Tilt and decentration of an intraocular lens implanted in the ciliary sulcus after capsular bag defect during cataract surgery

Authors: Thomas Sauer, Ulrich Mester

Published in: Graefe's Archive for Clinical and Experimental Ophthalmology | Issue 1/2013

Login to get access

Abstract

Purpose

To determine the impact of ciliary sulcus implantation after capsular bag defect on tilt and decentration of foldable intraocular lenses (IOL) measured by a novel device.

Methods

This prospective noncomparative single-center investigation included patients (N = 10) older than age 60 who underwent cataract surgery with a posterior capsular bag defect and implantation of a threepiece hydrophilic aberration neutral IOL (OphthalmoPro AC 7013) in the ciliary sulcus. All eyes (N = 10) were examined using the OPD-Scan II (Nidek). Spherical aberration (Z4,0), vertical coma (Z3,−1), and horizontal coma (Z3,1) were calculated for a 5.0 mm pupil. Lens tilt and decentration was studied using a new Purkinjemeter.

Results

The mean horizontal optic tilt was 7.68° ± 5.16 (SD) and the mean vertical optic tilt was 3.01° ± 2.44 (SD). Horizontal decentration was 0.4 ± 0.33 mm (SD) and vertical decentration was 0.31 ± 0.21 mm (SD). Corneal aberrations were partially compensated by IOL- related aberrations.

