The validity of a portable clinical force plate in assessment of static postural control: concurrent validity study

Authors: Samira Golriz, Jeffrey J Hebert, K Bo Foreman, Bruce F Walker

Published in: Chiropractic & Manual Therapies | Issue 1/2012

Login to get access

Abstract

Background

The broad use of force plates in clinical settings for postural control assessment suggests the need for instruments that are easy to use, affordable and readily available. In addition, these instruments of measurement should be reliable and valid as adequate reliability and validity are prerequisites to making correct inferences. The aim of this study was to examine the concurrent validity of postural control measures obtained with a clinical force plate.

Methods

Thirty-one healthy adults were recruited. Participants completed 1 set of 5 trials on each force plate. Postural control measures (centre of pressure [COP] average velocity and sway area) were collected and compared using the Midot Posture Scale Analyzer (clinical force plate) and the Accugait force plate (criterion measure). Intra class correlation coefficient (ICC), standard error of measurement , and paired t-tests were calculated and Bland-Altman plots were constructed to compare the force plates and assess consistency of measurement and agreement between them.

Results

The ICC values (ICC = 0.14-0.60) between the two force plates were lower than the acceptable value for both COP average velocity and sway area. There was significant difference (p > 0.05) in COP average velocity and sway area between the force plates. Examination of the plots revealed that there is less difference between the force plates in lower magnitudes of COP for average velocity and sway area however, the greater the average velocity and sway area, the greater the difference between the measures obtained from the two force plates.

