Unstable versus stable uncemented femoral stems: a radiological study of periprosthetic bone changes in two types of uncemented stems with different concepts of fixation

Authors: Henrik Bodén, Per Adolphson, Michael Öberg

Published in: Archives of Orthopaedic and Trauma Surgery | Issue 6/2004

Login to get access

Abstract

Introduction

Uncemented stems in total hip arthroplasty (THA) are used increasingly often because they are believed to offer a reliable long-term fixation. However, periprosthetic bone remodelling has been a worrying issue. A proximal demineralization has been noted in femurs with well-fixed stems, and it has been explained as by-passing of mechanical forces along the fixed implant (stress-shielding). Aseptic loosening has been a major problem in several uncemented series with earlier designs. The objective for this study was to investigate how the host bone adapts to a loose stem compared with a well-fixed stem after a long time.

Materials and methods

An investigation with dual-energy X-ray absorptiometry (DEXA), scintimetry and radiological assessment was carried out in 20 patients 8 years after a THA for arthrosis with two different uncemented stems. Ten patients received a stem coated with polytetrafluoroethylene (Anaform); all prostheses showed migration and were considered unstable. Ten patients received a hydroxyapatite-coated stem (Bi-Metric); no prosthesis migrated.

Results

Different remodelling patterns were seen. In the unstable group, the periprosthetic bone mineral density (BMD) was significantly reduced along the entire stem, while in the stable group only proximal bone loss was seen. The scintigraphic uptake was increased under the stem tip in both groups, and among unstable stems uptake was also increased in the calcar region.

Conclusion

The assessment of periprosthetic bone remodelling after uncemented THA with long-term observation shows a different host-bone response in stable versus unstable femoral implants. Prior to a femoral revision, measurement of the BMD could be beneficial; it may guide the surgeon when deciding which surgical technique to use.
Literature
1.
go back to reference Ang KC, Das De S, Goh JC, Low SL, Bose K (1997) Periprosthetic bone remodelling after cementless total hip replacement. A prospective comparison of two different implant designs. J Bone Joint Surg Br 79:675–679 Ang KC, Das De S, Goh JC, Low SL, Bose K (1997) Periprosthetic bone remodelling after cementless total hip replacement. A prospective comparison of two different implant designs. J Bone Joint Surg Br 79:675–679
2.
go back to reference Bourne RB, Rorabeck CH, Ghazal ME, Lee MH (1994) Pain in the thigh following total hip replacement with a porous-coated anatomic prosthesis for osteoarthrosis. A five-year follow-up study. J Bone Joint Surg Am 76:1464–1470PubMed Bourne RB, Rorabeck CH, Ghazal ME, Lee MH (1994) Pain in the thigh following total hip replacement with a porous-coated anatomic prosthesis for osteoarthrosis. A five-year follow-up study. J Bone Joint Surg Am 76:1464–1470PubMed
3.
go back to reference Bryan JM, Sumner DR, Hurwitz DE, Tompkins GS, Andriacchi TP, Galante JO (1996) Altered load history affects periprosthetic bone loss following cementless total hip arthroplasty. J Orthop Res 14:762–768PubMed Bryan JM, Sumner DR, Hurwitz DE, Tompkins GS, Andriacchi TP, Galante JO (1996) Altered load history affects periprosthetic bone loss following cementless total hip arthroplasty. J Orthop Res 14:762–768PubMed
4.
go back to reference Charnley J, Cupic Z (1973) The nine and ten year results of the low-friction arthroplasty of the hip. Clin Orthop 95:9–25PubMed Charnley J, Cupic Z (1973) The nine and ten year results of the low-friction arthroplasty of the hip. Clin Orthop 95:9–25PubMed
5.
go back to reference D’Antonio JA, Capello WN, Crothers OD, Jaffe WL, Manley MT (1992) Early clinical experience with hydroxyapatite-coated femoral implants. J Bone Joint Surg Am 74:995–1008PubMed D’Antonio JA, Capello WN, Crothers OD, Jaffe WL, Manley MT (1992) Early clinical experience with hydroxyapatite-coated femoral implants. J Bone Joint Surg Am 74:995–1008PubMed
6.
