Published in: Annals of Nuclear Medicine 7/2015

01-08-2015 | Technical Note

Validity of the mediastinum as a reference region to evaluate cardiac accumulation of iodine-123 metaiodobenzylguanidine

Authors: Yusuke Inoue, Yutaka Abe, Yuji Asano, Kei Kikuchi

Published in: Annals of Nuclear Medicine | Issue 7/2015

Login to get access

Abstract

Objective

The heart-to-mediastinum (H/M) ratio, the heart count normalized for the mediastinum count, is commonly used in cardiac 123I-metaiodobenzylguanidine (MIBG) imaging. However, there are reports describing age-dependent increases in the mediastinum count with or without correction for the injected dose (ID) and a poor correlation between the H/M ratio and heart count normalized for ID. We evaluated the validity of the mediastinum count as a reference in comparison with the ID.

Methods

Results of cardiac 123I-MIBG imaging in 200 patients were analyzed. The mean counts for the heart and mediastinum were estimated to calculate the H/M ratio. Additionally, the heart and mediastinum counts were normalized for ID measured with a dose calibrator. ID was corrected for body size represented by body weight, body surface area, or lean body mass.

Results

The coefficient of variance of the ID-normalized mediastinum count was reduced by correcting ID for body size. The indicators of body size showed significant negative correlations with age. Although a positive correlation was found between age and the ID-normalized mediastinum counts, the age-dependence was reduced by body size correction. There was a close correlation between the H/M ratio and ID-normalized heart counts, and body size correction improved the correlation.

