01-03-2014 | Original Article

Use of a vessel sealing system versus conventional electrocautery for lung parenchymal resection: a comparison of the clinicopathological outcomes in porcine lungs

Authors: Seiichiro Sugimoto, Shinichi Toyooka, Norichika Iga, Masashi Furukawa, Ryujiro Sugimoto, Kazuhiko Shien, Hitoshi Nishikawa, Junichi Soh, Masaomi Yamane, Takahiro Oto, Shinichiro Miyoshi

Published in: Surgery Today | Issue 3/2014

Login to get access

Abstract

Purpose

LigaSure, a vessel sealing system, has been shown to have excellent hemostatic properties; however, its use for lung parenchymal resection has been limited. We herein examined the hemostatic properties and potential for inducing histological lung injury of the LigaSure system in non-anatomic pulmonary resection to estimate the feasibility of its clinical application.

Methods

Non-anatomic pulmonary wedge resections of the right cranial, middle, and caudal lobes were performed in four pigs using the LigaSure system (Group A) or electrocautery (Group B). In each resection, the resection time, blood loss, and weight of the resected lung were measured. The thermal effect on the lung tissue was examined by means of intraoperative thermography and histology.

Results

A total of 12 lung wedge resections were performed in each group. For an equivalent length of operation and weight of the resected lung parenchyma, Group A showed significantly lower blood loss and lower maximum and minimum temperatures of the lung tissue, as assessed by thermography, than Group B. The degree of thermal injury as estimated by a histological examination was lower in Group A than in Group B.