go back to reference Dorr LD, Lewonowski K, Lucero M, Harris M, Wan Z (1997) Failure mechanisms of anatomic porous replacement I cementless total hip replacement. Clin Orthop 334:157–167CrossRefPubMed Dorr LD, Lewonowski K, Lucero M, Harris M, Wan Z (1997) Failure mechanisms of anatomic porous replacement I cementless total hip replacement. Clin Orthop 334:157–167CrossRefPubMed
7.
go back to reference Dorr LD, Takei GK, Conaty JP (1983) Total hip arthroplasties in patients less than forty-five years old. J Bone Joint Surg Am 65:474–479PubMed Dorr LD, Takei GK, Conaty JP (1983) Total hip arthroplasties in patients less than forty-five years old. J Bone Joint Surg Am 65:474–479PubMed
8.
go back to reference Engh CA, Bobyn JD (1988) The influence of stem size and extent of porous coating on femoral bone resorption after primary cementless hip arthroplasty. Clin Orthop 231:7–28PubMed Engh CA, Bobyn JD (1988) The influence of stem size and extent of porous coating on femoral bone resorption after primary cementless hip arthroplasty. Clin Orthop 231:7–28PubMed
9.
go back to reference Engh CA, Bobyn JD, Glassman AH (1987) Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results. J Bone Joint Surg Br 69:45–55PubMed Engh CA, Bobyn JD, Glassman AH (1987) Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results. J Bone Joint Surg Br 69:45–55PubMed
10.
go back to reference Engh CA, Massin P, Suthers KE (1990) Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. Clin Orthop 257:107–128PubMed Engh CA, Massin P, Suthers KE (1990) Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. Clin Orthop 257:107–128PubMed
11.
go back to reference Engh CA, McGovern TF, Bobyn JD, Harris WH (1992) A quantitative evaluation of periprosthetic bone-remodeling after cementless total hip arthroplasty. J Bone Joint Surg Am 74:1009–1020PubMed Engh CA, McGovern TF, Bobyn JD, Harris WH (1992) A quantitative evaluation of periprosthetic bone-remodeling after cementless total hip arthroplasty. J Bone Joint Surg Am 74:1009–1020PubMed
12.
go back to reference Engh CA, McGovern TF, Schmidt LM (1993) Roentgenographic densitometry of bone adjacent to a femoral prosthesis. Clin Orthop 292:177–190PubMed Engh CA, McGovern TF, Schmidt LM (1993) Roentgenographic densitometry of bone adjacent to a femoral prosthesis. Clin Orthop 292:177–190PubMed
13.
go back to reference Garcia-Cimbrelo E, Cruz-Pardos A, Cordero J, Sanchez-Sotelo J (2000) Low-friction arthroplasty in patients younger than 40 years old: 20- to 25-year results. J Arthroplasty 15:825–832CrossRefPubMed Garcia-Cimbrelo E, Cruz-Pardos A, Cordero J, Sanchez-Sotelo J (2000) Low-friction arthroplasty in patients younger than 40 years old: 20- to 25-year results. J Arthroplasty 15:825–832CrossRefPubMed
14.
go back to reference Geesink RG, De Groot K, Klein CP (1987) Chemical implant fixation using hydroxyl-apatite coatings: the development of a human total hip prosthesis for chemical fixation to bone using hydroxyl-apatite coatings on titanium substrates. Clin Orthop 225:147–170PubMed Geesink RG, De Groot K, Klein CP (1987) Chemical implant fixation using hydroxyl-apatite coatings: the development of a human total hip prosthesis for chemical fixation to bone using hydroxyl-apatite coatings on titanium substrates. Clin Orthop 225:147–170PubMed
15.
go back to reference Geesink RG, Hoefnagels NH (1995) Six-year results of hydroxyapatite-coated total hip replacement. J Bone Joint Surg Br 77:534–547PubMed Geesink RG, Hoefnagels NH (1995) Six-year results of hydroxyapatite-coated total hip replacement. J Bone Joint Surg Br 77:534–547PubMed
16.
go back to reference Gruen TA, McNeice GM, Amstutz HC (1979) ‘Modes of failure’ of cemented stem-type femoral components. A radiographic analysis of loosening. Clin Orthop 141:17–27PubMed Gruen TA, McNeice GM, Amstutz HC (1979) ‘Modes of failure’ of cemented stem-type femoral components. A radiographic analysis of loosening. Clin Orthop 141:17–27PubMed
17.
