Variability in low-flow oxygen delivery by nasal cannula evaluated in neonatal and infant airway replicas

Authors: Mozhgan Sabz, Scott Tavernini, Kineshta Pillay, Cole Christianson, Michelle Noga, Warren H. Finlay, Hossein Rouhani, Andrew R. Martin

Published in: Respiratory Research | Issue 1/2022

Login to get access

Abstract

Background

The nasal cannula is considered a trusted and effective means of administering low-flow oxygen and is widely used for neonates and infants requiring oxygen therapy, despite an understanding that oxygen concentrations delivered to patients are variable.

Methods

In the present study, realistic nasal airway replicas derived from medical scans of children less than 3 months old were used to measure the fraction of oxygen inhaled (FiO2) through nasal cannulas during low-flow oxygen delivery. Parameters influencing variability in FiO2 were evaluated, as was the hypothesis that measured FiO2 values could be predicted using a simple, flow-weighted calculation that assumes ideal mixing of oxygen with entrained room air. Tidal breathing through neonatal and infant nasal airway replicas was controlled using a lung simulator. Parameters for nasal cannula oxygen flow rate, nasal airway geometry, tidal volume, respiratory rate, inhalation/exhalation, or I:E ratio (ti/te), breath waveform, and cannula prong insertion position were varied to determine their effect on measured FiO2. In total, FiO2 was measured for 384 different parameter combinations, with each combination repeated in triplicate. Analysis of variance (ANOVA) was used to assess the influence of parameters on measured FiO2.

Results

Measured FiO2 was not appreciably affected by the breath waveform shape, the replica geometry, or the cannula position but was significantly influenced by the tidal volume, the inhalation time, and the nasal cannula flow rate.

