The prevalence and significance of renal perfusion defects in early kidney transplants quantified using 3D contrast enhanced ultrasound (CEUS)

Authors: Ben Stenberg, M. Wilkinson, S. Elliott, N. Caplan

Published in: European Radiology | Issue 11/2017

Login to get access

Abstract

Objectives

Vascular complications are one of the most common causes of early kidney transplant dysfunction. Contrast enhanced ultrasound increases sensitivity to vascular changes. The aim of this study was to assess the prevalence and size of vascular abnormalities in early renal transplants using 3D CEUS and the significance of perfusion defects on renal function.

Methods

Ninety-nine renal transplant patients underwent 3D CEUS after surgery to quantify perfusion defects as percentage total renal volume (TRV). Serum creatinine and estimated glomerular filtration rate (eGFR) were recorded up to 3 months post-surgery.

Results

Twenty participants had focal perfusion defects (0.2–43%TRV). There was a meaningful difference in patients with perfusion defects in eGFR at 1 month (90% CI 2.7–19.2 mL/min/1.73 m2) and 3 months (90% CI 1.9-19.6 mL/min/1.73 m2) and creatinine at 3 months (90% CI -56 – -8 μmol/L) using a predetermined clinical threshold. Perfusion defect size correlated well with both serum creatinine and eGFR at 3 months (R = 0.80, p ≤ 0.000 and 0.58, p = 0.038). No correlation was seen prior to 3 months.

Conclusions

Perfusion defects in kidney transplants were more common than expected and were highly likely to reduce renal function at 1-3 months, and the size of the defect affected the degree of functional change at 3 months.

