Validating optimal function of the closed loop stimulation sensor with high right septal ventricular electrode placement in ‘ablate and pace’ patients

Authors: John Silberbauer, Paul S. G. Hong, Rick A. Veasey, Nadeem A. Maddekar, Wasing Taggu, Nikhil R. Patel, Guy W. Lloyd, Neil Sulke

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 1/2009

Login to get access

Abstract

Purpose

The study aim was to validate the closed loop stimulation (CLS) vs. accelerometer (ACC) rate-responsive sensors with electrodes placed in the right ventricular high septal (RVHS) or right ventricular apical (RVA) lead positions in patients following ‘ablate and pace’ therapy for persistent atrial fibrillation.

Methods

‘Ablate and pace’ patients were randomised to either RVHS or RVA electrode placement with a dual sensor device. A double-blind crossover study comparing CLS vs. ACC rate-response pacing modes was undertaken. Subjects undertook cardiopulmonary testing with constant workload light exercise followed by a ramp protocol in addition to activity of daily living assessments.

Results

Twenty subjects (14 male; age, 74 ± 8 years) were studied. Heart rate increase was greater from lying to sitting with ACC. With mental stress, heart rate increase was greater with CLS. Peak heart rates were similar for stair ascent and descent in ACC mode. With CLS mode, however, the peak heart rate was significantly lower for stair descent. There was no difference between modes in mean response time, oxygen deficit, peak VO2, VO2 at anaerobic threshold, peak heart rate, total exercise time and total workload. CLS function was equally optimal at both electrode sites.

