Three dimensional three component whole heart cardiovascular magnetic resonance velocity mapping: comparison of flow measurements from 3D and 2D acquisitions

Authors: Lau Brix, Steffen Ringgaard, Allan Rasmusson, Thomas Sangild Sørensen, W Yong Kim

Published in: Journal of Cardiovascular Magnetic Resonance | Issue 1/2009

Login to get access

Abstract

Background

Two-dimensional, unidirectionally encoded, cardiovascular magnetic resonance (CMR) velocity mapping is an established technique for the quantification of blood flow in large vessels. However, it requires an operator to correctly align the planes of acquisition. If all three directional components of velocity are measured for each voxel of a 3D volume through the phases of the cardiac cycle, blood flow through any chosen plane can potentially be calculated retrospectively. The initial acquisition is then more time consuming but relatively operator independent.

Aims

To compare the curves and volumes of flow derived from conventional 2D and comprehensive 3D flow acquisitions in a steady state flow model, and in vivo through planes transecting the ascending aorta and pulmonary trunk in 10 healthy volunteers.

Methods

Using a 1.5 T Phillips Intera CMR system, 3D acquisitions used an anisotropic 3D segmented k-space phase contrast gradient echo sequence with a short EPI readout, with prospective ECG and diaphragm navigator gating. The 2D acquisitions used segmented k-space phase contrast with prospective ECG and diaphragm navigator gating. Quantitative flow analyses were performed retrospectively with dedicated software for both the in vivo and in vitro acquisitions.

Results

Analysis of in vitro data found the 3D technique to have overestimated the continuous flow rate by approximately 5% across the entire applied flow range. In vivo, the 2D and the 3D techniques yielded similar volumetric flow curves and measurements. Aortic flow: (mean ± SD), 2D = 89.5 ± 13.5 ml & 3D = 92.7 ± 17.5 ml. Pulmonary flow: 2D = 98.8 ± 18.4 ml & 3D = 94.9 ± 19.0 ml). Each in vivo 3D acquisition took about 8 minutes or more.