Key Points

Perfusion defects were more common than previously thought.
Perfusion defects could be quantified using 3D CEUS.
The presence of even small perfusion defects may affect kidney function.
Size of perfusion defects correlated with subsequent kidney function at 3 months.
Potentially useful in informing clinician expectations of kidney function post-surgery.
Literature
1.
go back to reference Gang S, Rajapurkar M (2009) Vascular complications following renal transplantation. JNRT 2:122–132 Gang S, Rajapurkar M (2009) Vascular complications following renal transplantation. JNRT 2:122–132
2.
go back to reference Rao KV, Kjellstrand CM (1983) Post-transplant acute renal failure: a review. Clin Exp Dial Apheresis 7:127–143CrossRef Rao KV, Kjellstrand CM (1983) Post-transplant acute renal failure: a review. Clin Exp Dial Apheresis 7:127–143CrossRef
3.
go back to reference Sanches A, Etchebehere ECSC, Mazzali M. The accuracy of 99mTc-DTPA scintigraphy in the evaluation of acute renal graft complications. Int Braz J Urol. 2003;29(6)CrossRef Sanches A, Etchebehere ECSC, Mazzali M. The accuracy of 99mTc-DTPA scintigraphy in the evaluation of acute renal graft complications. Int Braz J Urol. 2003;29(6)CrossRef
4.
go back to reference Harvey CJ, Sidhu PS, Bachmann Nielsen M (2013) Contrast enhanced ultrasound in renal transplants: applications and future directions. Ultraschall Med 34:319–321CrossRef Harvey CJ, Sidhu PS, Bachmann Nielsen M (2013) Contrast enhanced ultrasound in renal transplants: applications and future directions. Ultraschall Med 34:319–321CrossRef
5.
go back to reference Fernandez CP, Ripolles T, Martinez MJ et al (2013) Diagnosis of acute cortical necrosis in renal transplantation by contrast-enhanced ultrasound: a preliminary experience. Ultraschall Med 34:340–344PubMed Fernandez CP, Ripolles T, Martinez MJ et al (2013) Diagnosis of acute cortical necrosis in renal transplantation by contrast-enhanced ultrasound: a preliminary experience. Ultraschall Med 34:340–344PubMed
6.
go back to reference Piscaglia F, Nolsøe C, Dietrich CF et al (2012) The EFSUMB guidelines and recommendations on the clinical practice of contrast enhanced ultrasound (CEUS): update 2011 on non-hepatic applications. Ultraschall Med 33:33–59CrossRef Piscaglia F, Nolsøe C, Dietrich CF et al (2012) The EFSUMB guidelines and recommendations on the clinical practice of contrast enhanced ultrasound (CEUS): update 2011 on non-hepatic applications. Ultraschall Med 33:33–59CrossRef
7.
go back to reference Grzelak P, Kurnatowska I, Sapieha M (2011) Ultrasonographic evaluation of disturbances in the perfusion of renal graft parenchyma as a result of acute occlusion of supernumerary arteries – a new application for contrast- enhanced ultrasonography. Ann Transplant 16:23–29CrossRef Grzelak P, Kurnatowska I, Sapieha M (2011) Ultrasonographic evaluation of disturbances in the perfusion of renal graft parenchyma as a result of acute occlusion of supernumerary arteries – a new application for contrast- enhanced ultrasonography. Ann Transplant 16:23–29CrossRef
8.
go back to reference Stenberg B, Chandler C, Wyrley-Birch H, Elliott ST (2014) Post-operative 3-dimensional contrast-enhanced ultrasound (CEUS) versus Tc99m-DTPA in the detection of post-surgical perfusion defects in kidney transplants – preliminary findings. Ultraschall Med 35:273–278PubMed Stenberg B, Chandler C, Wyrley-Birch H, Elliott ST (2014) Post-operative 3-dimensional contrast-enhanced ultrasound (CEUS) versus Tc99m-DTPA in the detection of post-surgical perfusion defects in kidney transplants – preliminary findings. Ultraschall Med 35:273–278PubMed
9.
go back to reference Partik BL et al (2002) 3D versus 2D ultrasound: accuracy of volume measurement in human cadaver kidneys. Investig Radiol 37:489–495CrossRef Partik BL et al (2002) 3D versus 2D ultrasound: accuracy of volume measurement in human cadaver kidneys. Investig Radiol 37:489–495CrossRef
10.
go back to reference Stenberg B, Elliott ST (2010) Diagnosis of gallbladder problems using three-dimensional ultrasound. Eur Radiol 20:908–914CrossRef Stenberg B, Elliott ST (2010) Diagnosis of gallbladder problems using three-dimensional ultrasound. Eur Radiol 20:908–914CrossRef
11.
go back to reference Stenberg B. MSc Thesis: 3DUS stacked contour volume assessment of regular and irregular shapes. University of Leeds; 2009 Stenberg B. MSc Thesis: 3DUS stacked contour volume assessment of regular and irregular shapes. University of Leeds; 2009
12.
go back to reference Kot BC, Sin DM, Ying M (2009) Evaluation of the accuracy and reliability of two 3-dimensional sonography methods in volume measurement of small structures: an in vitro phantom study. J Clin Ultrasound 37:82–88CrossRef Kot BC, Sin DM, Ying M (2009) Evaluation of the accuracy and reliability of two 3-dimensional sonography methods in volume measurement of small structures: an in vitro phantom study. J Clin Ultrasound 37:82–88CrossRef
13.
go back to reference McArthur C, Baxter G (2012) Current and potential renal applications of contrast-enhanced ultrasound. Clin Radiol 67:909–922CrossRef McArthur C, Baxter G (2012) Current and potential renal applications of contrast-enhanced ultrasound. Clin Radiol 67:909–922CrossRef
14.
go back to reference Levey AS, Bosch JP, Lewis JB et al (1999) A more accurate method to estimate glomerular filtration rate from serum creatinine: a new prediction equation. Modification of Diet in Renal Disease Study Group. Ann Intern Med 130:461–470CrossRef Levey AS, Bosch JP, Lewis JB et al (1999) A more accurate method to estimate glomerular filtration rate from serum creatinine: a new prediction equation. Modification of Diet in Renal Disease Study Group. Ann Intern Med 130:461–470CrossRef
15.
go back to reference Hopkins WG, Marshall SW, Batterham AM, Hanin J (2009) Progressive statistics for studies in sports medicine and exercise science. Med Sci Sports Exerc 41:3–12CrossRef Hopkins WG, Marshall SW, Batterham AM, Hanin J (2009) Progressive statistics for studies in sports medicine and exercise science. Med Sci Sports Exerc 41:3–12CrossRef
18.
go back to reference Kanchanabat B, Siddins M, Coates T et al (2002) Segmental infarction with graft dysfunction:an emerging syndrome in renal transplantation. Nephrol Dial Transplant 17:123–128CrossRef Kanchanabat B, Siddins M, Coates T et al (2002) Segmental infarction with graft dysfunction:an emerging syndrome in renal transplantation. Nephrol Dial Transplant 17:123–128CrossRef
19.
go back to reference Ghfir I, Boumaaza O, Ben RaisN. Contribution of 99mTc-DTPA scintigraphy with diuretic test in the exploration of acute urinary obstruction in a transplant patient. Internet J Nucl Med. 2007;5(1). Ghfir I, Boumaaza O, Ben RaisN. Contribution of 99mTc-DTPA scintigraphy with diuretic test in the exploration of acute urinary obstruction in a transplant patient. Internet J Nucl Med. 2007;5(1).
20.
go back to reference Lebranchu Y, Halimi JM, Bock A et al (2005) Delayed graft function: risk factors, consequences and parameters affecting outcome-results from MOST. A multinational observational study. Transplant Proc 37:345–347CrossRef Lebranchu Y, Halimi JM, Bock A et al (2005) Delayed graft function: risk factors, consequences and parameters affecting outcome-results from MOST. A multinational observational study. Transplant Proc 37:345–347CrossRef
21.
go back to reference Perico N, Cattaneo D, Sayegh MH, Remuzzi G (2004) Delayed graft function in kidney transplantation. Lancet 364:1814–1827CrossRef Perico N, Cattaneo D, Sayegh MH, Remuzzi G (2004) Delayed graft function in kidney transplantation. Lancet 364:1814–1827CrossRef
22.
go back to reference Metcalfe W (2007) How does early chronic kidney disease progress? Nephrol Dial Transplant 22:ix26–ix30CrossRef Metcalfe W (2007) How does early chronic kidney disease progress? Nephrol Dial Transplant 22:ix26–ix30CrossRef
23.
go back to reference Ahmad I (2004) Biopsy of the transplanted kidney. Semin Interv Radiol 21:275–281CrossRef Ahmad I (2004) Biopsy of the transplanted kidney. Semin Interv Radiol 21:275–281CrossRef
Metadata
Title
The prevalence and significance of renal perfusion defects in early kidney transplants quantified using 3D contrast enhanced ultrasound (CEUS)
Authors
Ben Stenberg
M. Wilkinson
S. Elliott
N. Caplan
Publication date
01-11-2017
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 11/2017
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-017-4871-3

Other articles of this Issue 11/2017

European Radiology 11/2017 Go to the issue