Abstract

Background

Laparoscopy requires a reliable light source to provide adequate visualization. However, thermal energy is produced as a by-product from the optical cable. This study attempts to quantify the degree of possible thermal damage secondary to the fiber-optic light source.

Methods

Using a digital thermometer, temperature measurements were recorded at the tip of optical cables from five different light sources (Karl Storz, Inc., Tuttlingen, Germany). Temperature measurements were recorded with new and old bulbs. The tip of the cable was applied to surgical drapes and the time to charring was recorded. Subsequently, the tip of the optical cable was applied to a porcine model and tissue samples were obtained after varying amounts of time (5, 15, 30, 60, and 90 s). Sections of the damaged tissue were prepared for microscopic evaluation. Parameters for thermal injury included extent of epidermal, dermal, and subcutaneous fat damage and necrosis. The lateral extent and depth of injury were measured.

Results

The maximum temperature at the tip of the optical cable varied between 119.5°C and 268.6°C. When surgical drapes were exposed to the tip of the light source, the time to char was 3–6 s. The degree and volume of injury increased with longer exposure times, and significant injury was recorded with the optical cable 3 mm from the skin.

Conclusions

This study demonstrates that the temperature at the tip of the optical light cord can induce extensive damage. The by-product of light, heat, can produce immediate superficial tissue necrosis that can extend into the subcutaneous fat even when the optical tip is not in direct contact with the skin. In addition, our study shows the variation in temperature that exists between light sources and bulb status. Overall, surgeons must realize and respect the potential complications associated with optical technology.
Literature
1.
go back to reference Emam TA, Cuschieri A (2003) How safe is high-power ultrasonic dissection? Ann Surg 237(2):186–191PubMedCrossRef Emam TA, Cuschieri A (2003) How safe is high-power ultrasonic dissection? Ann Surg 237(2):186–191PubMedCrossRef
2.
go back to reference Kim FJ et al (2007) Temperature safety profile of laparoscopic devices: Harmonic ACE (ACE), Ligasure V (LV), and plasma trisector (PT). Surg Endosc 22(6):1464–1469 Kim FJ et al (2007) Temperature safety profile of laparoscopic devices: Harmonic ACE (ACE), Ligasure V (LV), and plasma trisector (PT). Surg Endosc 22(6):1464–1469
3.
go back to reference Tulikangas PK et al (2001) Gross and histologic characteristics of laparoscopic injuries with four different energy sources. Fertil Steril 75(4):806–810PubMedCrossRef Tulikangas PK et al (2001) Gross and histologic characteristics of laparoscopic injuries with four different energy sources. Fertil Steril 75(4):806–810PubMedCrossRef
4.
go back to reference Yavuz Y et al (2006) Are cold light sources really cold? Surg Laparosc Endosc Percutan Tech 16(5):370–376PubMedCrossRef Yavuz Y et al (2006) Are cold light sources really cold? Surg Laparosc Endosc Percutan Tech 16(5):370–376PubMedCrossRef
6.
go back to reference Cotran KC (1999) Robbins Pathologic Basis of Disease, 16th edn. pp 433–434 Cotran KC (1999) Robbins Pathologic Basis of Disease, 16th edn. pp 433–434
7.
go back to reference Lee RC et al (1995) Biophysical mechanisms of cell membrane damage in electrical shock. Semin Neurol 15(4):367–374PubMedCrossRef Lee RC et al (1995) Biophysical mechanisms of cell membrane damage in electrical shock. Semin Neurol 15(4):367–374PubMedCrossRef
8.
go back to reference Sabiston, Townsend (2004) Sabiston Textbook of Surgery, 17th edn. pp 570–571 Sabiston, Townsend (2004) Sabiston Textbook of Surgery, 17th edn. pp 570–571
Metadata
Title
Thermal injury secondary to laparoscopic fiber-optic cables
Authors
A. Katharine Hindle
Fred Brody
Vernon Hopkins
Greg Rosales
Florencia Gonzalez
Arnold Schwartz
Publication date
01-08-2009
Publisher
Springer-Verlag
Published in
Surgical Endoscopy / Issue 8/2009
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-008-0219-z

Other articles of this Issue 8/2009

Surgical Endoscopy 8/2009 Go to the issue