Conclusion

Flow measurements derived from the 3D and 2D acquisitions were comparable. Although time consuming, comprehensive 3D velocity acquisition could be relatively operator independent, and could potentially yield information on flow through several retrospectively chosen planes, for example in patients with congenital or valvular heart disease.
Appendix
Available only for authorised users
Literature
1.
go back to reference Lotz J, Meier C, Leppert A, Galanski M: Cardiovascular flow measurement with phase-contrast MR imaging: basic facts and implementation. Radiographics. 2002, 22: 651-671.CrossRefPubMed Lotz J, Meier C, Leppert A, Galanski M: Cardiovascular flow measurement with phase-contrast MR imaging: basic facts and implementation. Radiographics. 2002, 22: 651-671.CrossRefPubMed
2.
go back to reference Underwood SR, Firmin DN, Klipstein RH, Rees RS, Longmore DB: Magnetic resonance velocity mapping: clinical application of a new technique. Br Heart J. 1987, 57: 404-412. 10.1136/hrt.57.5.404.PubMedCentralCrossRefPubMed Underwood SR, Firmin DN, Klipstein RH, Rees RS, Longmore DB: Magnetic resonance velocity mapping: clinical application of a new technique. Br Heart J. 1987, 57: 404-412. 10.1136/hrt.57.5.404.PubMedCentralCrossRefPubMed
3.
go back to reference Pelc NJ, Herfkens RJ, Shimakawa A, Enzmann DR: Phase contrast cine magnetic resonance imaging. Magn Reson Q. 1991, 7: 229-254.PubMed Pelc NJ, Herfkens RJ, Shimakawa A, Enzmann DR: Phase contrast cine magnetic resonance imaging. Magn Reson Q. 1991, 7: 229-254.PubMed
4.
go back to reference Wigstrom L, Sjoqvist L, Wranne B: Temporally resolved 3D phase-contrast imaging. Magn Reson Med. 1996, 36: 800-803. 10.1002/mrm.1910360521.CrossRefPubMed Wigstrom L, Sjoqvist L, Wranne B: Temporally resolved 3D phase-contrast imaging. Magn Reson Med. 1996, 36: 800-803. 10.1002/mrm.1910360521.CrossRefPubMed
5.
go back to reference Pelc NJ, Bernstein MA, Shimakawa A, Glover GH: Encoding strategies for three-direction phase-contrast MR imaging of flow. J Magn Reson Imaging. 1991, 1: 405-413. 10.1002/jmri.1880010404.CrossRefPubMed Pelc NJ, Bernstein MA, Shimakawa A, Glover GH: Encoding strategies for three-direction phase-contrast MR imaging of flow. J Magn Reson Imaging. 1991, 1: 405-413. 10.1002/jmri.1880010404.CrossRefPubMed
6.
go back to reference Sorensen TS, Beerbaum P, Korperich H, Pedersen EM: Three-dimensional, isotropic MRI: a unified approach to quantification and visualization in congenital heart disease. Int J Cardiovasc Imaging. 2005, 21: 283-292. 10.1007/s10554-004-4018-x.CrossRefPubMed Sorensen TS, Beerbaum P, Korperich H, Pedersen EM: Three-dimensional, isotropic MRI: a unified approach to quantification and visualization in congenital heart disease. Int J Cardiovasc Imaging. 2005, 21: 283-292. 10.1007/s10554-004-4018-x.CrossRefPubMed
7.
go back to reference Kilner PJ, Yang GZ, Wilkes AJ, Mohiaddin RH, Firmin DN, Yacoub MH: Asymmetric redirection of flow through the heart. Nature. 2000, 404: 759-761. 10.1038/35008075.CrossRefPubMed Kilner PJ, Yang GZ, Wilkes AJ, Mohiaddin RH, Firmin DN, Yacoub MH: Asymmetric redirection of flow through the heart. Nature. 2000, 404: 759-761. 10.1038/35008075.CrossRefPubMed
8.
go back to reference Kozerke S, Hasenkam JM, Pedersen EM, Boesiger P: Visualization of flow patterns distal to aortic valve prostheses in humans using a fast approach for cine 3D velocity mapping. J Magn Reson Imaging. 2001, 13: 690-698. 10.1002/jmri.1097.CrossRefPubMed Kozerke S, Hasenkam JM, Pedersen EM, Boesiger P: Visualization of flow patterns distal to aortic valve prostheses in humans using a fast approach for cine 3D velocity mapping. J Magn Reson Imaging. 2001, 13: 690-698. 10.1002/jmri.1097.CrossRefPubMed
9.
go back to reference Bogren HG, Mohiaddin RH, Kilner PJ, Jimenez-Borreguero LJ, Yang GZ, Firmin DN: Blood flow patterns in the thoracic aorta studied with three-directional MR velocity mapping: the effects of age and coronary artery disease. J Magn Reson Imaging. 1997, 7: 784-793. 10.1002/jmri.1880070504.CrossRefPubMed Bogren HG, Mohiaddin RH, Kilner PJ, Jimenez-Borreguero LJ, Yang GZ, Firmin DN: Blood flow patterns in the thoracic aorta studied with three-directional MR velocity mapping: the effects of age and coronary artery disease. J Magn Reson Imaging. 1997, 7: 784-793. 10.1002/jmri.1880070504.CrossRefPubMed
