Skip to main content
Top
Published in: Trials 1/2022

Open Access 01-12-2022 | Ultrasound | Study protocol

Multicenter study of lumen-apposing metal stents with or without pigtail in endoscopic ultrasound-guided biliary drainage for malignant obstruction—BAMPI TRIAL: an open-label, randomized controlled trial protocol

Authors: Albert Garcia-Sumalla, Carme Loras, Vicente Sanchiz, Rafael Pedraza Sanz, Enrique Vazquez-Sequeiros, Jose Ramon Aparicio, Carlos de la Serna-Higuera, Daniel Luna-Rodriguez, Xavier Andujar, María Capilla, Tatiana Barberá, Jose Ramon Foruny-Olcina, Belen Martínez, Miguel Dura, Silvia Salord, Berta Laquente, Cristian Tebe, Sebastia Videla, Manuel Perez-Miranda, Joan B. Gornals, On behalf of Spanish Working Group on Endoscopic Ultrasound Guided Biliary Drainage

Published in: Trials | Issue 1/2022

Login to get access

Abstract

Background

It is unclear whether the insertion of an axis-orienting double-pigtail plastic stent (DPS) through biliary lumen-apposing meal stent (LAMS) in EUS-guided choledochoduodenostomy (CDS) improves the stent patency. The aim of this study is to determine whether this technical variant offers a clinical benefit in EUS-guided biliary drainage (BD) for the management of distal malignant biliary obstruction.

Methods/design

This is a multicenter open-label, randomized controlled trial with two parallel groups. Eighty-four patients with malignant biliary obstruction will undergo EUS-BD (CDS type) using LAMS in 7 tertiary hospitals in Spain and will be randomized to the LAMS and LAMS plus DPS groups. The primary endpoint is the rate of recurrent biliary obstruction, as a stent dysfunction parameter, detected during follow-up. Secondary endpoints: technical and clinical success (reduction in bilirubin > 50% within 14 days of stent placement), safety, and others (number of reinterventions, time to biliary obstruction, prognostic factors, survival rate).

Discussion

The BAMPI trial has been designed to determine whether the addition of a coaxial axis-orienting DPS through LAMS is superior to LAMS alone to prevent stent dysfunction.