go back to reference Halstead A, Jones CW, Rawlings RD (1979) A study of the reaction of human tissue to proplast. J Biomed Mater Res 13:121–134PubMed Halstead A, Jones CW, Rawlings RD (1979) A study of the reaction of human tissue to proplast. J Biomed Mater Res 13:121–134PubMed
18.
go back to reference Harris WH (1969) Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty—an end-result study using a new method of result evaluation. J Bone Joint Surg Am 51:737–755PubMed Harris WH (1969) Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty—an end-result study using a new method of result evaluation. J Bone Joint Surg Am 51:737–755PubMed
19.
go back to reference Havelin LI, Engeseater LB, Espehaug B, Furnes O, Lie SA, Vollset SE (2000) The Norwegian Arthroplasty Register: 11 years and 73,000 arthroplasties. Acta Orthop Scand 71:337–353CrossRefPubMed Havelin LI, Engeseater LB, Espehaug B, Furnes O, Lie SA, Vollset SE (2000) The Norwegian Arthroplasty Register: 11 years and 73,000 arthroplasties. Acta Orthop Scand 71:337–353CrossRefPubMed
20.
go back to reference Hedley AK, Gruen TAW, Borden LS, Hungerford DS, Habermann E, Kenna RV (1987) Two-year follow-up of the PCA noncemented total hip replacement. In: Brand RA (ed) The hip. Proceedings of the 14th Open Scientific Meeting of The Hip Society. CV Mosby, St. Louis, p 225 Hedley AK, Gruen TAW, Borden LS, Hungerford DS, Habermann E, Kenna RV (1987) Two-year follow-up of the PCA noncemented total hip replacement. In: Brand RA (ed) The hip. Proceedings of the 14th Open Scientific Meeting of The Hip Society. CV Mosby, St. Louis, p 225
21.
go back to reference Herberts P, Malchau H (2000) Long-term registration has improved the quality of hip replacement: a review of the Swedish THR Register comparing 160,000 cases. Acta Orthop Scand 71:111–121PubMed Herberts P, Malchau H (2000) Long-term registration has improved the quality of hip replacement: a review of the Swedish THR Register comparing 160,000 cases. Acta Orthop Scand 71:111–121PubMed
22.
go back to reference Homsy CA (2000) Soft porous PTFE-composite alloplasts: tissue-bonding characteristics. J Endourol 14:25–32PubMed Homsy CA (2000) Soft porous PTFE-composite alloplasts: tissue-bonding characteristics. J Endourol 14:25–32PubMed
23.
go back to reference Homsy CA, Cain TE, Kessler FB, Anderson MS, King JW (1972) Porous implant systems for prosthesis stabilization. Clin Orthop 89:220–235PubMed Homsy CA, Cain TE, Kessler FB, Anderson MS, King JW (1972) Porous implant systems for prosthesis stabilization. Clin Orthop 89:220–235PubMed
24.
go back to reference Jacobs JJ, Sumner DR, Galante JO (1993) Mechanisms of bone loss associated with total hip replacement. Orthop Clin North Am 24:583–590PubMed Jacobs JJ, Sumner DR, Galante JO (1993) Mechanisms of bone loss associated with total hip replacement. Orthop Clin North Am 24:583–590PubMed
25.
go back to reference Jacobsen S, Jensen FK, Poulsen K, Sturup J, Retpen JB (2003) Good performance of a titanium femoral component in cementless hip arthroplasty in younger patients: 97 arthroplasties followed for 5–11 years. Acta Orthop Scand 74:375–379PubMed Jacobsen S, Jensen FK, Poulsen K, Sturup J, Retpen JB (2003) Good performance of a titanium femoral component in cementless hip arthroplasty in younger patients: 97 arthroplasties followed for 5–11 years. Acta Orthop Scand 74:375–379PubMed
26.
go back to reference Jacobsson SA, Djerf K, Gillquist J, Hammerby S, Ivarsson I (1993) A prospective comparison of Butel and PCA hip arthroplasty. J Bone Joint Surg Br 75:624–629PubMed Jacobsson SA, Djerf K, Gillquist J, Hammerby S, Ivarsson I (1993) A prospective comparison of Butel and PCA hip arthroplasty. J Bone Joint Surg Br 75:624–629PubMed
27.