10.
go back to reference Hope TA, Markl M, Wigstrom L, Alley MT, Miller DC, Herfkens RJ: Comparison of flow patterns in ascending aortic aneurysms and volunteers using four-dimensional magnetic resonance velocity mapping. J Magn Reson Imaging. 2007, 26: 1471-1479. 10.1002/jmri.21082.CrossRefPubMed Hope TA, Markl M, Wigstrom L, Alley MT, Miller DC, Herfkens RJ: Comparison of flow patterns in ascending aortic aneurysms and volunteers using four-dimensional magnetic resonance velocity mapping. J Magn Reson Imaging. 2007, 26: 1471-1479. 10.1002/jmri.21082.CrossRefPubMed
11.
go back to reference Markl M, Harloff A, Bley TA, Zaitsev M, Jung B, Weigang E, Langer M, Hennig J, Frydrychowicz A: Time-resolved 3D MR velocity mapping at 3T: improved navigator-gated assessment of vascular anatomy and blood flow. J Magn Reson Imaging. 2007, 25: 824-831. 10.1002/jmri.20871.CrossRefPubMed Markl M, Harloff A, Bley TA, Zaitsev M, Jung B, Weigang E, Langer M, Hennig J, Frydrychowicz A: Time-resolved 3D MR velocity mapping at 3T: improved navigator-gated assessment of vascular anatomy and blood flow. J Magn Reson Imaging. 2007, 25: 824-831. 10.1002/jmri.20871.CrossRefPubMed
12.
go back to reference Buonocore MH: Visualizing blood flow patterns using streamlines, arrows, and particle paths. Magn Reson Med. 1998, 40: 210-226. 10.1002/mrm.1910400207.CrossRefPubMed Buonocore MH: Visualizing blood flow patterns using streamlines, arrows, and particle paths. Magn Reson Med. 1998, 40: 210-226. 10.1002/mrm.1910400207.CrossRefPubMed
13.
go back to reference Frydrychowicz A, Winterer JT, Zaitsev M, Jung B, Hennig J, Langer M, Markl M: Visualization of iliac and proximal femoral artery hemodynamics using time-resolved 3D phase contrast MRI at 3T. J Magn Reson Imaging. 2007, 25: 1085-1092. 10.1002/jmri.20900.CrossRefPubMed Frydrychowicz A, Winterer JT, Zaitsev M, Jung B, Hennig J, Langer M, Markl M: Visualization of iliac and proximal femoral artery hemodynamics using time-resolved 3D phase contrast MRI at 3T. J Magn Reson Imaging. 2007, 25: 1085-1092. 10.1002/jmri.20900.CrossRefPubMed
14.
go back to reference Bammer R, Hope TA, Aksoy M, Alley MT: Time-resolved 3D quantitative flow MRI of the major intracranial vessels: initial experience and comparative evaluation at 1.5 T and 3.0 T in combination with parallel imaging. Magn Reson Med. 2007, 57: 127-140. 10.1002/mrm.21109.PubMedCentralCrossRefPubMed Bammer R, Hope TA, Aksoy M, Alley MT: Time-resolved 3D quantitative flow MRI of the major intracranial vessels: initial experience and comparative evaluation at 1.5 T and 3.0 T in combination with parallel imaging. Magn Reson Med. 2007, 57: 127-140. 10.1002/mrm.21109.PubMedCentralCrossRefPubMed
15.
go back to reference Wetzel S, Meckel S, Frydrychowicz A, Bonati L, Radue EW, Scheffler K, Hennig J, Markl M: In vivo assessment and visualization of intracranial arterial hemodynamics with flow-sensitized 4D MR imaging at 3T. AJNR Am J Neuroradiol. 2007, 28: 433-438.PubMed Wetzel S, Meckel S, Frydrychowicz A, Bonati L, Radue EW, Scheffler K, Hennig J, Markl M: In vivo assessment and visualization of intracranial arterial hemodynamics with flow-sensitized 4D MR imaging at 3T. AJNR Am J Neuroradiol. 2007, 28: 433-438.PubMed
16.
go back to reference Bakker CJ, Hoogeveen RM, Viergever MA: Construction of a protocol for measuring blood flow by two-dimensional phase-contrast MRA. J Magn Reson Imaging. 1999, 9: 119-127. 10.1002/(SICI)1522-2586(199901)9:1<119::AID-JMRI16>3.0.CO;2-F.CrossRefPubMed Bakker CJ, Hoogeveen RM, Viergever MA: Construction of a protocol for measuring blood flow by two-dimensional phase-contrast MRA. J Magn Reson Imaging. 1999, 9: 119-127. 10.1002/(SICI)1522-2586(199901)9:1<119::AID-JMRI16>3.0.CO;2-F.CrossRefPubMed
17.
go back to reference Chatzimavroudis GP, Oshinski JN, Franch RH, Walker PG, Yoganathan AP, Pettigrew RI: Evaluation of the precision of magnetic resonance phase velocity mapping for blood flow measurements. J Cardiovasc Magn Reson. 2001, 3: 11-19. 10.1081/JCMR-100000142.CrossRefPubMed Chatzimavroudis GP, Oshinski JN, Franch RH, Walker PG, Yoganathan AP, Pettigrew RI: Evaluation of the precision of magnetic resonance phase velocity mapping for blood flow measurements. J Cardiovasc Magn Reson. 2001, 3: 11-19. 10.1081/JCMR-100000142.CrossRefPubMed