Trial registration

ClinicalTrials.​govNCT04595058. Registered on October 14, 2020.
Appendix
Available only for authorised users
Literature
1.
go back to reference Puga M, Pallarès N, Velásquez-Rodríguez J, García-Sumalla A, Consiglieri CF, Busquets J, et al. Endoscopic biliary drainage in unresectable biliary obstruction: the role of endoscopic ultrasound-guidance in a cohort study. Rev Esp Enferm Dig. 2019;111:683–9.CrossRef Puga M, Pallarès N, Velásquez-Rodríguez J, García-Sumalla A, Consiglieri CF, Busquets J, et al. Endoscopic biliary drainage in unresectable biliary obstruction: the role of endoscopic ultrasound-guidance in a cohort study. Rev Esp Enferm Dig. 2019;111:683–9.CrossRef
2.
go back to reference Chen YI, Callichurn K, Chatterjee A, Desilets E, Fergal D, Forbes N, et al. ELEMENT trial and for the Canadian Endoscopic Research Collaborative (CERC). ELEMENT TRIAL: study protocol for a randomized controlled trial on endoscopic ultrasound-guided biliary drainage of first intent with a lumen-apposing metal stent vs. endoscopic retrograde cholangio-pancreatography in the management of malignant distal biliary obstruction. Trials. 2019;20:696.CrossRef Chen YI, Callichurn K, Chatterjee A, Desilets E, Fergal D, Forbes N, et al. ELEMENT trial and for the Canadian Endoscopic Research Collaborative (CERC). ELEMENT TRIAL: study protocol for a randomized controlled trial on endoscopic ultrasound-guided biliary drainage of first intent with a lumen-apposing metal stent vs. endoscopic retrograde cholangio-pancreatography in the management of malignant distal biliary obstruction. Trials. 2019;20:696.CrossRef
3.
go back to reference Itoi T, Binmoeller KF. EUS-guided choledochoduodenostomy by using a biflanged lumen-apposing metal stent. Gastrointest Endosc. 2014;79:715.CrossRef Itoi T, Binmoeller KF. EUS-guided choledochoduodenostomy by using a biflanged lumen-apposing metal stent. Gastrointest Endosc. 2014;79:715.CrossRef
4.
go back to reference Gornals JB, Consiglieri C, Gallarreta V, Busquets J, Catala I, Laquente B. Single-session fluoroless endoscopic ultrasound-guided fine-needle aspiration and choledochoduodenostomy with a biliary lumen-apposing stent. Endoscopy. 2015;47:418–9.CrossRef Gornals JB, Consiglieri C, Gallarreta V, Busquets J, Catala I, Laquente B. Single-session fluoroless endoscopic ultrasound-guided fine-needle aspiration and choledochoduodenostomy with a biliary lumen-apposing stent. Endoscopy. 2015;47:418–9.CrossRef
5.
go back to reference Anderloni A, Anderloni A, Fugazza A, Troncone E, Auriemma F, Carrara S, et al. Single-stage EUS-guided choledochoduodenostomy using a lumen-apposing metal stent for malignant distal biliary obstruction. Gastrointest Endosc. 2019;89:69–76.CrossRef Anderloni A, Anderloni A, Fugazza A, Troncone E, Auriemma F, Carrara S, et al. Single-stage EUS-guided choledochoduodenostomy using a lumen-apposing metal stent for malignant distal biliary obstruction. Gastrointest Endosc. 2019;89:69–76.CrossRef
6.
go back to reference Jacques J, Privat J, Pinard F, Fumex F, Valats JC, Chaoui A, et al. Endoscopic ultrasound-guided choledochoduodenostomy with electrocautery-enhanced lumen-apposing stents: A retrospective analysis. Endoscopy. 2019;51:540–7.CrossRef Jacques J, Privat J, Pinard F, Fumex F, Valats JC, Chaoui A, et al. Endoscopic ultrasound-guided choledochoduodenostomy with electrocautery-enhanced lumen-apposing stents: A retrospective analysis. Endoscopy. 2019;51:540–7.CrossRef
7.