Conclusions

CLS rate adaptive pacing is appropriate for ‘ablate and pace’ patients, and this sensor functions equally well using RVA or RVHS lead positions.
Literature
1.
go back to reference Lau, C. P. (1993). The combination of sensors and algorithms. In C. P. Lau (Ed.), Rate adaptive cardiac pacing: Single and dual chamber (pp. 213–227). Mount Kisco: Futura. Lau, C. P. (1993). The combination of sensors and algorithms. In C. P. Lau (Ed.), Rate adaptive cardiac pacing: Single and dual chamber (pp. 213–227). Mount Kisco: Futura.
2.
go back to reference Schaldach, M., Ebner, E., Hutten, H., von Knorre, G. H., Niederlag, W., Rentsch, W., et al. (1992). Right ventricular conductance to establish closed-loop pacing. European Heart Journal, 13(Suppl E), 104–112.PubMed Schaldach, M., Ebner, E., Hutten, H., von Knorre, G. H., Niederlag, W., Rentsch, W., et al. (1992). Right ventricular conductance to establish closed-loop pacing. European Heart Journal, 13(Suppl E), 104–112.PubMed
3.
go back to reference Osswald, S., Cron, T., Gradel, C., Hilti, P., Lippert, M., Strobel, J., et al. (2000). Closed-loop stimulation using intracardiac impedance as a sensor principle: correlation of right ventricular dP/dtmax and intracardiac impedance during dobutamine stress test. Pacing and Clinical Electrophysiology, 23(10 Pt 1), 1502–1508.PubMedCrossRef Osswald, S., Cron, T., Gradel, C., Hilti, P., Lippert, M., Strobel, J., et al. (2000). Closed-loop stimulation using intracardiac impedance as a sensor principle: correlation of right ventricular dP/dtmax and intracardiac impedance during dobutamine stress test. Pacing and Clinical Electrophysiology, 23(10 Pt 1), 1502–1508.PubMedCrossRef
4.
go back to reference Cook, L., Tomczak, C., Busse, E., Tsang, J., Wojcik, W., & Haennel, R. (2005). Impact of a right ventricular impedance sensor on the cardiovascular responses to exercise in pacemaker dependent patients. Indian Pacing Electrophysiol J, 5(3), 160–174.PubMed Cook, L., Tomczak, C., Busse, E., Tsang, J., Wojcik, W., & Haennel, R. (2005). Impact of a right ventricular impedance sensor on the cardiovascular responses to exercise in pacemaker dependent patients. Indian Pacing Electrophysiol J, 5(3), 160–174.PubMed
5.
go back to reference Freedman, R. A., Hopper, D. L., Mah, J., Hummel, J., & Wilkoff, B. L. (2001). Assessment of pacemaker chronotropic response: implementation of the Wilkoff mathematical model. Pacing and Clinical Electrophysiology, 24(12), 1748–1754.PubMedCrossRef Freedman, R. A., Hopper, D. L., Mah, J., Hummel, J., & Wilkoff, B. L. (2001). Assessment of pacemaker chronotropic response: implementation of the Wilkoff mathematical model. Pacing and Clinical Electrophysiology, 24(12), 1748–1754.PubMedCrossRef
6.
go back to reference Carmouche, D. G., Bubien, R. S., & Kay, G. N. (1998). The effect of maximum heart rate on oxygen kinetics and exercise performance at low and high workloads. Pacing and Clinical Electrophysiology, 21(4 Pt 1), 679–686.PubMedCrossRef Carmouche, D. G., Bubien, R. S., & Kay, G. N. (1998). The effect of maximum heart rate on oxygen kinetics and exercise performance at low and high workloads. Pacing and Clinical Electrophysiology, 21(4 Pt 1), 679–686.PubMedCrossRef
7.
go back to reference Lewalter, T., MacCarter, D., Jung, W., Bauer, T., Schimpf, R., Manz, M., et al. (1995). The “low intensity treadmill exercise” protocol for appropriate rate adaptive programming of minute ventilation controlled pacemakers. Pacing and Clinical Electrophysiology, 18(7), 1374–1387.PubMedCrossRef Lewalter, T., MacCarter, D., Jung, W., Bauer, T., Schimpf, R., Manz, M., et al. (1995). The “low intensity treadmill exercise” protocol for appropriate rate adaptive programming of minute ventilation controlled pacemakers. Pacing and Clinical Electrophysiology, 18(7), 1374–1387.PubMedCrossRef
