Skip to main content
Top
Published in: Surgical Endoscopy 7/2011

01-07-2011 | Technique

The importance of precompression time for secure stapling with a linear stapler

Authors: Shinya Nakayama, Suguru Hasegawa, Satoshi Nagayama, Shigeru Kato, Koya Hida, Eiji Tanaka, Atsushi Itami, Hajime Kubo, Yoshiharu Sakai

Published in: Surgical Endoscopy | Issue 7/2011

Login to get access

Abstract

Background

One prerequisite for the safe stapling of anastomoses is the formation of perfectly aligned B-shaped staples. Using an animal model, this study investigated whether precompression enhances secure staple formation.

Methods

A 45-mm linear stapler with a blue cartridge (staple leg length, 3.5 mm) was used on three portions each of six porcine stomachs (cardia, center, and pylorus). Staple shape and height were examined according to the precompression time (0, 1, or 5 min) before firing. The shape was classified as optimal or suboptimal, and the heights of the staples were measured individually. The completeness of the proximal staple lines also was compared with the distal lines.

Results

The optimal staple rate in the 5-min group (52.7%) was significantly higher than in the 1-min group (28.7%; p < 0.001) or the 0-min group (17.1%; p = 0.002). The optimal staple rate in the 5-min group for the cardiac portion (mean wall thickness, 2.7 mm) reached 98.9%. However, this rate was less than 50% in the center (5.3 mm) and pylorus (4.2 mm) portions regardless of the precompression time. Compared with the root side, the top side of the cartridge had a lower optimal rate (45.3 vs. 18.7%; p < 0.001). A high correlation of completeness was observed between the distal and proximal sides of the stump.

Conclusions

When a linear stapling device is used, proper staple formation is correlated with intestinal wall thickness, and a sufficient amount of precompression time is effective for gaining the secure staple formation.
Literature
1.
go back to reference Ravitch MM, Brown IW, Daviglus GF (1959) Experimental and clinical use of the Soviet bronchus stapling instruments. Surgery 46:97–108PubMed Ravitch MM, Brown IW, Daviglus GF (1959) Experimental and clinical use of the Soviet bronchus stapling instruments. Surgery 46:97–108PubMed
3.
go back to reference Manufacturer’s instruction for use by Ethicon Endo-Surgery Inc., Cincinnati, OH, USA Manufacturer’s instruction for use by Ethicon Endo-Surgery Inc., Cincinnati, OH, USA
4.
go back to reference Wind J, Safiruddin F, van Berge Henegouwen MI, Slors JF, Bemelman WA (2008) Staple line failure using the Proximate 100-mm linear cutter. Dis Colon Rectum 51:1275–1278PubMedCrossRef Wind J, Safiruddin F, van Berge Henegouwen MI, Slors JF, Bemelman WA (2008) Staple line failure using the Proximate 100-mm linear cutter. Dis Colon Rectum 51:1275–1278PubMedCrossRef
5.
go back to reference Offodile AC II, Feingold DL, Nasar A, Whelan RL, Arnell TD (2010) High incidence of technical errors involving the EEA circular stapler: a single institution experience. J Am Coll Surg 210:331–335PubMedCrossRef Offodile AC II, Feingold DL, Nasar A, Whelan RL, Arnell TD (2010) High incidence of technical errors involving the EEA circular stapler: a single institution experience. J Am Coll Surg 210:331–335PubMedCrossRef
6.
go back to reference Morita K, Maeda N, Kawaoka T, Hiraki S, Kudo A, Fukuda S, Oka M (2008) Effects of the time interval between clamping and linear stapling for resection of porcine small intestine. Surg Endosc 22:750–756PubMedCrossRef Morita K, Maeda N, Kawaoka T, Hiraki S, Kudo A, Fukuda S, Oka M (2008) Effects of the time interval between clamping and linear stapling for resection of porcine small intestine. Surg Endosc 22:750–756PubMedCrossRef
7.
go back to reference Kawasaki K, Fujino Y, Kanemitsu K, Goto T, Kamigaki T, Kuroda D, Kuroda Y (2007) Experimental evaluation of the mechanical strength of stapling techniques. Surg Endosc 21:1796–1799PubMedCrossRef Kawasaki K, Fujino Y, Kanemitsu K, Goto T, Kamigaki T, Kuroda D, Kuroda Y (2007) Experimental evaluation of the mechanical strength of stapling techniques. Surg Endosc 21:1796–1799PubMedCrossRef
8.
go back to reference Goto T, Kawasaki K, Fujino Y, Kanemitsu K, Kamigaki T, Kuroda D, Suzuki Y, Kuroda Y (2007) Evaluation of the mechanical strength and patency of functional end-to-end anastomoses. Surg Endosc 21:1508–1511PubMedCrossRef Goto T, Kawasaki K, Fujino Y, Kanemitsu K, Kamigaki T, Kuroda D, Suzuki Y, Kuroda Y (2007) Evaluation of the mechanical strength and patency of functional end-to-end anastomoses. Surg Endosc 21:1508–1511PubMedCrossRef
9.
go back to reference Yo LS, Consten EC, Quarles van Ufford HM, Gooszen HG, Gagner M (2006) Buttressing of the staple line in gastrointestinal anastomoses: overview of new technology designed to reduce perioperative complications. Dig Surg 23:283–291PubMedCrossRef Yo LS, Consten EC, Quarles van Ufford HM, Gooszen HG, Gagner M (2006) Buttressing of the staple line in gastrointestinal anastomoses: overview of new technology designed to reduce perioperative complications. Dig Surg 23:283–291PubMedCrossRef
10.
go back to reference Ahmed S, Hanna GB, Cuschieri A (2004) Optimal angle between instrument shaft and handle for laparoscopic bowel suturing. Arch Surg 139:89–92PubMedCrossRef Ahmed S, Hanna GB, Cuschieri A (2004) Optimal angle between instrument shaft and handle for laparoscopic bowel suturing. Arch Surg 139:89–92PubMedCrossRef
11.
go back to reference Memon AA, Marks CG (1996) Stapled anastomoses in colorectal surgery: a prospective study. Eur J Surg 162:805–810PubMed Memon AA, Marks CG (1996) Stapled anastomoses in colorectal surgery: a prospective study. Eur J Surg 162:805–810PubMed
12.
Metadata
Title
The importance of precompression time for secure stapling with a linear stapler
Authors
Shinya Nakayama
Suguru Hasegawa
Satoshi Nagayama
Shigeru Kato
Koya Hida
Eiji Tanaka
Atsushi Itami
Hajime Kubo
Yoshiharu Sakai
Publication date
01-07-2011
Publisher
Springer-Verlag
Published in
Surgical Endoscopy / Issue 7/2011
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-010-1527-7

Other articles of this Issue 7/2011

Surgical Endoscopy 7/2011 Go to the issue