01-02-2017 | Original Research

Transcutaneous PtcCO2 measurement in combination with arterial blood gas analysis provides superior accuracy and reliability in ICU patients

Authors: Oliver Spelten, Fritz Fiedler, Robert Schier, Wolfgang A. Wetsch, Jochen Hinkelbein

Published in: Journal of Clinical Monitoring and Computing | Issue 1/2017

Login to get access

Abstract

Hyper or hypoventilation may have serious clinical consequences in critically ill patients and should be generally avoided, especially in neurosurgical patients. Therefore, monitoring of carbon dioxide partial pressure by intermittent arterial blood gas analysis (PaCO2) has become standard in intensive care units (ICUs). However, several additional methods are available to determine PCO2 including end-tidal (PetCO2) and transcutaneous (PtcCO2) measurements. The aim of this study was to compare the accuracy and reliability of different methods to determine PCO2 in mechanically ventilated patients on ICU. After approval of the local ethics committee PCO2 was determined in n = 32 ICU consecutive patients requiring mechanical ventilation: (1) arterial PaCO2 blood gas analysis with Radiometer ABL 625 (ABL; gold standard), (2) arterial PaCO2 analysis with Immediate Response Mobile Analyzer (IRMA), (3) end-tidal PetCO2 by a Propaq 106 EL monitor and (4) transcutaneous PtcCO2 determination by a Tina TCM4. Bland–Altman method was used for statistical analysis; p < 0.05 was considered statistically significant. Statistical analysis revealed good correlation between PaCO2 by IRMA and ABL (R2 = 0.766; p < 0.01) as well as between PtcCO2 and ABL (R2 = 0.619; p < 0.01), whereas correlation between PetCO2 and ABL was weaker (R2 = 0.405; p < 0.01). Bland–Altman analysis revealed a bias and precision of 2.0 ± 3.7 mmHg for the IRMA, 2.2 ± 5.7 mmHg for transcutaneous, and −5.5 ± 5.6 mmHg for end-tidal measurement. Arterial CO2 partial pressure by IRMA (PaCO2) and PtcCO2 provided greater accuracy compared to the reference measurement (ABL) than the end-tidal CO2 measurements in critically ill in mechanically ventilated patients patients.
Literature
3.
go back to reference Husain T, Gatward JJ, Hambidge OR, Asogan M, Southwood TJ. Strategies to prevent airway complications: a survey of adult intensive care units in Australia and New Zealand. Br J Anaesth. 2012;108(5):800–6. doi:10.1093/bja/aes030.CrossRefPubMed Husain T, Gatward JJ, Hambidge OR, Asogan M, Southwood TJ. Strategies to prevent airway complications: a survey of adult intensive care units in Australia and New Zealand. Br J Anaesth. 2012;108(5):800–6. doi:10.​1093/​bja/​aes030.CrossRefPubMed
4.
go back to reference Kannan S, Manji M. Survey of use of end-tidal carbon dioxide for confirming tracheal tube placement in intensive care units in the UK. Anaesthesia. 2003;58(5):476–9.CrossRefPubMed Kannan S, Manji M. Survey of use of end-tidal carbon dioxide for confirming tracheal tube placement in intensive care units in the UK. Anaesthesia. 2003;58(5):476–9.CrossRefPubMed
5.
go back to reference Janssens JP, Howarth-Frey C, Chevrolet JC, Abajo B, Rochat T. Transcutaneous PCO2 to monitor noninvasive mechanical ventilation in adults: assessment of a new transcutaneous PCO2 device. Chest. 1998;113(3):768–73.CrossRefPubMed Janssens JP, Howarth-Frey C, Chevrolet JC, Abajo B, Rochat T. Transcutaneous PCO2 to monitor noninvasive mechanical ventilation in adults: assessment of a new transcutaneous PCO2 device. Chest. 1998;113(3):768–73.CrossRefPubMed
7.
go back to reference Vincent JL, Baron JF, Reinhart K, Gattinoni L, Thijs L, Webb A, Meier-Hellmann A, Nollet G, Peres-Bota D, Investigators ABC. Anemia and blood transfusion in critically ill patients. JAMA J Am Med Assoc. 2002;288(12):1499–507.CrossRef Vincent JL, Baron JF, Reinhart K, Gattinoni L, Thijs L, Webb A, Meier-Hellmann A, Nollet G, Peres-Bota D, Investigators ABC. Anemia and blood transfusion in critically ill patients. JAMA J Am Med Assoc. 2002;288(12):1499–507.CrossRef
8.
go back to reference Belenkiy S, Ivey KM, Batchinsky AI, Langer T, Necsoiu C, Baker W, Salinas J, Cancio LC. Noninvasive carbon dioxide monitoring in a porcine model of acute lung injury due to smoke inhalation and burns. Shock. 2013;39(6):495–500. doi:10.1097/SHK.0b013e318292c331.CrossRefPubMed Belenkiy S, Ivey KM, Batchinsky AI, Langer T, Necsoiu C, Baker W, Salinas J, Cancio LC. Noninvasive carbon dioxide monitoring in a porcine model of acute lung injury due to smoke inhalation and burns. Shock. 2013;39(6):495–500. doi:10.​1097/​SHK.​0b013e318292c331​.CrossRefPubMed
