Transvaginal two-dimensional ultrasound measurement of cervical volume to predict the outcome of the induction of labour: a prospective observational study

Authors: S. R. Athulathmudali, M. Patabendige, S. K. Chandrasinghe, P. H. P. De Silva

Published in: BMC Pregnancy and Childbirth | Issue 1/2021

Login to get access

Abstract

Background

Assessing the likelihood of success of induction of labour using ultrasonically measured cervical volume is an important research question.

Method

A prospective observational study was carried out at North Colombo Teaching Hospital, Ragama, Sri Lanka. Pre-induction digital cervical assessment, transvaginal cervical length, and cervical volume measurements were performed. Inductions with singleton pregnancies at term were included. Basic demographic and clinical details, independent variables (Bishop score, cervical length and cervical volume), and dependent variables (frequency of delivery within 24 h and induction to delivery interval) were recorded. Vaginal delivery within 24 h was the primary outcome.

Results

We studied 100 pregnant women who had induction of labour. Median (IQR) Bishop score was 5 (3–6), mean (SD) cervical length was 3.6 (0.7) cm, and mean (SD) cervical volume was 27.5 (10.4) cm3. Cervical length was the best predictor for predicting the likelihood of vaginal delivery within 24 h [aOR – 12.12 (3.44, 42.71); < 0.001], and cervical volume also appeared to be a significant potential predictor [aOR-1.10 (1.01, 1.17); 0.01]. Cervical length was found to have the highest AUC (0.83) followed by the cervical volume (0.74). The best cut-off value for cervical volume in predicting the likelihood of vaginal delivery within 24 h was less than 28.5 cm3 with a sensitivity of 72% and specificity of 74%.