Conclusion

Findings of this study showed poor concurrent validity of the clinical force plate. This clinical force plate cannot be a replacement for known reliable and valid force plates and consequently measures obtained from this force plate should be treated with caution especially in a clinical population.
Appendix
Available only for authorised users
Literature
1.
go back to reference Westcott SL, Lowes LP, Richardson PK: Evaluation of postural stability in children: Current theories and assessment tools. Phys Ther. 1997, 77 (6): 629-645.PubMed Westcott SL, Lowes LP, Richardson PK: Evaluation of postural stability in children: Current theories and assessment tools. Phys Ther. 1997, 77 (6): 629-645.PubMed
2.
go back to reference McGuine TA: Balance as a predictor of ankle injuries in high school basketball players. Clin J Sport Med. 2000, 10 (4): 239-244. 10.1097/00042752-200010000-00003.CrossRefPubMed McGuine TA: Balance as a predictor of ankle injuries in high school basketball players. Clin J Sport Med. 2000, 10 (4): 239-244. 10.1097/00042752-200010000-00003.CrossRefPubMed
3.
go back to reference Lafond D: Postural control during prolonged standing in persons with chronic low back pain. Gait Posture. 2009, 29 (3): 421-427. 10.1016/j.gaitpost.2008.10.064.CrossRefPubMed Lafond D: Postural control during prolonged standing in persons with chronic low back pain. Gait Posture. 2009, 29 (3): 421-427. 10.1016/j.gaitpost.2008.10.064.CrossRefPubMed
4.
go back to reference Rogind H: Comparison of Kistler 9861A force platform and Chattecx Balance System® for measurement of postural sway: Correlation and test-retest reliability. Scand J Med Sci Sports. 2003, 13 (2): 106-114. 10.1034/j.1600-0838.2003.01139.x.CrossRefPubMed Rogind H: Comparison of Kistler 9861A force platform and Chattecx Balance System® for measurement of postural sway: Correlation and test-retest reliability. Scand J Med Sci Sports. 2003, 13 (2): 106-114. 10.1034/j.1600-0838.2003.01139.x.CrossRefPubMed
5.
go back to reference Prieto TE: Measures of postural steadiness: Differences between healthy young and elderly adults. IEEE Trans Biomed Eng. 1996, 43 (9): 956-966. 10.1109/10.532130.CrossRefPubMed Prieto TE: Measures of postural steadiness: Differences between healthy young and elderly adults. IEEE Trans Biomed Eng. 1996, 43 (9): 956-966. 10.1109/10.532130.CrossRefPubMed
6.
go back to reference Guskiewicz KM, Perrin DH: Research and clinical applications of assessing balance. J Sport Rehabil. 1996, 5 (1): 45-63. Guskiewicz KM, Perrin DH: Research and clinical applications of assessing balance. J Sport Rehabil. 1996, 5 (1): 45-63.
7.
go back to reference Portney LG, Watkins MP: Foundations of clinical research: Application to practice. 2008, Prentice Hall Health, Upper Saddle River, 3 Portney LG, Watkins MP: Foundations of clinical research: Application to practice. 2008, Prentice Hall Health, Upper Saddle River, 3
8.
go back to reference Golriz S: Assessment of the reliability of the Midot posture scale analyser for evaluation of the postural stability parameters. Chiropractic and Manual Therapies, 2012 May, . 2012 Golriz S: Assessment of the reliability of the Midot posture scale analyser for evaluation of the postural stability parameters. Chiropractic and Manual Therapies, 2012 May, . 2012
9.
go back to reference Malina RM: Height, mass and skeletal maturity of elite Portuguese soccer players aged 11–16 years. J Sports Sci. 2000, 18 (9): 685-693. 10.1080/02640410050120069.CrossRefPubMed Malina RM: Height, mass and skeletal maturity of elite Portuguese soccer players aged 11–16 years. J Sports Sci. 2000, 18 (9): 685-693. 10.1080/02640410050120069.CrossRefPubMed
10.
go back to reference Era P: Postural balance in a random sample of 7,979 subjects aged 30 years and over. Gerontology. 2006, 52 (4): 204-213. 10.1159/000093652.CrossRefPubMed Era P: Postural balance in a random sample of 7,979 subjects aged 30 years and over. Gerontology. 2006, 52 (4): 204-213. 10.1159/000093652.CrossRefPubMed
11.
go back to reference AMTI: Calibration validation manual. I.A. Advanced Mechanical Technology, Watertown AMTI: Calibration validation manual. I.A. Advanced Mechanical Technology, Watertown
12.
go back to reference Winter DA: Biomechanics and motor control of human movement. 2009, John Wiley & Sons INC, , 4CrossRef Winter DA: Biomechanics and motor control of human movement. 2009, John Wiley & Sons INC, , 4CrossRef
13.
go back to reference AMTI: Balance software for AMTI's AccuSway plus balance platfrom. 2006, I. A. Advanced Medical Technology, Watertown AMTI: Balance software for AMTI's AccuSway plus balance platfrom. 2006, I. A. Advanced Medical Technology, Watertown
14.
go back to reference Donner A, Eliasziw M: Sample size requirements for reliability studies. Stat Med. 1987, 6 (4): 441-448. 10.1002/sim.4780060404.CrossRefPubMed Donner A, Eliasziw M: Sample size requirements for reliability studies. Stat Med. 1987, 6 (4): 441-448. 10.1002/sim.4780060404.CrossRefPubMed
15.
go back to reference Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986, 1 (8476): 307-310.CrossRefPubMed Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986, 1 (8476): 307-310.CrossRefPubMed
16.
go back to reference Landis JR, Koch GG: The measurement of observer agreement for categorical data. Biometrics. 1977, 33 (1): 159-174. 10.2307/2529310.CrossRefPubMed Landis JR, Koch GG: The measurement of observer agreement for categorical data. Biometrics. 1977, 33 (1): 159-174. 10.2307/2529310.CrossRefPubMed
17.
go back to reference Blaszczyk JW: Assessment of postural instability in patients with Parkinson's disease. Exp Brain Res. 2007, 183 (1): 107-114. 10.1007/s00221-007-1024-y.CrossRefPubMed Blaszczyk JW: Assessment of postural instability in patients with Parkinson's disease. Exp Brain Res. 2007, 183 (1): 107-114. 10.1007/s00221-007-1024-y.CrossRefPubMed
18.
go back to reference Ruhe A, Fejer R, Walker B: Center of pressure excursion as a measure of balance performance in patients with non-specific low back pain compared to healthy controls: a systematic review of the literature. Eur Spine J. 2011, 20 (3): 358-368. 10.1007/s00586-010-1543-2.PubMedCentralCrossRefPubMed Ruhe A, Fejer R, Walker B: Center of pressure excursion as a measure of balance performance in patients with non-specific low back pain compared to healthy controls: a systematic review of the literature. Eur Spine J. 2011, 20 (3): 358-368. 10.1007/s00586-010-1543-2.PubMedCentralCrossRefPubMed
Metadata
Title
The validity of a portable clinical force plate in assessment of static postural control: concurrent validity study
Authors
Samira Golriz
Jeffrey J Hebert
K Bo Foreman
Bruce F Walker
Publication date
01-12-2012
Publisher
BioMed Central
Published in
Chiropractic & Manual Therapies / Issue 1/2012
Electronic ISSN: 2045-709X
DOI
https://doi.org/10.1186/2045-709X-20-15

Other articles of this Issue 1/2012

Chiropractic & Manual Therapies 1/2012 Go to the issue