go back to reference Jacobsson SA, Djerf K, Gillquist J, Svedberg J (1994) Tc-scintimetry in 56 cementless hip arthroplasties. A prospective, randomized comparison of two femoral components. Acta Orthop Scand 65:418–423PubMed Jacobsson SA, Djerf K, Gillquist J, Svedberg J (1994) Tc-scintimetry in 56 cementless hip arthroplasties. A prospective, randomized comparison of two femoral components. Acta Orthop Scand 65:418–423PubMed
28.
go back to reference Joshi AB, Porter ML, Trail IA, Hunt LP, Murphy JC, Hardinge K (1993) Long-term results of Charnley low-friction arthroplasty in young patients. J Bone Joint Surg Br 75:616–623PubMed Joshi AB, Porter ML, Trail IA, Hunt LP, Murphy JC, Hardinge K (1993) Long-term results of Charnley low-friction arthroplasty in young patients. J Bone Joint Surg Br 75:616–623PubMed
29.
go back to reference Kavanagh BF, Dewitz MA, Ilstrup DM, Stauffer RN, Coventry MB (1989) Charnley total hip arthroplasty with cement: Fifteen-year results. J Bone Joint Surg Am 71:1496–1503PubMed Kavanagh BF, Dewitz MA, Ilstrup DM, Stauffer RN, Coventry MB (1989) Charnley total hip arthroplasty with cement: Fifteen-year results. J Bone Joint Surg Am 71:1496–1503PubMed
30.
go back to reference Keet GG, Runne WC (1989) The Anaform endoprosthesis: a proplast-coated femoral endoprosthesis. Orthopedics 12:1185–1190PubMed Keet GG, Runne WC (1989) The Anaform endoprosthesis: a proplast-coated femoral endoprosthesis. Orthopedics 12:1185–1190PubMed
31.
go back to reference Khalily C, Whiteside LA (1998) Predictive value of early radiographic findings in cementless total hip arthroplasty femoral components: an 8- to 12-year follow-up. J Arthroplasty 13:768–773PubMed Khalily C, Whiteside LA (1998) Predictive value of early radiographic findings in cementless total hip arthroplasty femoral components: an 8- to 12-year follow-up. J Arthroplasty 13:768–773PubMed
32.
go back to reference Kilgus DJ, Shimaoka EE, Tipton JS, Eberle RW (1993) Dual-energy X-ray absorptiometry measurement of bone mineral density around porous-coated cementless femoral implants. Methods and preliminary results. J Bone Joint Surg Br 75:279–287 Kilgus DJ, Shimaoka EE, Tipton JS, Eberle RW (1993) Dual-energy X-ray absorptiometry measurement of bone mineral density around porous-coated cementless femoral implants. Methods and preliminary results. J Bone Joint Surg Br 75:279–287
33.
go back to reference Kim YH, Kim VE (1992) Results of the Harris-Galante cementless hip prosthesis. J Bone Joint Surg Br 74:83–87PubMed Kim YH, Kim VE (1992) Results of the Harris-Galante cementless hip prosthesis. J Bone Joint Surg Br 74:83–87PubMed
34.
go back to reference Kim YH, Kim VE (1993) Early migration of uncemented porous coated anatomic femoral component related to aseptic loosening. Clin Orthop 295:146–155PubMed Kim YH, Kim VE (1993) Early migration of uncemented porous coated anatomic femoral component related to aseptic loosening. Clin Orthop 295:146–155PubMed
35.
go back to reference Kiratli BJ, Checovich MM, McBeath AA, Wilson MA, Heiner JP (1996) Measurement of bone mineral density by dual-energy X-ray absorptiometry in patients with the Wisconsin hip, an uncemented femoral stem. J Arthroplasty 11:184–193PubMed Kiratli BJ, Checovich MM, McBeath AA, Wilson MA, Heiner JP (1996) Measurement of bone mineral density by dual-energy X-ray absorptiometry in patients with the Wisconsin hip, an uncemented femoral stem. J Arthroplasty 11:184–193PubMed
36.