Conclusions

Transvaginal sonographic measurement of cervical volume appears to be a potential novel predictor for the likelihood of vaginal delivery within 24 h of induction of labour. Cervical length is still more superior to cervical volume in predicting the likelihood of vaginal delivery. Bishop score was not a significant predictor in this context.
Literature
1.
go back to reference Tsakiridis I, Mamopoulos A, Athanasiadis A, Dagklis T. Induction of Labor: an Overview of Guidelines. Obstet Gynecol Surv. 2020;75(1):61–72. Tsakiridis I, Mamopoulos A, Athanasiadis A, Dagklis T. Induction of Labor: an Overview of Guidelines. Obstet Gynecol Surv. 2020;75(1):61–72.
2.
go back to reference Dhall K, Mittal SC, Kumar A. Evaluation of preinduction scoring systems. Aust New Zeal J Obstet Gynaecol. 1987;27(4):309–11.CrossRef Dhall K, Mittal SC, Kumar A. Evaluation of preinduction scoring systems. Aust New Zeal J Obstet Gynaecol. 1987;27(4):309–11.CrossRef
3.
go back to reference Hughey MJ, McElin TWBC. An evaluation of preinduction scoring systems. Obstet Gynecol. 1976;48:635–41.PubMed Hughey MJ, McElin TWBC. An evaluation of preinduction scoring systems. Obstet Gynecol. 1976;48:635–41.PubMed
4.
go back to reference Pandis GK, Papageorghiou AT, Ramanathan VG, Thompson MO, Nicolaides KH. Preinduction sonographic measurement of cervical length in the prediction of successful induction of labor. Ultrasound Obstet Gynecol. 2001;18:623–8.CrossRef Pandis GK, Papageorghiou AT, Ramanathan VG, Thompson MO, Nicolaides KH. Preinduction sonographic measurement of cervical length in the prediction of successful induction of labor. Ultrasound Obstet Gynecol. 2001;18:623–8.CrossRef
5.
go back to reference Sonek J, Shellhaas C. Cervical sonography: a review. Ultrasound Obstet Gynecol. 1998;11:71–8.CrossRef Sonek J, Shellhaas C. Cervical sonography: a review. Ultrasound Obstet Gynecol. 1998;11:71–8.CrossRef
7.
go back to reference Watson WJ, Stevens D, Welter S, Day D. Factors predicting successful labor induction. Obstet Gynecol. 1996;88(6):990–2.CrossRef Watson WJ, Stevens D, Welter S, Day D. Factors predicting successful labor induction. Obstet Gynecol. 1996;88(6):990–2.CrossRef
8.
go back to reference Park KH, Kim SN, Lee SY, Jeong EH, Jung HJ, Oh KJ. Comparison between sonographic cervical length and Bishop score in preinduction cervical assessment: a randomized trial. Ultrasound Obstet Gynecol. 2011;38(2):198–204.CrossRef Park KH, Kim SN, Lee SY, Jeong EH, Jung HJ, Oh KJ. Comparison between sonographic cervical length and Bishop score in preinduction cervical assessment: a randomized trial. Ultrasound Obstet Gynecol. 2011;38(2):198–204.CrossRef
9.
go back to reference Kyo HP. Transvaginal ultrasonographic cervical measurement in predicting failed labor induction and cesarean delivery for failure to progress in nulliparous women. J Korean Med Sci. 2007;22(4):722–7.CrossRef Kyo HP. Transvaginal ultrasonographic cervical measurement in predicting failed labor induction and cesarean delivery for failure to progress in nulliparous women. J Korean Med Sci. 2007;22(4):722–7.CrossRef
10.
go back to reference Paterson-Brown S, Fisk NM, Edmonds DK, Rodeck CH. Preinduction cervical assessment by Bishop’s score and transvaginal ultrasound. Eur J Obstet Gynecol Reprod Biol. 1991;40(1):17–23.CrossRef Paterson-Brown S, Fisk NM, Edmonds DK, Rodeck CH. Preinduction cervical assessment by Bishop’s score and transvaginal ultrasound. Eur J Obstet Gynecol Reprod Biol. 1991;40(1):17–23.CrossRef
11.
go back to reference Ezebialu IU, Eke AC, Eleje GU, Nwachukwu CE. Methods for assessing pre-induction cervical ripening. Cochrane Database Syst Rev. 2013;2013(10):CD010762. Ezebialu IU, Eke AC, Eleje GU, Nwachukwu CE. Methods for assessing pre-induction cervical ripening. Cochrane Database Syst Rev. 2013;2013(10):CD010762.
12.
go back to reference Kagan KO, Sonek J. How to measure cervical length. Ultrasound Obstet Gynecol. 2015;45(3):358–62.CrossRef Kagan KO, Sonek J. How to measure cervical length. Ultrasound Obstet Gynecol. 2015;45(3):358–62.CrossRef
14.
go back to reference Editor. Guideline on induction of labour 1. Sri Lanka J Obstet Gynaecol. 2014;35(4 December):136–8.CrossRef Editor. Guideline on induction of labour 1. Sri Lanka J Obstet Gynaecol. 2014;35(4 December):136–8.CrossRef
15.
go back to reference Laencina AMG, Sánchez FG, Gimenez JH, Martínez MS, Martínez JAV, Vizcaíno VM. Comparison of ultrasonographic cervical length and the Bishop score in predicting successful labor induction. Acta Obstet Gynecol Scand. 2007;86(7):799–804.CrossRef Laencina AMG, Sánchez FG, Gimenez JH, Martínez MS, Martínez JAV, Vizcaíno VM. Comparison of ultrasonographic cervical length and the Bishop score in predicting successful labor induction. Acta Obstet Gynecol Scand. 2007;86(7):799–804.CrossRef
16.
