The mechanical and clinical influences of prosthetic index structure in Morse taper implant-abutment connection: a scoping review

Authors: Liheng Shen, Chengzhi Dong, Jianping Chen, Xiaolei Bai, Fan Yang, Linhong Wang

Published in: BMC Oral Health | Issue 1/2023

Login to get access

Abstract

Aim

The implant-abutment connection is a crucial factor in determining the long-term stability of dental implants. The use of a prosthetic index structure in the Morse taper implant-abutment connection has been proposed as a potential solution to improve the accuracy of this connection. This study aimed to provide a scoping review of the mechanical and clinical effects of the prosthetic index structure in the Morse taper implant-abutment connection.

Methods

A systematic scoping review of articles related to "dental implants," "Morse taper," and "index" was conducted using PubMed/MEDLINE, Web of Science, Cochrane, and Scopus databases, as well as a comprehensive literature search by two independent reviewers. Relevant articles were selected for analysis and discussion, with a specific focus on investigating the impact of prosthetic index structure on the mechanical and clinical aspects of Morse taper implant-abutment connections.

Results

Finally, a total of 16 articles that met the inclusion criteria were included for data extraction and review. In vitro studies have demonstrated that the use of a prosthetic index structure in the Morse taper implant-abutment connection can affect stress distribution, biomechanical stability, and reverse torque values, which may reduce stress within cancellous bone and help limit crestal bone resorption. However, retrospective clinical studies have shown that this structure is also associated with a higher risk of mechanical complications, such as abutment fracture and abutment screw loosening.