go back to reference Kiratli BJ, Heiner JP, McBeath AA, Wilson MA (1992) Determination of bone mineral density by dual X-ray absorptiometry in patients with uncemented total hip arthroplasty. J Orthop Res 10:836–844PubMed Kiratli BJ, Heiner JP, McBeath AA, Wilson MA (1992) Determination of bone mineral density by dual X-ray absorptiometry in patients with uncemented total hip arthroplasty. J Orthop Res 10:836–844PubMed
37.
go back to reference Kröger H, Vanninen E, Overmyer M, Miettinen H, Rushton N, Suomalainen O (1997): Periprosthetic bone loss and regional bone turnover in uncemented total hip arthroplasty: a prospective study using high resolution single photon emission tomography and dual-energy X-ray absorptiometry. J Bone Miner Res 12:487–492 Kröger H, Vanninen E, Overmyer M, Miettinen H, Rushton N, Suomalainen O (1997): Periprosthetic bone loss and regional bone turnover in uncemented total hip arthroplasty: a prospective study using high resolution single photon emission tomography and dual-energy X-ray absorptiometry. J Bone Miner Res 12:487–492
38.
go back to reference Kröger H, Venesmaa P, Jurvelin J, Miettinen H, Suomalainen O, Alhava E (1998) Bone density at the proximal femur after total hip arthroplasty. Clin Orthop 352:66–74PubMed Kröger H, Venesmaa P, Jurvelin J, Miettinen H, Suomalainen O, Alhava E (1998) Bone density at the proximal femur after total hip arthroplasty. Clin Orthop 352:66–74PubMed
39.
go back to reference Kröger H, Miettinen H, Arnala I, Koski E, Rushton N, Suomalainen O (1996) Evaluation of periprosthetic bone using dual-energy x-ray absorptiometry: precision of the method and effect of operation on bone mineral density. J Bone Miner Res 11:1526–1530PubMed Kröger H, Miettinen H, Arnala I, Koski E, Rushton N, Suomalainen O (1996) Evaluation of periprosthetic bone using dual-energy x-ray absorptiometry: precision of the method and effect of operation on bone mineral density. J Bone Miner Res 11:1526–1530PubMed
40.
go back to reference Kärrholm J, Malchau H, Snorrason F, Herberts P (1994) Micromotion of femoral stems in total hip arthroplasty. A randomized study of cemented, hydroxyapatite-coated, and porous-coated stems with roentgen stereophotogrammetric analysis. J Bone Joint Surg Am 76:1692–1705PubMed Kärrholm J, Malchau H, Snorrason F, Herberts P (1994) Micromotion of femoral stems in total hip arthroplasty. A randomized study of cemented, hydroxyapatite-coated, and porous-coated stems with roentgen stereophotogrammetric analysis. J Bone Joint Surg Am 76:1692–1705PubMed
41.
go back to reference Maathuis PG, Visser JD (1996) High failure rate of soft-interface stem coating for fixation of femoral endoprostheses. J Arthroplasty 11:548–552PubMed Maathuis PG, Visser JD (1996) High failure rate of soft-interface stem coating for fixation of femoral endoprostheses. J Arthroplasty 11:548–552PubMed
42.
go back to reference Malchau H, Herberts P, Söderman P, Odén A (2000) Prognosis of total hip replacement. Update and validation of results from the Swedish National Hip Arthroplasty Registry 1979–1998. Presented at the 67th Annual Meeting of the American Academy of Orthopaedic Surgeons, Orlando, Florida, USA Malchau H, Herberts P, Söderman P, Odén A (2000) Prognosis of total hip replacement. Update and validation of results from the Swedish National Hip Arthroplasty Registry 1979–1998. Presented at the 67th Annual Meeting of the American Academy of Orthopaedic Surgeons, Orlando, Florida, USA
43.
go back to reference Malchau H, Kärrholm J, Wang YX, Herberts P (1995) Accuracy of migration analysis in hip arthroplasty. Digitized and conventional radiography compared to radiostereometry in 51 patients. Acta Orthop Scand 66:418–424PubMed Malchau H, Kärrholm J, Wang YX, Herberts P (1995) Accuracy of migration analysis in hip arthroplasty. Digitized and conventional radiography compared to radiostereometry in 51 patients. Acta Orthop Scand 66:418–424PubMed
44.