Conclusion

The results of this study indicate the validity of the mediastinum as a reference region and support the use of the H/M ratio as an index of cardiac accumulation of 123I-MIBG.
Literature
1.
go back to reference Flotats A, Carrió I, Agostini D, Le Guludec D, Marcassa C, Schaffers M, et al. Proposal for standardization of 123I-metaiodobenzylguanidine (MIBG) cardiac sympathetic imaging by the EANM Cardiovascular Committee and the European Council of Nuclear Cardiology. Eur J Nucl Med Mol Imaging. 2010;37:1802–12.PubMedCrossRef Flotats A, Carrió I, Agostini D, Le Guludec D, Marcassa C, Schaffers M, et al. Proposal for standardization of 123I-metaiodobenzylguanidine (MIBG) cardiac sympathetic imaging by the EANM Cardiovascular Committee and the European Council of Nuclear Cardiology. Eur J Nucl Med Mol Imaging. 2010;37:1802–12.PubMedCrossRef
2.
go back to reference Nakata T, Nakajima K, Yamashina S, Yamada T, Momose M, Kasama S, et al. A pooled analysis of multicenter cohort studies of 123I-mIBG imaging of sympathetic innervation for assessment of long-term prognosis in heart failure. JACC Cardiovasc Imaging. 2013;6:772–84.PubMedCrossRef Nakata T, Nakajima K, Yamashina S, Yamada T, Momose M, Kasama S, et al. A pooled analysis of multicenter cohort studies of 123I-mIBG imaging of sympathetic innervation for assessment of long-term prognosis in heart failure. JACC Cardiovasc Imaging. 2013;6:772–84.PubMedCrossRef
3.
go back to reference Jacobson AF, Senior R, Cerqueira MD, Wong ND, Thomas GS, Lopez VA, et al. Myocardial iodine-123 meta-iodobenzylguanidine imaging and cardiac events in heart failure. Results of the prospective ADMIRE-HF (AdreView Myocardial Imaging for Risk Evaluation in Heart Failure) study. J Am Coll Cardiol. 2010;55:2212–21.PubMedCrossRef Jacobson AF, Senior R, Cerqueira MD, Wong ND, Thomas GS, Lopez VA, et al. Myocardial iodine-123 meta-iodobenzylguanidine imaging and cardiac events in heart failure. Results of the prospective ADMIRE-HF (AdreView Myocardial Imaging for Risk Evaluation in Heart Failure) study. J Am Coll Cardiol. 2010;55:2212–21.PubMedCrossRef
4.
go back to reference Carrió I, Cowie MR, Yamazaki J, Udelson J, Camici PG. Cardiac sympathetic imaging with mIBG in heart failure. JACC Cardiovasc Imaging. 2010;3:92–100.PubMedCrossRef Carrió I, Cowie MR, Yamazaki J, Udelson J, Camici PG. Cardiac sympathetic imaging with mIBG in heart failure. JACC Cardiovasc Imaging. 2010;3:92–100.PubMedCrossRef
5.
go back to reference Treglia G, Cason E, Stefanelli A, Cocciolillo F, Di Giuda D, Fagioli G, et al. MIBG scintigraphy in differential diagnosis of Parkinsonism: a meta-analysis. Clin Auton Res. 2012;22:43–55.PubMedCrossRef Treglia G, Cason E, Stefanelli A, Cocciolillo F, Di Giuda D, Fagioli G, et al. MIBG scintigraphy in differential diagnosis of Parkinsonism: a meta-analysis. Clin Auton Res. 2012;22:43–55.PubMedCrossRef
6.
go back to reference Treglia G, Cason E. Diagnostic performance of myocardial innervation imaging using MIBG scintigraphy in differential diagnosis between dementia with lewy bodies and other dementias: a systematic review and a meta-analysis. J Neuroimaging. 2012;22:111–7.PubMedCrossRef Treglia G, Cason E. Diagnostic performance of myocardial innervation imaging using MIBG scintigraphy in differential diagnosis between dementia with lewy bodies and other dementias: a systematic review and a meta-analysis. J Neuroimaging. 2012;22:111–7.PubMedCrossRef
7.
go back to reference Fletcher AM, Motherwell DW, Small AD, McCurrach GM, Goodfield NE, Petrie MC, et al. I-123 MIBG cardiac uptake measurements: limitations of collimator choice and scatter correction in the clinical context. Nucl Med Commun. 2010;31:629–36.PubMed Fletcher AM, Motherwell DW, Small AD, McCurrach GM, Goodfield NE, Petrie MC, et al. I-123 MIBG cardiac uptake measurements: limitations of collimator choice and scatter correction in the clinical context. Nucl Med Commun. 2010;31:629–36.PubMed
8.
go back to reference Somsen GA, Borm JJ, Dubois EA, Schook MB, Van Der Wall EE, et al. Cardiac 123I-MIBG uptake is affected by variable uptake in reference regions: implications for interpretation in clinical studies. Nucl Med Commun. 1996;17:872–6.PubMedCrossRef Somsen GA, Borm JJ, Dubois EA, Schook MB, Van Der Wall EE, et al. Cardiac 123I-MIBG uptake is affected by variable uptake in reference regions: implications for interpretation in clinical studies. Nucl Med Commun. 1996;17:872–6.PubMedCrossRef
9.
go back to reference Jacobson AF, Chen J, Verdes L, Folks RD, Manatunga DN, Garcia EV. Impact of age on myocardial uptake of 123I-mIBG in older adult subjects without coronary heart disease. J Nucl Cardiol. 2013;20:406–14.PubMedCrossRef Jacobson AF, Chen J, Verdes L, Folks RD, Manatunga DN, Garcia EV. Impact of age on myocardial uptake of 123I-mIBG in older adult subjects without coronary heart disease. J Nucl Cardiol. 2013;20:406–14.PubMedCrossRef
10.