go back to reference Paik WH, Lee TH, Park DH, Choi JH, Kim SO, Jang S, et al. EUS-guided biliary drainage versus ERCP for the primary palliation of malignant biliary obstruction: a multicenter randomized clinical trial. Am J Gastroenterol. 2018;113:987–97.CrossRef Paik WH, Lee TH, Park DH, Choi JH, Kim SO, Jang S, et al. EUS-guided biliary drainage versus ERCP for the primary palliation of malignant biliary obstruction: a multicenter randomized clinical trial. Am J Gastroenterol. 2018;113:987–97.CrossRef
8.
go back to reference Park JK, Woo YS, Noh DH, Ju-il Y, Bae SY, Yun HS, et al. Efficacy of EUS-guided and ERCP-guided biliary drainage for malignant biliary obstruction: prospective randomized controlled study. Gastrointest Endosc. 2018;88:277–82.CrossRef Park JK, Woo YS, Noh DH, Ju-il Y, Bae SY, Yun HS, et al. Efficacy of EUS-guided and ERCP-guided biliary drainage for malignant biliary obstruction: prospective randomized controlled study. Gastrointest Endosc. 2018;88:277–82.CrossRef
9.
go back to reference Itonaga M, Kitano M, Yoshikawa T, Ashida R, Yamashita Y, Hatamaru K, et al. Comparison of endoscopic ultrasound-guided choledochoduodenostomy and endoscopic retrograde cholangiopancreatography in first-line biliary drainage for malignant distal bile duct obstruction: A multicenter randomized controlled trial. Med (Baltimore). 2021;100(12):e25268.CrossRef Itonaga M, Kitano M, Yoshikawa T, Ashida R, Yamashita Y, Hatamaru K, et al. Comparison of endoscopic ultrasound-guided choledochoduodenostomy and endoscopic retrograde cholangiopancreatography in first-line biliary drainage for malignant distal bile duct obstruction: A multicenter randomized controlled trial. Med (Baltimore). 2021;100(12):e25268.CrossRef
10.
go back to reference Krishnamoorthi R, Dasari CS, Thoguluva V, Priyan H, Jayaraj M, Law J, et al. Effectiveness and safety of EUS-guided choledochoduodenostomy using lumen-apposing metal stents ( LAMS): a systematic review and meta-analysis. Surg Endosc. 2020;34:2866–77.CrossRef Krishnamoorthi R, Dasari CS, Thoguluva V, Priyan H, Jayaraj M, Law J, et al. Effectiveness and safety of EUS-guided choledochoduodenostomy using lumen-apposing metal stents ( LAMS): a systematic review and meta-analysis. Surg Endosc. 2020;34:2866–77.CrossRef
11.
go back to reference Jain D, Shah M, Patel U, Sharma A, Singhal S. Endoscopic ultrasound guided choledocho-enterostomy by using lumen apposing metal stent in patients with failed endoscopic retrograde cholangiopancreatography: a literature review. Digestion. 2018;98:1–10.CrossRef Jain D, Shah M, Patel U, Sharma A, Singhal S. Endoscopic ultrasound guided choledocho-enterostomy by using lumen apposing metal stent in patients with failed endoscopic retrograde cholangiopancreatography: a literature review. Digestion. 2018;98:1–10.CrossRef
12.
go back to reference Mohan BP, Shakhatreh M, Garg R, Ponnada S, Navaneethan U, Adler DG. Efficacy and safety of endoscopic ultrasound-guided choledochoduodenostomy: a systematic review and metanalysis. J Clin Gastroenterol. 2019;53:243–50.CrossRef Mohan BP, Shakhatreh M, Garg R, Ponnada S, Navaneethan U, Adler DG. Efficacy and safety of endoscopic ultrasound-guided choledochoduodenostomy: a systematic review and metanalysis. J Clin Gastroenterol. 2019;53:243–50.CrossRef
13.
go back to reference Amato A, Sinagra E, Celsa C, Enea M, Buda A, Vieceli F, et al. Efficacy of lumen-apposing metal stents or self-expandable metal stents for endoscopic ultrasound-guided choledochoduodenostomy: a systematic review and meta-analysis. Endoscopy. 2021;53:1037–47.CrossRef Amato A, Sinagra E, Celsa C, Enea M, Buda A, Vieceli F, et al. Efficacy of lumen-apposing metal stents or self-expandable metal stents for endoscopic ultrasound-guided choledochoduodenostomy: a systematic review and meta-analysis. Endoscopy. 2021;53:1037–47.CrossRef
14.
