Published in: Journal of Orthopaedic Science 2/2013

01-03-2013 | Original Article

The influence of head diameter and wall thickness on deformations of metallic acetabular press-fit cups and UHMWPE liners: a finite element analysis

Authors: Paul Goebel, Daniel Kluess, Jan Wieding, Robert Souffrant, Horst Heyer, Manuela Sander, Rainer Bader

Published in: Journal of Orthopaedic Science | Issue 2/2013

Login to get access

Abstract

Background

To increase the range of motion of total hip endoprostheses, prosthetic heads need to be enlarged, which implies that the cup and/or liner thickness must decrease. This may have negative effects on the wear rate, because the acetabular cups and liners could deform during press-fit implantation and hip joint loading. We compared the metal cup and polyethylene liner deformations that occurred when different wall thicknesses were used in order to evaluate the resulting changes in the clearance of the articulating region.

Methods

A parametric finite element model utilized three cup and liner wall thicknesses to analyze cup and liner deformations after press-fit implantation into the pelvic bone. The resultant hip joint force during heel strike was applied while the femur was fixed, accounting for physiological muscle forces. The deformation behavior of the liner under joint loading was therefore assessed as a function of the head diameter and the resulting clearance.

Results

Press-fit implantation showed diametral cup deformations of 0.096, 0.034, and 0.014 mm for cup wall thicknesses of 3, 5, and 7 mm, respectively. The largest deformations (average 0.084 ± 0.003 mm) of liners with thicknesses of 4, 6, and 8 mm occurred with the smallest cup wall thickness (3 mm). The smallest liner deformation (0.011 mm) was obtained with largest cup and liner wall thicknesses. Under joint loading, liner deformations in thin-walled acetabular cups (3 mm) reduced the initial clearance by about 50 %.