go back to reference Mallory TH, Head WC, Lombardi AV Jr, Emerson RH Jr, Eberle RW, Mitchell MB (1996) Clinical and radiographic outcome of a cementless, titanium, plasma spray-coated total hip arthroplasty femoral component. Justification for continuance of use. J Arthroplasty 11:653–666PubMed Mallory TH, Head WC, Lombardi AV Jr, Emerson RH Jr, Eberle RW, Mitchell MB (1996) Clinical and radiographic outcome of a cementless, titanium, plasma spray-coated total hip arthroplasty femoral component. Justification for continuance of use. J Arthroplasty 11:653–666PubMed
45.
go back to reference Maloney WJ, Smith RL (1996) Periprosthetic osteolysis in total hip arthroplasty: the role of particulate wear debris. Instr Course Lect 45:171–182PubMed Maloney WJ, Smith RL (1996) Periprosthetic osteolysis in total hip arthroplasty: the role of particulate wear debris. Instr Course Lect 45:171–182PubMed
46.
go back to reference Marmor L (1976) Femoral loosening in total hip replacement. Clin Orthop 121:116–119PubMed Marmor L (1976) Femoral loosening in total hip replacement. Clin Orthop 121:116–119PubMed
47.
go back to reference Martini F, Lebherz C, Mayer F, Leichtle U, Kremling E, Sell S (2000) Precision of the measurements of periprosthetic bone mineral density in hips with a custom-made femoral stem. J Bone Joint Surg Br 82:1065–1071CrossRefPubMed Martini F, Lebherz C, Mayer F, Leichtle U, Kremling E, Sell S (2000) Precision of the measurements of periprosthetic bone mineral density in hips with a custom-made femoral stem. J Bone Joint Surg Br 82:1065–1071CrossRefPubMed
48.
go back to reference Mazess R, Collick B, Trempe J, Barden H, Hanson J (1989) Performance evaluation of a dual-energy X-ray bone densitometer. Calcif Tissue Int 44:228–232PubMed Mazess R, Collick B, Trempe J, Barden H, Hanson J (1989) Performance evaluation of a dual-energy X-ray bone densitometer. Calcif Tissue Int 44:228–232PubMed
49.
go back to reference McCarthy CK, Steinberg GG, Agren M, Leahey D, Wyman E, Baran DT (1991) Quantifying bone loss from the proximal femur after total hip arthroplasty. J Bone Joint Surg Br 73:774–778 McCarthy CK, Steinberg GG, Agren M, Leahey D, Wyman E, Baran DT (1991) Quantifying bone loss from the proximal femur after total hip arthroplasty. J Bone Joint Surg Br 73:774–778
50.
go back to reference McLaughlin JR, Lee KR (1997) Total hip arthroplasty with an uncemented femoral component. Excellent result at ten-year follow-up. J Bone Joint Surg Br 79:900–907CrossRefPubMed McLaughlin JR, Lee KR (1997) Total hip arthroplasty with an uncemented femoral component. Excellent result at ten-year follow-up. J Bone Joint Surg Br 79:900–907CrossRefPubMed
51.
go back to reference McNally SA, Shepperd JA, Mann CV, Walczak JP (2000) The results at nine to twelve years of the use of a hydroxyapatite-coated femoral stem. J Bone Joint Surg Br 82:378–382CrossRefPubMed McNally SA, Shepperd JA, Mann CV, Walczak JP (2000) The results at nine to twelve years of the use of a hydroxyapatite-coated femoral stem. J Bone Joint Surg Br 82:378–382CrossRefPubMed
52.
go back to reference Meding JB, Keating EM, Ritter MA, Faris PM, Berend ME (2004) Minimum ten-year follow-up of a straight-stemmed, plasma-sprayed, titanium-alloy, uncemented femoral component in primary total hip arthroplasty. J Bone J Surg Am 86:92–97 Meding JB, Keating EM, Ritter MA, Faris PM, Berend ME (2004) Minimum ten-year follow-up of a straight-stemmed, plasma-sprayed, titanium-alloy, uncemented femoral component in primary total hip arthroplasty. J Bone J Surg Am 86:92–97
53.