go back to reference Han SY, Kim SO, So H, Shin E, Kim DU, Park DH. EUS-guided biliary drainage versus ERCP for first-line palliation of malignant distal biliary obstruction: A systematic review and meta-analysis. Sci Rep. 2019;9(1). Han SY, Kim SO, So H, Shin E, Kim DU, Park DH. EUS-guided biliary drainage versus ERCP for first-line palliation of malignant distal biliary obstruction: A systematic review and meta-analysis. Sci Rep. 2019;9(1).
15.
go back to reference Tsuchiya T, Teoh AYB, Itoi T, Yamao K, Hara K, Nakai Y, et al. Long-term outcomes of EUS-guided choledochoduodenostomy using a lumen-apposing metal stent for malignant distal biliary obstruction: a prospective multicenter study. Gastrointest Endosc. 2018;87:1138–46.CrossRef Tsuchiya T, Teoh AYB, Itoi T, Yamao K, Hara K, Nakai Y, et al. Long-term outcomes of EUS-guided choledochoduodenostomy using a lumen-apposing metal stent for malignant distal biliary obstruction: a prospective multicenter study. Gastrointest Endosc. 2018;87:1138–46.CrossRef
16.
go back to reference Garcia-Sumalla A, Bazaga S, Gornals JB. EUS-guided biliary rendezvous as an emergent rescue after failed choledochoduodenostomy using a lumen-apposing metal stent. VideoGIE. 2021;6:263–5.CrossRef Garcia-Sumalla A, Bazaga S, Gornals JB. EUS-guided biliary rendezvous as an emergent rescue after failed choledochoduodenostomy using a lumen-apposing metal stent. VideoGIE. 2021;6:263–5.CrossRef
18.
go back to reference Gornals JB, Consiglieri CF, Bergamino MA. Double pigtail for preventing ascending cholangitis after endoscopic ultrasonography-guided choledochoduodenostomy with lumen-apposing metal stent. Dig Endosc. 2016;28:100.CrossRef Gornals JB, Consiglieri CF, Bergamino MA. Double pigtail for preventing ascending cholangitis after endoscopic ultrasonography-guided choledochoduodenostomy with lumen-apposing metal stent. Dig Endosc. 2016;28:100.CrossRef
19.
go back to reference Garcia-Sumalla A, Loras C, Guarner-Argente C, Velasquez-Rodriguez JG, Andujar X, Salord S, et al. Is a coaxial plastic stent within a lumen-apposing metal stent useful for the management of distal malignant biliary obstruction? Surg Endosc. 2021;35:4873–81.CrossRef Garcia-Sumalla A, Loras C, Guarner-Argente C, Velasquez-Rodriguez JG, Andujar X, Salord S, et al. Is a coaxial plastic stent within a lumen-apposing metal stent useful for the management of distal malignant biliary obstruction? Surg Endosc. 2021;35:4873–81.CrossRef
21.
go back to reference Cotton PB, Eisen GM, Aabakken L, Baron TH, Hutter MM, Jacobson BC, et al. A lexicon for endoscopic adverse events: report of an ASGE workshop. Gastrointest Endosc. 2010;71:446–54.CrossRef Cotton PB, Eisen GM, Aabakken L, Baron TH, Hutter MM, Jacobson BC, et al. A lexicon for endoscopic adverse events: report of an ASGE workshop. Gastrointest Endosc. 2010;71:446–54.CrossRef
22.
go back to reference On W, Paranandi B, Smith AM, Venkatachalapathy SV, James MW, Aithal GP, et al. EUS-guided choledochoduodenostomy with electrocautery-enhanced lumen-apposing metal stents in patients with malignant distal biliary obstruction: multicenter collaboration from the United Kingdom and Ireland. Gastrointest Endosc. 2021;S0016-5107(21):01702–8. https://doi.org/10.1016/j.gie.2021.09.040.CrossRef On W, Paranandi B, Smith AM, Venkatachalapathy SV, James MW, Aithal GP, et al. EUS-guided choledochoduodenostomy with electrocautery-enhanced lumen-apposing metal stents in patients with malignant distal biliary obstruction: multicenter collaboration from the United Kingdom and Ireland. Gastrointest Endosc. 2021;S0016-5107(21):01702–8. https://​doi.​org/​10.​1016/​j.​gie.​2021.​09.​040.CrossRef
23.