Conclusions

Therefore, the clinical trade-off between preventing crestal bone resorption and mechanical complications must be carefully considered when selecting appropriate abutments. The findings suggest that this structure can improve the accuracy and stability of the implant-abutment connection, but its use should be carefully evaluated in clinical practice.
Appendix
Available only for authorised users
Literature
5.
go back to reference Galindo-Moreno P, Concha-Jeronimo A, Lopez-Chaichio L, Rodriguez-Alvarez R, Sanchez-Fernandez E, Padial-Molina M. Marginal Bone Loss around Implants with Internal Hexagonal and Internal Conical Connections: A 12-Month Randomized Pilot Study. J Clin Med. 2021;10(22):5427. https://doi.org/10.3390/jcm10225427 Galindo-Moreno P, Concha-Jeronimo A, Lopez-Chaichio L, Rodriguez-Alvarez R, Sanchez-Fernandez E, Padial-Molina M. Marginal Bone Loss around Implants with Internal Hexagonal and Internal Conical Connections: A 12-Month Randomized Pilot Study. J Clin Med. 2021;10(22):5427. https://​doi.​org/​10.​3390/​jcm10225427
7.
go back to reference Larrucea Verdugo C, Jaramillo Núñez G, Acevedo Avila A, Larrucea San Martín C. Microleakage of the prosthetic abutment/implant interface with internal and external connection: in vitro study. Clin Oral Implants Res. 2014;25(9):1078–83. https://doi.org/10.1111/clr.12217. Larrucea Verdugo C, Jaramillo Núñez G, Acevedo Avila A, Larrucea San Martín C. Microleakage of the prosthetic abutment/implant interface with internal and external connection: in vitro study. Clin Oral Implants Res. 2014;25(9):1078–83. https://​doi.​org/​10.​1111/​clr.​12217.
10.
go back to reference Ding TA, Woody RD, Higginbottom FL, Miller BH. Evaluation of the ITI Morse taper implant/abutment design with an internal modification. Int J Oral Maxillofac Implants. 2003;18(6):865–72.PubMed Ding TA, Woody RD, Higginbottom FL, Miller BH. Evaluation of the ITI Morse taper implant/abutment design with an internal modification. Int J Oral Maxillofac Implants. 2003;18(6):865–72.PubMed
13.
go back to reference Martins CM, Ramos EV, Kreve S, et al. Reverse torque evaluation in indexed and nonindexed abutments of Morse Taper implants in a mechanical fatigue test. Dent Res J (Isfahan). 2019;16(2):110–6.CrossRefPubMed Martins CM, Ramos EV, Kreve S, et al. Reverse torque evaluation in indexed and nonindexed abutments of Morse Taper implants in a mechanical fatigue test. Dent Res J (Isfahan). 2019;16(2):110–6.CrossRefPubMed
21.
go back to reference Sutter FHPW. The new restorative concept of the Iti dental implant system: design and engineering. Int J Periodontics Restorative Dent. 1993;13(5):408–31. Sutter FHPW. The new restorative concept of the Iti dental implant system: design and engineering. Int J Periodontics Restorative Dent. 1993;13(5):408–31.
26.
go back to reference da Silva-Neto JP, Nóbilo MA, Penatti MP, Simamoto PC Jr, das Neves FD. Influence of methodologic aspects on the results of implant-abutment interface microleakage tests: a critical review of in vitro studies. Int J Oral Maxillofac Implants. 2012;27(4):793–800. da Silva-Neto JP, Nóbilo MA, Penatti MP, Simamoto PC Jr, das Neves FD. Influence of methodologic aspects on the results of implant-abutment interface microleakage tests: a critical review of in vitro studies. Int J Oral Maxillofac Implants. 2012;27(4):793–800.
28.
go back to reference Zancopé K, Dias Resende CC, Castro CG, Salatti RC, Domingues das Neves F. Influence of the Prosthetic Index on Fracture Resistance of Morse Taper Dental Implants. Int J Oral Maxillofac Implants. 2017;32(6):1333–1337. https://doi.org/10.11607/jomi.4658 Zancopé K, Dias Resende CC, Castro CG, Salatti RC, Domingues das Neves F. Influence of the Prosthetic Index on Fracture Resistance of Morse Taper Dental Implants. Int J Oral Maxillofac Implants. 2017;32(6):1333–1337. https://​doi.​org/​10.​11607/​jomi.​4658
31.
go back to reference Zancopé K, Resende CC, Tavares LN, Neves FD. Influence of indexed abutments on the fracture resistance of internal conical dental implants. Gen Dent. 2017;65(1):e14–6.PubMed Zancopé K, Resende CC, Tavares LN, Neves FD. Influence of indexed abutments on the fracture resistance of internal conical dental implants. Gen Dent. 2017;65(1):e14–6.PubMed
33.
go back to reference Hyun DG, Kwon HB, Lim YJ, Koak JY, Kim MJ. The Influence of a Positioning Hex on Abutment Rotation in Tapered Internal Implants: A 3D Finite Element Model Study. Int J Oral Maxillofac Implants. 2020;35(2):281–288. https://doi.org/10.11607/jomi.7673 Hyun DG, Kwon HB, Lim YJ, Koak JY, Kim MJ. The Influence of a Positioning Hex on Abutment Rotation in Tapered Internal Implants: A 3D Finite Element Model Study. Int J Oral Maxillofac Implants. 2020;35(2):281–288. https://​doi.​org/​10.​11607/​jomi.​7673
34.
35.
go back to reference Nokar S, Hajimiragha H, Sadighpour L, Mostafavi AS. Evaluation of reverse torque values and failure loads of three different abutment designs with internal connections. Dent Res J (Isfahan). 2020;17:439–46.CrossRefPubMed Nokar S, Hajimiragha H, Sadighpour L, Mostafavi AS. Evaluation of reverse torque values and failure loads of three different abutment designs with internal connections. Dent Res J (Isfahan). 2020;17:439–46.CrossRefPubMed
Metadata
Title
The mechanical and clinical influences of prosthetic index structure in Morse taper implant-abutment connection: a scoping review
Authors
Liheng Shen
Chengzhi Dong
Jianping Chen
Xiaolei Bai
Fan Yang
Linhong Wang
Publication date
01-12-2023
Publisher
BioMed Central
Published in
BMC Oral Health / Issue 1/2023
Electronic ISSN: 1472-6831
DOI
https://doi.org/10.1186/s12903-023-03545-3

Other articles of this Issue 1/2023

BMC Oral Health 1/2023 Go to the issue