18.
go back to reference Markl M, Chan FP, Alley MT, Wedding KL, Draney MT, Elkins CJ, Parker DW, Wicker R, Taylor CA, Herfkens RJ, Pelc NJ: Time-resolved three-dimensional phase-contrast MRI. J Magn Reson Imaging. 2003, 17: 499-506. 10.1002/jmri.10272.CrossRefPubMed Markl M, Chan FP, Alley MT, Wedding KL, Draney MT, Elkins CJ, Parker DW, Wicker R, Taylor CA, Herfkens RJ, Pelc NJ: Time-resolved three-dimensional phase-contrast MRI. J Magn Reson Imaging. 2003, 17: 499-506. 10.1002/jmri.10272.CrossRefPubMed
19.
go back to reference Markl M, Draney MT, Hope MD, Levin JM, Chan FP, Alley MT, Pelc NJ, Herfkens RJ: Time-resolved 3-dimensional velocity mapping in the thoracic aorta: visualization of 3-directional blood flow patterns in healthy volunteers and patients. J Comput Assist Tomogr. 2004, 28: 459-468. 10.1097/00004728-200407000-00005.CrossRefPubMed Markl M, Draney MT, Hope MD, Levin JM, Chan FP, Alley MT, Pelc NJ, Herfkens RJ: Time-resolved 3-dimensional velocity mapping in the thoracic aorta: visualization of 3-directional blood flow patterns in healthy volunteers and patients. J Comput Assist Tomogr. 2004, 28: 459-468. 10.1097/00004728-200407000-00005.CrossRefPubMed
20.
go back to reference Sørensen TS, Körperich H, Greil GF, Eichhorn J, Barth P, Meyer H, Pedersen EM, Beerbaum P: Operator-independent isotropic three-dimensional magnetic resonance imaging for morphology in congenital heart disease: a validation study. Circulation. 2004, 110: 163-169. 10.1161/01.CIR.0000134282.35183.AD.CrossRefPubMed Sørensen TS, Körperich H, Greil GF, Eichhorn J, Barth P, Meyer H, Pedersen EM, Beerbaum P: Operator-independent isotropic three-dimensional magnetic resonance imaging for morphology in congenital heart disease: a validation study. Circulation. 2004, 110: 163-169. 10.1161/01.CIR.0000134282.35183.AD.CrossRefPubMed
21.
go back to reference Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P: SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999, 42: 952-962. 10.1002/(SICI)1522-2594(199911)42:5<952::AID-MRM16>3.0.CO;2-S.CrossRefPubMed Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P: SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999, 42: 952-962. 10.1002/(SICI)1522-2594(199911)42:5<952::AID-MRM16>3.0.CO;2-S.CrossRefPubMed
22.
go back to reference Kozerke S, Tsao J: Reduced data acquisition methods in cardiac imaging. Top Magn Reson Imaging. 2004, 15: 161-168. 10.1097/01.rmr.0000132789.84706.9d.CrossRefPubMed Kozerke S, Tsao J: Reduced data acquisition methods in cardiac imaging. Top Magn Reson Imaging. 2004, 15: 161-168. 10.1097/01.rmr.0000132789.84706.9d.CrossRefPubMed
23.
go back to reference Baltes C, Kozerke S, Hansen MS, Pruessmann KP, Tsao J, Boesiger P: Accelerating cine phase-contrast flow measurements using k-t BLAST and k-t SENSE. Magn Reson Med. 2005, 54: 1430-1438. 10.1002/mrm.20730.CrossRefPubMed Baltes C, Kozerke S, Hansen MS, Pruessmann KP, Tsao J, Boesiger P: Accelerating cine phase-contrast flow measurements using k-t BLAST and k-t SENSE. Magn Reson Med. 2005, 54: 1430-1438. 10.1002/mrm.20730.CrossRefPubMed
24.
go back to reference Hofman MB, Visser FC, van Rossum AC, Vink QM, Sprenger M, Westerhof N: In vivo validation of magnetic resonance blood volume flow measurements with limited spatial resolution in small vessels. Magn Reson Med. 1995, 33: 778-784. 10.1002/mrm.1910330606.CrossRefPubMed Hofman MB, Visser FC, van Rossum AC, Vink QM, Sprenger M, Westerhof N: In vivo validation of magnetic resonance blood volume flow measurements with limited spatial resolution in small vessels. Magn Reson Med. 1995, 33: 778-784. 10.1002/mrm.1910330606.CrossRefPubMed
Metadata
Title
Three dimensional three component whole heart cardiovascular magnetic resonance velocity mapping: comparison of flow measurements from 3D and 2D acquisitions
Authors
Lau Brix
Steffen Ringgaard
Allan Rasmusson
Thomas Sangild Sørensen
W Yong Kim
Publication date
01-12-2009
Publisher
BioMed Central
Published in
Journal of Cardiovascular Magnetic Resonance / Issue 1/2009
Electronic ISSN: 1532-429X
DOI
https://doi.org/10.1186/1532-429X-11-3

Other articles of this Issue 1/2009

Journal of Cardiovascular Magnetic Resonance 1/2009 Go to the issue