Published in:
01-02-2020 | Ultrasound | Ultrasound
Learning curve of liver stiffness measurement using a new hybrid machine composed of transient elastography interfaced with ultrasound
Authors:
Christophe Cassinotto, Marie-Ange Pierredon-Foulongne, Ali Belgour, Julien Delicque, Laure Escal, Nicolas Molinari, Sophie Anselme, Tony Jacq, Ugo Chamard-Champliaud, Bastien Nicolan, Carole Allimant, Boris Guiu
Published in:
European Radiology
|
Issue 2/2020
Login to get access
Abstract
Objectives
To assess the learning curve for performing reliable liver stiffness measurements using a new hybrid machine composed of transient elastography (TE) interfaced with an ultrasound device for radiographers and radiologists with different levels of expertise in ultrasound imaging.
Methods
Ten novice operators who had never performed TE measurements were prospectively evaluated from April to October 2018: senior radiologists, young radiologists, fellows, radiographers, and residents, with different levels of experience in abdominal ultrasound imaging. All operators had a short theoretical training followed by a training session under supervision in three patients. Then, each operator had to perform TE in 50 consecutive patients with chronic liver disease, using beforehand ultrasound examination to select measurement area in the right liver lobe, and if needed, the XL probe. Percentages of failures and reliable measurements were compared.
Results
The rates of failures of measurements, poorly reliable, reliable, and very reliable results, were of 4.2% (21/500), 2.4% (12/500), 47.6% (238/500), and 45.8% (229/500), respectively. The rates of reliable plus very reliable results were excellent, ranging from 91 to 96% among all the subgroups. The rates of very reliable, reliable, and unreliable results did not differ between operator subgroups and especially between junior radiologists, senior radiologists, and radiographers. No breaking point was observed in the interquartile range/median values over time.
Conclusion
TE interfaced with ultrasound in this hybrid machine presents no learning curve effect. After a short initial training session, a novice observer is able to perform high rates of reliable and very reliable TE measurements.
Key Points
• When performing liver stiffness measurements using a new hybrid machine composed of transient elastography interfaced with ultrasound, the rate of failures of measurements is very low, below 5%.
• After a short training session and using ultrasound planning, a novice operator, whatever its expertise in ultrasound imaging, is capable of performing high rates of reliable and very reliable measurements.
• No learning curve is needed for performing reliable liver stiffness measurements using this new hybrid machine.