go back to reference Gehan EA, George SL. Estimation of human body surface area from height and weight. Cancer Chemother Rep. 1970;54:225–35.PubMed Gehan EA, George SL. Estimation of human body surface area from height and weight. Cancer Chemother Rep. 1970;54:225–35.PubMed
11.
go back to reference James W, Waterlow J. UK Department of Health and Social Security/Medical Research Council Group on obesity research. Research on obesity: a report of the DHSS/MRC Group. London, England: H.M.S.O. 1976. James W, Waterlow J. UK Department of Health and Social Security/Medical Research Council Group on obesity research. Research on obesity: a report of the DHSS/MRC Group. London, England: H.M.S.O. 1976.
12.
go back to reference Ito H, Ohshima A, Ohto N, Ogasawara M, Tsuzuki M, Takao K, et al. Relation between body composition and age in healthy Japanese subjects. Eur J Clin Nutr. 2001;55:462–70.PubMedCrossRef Ito H, Ohshima A, Ohto N, Ogasawara M, Tsuzuki M, Takao K, et al. Relation between body composition and age in healthy Japanese subjects. Eur J Clin Nutr. 2001;55:462–70.PubMedCrossRef
13.
go back to reference Green B, Duffull S. Caution when lean body weight is used as a size descriptor for obese subjects. Clin Pharmacol Ther. 2002;72:743–4.PubMedCrossRef Green B, Duffull S. Caution when lean body weight is used as a size descriptor for obese subjects. Clin Pharmacol Ther. 2002;72:743–4.PubMedCrossRef
14.
go back to reference Tahari AK, Chien D, Azadi JR, Wahl RL. Optimum lean body formulation for correction of standardized uptake value in PET imaging. J Nucl Med. 2014;55:1481–4.PubMedCentralPubMedCrossRef Tahari AK, Chien D, Azadi JR, Wahl RL. Optimum lean body formulation for correction of standardized uptake value in PET imaging. J Nucl Med. 2014;55:1481–4.PubMedCentralPubMedCrossRef
15.
go back to reference Inoue Y, Suzuki A, Shirouzu I, Machida T, Yoshizawa Y, Akita F, et al. Effect of collimator choice on quantitative assessment of cardiac iodine 123 MIBG uptake. J Nucl Cardiol. 2003;10:623–32.PubMedCrossRef Inoue Y, Suzuki A, Shirouzu I, Machida T, Yoshizawa Y, Akita F, et al. Effect of collimator choice on quantitative assessment of cardiac iodine 123 MIBG uptake. J Nucl Cardiol. 2003;10:623–32.PubMedCrossRef
16.
go back to reference Inoue Y, Abe Y, Itoh Y, Asano Y, Kikuchi K, Sakamoto Y, et al. Acquisition protocols and correction methods for estimation of the heart-to-mediastinum ratio in 123I-metaiodobenzylguanidine cardiac sympathetic imaging. J Nucl Med. 2013;54:707–13.PubMedCrossRef Inoue Y, Abe Y, Itoh Y, Asano Y, Kikuchi K, Sakamoto Y, et al. Acquisition protocols and correction methods for estimation of the heart-to-mediastinum ratio in 123I-metaiodobenzylguanidine cardiac sympathetic imaging. J Nucl Med. 2013;54:707–13.PubMedCrossRef
17.
go back to reference Inoue Y, Abe Y, Asano Y, Kikuchi K, Matsunaga K, Iizuka T, et al. An improved method for estimating the heart-to-mediastinum ratio from cardiac sympathetic nerve imaging with low-energy high-resolution collimators. J Nucl Cardiol. 2014;21:614–21.PubMedCrossRef Inoue Y, Abe Y, Asano Y, Kikuchi K, Matsunaga K, Iizuka T, et al. An improved method for estimating the heart-to-mediastinum ratio from cardiac sympathetic nerve imaging with low-energy high-resolution collimators. J Nucl Cardiol. 2014;21:614–21.PubMedCrossRef
18.
go back to reference Inoue Y. 123I-MIBG cardiac imaging: acquisition protocols and correction methods for quantitative evaluation. Curr Cardiovasc Imaging Rep. 2015;8:9306.CrossRef Inoue Y. 123I-MIBG cardiac imaging: acquisition protocols and correction methods for quantitative evaluation. Curr Cardiovasc Imaging Rep. 2015;8:9306.CrossRef
19.
go back to reference Jacobson AF, Centofanti R, Babalola OI, Dean B. Survey of the performance of commercial dose calibrators for measurement of 123I activity. J Nucl Med Technol. 2011;39:302–6.PubMedCrossRef Jacobson AF, Centofanti R, Babalola OI, Dean B. Survey of the performance of commercial dose calibrators for measurement of 123I activity. J Nucl Med Technol. 2011;39:302–6.PubMedCrossRef
20.
go back to reference Inoue Y, Shirouzu I, Machida T, Yoshizawa Y, Akita F, Doi I, et al. Physical characteristics of low and medium energy collimators for 123I imaging and simultaneous dual-isotope imaging. Nucl Med Commun. 2003;24:1195–202.PubMedCrossRef Inoue Y, Shirouzu I, Machida T, Yoshizawa Y, Akita F, Doi I, et al. Physical characteristics of low and medium energy collimators for 123I imaging and simultaneous dual-isotope imaging. Nucl Med Commun. 2003;24:1195–202.PubMedCrossRef
Metadata
Title
Validity of the mediastinum as a reference region to evaluate cardiac accumulation of iodine-123 metaiodobenzylguanidine
Authors
Yusuke Inoue
Yutaka Abe
Yuji Asano
Kei Kikuchi
Publication date
01-08-2015
Publisher
Springer Japan
Published in
Annals of Nuclear Medicine / Issue 7/2015
Print ISSN: 0914-7187
Electronic ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-015-0989-z

Other articles of this Issue 7/2015

Annals of Nuclear Medicine 7/2015 Go to the issue