Conclusions

The flow-weighted calculation overpredicted FiO2 for measured values above 60%, but an empirical correction to the calculation provided good agreement with measured FiO2 across the full range of experimental data.
Literature
1.
go back to reference World Health Organization. Oxygen therapy for children: a manual for health workers. 2016. ISBN 978-92-4-154955-4. World Health Organization. Oxygen therapy for children: a manual for health workers. 2016. ISBN 978-92-4-154955-4.
2.
go back to reference Walsh BK, Smallwood CD. Pediatric oxygen therapy: a review and update. Respir Care. 2017;62(6):645–61.CrossRefPubMed Walsh BK, Smallwood CD. Pediatric oxygen therapy: a review and update. Respir Care. 2017;62(6):645–61.CrossRefPubMed
3.
go back to reference Kepreotes E, et al. High-flow warm humidified oxygen versus standard low-flow nasal cannula oxygen for moderate bronchiolitis (HFWHO RCT): an open, phase 4, randomised controlled trial. Lancet. 2017;389(10072):930–9.CrossRefPubMed Kepreotes E, et al. High-flow warm humidified oxygen versus standard low-flow nasal cannula oxygen for moderate bronchiolitis (HFWHO RCT): an open, phase 4, randomised controlled trial. Lancet. 2017;389(10072):930–9.CrossRefPubMed
4.
go back to reference Pons-Odena M, et al. SpO2/FiO2 as a predictor of non-invasive ventilation failure in children with hypoxemic respiratory insufficiency. J Pediatr Intensive Care. 2013;2(03):111–9.CrossRefPubMedPubMedCentral Pons-Odena M, et al. SpO2/FiO2 as a predictor of non-invasive ventilation failure in children with hypoxemic respiratory insufficiency. J Pediatr Intensive Care. 2013;2(03):111–9.CrossRefPubMedPubMedCentral
5.
go back to reference Leteurtre S, et al. Assessment of the pediatric index of mortality 2 with the Pao2/Fio2 ratio derived from the Spo2/Fio2 ratio: a prospective pilot study in a French pediatric intensive care unit. Pediatr Crit Care Med. 2011;12(4):e184–6.CrossRefPubMed Leteurtre S, et al. Assessment of the pediatric index of mortality 2 with the Pao2/Fio2 ratio derived from the Spo2/Fio2 ratio: a prospective pilot study in a French pediatric intensive care unit. Pediatr Crit Care Med. 2011;12(4):e184–6.CrossRefPubMed
6.
go back to reference Kapadia V, Wyckoff MH. Oxygen therapy in the delivery room: what is the right dose? Clin Perinatol. 2018;45(2):293–306.CrossRefPubMed Kapadia V, Wyckoff MH. Oxygen therapy in the delivery room: what is the right dose? Clin Perinatol. 2018;45(2):293–306.CrossRefPubMed
7.
go back to reference Kuluz JW, et al. The fraction of inspired oxygen in infants receiving oxygen via nasal cannula often exceeds safe levels. Respir Care. 2001;46(9):897–901.PubMed Kuluz JW, et al. The fraction of inspired oxygen in infants receiving oxygen via nasal cannula often exceeds safe levels. Respir Care. 2001;46(9):897–901.PubMed
8.
go back to reference Fan LL, Voyles JB. Determination of inspired oxygen delivered by nasal cannula in infants with chronic lung disease. J Pediatr. 1983;103(6):923–5.CrossRefPubMed Fan LL, Voyles JB. Determination of inspired oxygen delivered by nasal cannula in infants with chronic lung disease. J Pediatr. 1983;103(6):923–5.CrossRefPubMed
9.
go back to reference Sung V, et al. Estimating inspired oxygen concentration delivered by nasal prongs in children with bronchiolitis. J Paediatr Child Health. 2008;44(1–2):14–8.PubMed Sung V, et al. Estimating inspired oxygen concentration delivered by nasal prongs in children with bronchiolitis. J Paediatr Child Health. 2008;44(1–2):14–8.PubMed
10.
go back to reference Benaron DA, Benitz WE. Maximizing the stability of oxygen delivered via nasal cannula. Arch Pediatr Adolesc Med. 1994;148(3):294–300.CrossRefPubMed Benaron DA, Benitz WE. Maximizing the stability of oxygen delivered via nasal cannula. Arch Pediatr Adolesc Med. 1994;148(3):294–300.CrossRefPubMed
11.
go back to reference Finer NN, Bates R, Tomat P. Low flow oxygen delivery via nasal cannula to neonates. Pediatr Pulmonol. 1996;21(1):48–51.CrossRefPubMed Finer NN, Bates R, Tomat P. Low flow oxygen delivery via nasal cannula to neonates. Pediatr Pulmonol. 1996;21(1):48–51.CrossRefPubMed
12.
go back to reference Vain NE, et al. Regulation of oxygen concentration delivered to infants via nasal cannulas. Am J Dis Child. 1989;143(12):1458–60.PubMed Vain NE, et al. Regulation of oxygen concentration delivered to infants via nasal cannulas. Am J Dis Child. 1989;143(12):1458–60.PubMed
13.
go back to reference Duong K, et al. Comparison of airway pressures and expired gas washout for nasal high flow versus CPAP in child airway replicas. Respir Res. 2021;22(1):1–13.CrossRef Duong K, et al. Comparison of airway pressures and expired gas washout for nasal high flow versus CPAP in child airway replicas. Respir Res. 2021;22(1):1–13.CrossRef
