Published in:
01-10-2013 | Clinical Investigation
Treatment of Visceral Aneurysm Using Multilayer Stent: Two-Year Follow-Up Results in Five Consecutive Patients
Authors:
Alberto Balderi, Alberto Antonietti, Fulvio Pedrazzini, Davide Sortino, Claudia Vinay, Maurizio Grosso
Published in:
CardioVascular and Interventional Radiology
|
Issue 5/2013
Login to get access
Abstract
Purpose
The present study was performed to analyze the midterm results (five consecutive patients, 2-year follow-up) of the endovascular management of visceral artery aneurysms using the Cardiatis Multilayer Flow Modulator (CMFM) (Cardiatis, Isnes, Belgium), a self-expandable stent.
Materials and Methods
From August 2009 to January 2011, we implanted five CMFMs in five patients (all men; mean age 73 years) to treat two common hepatic artery aneurysms, one celiac trunk aneurysm, one splenic artery aneurysm, and one superior mesenteric artery aneurysm (diameter 25–81 mm). The primary end point was technical success. The secondary end point was stent patency, absence of aneurysm rupture or reperfusion, and shrinking of the sac at 6-, 12-, and 24-month follow-up using computed tomography angiography. Follow-up ranged from 24 to 48 months (mean 31.2).
Results
Technical success was achieved in all patients. Complete exclusion of the aneurysm with sac shrinking was achieved in two patients. Two stents became occluded at 6- and 24-month follow-up, respectively; both patients were asymptomatic and were not retreated. One patient developed sac reperfusion due to incomplete aneurysm exclusion.
Conclusion
Long-term results in a wider population are needed to validate the effectiveness of the CMFM.