go back to reference el Chafic AH, Shah JN, Hamerski C, Binmoeller KF, Irani S, James TW, et al. EUS-guided choledochoduodenostomy for distal malignant biliary obstruction using electrocautery-enhanced lumen-apposing metal stents: First US, multicenter experience. Dig Dis Sci. 2019;64:3321–7.CrossRef el Chafic AH, Shah JN, Hamerski C, Binmoeller KF, Irani S, James TW, et al. EUS-guided choledochoduodenostomy for distal malignant biliary obstruction using electrocautery-enhanced lumen-apposing metal stents: First US, multicenter experience. Dig Dis Sci. 2019;64:3321–7.CrossRef
24.
go back to reference Ginestet C, Sanglier F, Hummel V, Rouchaud A, Legros R, Lepetit H, et al. EUS-guided biliary drainage with electrocautery-enhanced lumen-apposing metal stent placement should replace PTBD after ERCP failure in patients with distal tumoral biliary obstruction: a large real-life study. Surg Endosc. 2021. https://doi.org/10.1007/s00464-021-08653-1. Ginestet C, Sanglier F, Hummel V, Rouchaud A, Legros R, Lepetit H, et al. EUS-guided biliary drainage with electrocautery-enhanced lumen-apposing metal stent placement should replace PTBD after ERCP failure in patients with distal tumoral biliary obstruction: a large real-life study. Surg Endosc. 2021. https://​doi.​org/​10.​1007/​s00464-021-08653-1.
25.
go back to reference de Benito SM, Nájera-Muñoz R, de la Serna-Higuera C, Fuentes-Valenzuela E, Fanjul I, Chavarría C, et al. Lumen apposing metal stents versus tubular self-expandable metal stents for endoscopic ultrasound-guided choledochoduodenostomy in malignant biliary obstruction. Surg Endosc. 2021;35:6754–62.CrossRef de Benito SM, Nájera-Muñoz R, de la Serna-Higuera C, Fuentes-Valenzuela E, Fanjul I, Chavarría C, et al. Lumen apposing metal stents versus tubular self-expandable metal stents for endoscopic ultrasound-guided choledochoduodenostomy in malignant biliary obstruction. Surg Endosc. 2021;35:6754–62.CrossRef
26.
go back to reference Puga M, Consiglieri CF, Busquets J, Pallarès N, Secanella L, Peláez N, et al. Safety of lumen-apposing stent with or without coaxial plastic stent for endoscopic ultrasound-guided drainage of pancreatic fluid collections: a retrospective study. Endoscopy. 2018;50:1022–6.CrossRef Puga M, Consiglieri CF, Busquets J, Pallarès N, Secanella L, Peláez N, et al. Safety of lumen-apposing stent with or without coaxial plastic stent for endoscopic ultrasound-guided drainage of pancreatic fluid collections: a retrospective study. Endoscopy. 2018;50:1022–6.CrossRef
27.
go back to reference Chin JY-L, Seleq S, Weilert F. Safety and outcomes of endoscopic ultrasound-guided drainage for malignant biliary obstruction using cautery-enabled lumen-apposing metal stent. Endosc Int Open. 2020;08:E1633–8.CrossRef Chin JY-L, Seleq S, Weilert F. Safety and outcomes of endoscopic ultrasound-guided drainage for malignant biliary obstruction using cautery-enabled lumen-apposing metal stent. Endosc Int Open. 2020;08:E1633–8.CrossRef
Metadata
Title
Multicenter study of lumen-apposing metal stents with or without pigtail in endoscopic ultrasound-guided biliary drainage for malignant obstruction—BAMPI TRIAL: an open-label, randomized controlled trial protocol
Authors
Albert Garcia-Sumalla
Carme Loras
Vicente Sanchiz
Rafael Pedraza Sanz
Enrique Vazquez-Sequeiros
Jose Ramon Aparicio
Carlos de la Serna-Higuera
Daniel Luna-Rodriguez
Xavier Andujar
María Capilla
Tatiana Barberá
Jose Ramon Foruny-Olcina
Belen Martínez
Miguel Dura
Silvia Salord
Berta Laquente
Cristian Tebe
Sebastia Videla
Manuel Perez-Miranda
Joan B. Gornals
On behalf of Spanish Working Group on Endoscopic Ultrasound Guided Biliary Drainage
Publication date
01-12-2022
Publisher
BioMed Central
Published in
Trials / Issue 1/2022
Electronic ISSN: 1745-6215
DOI
https://doi.org/10.1186/s13063-022-06106-1

Other articles of this Issue 1/2022

Trials 1/2022 Go to the issue