8.
go back to reference Gibbons, R. J., Balady, G. J., Bricker, J. T., Chaitman, B. R., Fletcher, G. F., Froelicher, V. F., et al. (2002). ACC/AHA 2002 guideline update for exercise testing: summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee to Update the 1997 Exercise Testing Guidelines). Circulation, 106(14), 1883–1892.PubMedCrossRef Gibbons, R. J., Balady, G. J., Bricker, J. T., Chaitman, B. R., Fletcher, G. F., Froelicher, V. F., et al. (2002). ACC/AHA 2002 guideline update for exercise testing: summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee to Update the 1997 Exercise Testing Guidelines). Circulation, 106(14), 1883–1892.PubMedCrossRef
9.
go back to reference Wasserman, K., Hansen, J. E., Sue, D. Y., Whipp, B. J., & Casaburi, R. (1999). Principles of exercise testing and interpretation (3rd ed.). Baltimore: Lippincott Williams & Wilkins. Wasserman, K., Hansen, J. E., Sue, D. Y., Whipp, B. J., & Casaburi, R. (1999). Principles of exercise testing and interpretation (3rd ed.). Baltimore: Lippincott Williams & Wilkins.
10.
go back to reference Wasserman, K. (2002). Cardiopulmonary exercise testing and cardiovascular health. Armonk: Futura. Wasserman, K. (2002). Cardiopulmonary exercise testing and cardiovascular health. Armonk: Futura.
11.
go back to reference Beaver, W. L., Wasserman, K., & Whipp, B. J. (1986). A new method for detecting anaerobic threshold by gas exchange. Journal of Applied Physiology, 60(6), 2020–2027.PubMed Beaver, W. L., Wasserman, K., & Whipp, B. J. (1986). A new method for detecting anaerobic threshold by gas exchange. Journal of Applied Physiology, 60(6), 2020–2027.PubMed
12.
go back to reference Wasserman, K. (1996). Exercise gas exchange in heart disease. Armonk: Futura. Wasserman, K. (1996). Exercise gas exchange in heart disease. Armonk: Futura.
13.
go back to reference Franklin, B. A., Whaley, M. H., & Howley, E. T. (2000). ACSM’s guidelines for exercise testing and prescription (6th ed., pp. 300–312). Philadelphia, PA: Lippincott Williams & Wilkins. Franklin, B. A., Whaley, M. H., & Howley, E. T. (2000). ACSM’s guidelines for exercise testing and prescription (6th ed., pp. 300–312). Philadelphia, PA: Lippincott Williams & Wilkins.
14.
go back to reference Benditt, D. G., Mianulli, M., Lurie, K., Sakaguchi, S., & Adler, S. (1994). Multiple-sensor systems for physiologic cardiac pacing. Annals of Internal Medicine, 121(12), 960–968.PubMed Benditt, D. G., Mianulli, M., Lurie, K., Sakaguchi, S., & Adler, S. (1994). Multiple-sensor systems for physiologic cardiac pacing. Annals of Internal Medicine, 121(12), 960–968.PubMed
15.
go back to reference Dailey, S. M., Bubien, R. S., & Kay, G. N. (1994). Effect of chronotropic response pattern on oxygen kinetics. Pacing and Clinical Electrophysiology, 17(12 Pt 1), 2307–2314.PubMedCrossRef Dailey, S. M., Bubien, R. S., & Kay, G. N. (1994). Effect of chronotropic response pattern on oxygen kinetics. Pacing and Clinical Electrophysiology, 17(12 Pt 1), 2307–2314.PubMedCrossRef
16.
go back to reference Whipp, B. J. (1971). Rate constant for the kinetics of oxygen uptake during light exercise. Journal of Applied Physiology, 30(2), 261–263.PubMed Whipp, B. J. (1971). Rate constant for the kinetics of oxygen uptake during light exercise. Journal of Applied Physiology, 30(2), 261–263.PubMed
17.
go back to reference Schneider, D. A., Phillips, S. E., & Stoffolano, S. (1993). The simplified V-slope method of detecting the gas exchange threshold. Medicine and Science in Sports and Exercise, 25(10), 1180–1184.PubMed Schneider, D. A., Phillips, S. E., & Stoffolano, S. (1993). The simplified V-slope method of detecting the gas exchange threshold. Medicine and Science in Sports and Exercise, 25(10), 1180–1184.PubMed
18.