go back to reference Moilanen T, Scott G, Newell M, Garvie N, Freeman MAR (1997) Bone scintigraphic appearance of asymptomatic hydroxyapatite-coated hip arthroplasties. J Arthroplasty 12:380–386PubMed Moilanen T, Scott G, Newell M, Garvie N, Freeman MAR (1997) Bone scintigraphic appearance of asymptomatic hydroxyapatite-coated hip arthroplasties. J Arthroplasty 12:380–386PubMed
54.
go back to reference Morrey BF, Adams RA, Kessler M (2000) A conservative femoral replacement for total hip arthroplasty. A prospective study. J Bone Joint Surg Br 82:952–958CrossRefPubMed Morrey BF, Adams RA, Kessler M (2000) A conservative femoral replacement for total hip arthroplasty. A prospective study. J Bone Joint Surg Br 82:952–958CrossRefPubMed
55.
go back to reference Oswald SG, Van Nostrand D, Savory CG, Callaghan JJ (1989) Three-phase bone scan and indium white blood cell scintigraphy following porous coated hip arthroplasty: a prospective study of the prosthetic tip. J Nucl Med 30:1321–1331PubMed Oswald SG, Van Nostrand D, Savory CG, Callaghan JJ (1989) Three-phase bone scan and indium white blood cell scintigraphy following porous coated hip arthroplasty: a prospective study of the prosthetic tip. J Nucl Med 30:1321–1331PubMed
56.
go back to reference Pellicci PM, Wilson PD Jr, Sledge CB, Salvati EA, Ranawat CS, Poss R (1982) Revision total hip arthroplasty. Clin Orthop 170:34–41PubMed Pellicci PM, Wilson PD Jr, Sledge CB, Salvati EA, Ranawat CS, Poss R (1982) Revision total hip arthroplasty. Clin Orthop 170:34–41PubMed
57.
go back to reference Rahmy AI, Tonino AJ, Tan WD (1994) Quantitative analysis of technetium-99m-methylene disphosphonate uptake in unilateral hydroxyapatite-coated total hip prostheses: first year follow-up. J Nucl Med 35:1788–1791PubMed Rahmy AI, Tonino AJ, Tan WD (1994) Quantitative analysis of technetium-99m-methylene disphosphonate uptake in unilateral hydroxyapatite-coated total hip prostheses: first year follow-up. J Nucl Med 35:1788–1791PubMed
58.
go back to reference Rosenthall L, Ghazal ME, Brooks CE (1991) Quantitative analysis of radiophosphate uptakes in asymptomatic porous-coated hip endoprostheses. J Nucl Med 32:1391–1393PubMed Rosenthall L, Ghazal ME, Brooks CE (1991) Quantitative analysis of radiophosphate uptakes in asymptomatic porous-coated hip endoprostheses. J Nucl Med 32:1391–1393PubMed
59.
go back to reference Santavirta S, Konttinen YT, Bergroth V, Eskola A, Tallroth K, Lindholm TS (1990) Aggressive granulomatous lesions associated with hip arthroplasty. Immunopathological studies. J Bone Joint Surg Am 72:252–258PubMed Santavirta S, Konttinen YT, Bergroth V, Eskola A, Tallroth K, Lindholm TS (1990) Aggressive granulomatous lesions associated with hip arthroplasty. Immunopathological studies. J Bone Joint Surg Am 72:252–258PubMed
60.
go back to reference Schicha H, Perner K, Voth E, Reith HG, Willert HG, Emrich D (1986) Zementfreie Implantation von Zweimüller-Endler-Totalendoprothesen der Hüfte- Klinische, röntgenologische und szintigraphische Verlaufskontrolle über zwei Jahre. Nuklearmedizin 25:55–60PubMed Schicha H, Perner K, Voth E, Reith HG, Willert HG, Emrich D (1986) Zementfreie Implantation von Zweimüller-Endler-Totalendoprothesen der Hüfte- Klinische, röntgenologische und szintigraphische Verlaufskontrolle über zwei Jahre. Nuklearmedizin 25:55–60PubMed
61.
go back to reference Scott DF, Jaffe WL (1996) Host-bone response to porous-coated cobalt-chrome and hydroxyapatite-coated titanium femoral components in hip arthroplasty. Dual-energy X-ray absorptiometry analysis of paired bilateral cases at 5 to 7 years. J Arthroplasty 11:429–437PubMed Scott DF, Jaffe WL (1996) Host-bone response to porous-coated cobalt-chrome and hydroxyapatite-coated titanium femoral components in hip arthroplasty. Dual-energy X-ray absorptiometry analysis of paired bilateral cases at 5 to 7 years. J Arthroplasty 11:429–437PubMed
62.