14.
go back to reference Pillay K, et al. Inhaled nitric oxide: in vitro analysis of continuous flow noninvasive delivery via nasal cannula. Respir Care. 2021;66(2):228–39.CrossRefPubMed Pillay K, et al. Inhaled nitric oxide: in vitro analysis of continuous flow noninvasive delivery via nasal cannula. Respir Care. 2021;66(2):228–39.CrossRefPubMed
15.
go back to reference Wilkins JV, et al. Mechanistic understanding of high flow nasal cannula therapy and pressure support with an in vitro infant model. Ann Biomed Eng. 2020;48(2):624–33.CrossRefPubMed Wilkins JV, et al. Mechanistic understanding of high flow nasal cannula therapy and pressure support with an in vitro infant model. Ann Biomed Eng. 2020;48(2):624–33.CrossRefPubMed
16.
go back to reference Moore C, et al. The influence of flowrate and gas density on positive airway pressure for high flow nasal cannula applied to infant airway replicas. J Biomech. 2020;112: 110022.CrossRefPubMed Moore C, et al. The influence of flowrate and gas density on positive airway pressure for high flow nasal cannula applied to infant airway replicas. J Biomech. 2020;112: 110022.CrossRefPubMed
17.
go back to reference Chen JZ, et al. Comparison of pulsed versus continuous oxygen delivery using realistic adult nasal airway replicas. Int J Chron Obstruct Pulmon Dis. 2017;12:2559.CrossRefPubMedPubMedCentral Chen JZ, et al. Comparison of pulsed versus continuous oxygen delivery using realistic adult nasal airway replicas. Int J Chron Obstruct Pulmon Dis. 2017;12:2559.CrossRefPubMedPubMedCentral
18.
go back to reference Katz I, et al. Dose variability of supplemental oxygen therapy with open patient interfaces based on in vitro measurements using a physiologically realistic upper airway model. Respir Res. 2019;20(1):1–12.CrossRef Katz I, et al. Dose variability of supplemental oxygen therapy with open patient interfaces based on in vitro measurements using a physiologically realistic upper airway model. Respir Res. 2019;20(1):1–12.CrossRef
19.
go back to reference Chen DY, et al. A pilot study of heated and humidified low flow oxygen therapy: an assessment in infants with mild and moderate bronchiolitis (HHOT AIR study). Pediatr Pulmonol. 2019;54(5):620–7.CrossRefPubMed Chen DY, et al. A pilot study of heated and humidified low flow oxygen therapy: an assessment in infants with mild and moderate bronchiolitis (HHOT AIR study). Pediatr Pulmonol. 2019;54(5):620–7.CrossRefPubMed
21.
go back to reference Martin DC. Contemporary portable oxygen concentrators and diverse breathing behaviours–a bench comparison. BMC Pulm Med. 2019;19(1):1–11.CrossRef Martin DC. Contemporary portable oxygen concentrators and diverse breathing behaviours–a bench comparison. BMC Pulm Med. 2019;19(1):1–11.CrossRef
22.
go back to reference Moore CP, et al. Correlation of high flow nasal cannula outlet area with gas clearance and pressure in adult upper airway replicas. Clin Biomech. 2019;66:66–73.CrossRef Moore CP, et al. Correlation of high flow nasal cannula outlet area with gas clearance and pressure in adult upper airway replicas. Clin Biomech. 2019;66:66–73.CrossRef
23.
go back to reference Moore C, et al. High flow nasal cannula: Influence of gas type and flow rate on airway pressure and CO2 clearance in adult nasal airway replicas. Clin Biomech. 2019;65:73–80.CrossRef Moore C, et al. High flow nasal cannula: Influence of gas type and flow rate on airway pressure and CO2 clearance in adult nasal airway replicas. Clin Biomech. 2019;65:73–80.CrossRef
25.
go back to reference Christianson CD, et al. In vitro evaluation of a nasal interface used to improve delivery from a portable oxygen concentrator. J Med Devices. 2022;16(1): 011101.CrossRef Christianson CD, et al. In vitro evaluation of a nasal interface used to improve delivery from a portable oxygen concentrator. J Med Devices. 2022;16(1): 011101.CrossRef
26.
go back to reference Tavernini S, et al. Deposition of micrometer-sized aerosol particles in neonatal nasal airway replicas. Aerosol Sci Technol. 2018;52(4):407–19.CrossRef Tavernini S, et al. Deposition of micrometer-sized aerosol particles in neonatal nasal airway replicas. Aerosol Sci Technol. 2018;52(4):407–19.CrossRef
27.
go back to reference Clarke JR, Aston H, Silverman M. Evaluation of a tidal expiratory flow index in healthy and diseased infants. Pediatr Pulmonol. 1994;17(5):285–90.CrossRefPubMed Clarke JR, Aston H, Silverman M. Evaluation of a tidal expiratory flow index in healthy and diseased infants. Pediatr Pulmonol. 1994;17(5):285–90.CrossRefPubMed
28.