go back to reference Stroop, J. R. (1935). The basis of Ligon’s theory. American Journal of Psychology, 47, 499–504.CrossRef Stroop, J. R. (1935). The basis of Ligon’s theory. American Journal of Psychology, 47, 499–504.CrossRef
19.
go back to reference Rosenqvist, M., Isaaz, K., Botvinick, E. H., Dae, M. W., Cockrell, J., Abbott, J. A., et al. (1991). Relative importance of activation sequence compared to atrioventricular synchrony in left ventricular function. American Journal of Cardiology, 67(2), 148–156.PubMedCrossRef Rosenqvist, M., Isaaz, K., Botvinick, E. H., Dae, M. W., Cockrell, J., Abbott, J. A., et al. (1991). Relative importance of activation sequence compared to atrioventricular synchrony in left ventricular function. American Journal of Cardiology, 67(2), 148–156.PubMedCrossRef
20.
go back to reference Tantengco, M. V., Thomas, R. L., & Karpawich, P. P. (2001). Left ventricular dysfunction after long-term right ventricular apical pacing in the young. Journal of the American College of Cardiology, 37(8), 2093–2100.PubMedCrossRef Tantengco, M. V., Thomas, R. L., & Karpawich, P. P. (2001). Left ventricular dysfunction after long-term right ventricular apical pacing in the young. Journal of the American College of Cardiology, 37(8), 2093–2100.PubMedCrossRef
21.
go back to reference Wilkoff, B. L., Cook, J. R., Epstein, A. E., Greene, H. L., Hallstrom, A. P., Hsia, H., et al. (2002). Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial. JAMA, 288(24), 3115–3123.PubMedCrossRef Wilkoff, B. L., Cook, J. R., Epstein, A. E., Greene, H. L., Hallstrom, A. P., Hsia, H., et al. (2002). Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial. JAMA, 288(24), 3115–3123.PubMedCrossRef
22.
go back to reference Sweeney, M. O., Bank, A. J., Nsah, E., Koullick, M., Zeng, Q. C., Hettrick, D., et al. (2007). Minimizing ventricular pacing to reduce atrial fibrillation in sinus-node disease. New England Journal of Medicine, 357(10), 1000–1008.PubMedCrossRef Sweeney, M. O., Bank, A. J., Nsah, E., Koullick, M., Zeng, Q. C., Hettrick, D., et al. (2007). Minimizing ventricular pacing to reduce atrial fibrillation in sinus-node disease. New England Journal of Medicine, 357(10), 1000–1008.PubMedCrossRef
23.
go back to reference Brady, P. A., & Hammill, S. C. (2003). Ventricular-based pacing: one site fits all? Journal of Cardiovascular Electrophysiology, 14(11), 1187–1188.PubMedCrossRef Brady, P. A., & Hammill, S. C. (2003). Ventricular-based pacing: one site fits all? Journal of Cardiovascular Electrophysiology, 14(11), 1187–1188.PubMedCrossRef
24.
go back to reference McGavigan, A. D., Roberts-Thomson, K. C., Hillock, R. J., Stevenson, I. H., & Mond, H. G. (2006). Right ventricular outflow tract pacing: radiographic and electrocardiographic correlates of lead position. Pacing and Clinical Electrophysiology, 29(10), 1063–1068.PubMedCrossRef McGavigan, A. D., Roberts-Thomson, K. C., Hillock, R. J., Stevenson, I. H., & Mond, H. G. (2006). Right ventricular outflow tract pacing: radiographic and electrocardiographic correlates of lead position. Pacing and Clinical Electrophysiology, 29(10), 1063–1068.PubMedCrossRef
25.
go back to reference Sweeney, M. O., & Prinzen, F. W. (2006). A new paradigm for physiologic ventricular pacing. Journal of the American College of Cardiology, 47(2), 282–288.PubMedCrossRef Sweeney, M. O., & Prinzen, F. W. (2006). A new paradigm for physiologic ventricular pacing. Journal of the American College of Cardiology, 47(2), 282–288.PubMedCrossRef
26.
go back to reference Victor, F., Leclercq, C., Mabo, P., Pavin, D., Deviller, A., de Place, C., et al. (1999). Optimal right ventricular pacing site in chronically implanted patients: a prospective randomized crossover comparison of apical and outflow tract pacing. Journal of the American College of Cardiology, 33(2), 311–316.PubMedCrossRef Victor, F., Leclercq, C., Mabo, P., Pavin, D., Deviller, A., de Place, C., et al. (1999). Optimal right ventricular pacing site in chronically implanted patients: a prospective randomized crossover comparison of apical and outflow tract pacing. Journal of the American College of Cardiology, 33(2), 311–316.PubMedCrossRef