Conclusions

Both tilt and decentration of sulcus-fixated IOLs exceeded the tolerable amounts evaluated for aspheric IOLs in eye model experiments. However, these spheric IOLs showed mitigation rather than accentuation of corneal wavefront aberrations in the rare event of capsular bag defect during cataract surgery.
Literature
1.
go back to reference Zuberbühler B, Menapace R, Neuhann T (2008) Kataraktchirurgie. Springer, Heidelberg Zuberbühler B, Menapace R, Neuhann T (2008) Kataraktchirurgie. Springer, Heidelberg
2.
go back to reference Mackool RJ, Russell RS (1995) Intracapsular posterior chamber intraocular lens insertion with posterior capsular tears or zonular instability. J Cataract Refract Surg 21(4):376–377PubMed Mackool RJ, Russell RS (1995) Intracapsular posterior chamber intraocular lens insertion with posterior capsular tears or zonular instability. J Cataract Refract Surg 21(4):376–377PubMed
3.
go back to reference Smiddy WE, Avery R (1991) Posterior chamber IOL implantation with suboptimal posterior capsular support. Ophthalmic Surg 22(1):16–19PubMed Smiddy WE, Avery R (1991) Posterior chamber IOL implantation with suboptimal posterior capsular support. Ophthalmic Surg 22(1):16–19PubMed
4.
go back to reference Wang L, Dai E, Koch DD, Nathoo A (2003) Optical aberrations of the human anterior cornea. J Cataract Refract Surg 29(8):1514–1521PubMedCrossRef Wang L, Dai E, Koch DD, Nathoo A (2003) Optical aberrations of the human anterior cornea. J Cataract Refract Surg 29(8):1514–1521PubMedCrossRef
5.
go back to reference Artal P, Berrio E, Guirao A, Piers P (2002) Contribution of the cornea and internal surfaces to the change of ocular aberrations with age. J Opt Soc Am A Opt Image Sci Vis 19(1):137–143PubMedCrossRef Artal P, Berrio E, Guirao A, Piers P (2002) Contribution of the cornea and internal surfaces to the change of ocular aberrations with age. J Opt Soc Am A Opt Image Sci Vis 19(1):137–143PubMedCrossRef
6.
go back to reference Oshika T, Klyce SD, Applegate RA, Howland HC (1999) Changes in corneal wavefront aberrations with aging. Invest Ophthalmol Vis Sci 40(7):1351–1355PubMed Oshika T, Klyce SD, Applegate RA, Howland HC (1999) Changes in corneal wavefront aberrations with aging. Invest Ophthalmol Vis Sci 40(7):1351–1355PubMed
7.
go back to reference Smith G, Cox MJ, Calver R, Garner LF (2001) The spherical aberration of the crystalline lens of the human eye. Vision Res 41(2):235–243PubMedCrossRef Smith G, Cox MJ, Calver R, Garner LF (2001) The spherical aberration of the crystalline lens of the human eye. Vision Res 41(2):235–243PubMedCrossRef
8.
go back to reference Mester U, Dillinger P, Anterist N (2003) Impact of a modified optic design on visual function: clinical comparative study. J Cataract Refract Surg 29(4):652–660PubMedCrossRef Mester U, Dillinger P, Anterist N (2003) Impact of a modified optic design on visual function: clinical comparative study. J Cataract Refract Surg 29(4):652–660PubMedCrossRef
9.
go back to reference Packer M, Fine IH, Hoffman RS (2004) Wavefront technology in cataract surgery. Curr Opin Ophthalmol 15(1):56–60PubMedCrossRef Packer M, Fine IH, Hoffman RS (2004) Wavefront technology in cataract surgery. Curr Opin Ophthalmol 15(1):56–60PubMedCrossRef
10.
go back to reference Millodot M, Sivak J (1979) Contribution of the cornea and lens to the spherical aberration of the eye. Vision Res 19(6):685–687PubMedCrossRef Millodot M, Sivak J (1979) Contribution of the cornea and lens to the spherical aberration of the eye. Vision Res 19(6):685–687PubMedCrossRef
11.
go back to reference Guirao A, Redondo M, Artal P (2000) Optical aberrations of the human cornea as a function of age. J Opt Soc Am A Opt Image Sci Vis 17(10):1697–1702PubMedCrossRef Guirao A, Redondo M, Artal P (2000) Optical aberrations of the human cornea as a function of age. J Opt Soc Am A Opt Image Sci Vis 17(10):1697–1702PubMedCrossRef
12.
go back to reference Barbero S, Marcos S, Jimenez-Alfaro I (2003) Optical aberrations of intraocular lenses measured in vivo and in vitro. J Opt Soc Am A Opt Image Sci Vis 20(10):1841–1851PubMedCrossRef Barbero S, Marcos S, Jimenez-Alfaro I (2003) Optical aberrations of intraocular lenses measured in vivo and in vitro. J Opt Soc Am A Opt Image Sci Vis 20(10):1841–1851PubMedCrossRef
13.
go back to reference Glasser A, Campbell MC (1998) Presbyopia and the optical changes in the human crystalline lens with age. Vision Res 38(2):209–229PubMedCrossRef Glasser A, Campbell MC (1998) Presbyopia and the optical changes in the human crystalline lens with age. Vision Res 38(2):209–229PubMedCrossRef
14.
go back to reference Holladay JT, Piers PA, Koranyi G, van der Mooren M, Norrby NE (2002) A new intraocular lens design to reduce spherical aberration of pseudophakic eyes. J Refract Surg 18(6):683–691PubMed Holladay JT, Piers PA, Koranyi G, van der Mooren M, Norrby NE (2002) A new intraocular lens design to reduce spherical aberration of pseudophakic eyes. J Refract Surg 18(6):683–691PubMed
15.
go back to reference Guirao A, Redondo M, Geraghty E, Piers P, Norrby S, Artal P (2002) Corneal optical aberrations and retinal image quality in patients in whom monofocal intraocular lenses were implanted. Arch Ophthalmol 120(9):1143–1151PubMedCrossRef Guirao A, Redondo M, Geraghty E, Piers P, Norrby S, Artal P (2002) Corneal optical aberrations and retinal image quality in patients in whom monofocal intraocular lenses were implanted. Arch Ophthalmol 120(9):1143–1151PubMedCrossRef
16.
17.