Conclusion

Our study suggests that the LigaSure system may be superior to conventional electrocautery, indicating its clinical usefulness for non-anatomic pulmonary resection.
Literature
1.
go back to reference Oey I, Waller DA. Metalloptysis: a late complication of lung volume reduction surgery. Ann Thorac Surg. 2001;71:1694–5.PubMedCrossRef Oey I, Waller DA. Metalloptysis: a late complication of lung volume reduction surgery. Ann Thorac Surg. 2001;71:1694–5.PubMedCrossRef
2.
go back to reference Entezari K, Hoffmann P, Goris M, Peltier A, Van Velthoven R. A review of currently available vessel sealing systems. Minim Invasive Ther Allied Technol. 2007;16:52–7.PubMedCrossRef Entezari K, Hoffmann P, Goris M, Peltier A, Van Velthoven R. A review of currently available vessel sealing systems. Minim Invasive Ther Allied Technol. 2007;16:52–7.PubMedCrossRef
3.
go back to reference Shigemura N, Akashi A, Nakagiri T, Ohta M, Matsuda H. A new tissue-sealing technique using the ligasure system for nonanatomical pulmonary resection: preliminary results of sutureless and stapleless thoracoscopic surgery. Ann Thorac Surg. 2004;77:1415–8.PubMedCrossRef Shigemura N, Akashi A, Nakagiri T, Ohta M, Matsuda H. A new tissue-sealing technique using the ligasure system for nonanatomical pulmonary resection: preliminary results of sutureless and stapleless thoracoscopic surgery. Ann Thorac Surg. 2004;77:1415–8.PubMedCrossRef
4.
go back to reference Santini M, Vicidomini G, Baldi A, Gallo G, Laperuta P, Busiello L, et al. Use of an electrothermal bipolar tissue sealing system in lung surgery. Eur J Cardiothorac Surg. 2006;29:226–30.PubMedCrossRef Santini M, Vicidomini G, Baldi A, Gallo G, Laperuta P, Busiello L, et al. Use of an electrothermal bipolar tissue sealing system in lung surgery. Eur J Cardiothorac Surg. 2006;29:226–30.PubMedCrossRef
5.
go back to reference Kovacs O, Szanto Z, Krasznai G, Herr G. Comparing bipolar electrothermal device and endostapler in endoscopic lung wedge resection. Interact Cardiovasc Thorac Surg. 2009;9:11–4.PubMedCrossRef Kovacs O, Szanto Z, Krasznai G, Herr G. Comparing bipolar electrothermal device and endostapler in endoscopic lung wedge resection. Interact Cardiovasc Thorac Surg. 2009;9:11–4.PubMedCrossRef
6.
go back to reference Schuchert MJ, Abbas G, Pettiford BL, Luketich JD, Landreneau RJ. Preliminary results of anatomic lung resection using energy-based tissue and vessel coagulative fusion technology. J Thorac Cardiovasc Surg. 2010;140:1168–73.PubMedCrossRef Schuchert MJ, Abbas G, Pettiford BL, Luketich JD, Landreneau RJ. Preliminary results of anatomic lung resection using energy-based tissue and vessel coagulative fusion technology. J Thorac Cardiovasc Surg. 2010;140:1168–73.PubMedCrossRef
7.
go back to reference Kennedy JS, Stranahan PL, Taylor KD, Chandler JG. High-burst-strength, feedback-controlled bipolar vessel sealing. Surg Endosc. 1998;12:876–8.PubMedCrossRef Kennedy JS, Stranahan PL, Taylor KD, Chandler JG. High-burst-strength, feedback-controlled bipolar vessel sealing. Surg Endosc. 1998;12:876–8.PubMedCrossRef
8.
go back to reference Tirabassi M. Quantitation of lung sealing in the survival swine model. J Pediatr Surg. 2004;39:387–90.PubMedCrossRef Tirabassi M. Quantitation of lung sealing in the survival swine model. J Pediatr Surg. 2004;39:387–90.PubMedCrossRef
9.
go back to reference Landman J, Kerbl K, Rehman J, Andreoni C, Humphrey PA, Collyer W, et al. Evaluation of a vessel sealing system, bipolar electrosurgery, harmonic scalpel, titanium clips, endoscopic gastrointestinal anastomosis vascular staples and sutures for arterial and venous ligation in a porcine model. J Urol. 2003;169:697–700.PubMedCrossRef Landman J, Kerbl K, Rehman J, Andreoni C, Humphrey PA, Collyer W, et al. Evaluation of a vessel sealing system, bipolar electrosurgery, harmonic scalpel, titanium clips, endoscopic gastrointestinal anastomosis vascular staples and sutures for arterial and venous ligation in a porcine model. J Urol. 2003;169:697–700.PubMedCrossRef
10.
go back to reference Huang W, Yen RT, McLaurine M, Bledsoe G. Morphometry of the human pulmonary vasculature. J Appl Physiol. 1996;81:2123–33.PubMed Huang W, Yen RT, McLaurine M, Bledsoe G. Morphometry of the human pulmonary vasculature. J Appl Physiol. 1996;81:2123–33.PubMed
11.
go back to reference Macario A, Dexter F, Sypal J, Cosgriff N, Heniford BT. Operative time and other outcomes of the electrothermal bipolar vessel sealing system (LigaSure) versus other methods for surgical hemostasis: a meta-analysis. Surg Innov. 2008;15:284–91.PubMedCrossRef Macario A, Dexter F, Sypal J, Cosgriff N, Heniford BT. Operative time and other outcomes of the electrothermal bipolar vessel sealing system (LigaSure) versus other methods for surgical hemostasis: a meta-analysis. Surg Innov. 2008;15:284–91.PubMedCrossRef
12.
go back to reference Lamberton GR, Hsi RS, Jin DH, Lindler TU, Jellison FC, Baldwin DD. Prospective comparison of four laparoscopic vessel ligation devices. J Endourol. 2008;22:2307–12.PubMedCrossRef Lamberton GR, Hsi RS, Jin DH, Lindler TU, Jellison FC, Baldwin DD. Prospective comparison of four laparoscopic vessel ligation devices. J Endourol. 2008;22:2307–12.PubMedCrossRef
13.
go back to reference Nagamatsu Y, Iwasaki Y, Kashihara M, Nishi T, Yoshiyama K, Yamana H, et al. Selection of pulmonary resection procedures to reduce postoperative complications in 200 patients. Surg Today. 2011;41:780–6.PubMedCrossRef Nagamatsu Y, Iwasaki Y, Kashihara M, Nishi T, Yoshiyama K, Yamana H, et al. Selection of pulmonary resection procedures to reduce postoperative complications in 200 patients. Surg Today. 2011;41:780–6.PubMedCrossRef
14.
go back to reference Okada M, Mimura T, Ikegaki J, Katoh H, Itoh H, Tsubota N. A novel video-assisted anatomic segmentectomy technique: selective segmental inflation via bronchofiberoptic jet followed by cautery cutting. J Thorac Cardiovasc Surg. 2007;133:753–8.PubMedCrossRef Okada M, Mimura T, Ikegaki J, Katoh H, Itoh H, Tsubota N. A novel video-assisted anatomic segmentectomy technique: selective segmental inflation via bronchofiberoptic jet followed by cautery cutting. J Thorac Cardiovasc Surg. 2007;133:753–8.PubMedCrossRef
Metadata
Title
Use of a vessel sealing system versus conventional electrocautery for lung parenchymal resection: a comparison of the clinicopathological outcomes in porcine lungs
Authors
Seiichiro Sugimoto
Shinichi Toyooka
Norichika Iga
Masashi Furukawa
Ryujiro Sugimoto
Kazuhiko Shien
Hitoshi Nishikawa
Junichi Soh
Masaomi Yamane
Takahiro Oto
Shinichiro Miyoshi
Publication date
01-03-2014
Publisher
Springer Japan
Published in
Surgery Today / Issue 3/2014
Print ISSN: 0941-1291
Electronic ISSN: 1436-2813
DOI
https://doi.org/10.1007/s00595-013-0545-1

Other articles of this Issue 3/2014

Surgery Today 3/2014 Go to the issue

Letter to the Editor

Author’s response