Open Access 01-12-2022 | Transesophageal Echocardiography | Case report

Venous cannula occlusion during cardiopulmonary bypass recognized by ultrasonography of the internal jugular vein

Authors: Soshi Narasaki, Hirotsugu Miyoshi, Ryuji Nakamura, Ayako Sumii, Tomoyuki Watanabe, Sachiko Otsuki, Yasuo M. Tsutsumi

Published in: JA Clinical Reports | Issue 1/2022

Login to get access

Abstract

Background

Occlusion or malposition of the venous cannula during cardiopulmonary bypass (CPB) increases central venous pressure (CVP). When high CVP is measured, we need to determine if it is actually high or if it is measured due to catheter occlusion or technical problems with the measurement.

Case presentation

We experienced a case of excessively high CVP due to malposition of the venous cannula during CPB. A 78-year-old woman underwent an aortic arch replacement for acute aortic dissection. During CPB, CVP increased up to 78 mmHg, and the time above 50 mmHg was 48 min. In this case, ultrasonography of the internal jugular vein (IJV) was useful to confirm high CVP.

Conclusions

Ultrasonography is now a familiar diagnostic tool and can be used at any time. We should consider ultrasonography as the first choice for diagnosing the cause of high CVP during CPB.
Literature
1.
go back to reference Tovedal T, Myrdal G, Jonsson O, et al. Experimental treatment of superior venous congestion during cardiopulmonary bypass. Eur J Cardiothorac Surg. 2013;44:239–44.CrossRef Tovedal T, Myrdal G, Jonsson O, et al. Experimental treatment of superior venous congestion during cardiopulmonary bypass. Eur J Cardiothorac Surg. 2013;44:239–44.CrossRef
2.
go back to reference Donahue SP, Wood JP, Patel BM, et al. Correlation of sonographic measurements of the internal jugular vein with central venous pressure. Am J Emerg Med. 2009;27:851–5.CrossRef Donahue SP, Wood JP, Patel BM, et al. Correlation of sonographic measurements of the internal jugular vein with central venous pressure. Am J Emerg Med. 2009;27:851–5.CrossRef
3.
go back to reference Hossein-Nejad H, Mohammadinejad P, Ahmadi F. Internal jugular vein/common carotid artery cross-sectional area ratio and central venous pressure. J Clin Ultrasound. 2016;44:312–8.CrossRef Hossein-Nejad H, Mohammadinejad P, Ahmadi F. Internal jugular vein/common carotid artery cross-sectional area ratio and central venous pressure. J Clin Ultrasound. 2016;44:312–8.CrossRef
4.
go back to reference Marymont J, Murphy GS. Intraoperative monitoring with transesophageal echocardiography: indications, risks, and training. Anesthesiol Clin. 2006;24:737–53.CrossRef Marymont J, Murphy GS. Intraoperative monitoring with transesophageal echocardiography: indications, risks, and training. Anesthesiol Clin. 2006;24:737–53.CrossRef
5.
go back to reference Fayad A, Shillcutt SK. Perioperative transesophageal echocardiography for non-cardiac surgery. Can J Anaesth. 2018;65:381–98.CrossRef Fayad A, Shillcutt SK. Perioperative transesophageal echocardiography for non-cardiac surgery. Can J Anaesth. 2018;65:381–98.CrossRef
6.
go back to reference Sharma M, Pathak A, Shoukat A, Rameshbabu CS. Blind areas of cardiac imaging during transesophageal echocardiogram/endoscopic ultrasound. Endosc Ultrasound. 2017;6:352–4.CrossRef Sharma M, Pathak A, Shoukat A, Rameshbabu CS. Blind areas of cardiac imaging during transesophageal echocardiogram/endoscopic ultrasound. Endosc Ultrasound. 2017;6:352–4.CrossRef
Metadata
Title
Venous cannula occlusion during cardiopulmonary bypass recognized by ultrasonography of the internal jugular vein
Authors
Soshi Narasaki
Hirotsugu Miyoshi
Ryuji Nakamura
Ayako Sumii
Tomoyuki Watanabe
Sachiko Otsuki
Yasuo M. Tsutsumi
Publication date
01-12-2022
Publisher
Springer Berlin Heidelberg
Published in
JA Clinical Reports / Issue 1/2022
Electronic ISSN: 2363-9024
DOI
https://doi.org/10.1186/s40981-022-00519-2

Other articles of this Issue 1/2022

JA Clinical Reports 1/2022 Go to the issue