27.
go back to reference Mera, F., DeLurgio, D. B., Patterson, R. E., Merlino, J. D., Wade, M. E., & Leon, A. R. (1999). A comparison of ventricular function during high right ventricular septal and apical pacing after his-bundle ablation for refractory atrial fibrillation. Pacing and Clinical Electrophysiology, 22(8), 1234–1239.PubMedCrossRef Mera, F., DeLurgio, D. B., Patterson, R. E., Merlino, J. D., Wade, M. E., & Leon, A. R. (1999). A comparison of ventricular function during high right ventricular septal and apical pacing after his-bundle ablation for refractory atrial fibrillation. Pacing and Clinical Electrophysiology, 22(8), 1234–1239.PubMedCrossRef
28.
go back to reference Tse, H. F., Yu, C., Wong, K. K., Tsang, V., Leung, Y. L., Ho, W. Y., et al. (2002). Functional abnormalities in patients with permanent right ventricular pacing: the effect of sites of electrical stimulation. Journal of the American College of Cardiology, 40(8), 1451–1458.PubMedCrossRef Tse, H. F., Yu, C., Wong, K. K., Tsang, V., Leung, Y. L., Ho, W. Y., et al. (2002). Functional abnormalities in patients with permanent right ventricular pacing: the effect of sites of electrical stimulation. Journal of the American College of Cardiology, 40(8), 1451–1458.PubMedCrossRef
29.
go back to reference Bourke, J. P., Hawkins, T., Keavey, P., Tynan, M., Jamieson, S., Behulova, R., et al. (2002). Evolution of ventricular function during permanent pacing from either right ventricular apex or outflow tract following AV-junctional ablation for atrial fibrillation. Europace, 4(3), 219–228.PubMedCrossRef Bourke, J. P., Hawkins, T., Keavey, P., Tynan, M., Jamieson, S., Behulova, R., et al. (2002). Evolution of ventricular function during permanent pacing from either right ventricular apex or outflow tract following AV-junctional ablation for atrial fibrillation. Europace, 4(3), 219–228.PubMedCrossRef
30.
go back to reference Stambler, B. S., Ellenbogen, K., Zhang, X., Porter, T. R., Xie, F., Malik, R., et al. (2003). Right ventricular outflow versus apical pacing in pacemaker patients with congestive heart failure and atrial fibrillation. Journal of Cardiovascular Electrophysiology, 14(11), 1180–1186.PubMedCrossRef Stambler, B. S., Ellenbogen, K., Zhang, X., Porter, T. R., Xie, F., Malik, R., et al. (2003). Right ventricular outflow versus apical pacing in pacemaker patients with congestive heart failure and atrial fibrillation. Journal of Cardiovascular Electrophysiology, 14(11), 1180–1186.PubMedCrossRef
31.
go back to reference Alt, E., Combs, W., Willhaus, R., Condie, C., Bambl, E., Fotuhi, P., et al. (1998). A comparative study of activity and dual sensor: activity and minute ventilation pacing responses to ascending and descending stairs. Pacing and Clinical Electrophysiology, 21(10), 1862–1868.PubMedCrossRef Alt, E., Combs, W., Willhaus, R., Condie, C., Bambl, E., Fotuhi, P., et al. (1998). A comparative study of activity and dual sensor: activity and minute ventilation pacing responses to ascending and descending stairs. Pacing and Clinical Electrophysiology, 21(10), 1862–1868.PubMedCrossRef
32.
go back to reference Witte, J., Reibis, R., Pichlmaier, A. M., Ebner, E., Malinowski, K., Rödiger, W., et al. (1996). ANS-controlled rate-adaptive pacing - a clinical evaluation. European Journal of Cardiac pacing and Electrophysiology, 6, 53–59. Witte, J., Reibis, R., Pichlmaier, A. M., Ebner, E., Malinowski, K., Rödiger, W., et al. (1996). ANS-controlled rate-adaptive pacing - a clinical evaluation. European Journal of Cardiac pacing and Electrophysiology, 6, 53–59.
