollicis longus following fixation of a distal radius fracture. Injury. 1989;20:375–6.PubMedCrossRef Wong-Chung J, Quinlan W. Rupture of extensor pollicis longus following fixation of a distal radius fracture. Injury. 1989;20:375–6.PubMedCrossRef
Metadata
Title
Volar locking plate fixation of distal radius fractures: use of an intra-operative ‘carpal shoot through’ view to identify dorsal compartment and distal radioulnar joint screw penetration
Authors
Daniel Marsland
Chris M. Hobbs
Philip S. Sauvé
Publication date
01-12-2014
Publisher
Springer US
Published in
HAND / Issue 4/2014
Print ISSN: 1558-9447
Electronic ISSN: 1558-9455
DOI
https://doi.org/10.1007/s11552-014-9607-0

Other articles of this Issue 4/2014

HAND 4/2014 Go to the issue