Conclusion

Acetabular press-fit cups with wall thicknesses of ≤5 mm should only be used in combination with polyethylene liners >6 mm thick in order to minimize the reduction in clearance.
Literature
1.
go back to reference Krukemeyer M, Möllenhoff G. Endoprothetik. Leitfäden für Praktiker. Berlin: de Gruyter; 2009. Krukemeyer M, Möllenhoff G. Endoprothetik. Leitfäden für Praktiker. Berlin: de Gruyter; 2009.
2.
go back to reference Widmer KH, Zurfluh B, Morscher EW. Load transfer and fixation mode of press-fit acetabular sockets. J Arthroplast. 2002;17(7):926–35.CrossRef Widmer KH, Zurfluh B, Morscher EW. Load transfer and fixation mode of press-fit acetabular sockets. J Arthroplast. 2002;17(7):926–35.CrossRef
3.
go back to reference Maloney WJ. Orthopaedic crossfire—larger femoral heads: a triumph of hope over reason! In opposition. J Arthroplast. 2003;18(3 Suppl 1):85–7.CrossRef Maloney WJ. Orthopaedic crossfire—larger femoral heads: a triumph of hope over reason! In opposition. J Arthroplast. 2003;18(3 Suppl 1):85–7.CrossRef
4.
go back to reference Hermida JC, Bergula A, Chen P, Colwell CW Jr, D’Lima DD. Comparison of the wear rates of twenty-eight and thirty-two-millimeter femoral heads on cross-linked polyethylene acetabular cups in a wear simulator. J Bone Joint Surg Am. 2003;85-A(12):2325–31.PubMed Hermida JC, Bergula A, Chen P, Colwell CW Jr, D’Lima DD. Comparison of the wear rates of twenty-eight and thirty-two-millimeter femoral heads on cross-linked polyethylene acetabular cups in a wear simulator. J Bone Joint Surg Am. 2003;85-A(12):2325–31.PubMed
5.
go back to reference Kluess D, Martin H, Mittelmeier W, Schmitz KP, Bader R. Influence of femoral head size on impingement, dislocation and stress distribution in total hip replacement. Med Eng Phys. 2007;29(4):465–71.PubMedCrossRef Kluess D, Martin H, Mittelmeier W, Schmitz KP, Bader R. Influence of femoral head size on impingement, dislocation and stress distribution in total hip replacement. Med Eng Phys. 2007;29(4):465–71.PubMedCrossRef
6.
go back to reference Teoh SH, Chan WH, Thampuran R. An elasto-plastic finite element model for polyethylene wear in total hip arthroplasty. J Biomech. 2002;35(3):323–30.PubMedCrossRef Teoh SH, Chan WH, Thampuran R. An elasto-plastic finite element model for polyethylene wear in total hip arthroplasty. J Biomech. 2002;35(3):323–30.PubMedCrossRef
7.
go back to reference Ong KL, Rundell S, Liepins I, Laurent R, Markel D, Kurtz SM. Biomechanical modeling of acetabular component polyethylene stresses, fracture risk, and wear rate following press-fit implantation. J Orthop Res. 2009;27(11):1467–72.PubMedCrossRef Ong KL, Rundell S, Liepins I, Laurent R, Markel D, Kurtz SM. Biomechanical modeling of acetabular component polyethylene stresses, fracture risk, and wear rate following press-fit implantation. J Orthop Res. 2009;27(11):1467–72.PubMedCrossRef
8.
go back to reference Mai MT, Schmalzried TP, Dorey FJ, Campbell PA, Amstutz HC. The contribution of frictional torque to loosening at the cement–bone interface in Tharies hip replacements. J Bone Joint Surg Am. 1996;78(4):505–11. Mai MT, Schmalzried TP, Dorey FJ, Campbell PA, Amstutz HC. The contribution of frictional torque to loosening at the cement–bone interface in Tharies hip replacements. J Bone Joint Surg Am. 1996;78(4):505–11.
9.
go back to reference Hamilton WG, Calendine CL, Beykirch SE, Hopper RH Jr, Engh CA. Acetabular fixation options: first-generation modular cup curtain calls and caveats. J Arthroplast. 2007;22(4 Suppl 1):75–81.CrossRef Hamilton WG, Calendine CL, Beykirch SE, Hopper RH Jr, Engh CA. Acetabular fixation options: first-generation modular cup curtain calls and caveats. J Arthroplast. 2007;22(4 Suppl 1):75–81.CrossRef
10.
go back to reference Jin ZM, Meakins S, Morlock MM, Parsons P, Hardaker C, Flett M, Isaac G. Deformation of press-fitted metallic resurfacing cups. Part 1: experimental simulation. Proc Inst Mech Eng H. 2006;220(2):299–309.CrossRef Jin ZM, Meakins S, Morlock MM, Parsons P, Hardaker C, Flett M, Isaac G. Deformation of press-fitted metallic resurfacing cups. Part 1: experimental simulation. Proc Inst Mech Eng H. 2006;220(2):299–309.CrossRef
11.
go back to reference Schmidig G, Patel A, Liepins I, Thakore M, Markel DC. The effects of acetabular shell deformation and liner thickness on frictional torque in ultrahigh-molecular-weight polyethylene acetabular bearings. J Arthroplast. 2010;25(4):644–53.CrossRef Schmidig G, Patel A, Liepins I, Thakore M, Markel DC. The effects of acetabular shell deformation and liner thickness on frictional torque in ultrahigh-molecular-weight polyethylene acetabular bearings. J Arthroplast. 2010;25(4):644–53.CrossRef