9.
go back to reference Shoemaker WC, Wo CC, Yu S, Farjam F, Thangathurai D. Invasive and noninvasive haemodynamic monitoring of acutely ill sepsis and septic shock patients in the emergency department. Eur J Emerg Med. 2000;7(3):169–75.CrossRefPubMed Shoemaker WC, Wo CC, Yu S, Farjam F, Thangathurai D. Invasive and noninvasive haemodynamic monitoring of acutely ill sepsis and septic shock patients in the emergency department. Eur J Emerg Med. 2000;7(3):169–75.CrossRefPubMed
10.
go back to reference Tatevossian RG, Wo CC, Velmahos GC, Demetriades D, Shoemaker WC. Transcutaneous oxygen and CO2 as early warning of tissue hypoxia and hemodynamic shock in critically ill emergency patients. Crit Care Med. 2000;28(7):2248–53.CrossRefPubMed Tatevossian RG, Wo CC, Velmahos GC, Demetriades D, Shoemaker WC. Transcutaneous oxygen and CO2 as early warning of tissue hypoxia and hemodynamic shock in critically ill emergency patients. Crit Care Med. 2000;28(7):2248–53.CrossRefPubMed
11.
go back to reference Hinkelbein J, Floss F, Denz C, Krieter H. Accuracy and precision of three different methods to determine PCO2 (Paco2 vs. Petco2 vs. Ptcco2) during interhospital ground transport of critically ill and ventilated adults. J Trauma. 2008;65(1):10–8. doi:10.1097/TA.0b013e31815eba83.CrossRefPubMed Hinkelbein J, Floss F, Denz C, Krieter H. Accuracy and precision of three different methods to determine PCO2 (Paco2 vs. Petco2 vs. Ptcco2) during interhospital ground transport of critically ill and ventilated adults. J Trauma. 2008;65(1):10–8. doi:10.​1097/​TA.​0b013e31815eba83​.CrossRefPubMed
12.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307–10.CrossRefPubMed Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307–10.CrossRefPubMed
14.
go back to reference Wahr JA, Lau W, Tremper KK, Hallock L, Smith K. Accuracy and precision of a new, portable, handheld blood gas analyzer, the IRMA. J Clin Monit. 1996;12(4):317–24.CrossRefPubMed Wahr JA, Lau W, Tremper KK, Hallock L, Smith K. Accuracy and precision of a new, portable, handheld blood gas analyzer, the IRMA. J Clin Monit. 1996;12(4):317–24.CrossRefPubMed
15.
17.
go back to reference Russell GB, Graybeal JM. End-tidal carbon dioxide as an indicator of arterial carbon dioxide in neurointensive care patients. J Neurosurg Anesthesiol. 1992;4(4):245–9.CrossRefPubMed Russell GB, Graybeal JM. End-tidal carbon dioxide as an indicator of arterial carbon dioxide in neurointensive care patients. J Neurosurg Anesthesiol. 1992;4(4):245–9.CrossRefPubMed
18.
go back to reference Christensen MA, Bloom J, Sutton KR. Comparing arterial and end-tidal carbon dioxide values in hyperventilated neurosurgical patients. Am J Crit Care. 1995;4(2):116–21.PubMed Christensen MA, Bloom J, Sutton KR. Comparing arterial and end-tidal carbon dioxide values in hyperventilated neurosurgical patients. Am J Crit Care. 1995;4(2):116–21.PubMed
19.
go back to reference Seguin P, Bleichner JP, Branger B, Guillou YM, Feuillu A, Malledant Y. The measurement of end-tidal carbon dioxide (PETCO2) is not a significant parameter to monitor in patients with severe traumatic brain injury. Can J Anaesth. 2001;48(4):396–400.CrossRefPubMed Seguin P, Bleichner JP, Branger B, Guillou YM, Feuillu A, Malledant Y. The measurement of end-tidal carbon dioxide (PETCO2) is not a significant parameter to monitor in patients with severe traumatic brain injury. Can J Anaesth. 2001;48(4):396–400.CrossRefPubMed
20.
go back to reference Barton C, Callaham M. Lack of correlation between end-tidal carbon dioxide concentrations and Paco2 in cardiac arrest. Crit Care Med. 1991;19(1):108–10.CrossRefPubMed Barton C, Callaham M. Lack of correlation between end-tidal carbon dioxide concentrations and Paco2 in cardiac arrest. Crit Care Med. 1991;19(1):108–10.CrossRefPubMed
Metadata
Title
Transcutaneous PtcCO2 measurement in combination with arterial blood gas analysis provides superior accuracy and reliability in ICU patients
Authors
Oliver Spelten
Fritz Fiedler
Robert Schier
Wolfgang A. Wetsch
Jochen Hinkelbein
Publication date
01-02-2017
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 1/2017
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-015-9810-8

Other articles of this Issue 1/2017

Journal of Clinical Monitoring and Computing 1/2017 Go to the issue