go back to reference Sumner DR, Galante JO (1992) Determinants of stress shielding: design versus materials versus interface. Clin Orthop 274:202–212PubMed Sumner DR, Galante JO (1992) Determinants of stress shielding: design versus materials versus interface. Clin Orthop 274:202–212PubMed
63.
go back to reference Søballe K, Hansen ES, Brockstedt-Rasmussen H, Hjortdal VE, Juhl GI, Pedersen CM, Hvid I, Bünger C (1991) Gap healing enhanced by hydroxyapatite coating in dogs. Clin Orthop 272:300–307PubMed Søballe K, Hansen ES, Brockstedt-Rasmussen H, Hjortdal VE, Juhl GI, Pedersen CM, Hvid I, Bünger C (1991) Gap healing enhanced by hydroxyapatite coating in dogs. Clin Orthop 272:300–307PubMed
64.
go back to reference Trevisan C, Bigoni M, Randelli G, Marinoni EC, Peretti G, Ortolani S (1997) Periprosthetic bone density around fully hydroxyapatite coated femoral stem. Clin Orthop 340:109–117PubMed Trevisan C, Bigoni M, Randelli G, Marinoni EC, Peretti G, Ortolani S (1997) Periprosthetic bone density around fully hydroxyapatite coated femoral stem. Clin Orthop 340:109–117PubMed
65.
go back to reference Venesmaa P, Kröger H, Miettinen H, Jurvelin J, Suomalainen O, Alhava E (2000) Bone loss around failed femoral implant measured by dual-energy X-ray absorptiometry. J Orthop Sci 5:380–384CrossRefPubMed Venesmaa P, Kröger H, Miettinen H, Jurvelin J, Suomalainen O, Alhava E (2000) Bone loss around failed femoral implant measured by dual-energy X-ray absorptiometry. J Orthop Sci 5:380–384CrossRefPubMed
66.
go back to reference Venesmaa PK, Kröger HP, Miettinen HJ, Jurvelin JS, Suomalainen OT, Alhava EM (2001) Monitoring of periprosthetic BMD after uncemented total hip arthroplasty with dual-energy X-ray absorptiometry--a 3-year follow-up study. J Bone Miner Res 16:1056–1061PubMed Venesmaa PK, Kröger HP, Miettinen HJ, Jurvelin JS, Suomalainen OT, Alhava EM (2001) Monitoring of periprosthetic BMD after uncemented total hip arthroplasty with dual-energy X-ray absorptiometry--a 3-year follow-up study. J Bone Miner Res 16:1056–1061PubMed
67.
go back to reference Wixson RL, Stulberg SD, Mehlhoff M (1988) Results of cemented, uncemented, and hybrid total hip replacements at two to four years. Presented at the 55th Annual Meeting of the American Academy of Orthopaedic Surgeons, Atlanta, Georgia, USA Wixson RL, Stulberg SD, Mehlhoff M (1988) Results of cemented, uncemented, and hybrid total hip replacements at two to four years. Presented at the 55th Annual Meeting of the American Academy of Orthopaedic Surgeons, Atlanta, Georgia, USA
68.
go back to reference Wixson RL, Stulberg SD, Van Flandern GJ, Puri L (1997) Maintenance of proximal bone mass with an uncemented femoral stem analysis with dual-energy X-ray absorptiometry. J Arthroplasty 12:365–372 Wixson RL, Stulberg SD, Van Flandern GJ, Puri L (1997) Maintenance of proximal bone mass with an uncemented femoral stem analysis with dual-energy X-ray absorptiometry. J Arthroplasty 12:365–372
Metadata
Title
Unstable versus stable uncemented femoral stems: a radiological study of periprosthetic bone changes in two types of uncemented stems with different concepts of fixation
Authors
Henrik Bodén
Per Adolphson
Michael Öberg
Publication date
01-07-2004
Publisher
Springer-Verlag
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 6/2004
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-004-0666-5

Other articles of this Issue 6/2004

Archives of Orthopaedic and Trauma Surgery 6/2004 Go to the issue