go back to reference Kershner RM (2003) Retinal image contrast and functional visual performance with aspheric, silicone, and acrylic intraocular lenses. Prospective evaluation. J Cataract Refract Surg 29(9):1684–1694PubMedCrossRef Kershner RM (2003) Retinal image contrast and functional visual performance with aspheric, silicone, and acrylic intraocular lenses. Prospective evaluation. J Cataract Refract Surg 29(9):1684–1694PubMedCrossRef
18.
go back to reference Kennis H, Huygens M, Callebaut F (2004) Comparing the contrast sensitivity of a modified prolate anterior surface IOL and of two spherical IOLs. Bull Soc Belge Ophtalmol 294:49–58PubMed Kennis H, Huygens M, Callebaut F (2004) Comparing the contrast sensitivity of a modified prolate anterior surface IOL and of two spherical IOLs. Bull Soc Belge Ophtalmol 294:49–58PubMed
19.
go back to reference Bellucci R, Scialdone A, Buratto L, Morselli S, Chierego C, Criscuoli A, Moretti G, Piers P (2005) Visual acuity and contrast sensitivity comparison between tecnis and AcrySof SA60AT intraocular lenses: a multicenter randomized study. J Cataract Refract Surg 31(4):712–717PubMedCrossRef Bellucci R, Scialdone A, Buratto L, Morselli S, Chierego C, Criscuoli A, Moretti G, Piers P (2005) Visual acuity and contrast sensitivity comparison between tecnis and AcrySof SA60AT intraocular lenses: a multicenter randomized study. J Cataract Refract Surg 31(4):712–717PubMedCrossRef
20.
go back to reference Sandoval HP, Fernandez de Castro LE, Vroman DT, Solomon KD (2008) Comparison of visual outcomes, photopic contrast sensitivity, wavefront analysis, and patient satisfaction following cataract extraction and IOL implantation: aspheric vs spherical acrylic lenses. Eye 22(12):1469–1475PubMedCrossRef Sandoval HP, Fernandez de Castro LE, Vroman DT, Solomon KD (2008) Comparison of visual outcomes, photopic contrast sensitivity, wavefront analysis, and patient satisfaction following cataract extraction and IOL implantation: aspheric vs spherical acrylic lenses. Eye 22(12):1469–1475PubMedCrossRef
21.
go back to reference Tzelikis PF, Akaishi L, Trindade FC, Boteon JE (2007) Ocular aberrations and contrast sensitivity after cataract surgery with AcrySof IQ intraocular lens implantation clinical comparative study. J Cataract Refract Surg 33(11):1918–1924PubMedCrossRef Tzelikis PF, Akaishi L, Trindade FC, Boteon JE (2007) Ocular aberrations and contrast sensitivity after cataract surgery with AcrySof IQ intraocular lens implantation clinical comparative study. J Cataract Refract Surg 33(11):1918–1924PubMedCrossRef
22.
go back to reference Piers PA, Weeber HA, Artal P, Norrby S (2007) Theoretical comparison of aberration-correcting customized and aspheric intraocular lenses. J Refract Surg 23(4):374–384PubMed Piers PA, Weeber HA, Artal P, Norrby S (2007) Theoretical comparison of aberration-correcting customized and aspheric intraocular lenses. J Refract Surg 23(4):374–384PubMed
23.
go back to reference Pieh S, Fiala W, Malz A, Stork W (2009) In vitro strehl ratios with spherical, aberration-free, average, and customized spherical aberration-correcting intraocular lenses. Invest Ophthalmol Vis Sci 50(3):1264–1270PubMedCrossRef Pieh S, Fiala W, Malz A, Stork W (2009) In vitro strehl ratios with spherical, aberration-free, average, and customized spherical aberration-correcting intraocular lenses. Invest Ophthalmol Vis Sci 50(3):1264–1270PubMedCrossRef
24.
go back to reference Eppig T, Scholz K, Loffler A, Messner A, Langenbucher A (2009) Effect of decentration and tilt on the image quality of aspheric intraocular lens designs in a model eye. J Cataract Refract Surg 35(6):1091–1100PubMedCrossRef Eppig T, Scholz K, Loffler A, Messner A, Langenbucher A (2009) Effect of decentration and tilt on the image quality of aspheric intraocular lens designs in a model eye. J Cataract Refract Surg 35(6):1091–1100PubMedCrossRef
25.
go back to reference Mester U, Sauer T, Kaymak H (2009) Decentration and tilt of a single-piece aspheric intraocular lens compared with the lens position in young phakic eyes. J Cataract Refract Surg 35(3):485–490PubMedCrossRef Mester U, Sauer T, Kaymak H (2009) Decentration and tilt of a single-piece aspheric intraocular lens compared with the lens position in young phakic eyes. J Cataract Refract Surg 35(3):485–490PubMedCrossRef
26.
go back to reference Baumeister M, Buhren J, Kohnen T (2009) Tilt and decentration of spherical and aspheric intraocular lenses: effect on higher-order aberrations. J Cataract Refract Surg 35(6):1006–1012PubMedCrossRef Baumeister M, Buhren J, Kohnen T (2009) Tilt and decentration of spherical and aspheric intraocular lenses: effect on higher-order aberrations. J Cataract Refract Surg 35(6):1006–1012PubMedCrossRef
27.
go back to reference Schaeffel F (2008) Binocular lens tilt and decentration measurements in healthy subjects with phakic eyes. Invest Ophthalmol Vis Sci 49(5):2216–2222PubMedCrossRef Schaeffel F (2008) Binocular lens tilt and decentration measurements in healthy subjects with phakic eyes. Invest Ophthalmol Vis Sci 49(5):2216–2222PubMedCrossRef
Metadata
Title
Tilt and decentration of an intraocular lens implanted in the ciliary sulcus after capsular bag defect during cataract surgery
Authors
Thomas Sauer
Ulrich Mester
Publication date
01-01-2013
Publisher
Springer-Verlag
Published in
Graefe's Archive for Clinical and Experimental Ophthalmology / Issue 1/2013
Print ISSN: 0721-832X
Electronic ISSN: 1435-702X
DOI
https://doi.org/10.1007/s00417-012-2141-2

Other articles of this Issue 1/2013

Graefe's Archive for Clinical and Experimental Ophthalmology 1/2013 Go to the issue