go back to reference Gabriel R, Darnaud T, Chalot F, Gonzalez N, Leymarie F, Quereux C. Transvaginal sonography of the uterine cervix prior to labor induction. Ultrasound Obstet Gynecol. 2002;19(3):254–7.CrossRef Gabriel R, Darnaud T, Chalot F, Gonzalez N, Leymarie F, Quereux C. Transvaginal sonography of the uterine cervix prior to labor induction. Ultrasound Obstet Gynecol. 2002;19(3):254–7.CrossRef
17.
go back to reference Saccone G, Simonetti B, Berghella V. Transvaginal ultrasound cervical length for prediction of spontaneous labour at term: a systematic review and meta-analysis. BJOG. 2016;123(1):16–22.CrossRef Saccone G, Simonetti B, Berghella V. Transvaginal ultrasound cervical length for prediction of spontaneous labour at term: a systematic review and meta-analysis. BJOG. 2016;123(1):16–22.CrossRef
18.
go back to reference Verhoeven CJM, Opmeer BC, Oei SG, Latour V, Van Der Post JAM, Mol BWJ. Transvaginal sonographic assessment of cervical length and wedging for predicting outcome of labor induction at term: a systematic review and meta-analysis. Ultrasound Obstet Gynecol. 2013;42(5):500–8.CrossRef Verhoeven CJM, Opmeer BC, Oei SG, Latour V, Van Der Post JAM, Mol BWJ. Transvaginal sonographic assessment of cervical length and wedging for predicting outcome of labor induction at term: a systematic review and meta-analysis. Ultrasound Obstet Gynecol. 2013;42(5):500–8.CrossRef
19.
go back to reference Tan PC, Vallikkannu N, Suguna S, Quek KF, Hassan J. Transvaginal sonographic measurement of cervical length vs. Bishop score in labor induction at term: tolerability and prediction of cesarean delivery. Ultrasound Obstet Gynecol. 2007;29(5):568–73.CrossRef Tan PC, Vallikkannu N, Suguna S, Quek KF, Hassan J. Transvaginal sonographic measurement of cervical length vs. Bishop score in labor induction at term: tolerability and prediction of cesarean delivery. Ultrasound Obstet Gynecol. 2007;29(5):568–73.CrossRef
20.
go back to reference Dilek TUK, Gurbuz A, Yazici G, Arslan M, Gulhan S, Pata O, et al. Comparison of cervical volume and cervical length to predict preterm delivery by transvaginal ultrasound. Am J Perinatol. 2006;23(3):167–71.CrossRef Dilek TUK, Gurbuz A, Yazici G, Arslan M, Gulhan S, Pata O, et al. Comparison of cervical volume and cervical length to predict preterm delivery by transvaginal ultrasound. Am J Perinatol. 2006;23(3):167–71.CrossRef
21.
go back to reference Jo YS, Lee GSR, Kim N, Jang DG, Kim SJ, Lee Y. Clinical efficacy of cervical length and volume for prediction of labor onset in VBAC candidates. Int J Med Sci. 2012;9(9):738–42.CrossRef Jo YS, Lee GSR, Kim N, Jang DG, Kim SJ, Lee Y. Clinical efficacy of cervical length and volume for prediction of labor onset in VBAC candidates. Int J Med Sci. 2012;9(9):738–42.CrossRef
22.
go back to reference Jo YS, Jang DG, Kim N, Kim SJ, Lee G. Comparison of cervical parameters by three-dimensional ultrasound according to parity and previous delivery mode. Int J Med Sci. 2011;8(8):673–8.CrossRef Jo YS, Jang DG, Kim N, Kim SJ, Lee G. Comparison of cervical parameters by three-dimensional ultrasound according to parity and previous delivery mode. Int J Med Sci. 2011;8(8):673–8.CrossRef
23.
go back to reference Park IY, Kwon JY, Kwon JY, Hong SC, Choi HM, Kwon HS, et al. Usefulness of cervical volume by three-dimensional ultrasound in identifying the risk for preterm birth. Ultrasound Med Biol. 2011;37(7):1039–45.CrossRef Park IY, Kwon JY, Kwon JY, Hong SC, Choi HM, Kwon HS, et al. Usefulness of cervical volume by three-dimensional ultrasound in identifying the risk for preterm birth. Ultrasound Med Biol. 2011;37(7):1039–45.CrossRef
24.
go back to reference Hoesli IM, Surbek DV, Tercanli S, Holzgreve W. Three dimensional volume measurement of the cervix during pregnancy compared to conventional 2D-sonography. Int J Gynecol Obstet. 1999;64(2):115–9.CrossRef Hoesli IM, Surbek DV, Tercanli S, Holzgreve W. Three dimensional volume measurement of the cervix during pregnancy compared to conventional 2D-sonography. Int J Gynecol Obstet. 1999;64(2):115–9.CrossRef
25.
go back to reference Rovas L, Sladkevicius P, Strobel E, De Smet F, De Moor B, Valentin L. Three-dimensional ultrasound assessment of the cervix for predicting time to spontaneous onset of labor and time to delivery in prolonged pregnancy. Ultrasound Obstet Gynecol. 2006;28(3):306–11.CrossRef Rovas L, Sladkevicius P, Strobel E, De Smet F, De Moor B, Valentin L. Three-dimensional ultrasound assessment of the cervix for predicting time to spontaneous onset of labor and time to delivery in prolonged pregnancy. Ultrasound Obstet Gynecol. 2006;28(3):306–11.CrossRef
Metadata
Title
Transvaginal two-dimensional ultrasound measurement of cervical volume to predict the outcome of the induction of labour: a prospective observational study
Authors
S. R. Athulathmudali
M. Patabendige
S. K. Chandrasinghe
P. H. P. De Silva
Publication date
01-12-2021
Publisher
BioMed Central
Published in
BMC Pregnancy and Childbirth / Issue 1/2021
Electronic ISSN: 1471-2393
DOI
https://doi.org/10.1186/s12884-021-03929-9

Other articles of this Issue 1/2021

BMC Pregnancy and Childbirth 1/2021 Go to the issue