go back to reference Phelan P, Williams H, Freeman M. The disturbances of ventilation in acute viral bronchiolitis. J Paediatr Child Health. 1968;4(2):96–104.CrossRef Phelan P, Williams H, Freeman M. The disturbances of ventilation in acute viral bronchiolitis. J Paediatr Child Health. 1968;4(2):96–104.CrossRef
29.
go back to reference Schmalisch G, et al. Influence of preterm onset of inspiration on tidal breathing parameters in infants with and without CLD. Respir Physiol Neurobiol. 2003;135(1):39–46.CrossRefPubMed Schmalisch G, et al. Influence of preterm onset of inspiration on tidal breathing parameters in infants with and without CLD. Respir Physiol Neurobiol. 2003;135(1):39–46.CrossRefPubMed
30.
go back to reference Paetow U, Windstetter D, Schmalisch G. Variability of tidal breathing low-volume loops in healthy and sick newborns. Am J Perinatol. 1999;16(10):0549–60.CrossRef Paetow U, Windstetter D, Schmalisch G. Variability of tidal breathing low-volume loops in healthy and sick newborns. Am J Perinatol. 1999;16(10):0549–60.CrossRef
31.
go back to reference Dezateux C, et al. Infant respiratory function after RSV-proven bronchiolitis. Am J Respir Crit Care Med. 1997;155(4):1349–55.CrossRefPubMed Dezateux C, et al. Infant respiratory function after RSV-proven bronchiolitis. Am J Respir Crit Care Med. 1997;155(4):1349–55.CrossRefPubMed
32.
go back to reference Zhou S, Chatburn RL. Effect of the anatomic reservoir on low-flow oxygen delivery via nasal cannula: constant flow versus pulse flow with portable oxygen concentrator. Respir Care. 2014;59(8):1199–209.CrossRefPubMed Zhou S, Chatburn RL. Effect of the anatomic reservoir on low-flow oxygen delivery via nasal cannula: constant flow versus pulse flow with portable oxygen concentrator. Respir Care. 2014;59(8):1199–209.CrossRefPubMed
33.
go back to reference Lawrence I, Lin K. A concordance correlation coefficient to evaluate reproducibility. Biometrics. 1989;45(1):255–68.CrossRef Lawrence I, Lin K. A concordance correlation coefficient to evaluate reproducibility. Biometrics. 1989;45(1):255–68.CrossRef
35.
go back to reference McBride G. A proposal for strength-of-agreement criteria for Lin’s concordance correlation coefficient. NIWA client report: HAM2005–062, 2005. 45: p. 307–310. McBride G. A proposal for strength-of-agreement criteria for Lin’s concordance correlation coefficient. NIWA client report: HAM2005–062, 2005. 45: p. 307–310.
36.
go back to reference Duprez F, et al. A new formula for predicting the fraction of delivered oxygen during low-flow oxygen therapy. Respir Care. 2018;63(12):1528–34.CrossRefPubMed Duprez F, et al. A new formula for predicting the fraction of delivered oxygen during low-flow oxygen therapy. Respir Care. 2018;63(12):1528–34.CrossRefPubMed
37.
go back to reference Chirico G, Beccagutti F. Nasal obstruction in neonates and infants. Minerva Pediatr. 2010;62(5):499–505.PubMed Chirico G, Beccagutti F. Nasal obstruction in neonates and infants. Minerva Pediatr. 2010;62(5):499–505.PubMed
38.
go back to reference Gnagi SH, Schraff SA. Nasal obstruction in newborns. Pediatr Clin. 2013;60(4):903–22. Gnagi SH, Schraff SA. Nasal obstruction in newborns. Pediatr Clin. 2013;60(4):903–22.
39.
go back to reference Bergeson PS, Shaw JC. Are infants really obligatory nasal breathers? Clin Pediatr. 2001;40(10):567–9.CrossRef Bergeson PS, Shaw JC. Are infants really obligatory nasal breathers? Clin Pediatr. 2001;40(10):567–9.CrossRef
40.
go back to reference Wei J, et al. Low re-inhalation of the exhaled flow during normal nasal breathing in a pediatric airway replica. Build Environ. 2016;97:40–7.CrossRefPubMed Wei J, et al. Low re-inhalation of the exhaled flow during normal nasal breathing in a pediatric airway replica. Build Environ. 2016;97:40–7.CrossRefPubMed
Metadata
Title
Variability in low-flow oxygen delivery by nasal cannula evaluated in neonatal and infant airway replicas
Authors
Mozhgan Sabz
Scott Tavernini
Kineshta Pillay
Cole Christianson
Michelle Noga
Warren H. Finlay
Hossein Rouhani
Andrew R. Martin
Publication date
01-12-2022
Publisher
BioMed Central
Published in
Respiratory Research / Issue 1/2022
Electronic ISSN: 1465-993X
DOI
https://doi.org/10.1186/s12931-022-02260-x

Other articles of this Issue 1/2022

Respiratory Research 1/2022 Go to the issue
Live Webinar | 27-06-2024 | 18:00 (CEST)

Keynote webinar | Spotlight on medication adherence

Live: Thursday 27th June 2024, 18:00-19:30 (CEST)

WHO estimates that half of all patients worldwide are non-adherent to their prescribed medication. The consequences of poor adherence can be catastrophic, on both the individual and population level.

Join our expert panel to discover why you need to understand the drivers of non-adherence in your patients, and how you can optimize medication adherence in your clinics to drastically improve patient outcomes.