12.
go back to reference Yew A, Jin ZM, Donn A, Morlock MM, Isaac G. Deformation of press-fitted metallic resurfacing cups. Part 2: finite element simulation. Proc Inst Mech Eng H. 2006;220(2):311–9.PubMedCrossRef Yew A, Jin ZM, Donn A, Morlock MM, Isaac G. Deformation of press-fitted metallic resurfacing cups. Part 2: finite element simulation. Proc Inst Mech Eng H. 2006;220(2):311–9.PubMedCrossRef
13.
go back to reference Fritsche A, Bialek K, Mittelmeier W, Simnacher M, Fethke K, Wree A, Bader R. Experimental investigations of the insertion and deformation behavior of press-fit and threaded acetabular cups for total hip replacement. J Orthop Sci. 2008;13(3):240–7.PubMedCrossRef Fritsche A, Bialek K, Mittelmeier W, Simnacher M, Fethke K, Wree A, Bader R. Experimental investigations of the insertion and deformation behavior of press-fit and threaded acetabular cups for total hip replacement. J Orthop Sci. 2008;13(3):240–7.PubMedCrossRef
14.
go back to reference Crowninshield RD, Maloney WJ, Wentz DH, Humphrey SM, Blanchard CR. Biomechanics of large femoral heads: what they do and don’t do. Clin Orthop Relat Res. 2004;429:102–7.PubMedCrossRef Crowninshield RD, Maloney WJ, Wentz DH, Humphrey SM, Blanchard CR. Biomechanics of large femoral heads: what they do and don’t do. Clin Orthop Relat Res. 2004;429:102–7.PubMedCrossRef
15.
go back to reference Bergmann G, Deuretzbacher G, Heller M, Graichen F, Rohlmann A, Strauss J, Duda GN. Hip contact forces and gait patterns from routine activities. J Biomech. 2001;34(7):859–71.PubMedCrossRef Bergmann G, Deuretzbacher G, Heller M, Graichen F, Rohlmann A, Strauss J, Duda GN. Hip contact forces and gait patterns from routine activities. J Biomech. 2001;34(7):859–71.PubMedCrossRef
16.
go back to reference Dalstra M, Huiskes R, van Erning L. Development and validation of a three-dimensional finite element model of the pelvic bone. J Biomech Eng. 1995;117(3):272–8.PubMedCrossRef Dalstra M, Huiskes R, van Erning L. Development and validation of a three-dimensional finite element model of the pelvic bone. J Biomech Eng. 1995;117(3):272–8.PubMedCrossRef
17.
go back to reference Garcia JM, Doblare M, Seral B, Seral F, Palanca D, Gracia L. Three-dimensional finite element analysis of several internal and external pelvis fixations. J Biomech Eng. 2000;122(5):516–22.PubMedCrossRef Garcia JM, Doblare M, Seral B, Seral F, Palanca D, Gracia L. Three-dimensional finite element analysis of several internal and external pelvis fixations. J Biomech Eng. 2000;122(5):516–22.PubMedCrossRef
18.
go back to reference ASTM International. ASTM-D695. Standard test method for compressive properties of rigid plastics. West Conshohocken: ASTM International; 2010. ASTM International. ASTM-D695. Standard test method for compressive properties of rigid plastics. West Conshohocken: ASTM International; 2010.
19.
go back to reference ASTM International. ASTM-D1621. Standard test method for compressive properties of rigid cellular plastics. West Conshohocken: ASTM International; 2010. ASTM International. ASTM-D1621. Standard test method for compressive properties of rigid cellular plastics. West Conshohocken: ASTM International; 2010.
20.
go back to reference Zhang C, Leng Y, Chen J. Elastic and plastic behavior of plasma-sprayed hydroxyapatite coatings on a Ti-6Al-4V substrate. Biomaterials. 2001;22(11):1357–63.PubMedCrossRef Zhang C, Leng Y, Chen J. Elastic and plastic behavior of plasma-sprayed hydroxyapatite coatings on a Ti-6Al-4V substrate. Biomaterials. 2001;22(11):1357–63.PubMedCrossRef
21.
go back to reference Burger W, Richter HG. High strength and toughness alumina matrix composites by transformation toughening and “in situ” platelet reinforcement (ZPTA)—the new generation of bioceramic. Bioceramics. 2001;13:545–8. Burger W, Richter HG. High strength and toughness alumina matrix composites by transformation toughening and “in situ” platelet reinforcement (ZPTA)—the new generation of bioceramic. Bioceramics. 2001;13:545–8.
22.
go back to reference Kluess D, Souffrant R, Mittelmeier W, Wree A, Schmitz KP, Bader R. A convenient approach for finite-element-analyses of orthopaedic implants in bone contact: modeling and experimental validation. Comput Methods Programs Biomed. 2009;95(1):23–30.PubMedCrossRef Kluess D, Souffrant R, Mittelmeier W, Wree A, Schmitz KP, Bader R. A convenient approach for finite-element-analyses of orthopaedic implants in bone contact: modeling and experimental validation. Comput Methods Programs Biomed. 2009;95(1):23–30.PubMedCrossRef