33.
go back to reference Lewalter, T., Rickli, H., MacCarter, D., Jung, W., Schimpf, R., Schwartze, P., et al. (1997). Oxygen uptake kinetics during low intensity exercise: relevance for rate adaptive pacemaker programming. Heart, 77(2), 168–172.PubMed Lewalter, T., Rickli, H., MacCarter, D., Jung, W., Schimpf, R., Schwartze, P., et al. (1997). Oxygen uptake kinetics during low intensity exercise: relevance for rate adaptive pacemaker programming. Heart, 77(2), 168–172.PubMed
34.
go back to reference Lau, C. P., Rushby, J., Leigh-Jones, M., Tam, C. Y., Poloniecki, J., Ingram, A., et al. (1989). Symptomatology and quality of life in patients with rate-responsive pacemakers: a double-blind, randomized, crossover study. Clinical Cardiology, 12(9), 505–512.PubMedCrossRef Lau, C. P., Rushby, J., Leigh-Jones, M., Tam, C. Y., Poloniecki, J., Ingram, A., et al. (1989). Symptomatology and quality of life in patients with rate-responsive pacemakers: a double-blind, randomized, crossover study. Clinical Cardiology, 12(9), 505–512.PubMedCrossRef
35.
go back to reference Sulke, N., Chambers, J., Dritsas, A., & Sowton, E. (1991). A randomized double-blind crossover comparison of four rate-responsive pacing modes. Journal of the American College of Cardiology, 17(3), 696–706.PubMedCrossRef Sulke, N., Chambers, J., Dritsas, A., & Sowton, E. (1991). A randomized double-blind crossover comparison of four rate-responsive pacing modes. Journal of the American College of Cardiology, 17(3), 696–706.PubMedCrossRef
36.
go back to reference Benditt, D. G., Mianulli, M., Fetter, J., Benson, D. W., Jr., Dunnigan, A., Molina, E., et al. (1987). Single-chamber cardiac pacing with activity-initiated chronotropic response: evaluation by cardiopulmonary exercise testing. Circulation, 75(1), 184–191.PubMed Benditt, D. G., Mianulli, M., Fetter, J., Benson, D. W., Jr., Dunnigan, A., Molina, E., et al. (1987). Single-chamber cardiac pacing with activity-initiated chronotropic response: evaluation by cardiopulmonary exercise testing. Circulation, 75(1), 184–191.PubMed
37.
go back to reference Faerestrand, S., Breivik, K., & Ohm, O. J. (1987). Assessment of the work capacity and relationship between rate response and exercise tolerance associated with activity-sensing rate-responsive ventricular pacing. Pacing and Clinical Electrophysiology, 10(6), 1277–1290.PubMedCrossRef Faerestrand, S., Breivik, K., & Ohm, O. J. (1987). Assessment of the work capacity and relationship between rate response and exercise tolerance associated with activity-sensing rate-responsive ventricular pacing. Pacing and Clinical Electrophysiology, 10(6), 1277–1290.PubMedCrossRef
38.
go back to reference Kay, G. N., Bubien, R. S., Epstein, A. E., & Plumb, V. J. (1989). Rate-modulated cardiac pacing based on transthoracic impedance measurements of minute ventilation: correlation with exercise gas exchange. Journal of the American College of Cardiology, 14(5), 1283–1289.PubMedCrossRef Kay, G. N., Bubien, R. S., Epstein, A. E., & Plumb, V. J. (1989). Rate-modulated cardiac pacing based on transthoracic impedance measurements of minute ventilation: correlation with exercise gas exchange. Journal of the American College of Cardiology, 14(5), 1283–1289.PubMedCrossRef
Metadata
Title
Validating optimal function of the closed loop stimulation sensor with high right septal ventricular electrode placement in ‘ablate and pace’ patients
Authors
John Silberbauer
Paul S. G. Hong
Rick A. Veasey
Nadeem A. Maddekar
Wasing Taggu
Nikhil R. Patel
Guy W. Lloyd
Neil Sulke
Publication date
01-10-2009
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 1/2009
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-009-9426-3

Other articles of this Issue 1/2009

Journal of Interventional Cardiac Electrophysiology 1/2009 Go to the issue