23.
go back to reference Speirs AD, Heller MO, Duda GN, Taylor WR. Physiologically based boundary conditions in finite element modelling. J Biomech. 2007;40(10):2318–23.PubMedCrossRef Speirs AD, Heller MO, Duda GN, Taylor WR. Physiologically based boundary conditions in finite element modelling. J Biomech. 2007;40(10):2318–23.PubMedCrossRef
24.
go back to reference Heller MO, Bergmann G, Kassi JP, Claes L, Haas NP, Duda GN. Determination of muscle loading at the hip joint for use in pre-clinical testing. J Biomech. 2005;38(5):1155–63.PubMedCrossRef Heller MO, Bergmann G, Kassi JP, Claes L, Haas NP, Duda GN. Determination of muscle loading at the hip joint for use in pre-clinical testing. J Biomech. 2005;38(5):1155–63.PubMedCrossRef
25.
go back to reference Heiner AD. Structural properties of fourth-generation composite femurs and tibias. J Biomech. 2008;41(15):3282–4.PubMedCrossRef Heiner AD. Structural properties of fourth-generation composite femurs and tibias. J Biomech. 2008;41(15):3282–4.PubMedCrossRef
26.
go back to reference Kim JE, Li Z, Ito Y, Huber CD, Shih AM, Eberhardt AW, Yang KH, King AI, Soni BK. Finite element model development of a child pelvis with optimization-based material identification. J Biomech. 2009;42(13):2191–5.PubMedCrossRef Kim JE, Li Z, Ito Y, Huber CD, Shih AM, Eberhardt AW, Yang KH, King AI, Soni BK. Finite element model development of a child pelvis with optimization-based material identification. J Biomech. 2009;42(13):2191–5.PubMedCrossRef
27.
go back to reference Polgar K, Viceconti M, O’Connor JJ. A comparison between automatically generated linear and parabolic tetrahedra when used to mesh a human femur. Proc Inst Mech Eng H. 2001;215(1):85–94.PubMedCrossRef Polgar K, Viceconti M, O’Connor JJ. A comparison between automatically generated linear and parabolic tetrahedra when used to mesh a human femur. Proc Inst Mech Eng H. 2001;215(1):85–94.PubMedCrossRef
28.
go back to reference Jin ZM, Dowson D, Fisher J. Analysis of fluid film lubrication in artificial hip joint replacements with surfaces of high elastic modulus. Proc Inst Mech Eng H. 1997;211(3):247–56.PubMedCrossRef Jin ZM, Dowson D, Fisher J. Analysis of fluid film lubrication in artificial hip joint replacements with surfaces of high elastic modulus. Proc Inst Mech Eng H. 1997;211(3):247–56.PubMedCrossRef
29.
go back to reference Wang A, Essner A, Klein R. Effect of contact stress on friction and wear of ultra-high molecular weight polyethylene in total hip replacement. Proc Inst Mech Eng H. 2001;215(2):133–9.PubMed Wang A, Essner A, Klein R. Effect of contact stress on friction and wear of ultra-high molecular weight polyethylene in total hip replacement. Proc Inst Mech Eng H. 2001;215(2):133–9.PubMed
30.
go back to reference Bader R, Scholz R, Steinhauser E, Zimmermann S, Busch R, Mittelmeier W. The influence of head and neck geometry on stability of total hip replacement: a mechanical test study. Acta Orthop Scand. 2004;75(4):415–21.PubMedCrossRef Bader R, Scholz R, Steinhauser E, Zimmermann S, Busch R, Mittelmeier W. The influence of head and neck geometry on stability of total hip replacement: a mechanical test study. Acta Orthop Scand. 2004;75(4):415–21.PubMedCrossRef
31.
go back to reference Waewsawangwong W, Goodman SB. Unexpected failure of highly cross-linked polyethylene acetabular liner. J Arthroplast. 2012;27(2):323.e1–4. Waewsawangwong W, Goodman SB. Unexpected failure of highly cross-linked polyethylene acetabular liner. J Arthroplast. 2012;27(2):323.e1–4.
32.
go back to reference Tower SS, Currier JH, Currier BH, Lyford KA, Van Citters DW, Mayor MB. Rim cracking of the cross-linked longevity polyethylene acetabular liner after total hip arthroplasty. J Bone Joint Surg Am. 2007;89(10):2212–7.PubMedCrossRef Tower SS, Currier JH, Currier BH, Lyford KA, Van Citters DW, Mayor MB. Rim cracking of the cross-linked longevity polyethylene acetabular liner after total hip arthroplasty. J Bone Joint Surg Am. 2007;89(10):2212–7.PubMedCrossRef
Metadata
Title
The influence of head diameter and wall thickness on deformations of metallic acetabular press-fit cups and UHMWPE liners: a finite element analysis
Authors
Paul Goebel
Daniel Kluess
Jan Wieding
Robert Souffrant
Horst Heyer
Manuela Sander
Rainer Bader
Publication date
01-03-2013
Publisher
Springer Japan
Published in
Journal of Orthopaedic Science / Issue 2/2013
Print ISSN: 0949-2658
Electronic ISSN: 1436-2023
DOI
https://doi.org/10.1007/s00776-012-0340-7